Gas-phase separation using a trapped ion mobility spectrometer
Open accessThe DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
In the present work we describe the principles of operation, versatility and applicability of a trapped ion mobility spectrometer (TIMS) analyzer for fast, gas-phase separation of molecular ions based on their size-to-charge ratio. Mobility-based separation using a TIMS device is shown for a series for isobar pairs. In a TIMS device, mobility resolution depends on the bath gas velocity and analysis scan speed, with the particularity that the mobility separation can be easily tuned from low to high resolution (R > 50) in accordance with the analytical challenge . In contrast to traditional drift tube IMS analyzer, a TIMS device can be easily integrated in a mass spectrometer without a noticeable loss in ion transmission or sensitivity, thus providing a powerful separation platform prior to mass analysis.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Ion mobility–mass spectrometry
Kanu, A. B., Dwivedi, P., Tam, M., Matz, L., & Hill, H. H. (2008) · Journal of Mass Spectrometry · doi:10.1002/jms.1383 · Cited by 1,154
Structural Information from Ion Mobility Measurements: Effects of the Long-Range Potential
Mesleh, M. F., Hunter, J. M., Shvartsburg, A. A., Schatz, G. C., & Jarrold, M. F. (1996) · The Journal of Physical Chemistry · doi:10.1021/jp961623v · Cited by 1,043
The mobility and diffusion of ions in gases
McDaniel, E. W., & Mason, E. A. (1973) · CERN Document Server (European Organization for Nuclear Research) · Cited by 942
An exact hard-spheres scattering model for the mobilities of polyatomic ions
Shvartsburg, A. A., & Jarrold, M. F. (1996) · Chemical Physics Letters · doi:10.1016/0009-2614(96)00941-4 · Cited by 819
Pringle, S. D., Giles, K., Wildgoose, J. L., Williams, J. P., Slade, S. E., Thalassinos, K., Bateman, R. H., Bowers, M. T., & Scrivens, J. H. (2007) · International Journal of Mass Spectrometry · doi:10.1016/j.ijms.2006.07.021 · Cited by 808
Silicon cluster ions: Evidence for a structural transition
Jarrold, M. F., & Constant, V. A. (1991) · Physical Review Letters · doi:10.1103/physrevlett.67.2994 · Cited by 380
Kolakowski, B. M., & Mester, Z. (2007) · The Analyst · doi:10.1039/b706039d · Cited by 351
Ion mobility–mass spectrometry: a new paradigm for proteomics
McLean, J. A., Ruotolo, B. T., Gillig, K. J., & Russell, D. H. (2005) · International Journal of Mass Spectrometry · doi:10.1016/j.ijms.2004.10.003 · Cited by 309
High-resolution ion mobility measurements
Dugourd, P., Hudgins, R. R., Clemmer, D. E., & Jarrold, M. F. (1997) · Review of Scientific Instruments · doi:10.1063/1.1147873 · Cited by 299
Gaseous lon mobility in electric fields of arbitrary strength
Viehland, L. A., & Mason, E. A. (1975) · Annals of Physics · doi:10.1016/0003-4916(75)90233-x · Cited by 293
Koeniger, S. L., Merenbloom, S. I., Valentine, S. J., Jarrold, M. F., Udseth, H. R., Smith, R. D., & Clemmer, D. E. (2006) · Analytical Chemistry · doi:10.1021/ac051060w · Cited by 272
A new, higher resolution, ion mobility mass spectrometer
Kemper, P. R., Dupuis, N. F., & Bowers, M. T. (2009) · International Journal of Mass Spectrometry · doi:10.1016/j.ijms.2009.01.012 · Cited by 195
Petroleum Crude Oil Characterization by IMS-MS and FTICR MS
Fernandez-Lima, F. A., Becker, C., McKenna, A. M., Rodgers, R. P., Marshall, A. G., & Russell, D. H. (2009) · Analytical Chemistry · doi:10.1021/ac901594f · Cited by 188
Self-citationMapping the human plasma proteome by SCX-LC-IMS-MS
Liu, X., Valentine, S. J., Plasencia, M. D., Trimpin, S., Naylor, S., & Clemmer, D. E. (2007) · Journal of the American Society for Mass Spectrometry · doi:10.1016/j.jasms.2007.04.012 · Cited by 187
Metabolic profiling by ion mobility mass spectrometry (IMMS)
Dwivedi, P., Wu, P., Klopsch, S. J., Puzon, G. J., Xun, L., & Hill, H. H. (2008) · Metabolomics · doi:10.1007/s11306-007-0093-z · Cited by 160
High-Resolution Ion Cyclotron Mobility Spectrometry
Merenbloom, S. I., Glaskin, R. S., Henson, Z. B., & Clemmer, D. E. (2009) · Analytical Chemistry · doi:10.1021/ac801880a · Cited by 149
An Ion Trap Interface for ESI−Ion Mobility Experiments
Hoaglund‐Hyzer, C. S., Valentine, S. J., & Clemmer, D. E. (1997) · Analytical Chemistry · doi:10.1021/ac970526a · Cited by 117
Zeleny, J. (1898) · The London Edinburgh and Dublin Philosophical Magazine and Journal of Science · doi:10.1080/14786449808621173 · Cited by 74
Cited by
Meier, F., Brunner, A.-D., Koch, S., Koch, H., Lubeck, M., Krause, M., Goedecke, N., Decker, J., Kosinski, T., Park, M. A., et al. (2018) · Molecular & Cellular Proteomics · doi:10.1074/mcp.tir118.000900 · Cited by 965
Self-citationdiaPASEF: parallel accumulation–serial fragmentation combined with data-independent acquisition
Meier, F., Brunner, A.-D., Frank, M., Ha, A., Bludau, I., Voytik, E., Kaspar-Schoenefeld, S., Lubeck, M., Raether, O., Bache, N., et al. (2020) · Nature Methods · doi:10.1038/s41592-020-00998-0 · Cited by 897
Recommendations for reporting ion mobility Mass Spectrometry measurements
Gabelica, V., Shvartsburg, A. A., Afonso, C., Barran, P., Benesch, J. L. P., Bleiholder, C., Bowers, M. T., Bilbao, A., Bush, M. F., Campbell, J. L., et al. (2019) · Mass Spectrometry Reviews · doi:10.1002/mas.21585 · Cited by 493
“In many TIMS systems (Fernandez-Lima et al., 2011; Michelmann et al., 2015; Bleiholder, 2016) and in the current tandem-TIMS implementation, s(z)1⁄4 z/z0 but s(z) can be nonlinear in other systems.”
Self-citationReview on ion mobility spectrometry. Part 1: current instrumentation
Cumeras, R., Figueras, E., Davis, C. E., Baumbach, J. I., & Gràcia, I. (2015) · The Analyst · doi:10.1039/c4an01100g · Cited by 459
Ion Mobility-Mass Spectrometry: Time-Dispersive Instrumentation
May, J. C., & McLean, J. A. (2015) · Analytical Chemistry · doi:10.1021/ac504720m · Cited by 435
Metadata
- Authors
- Fernandez-Lima, F., Kaplan, D. A., Suetering, J., & Park, M. A.
- Venue
- International Journal for Ion Mobility Spectrometry
- Published
- 2011-06-25
- Volume
- 14
- Issue
- 2-3
- Pages
- 93-98
- ISSN
- 1435-6163 (ISSN-L)
- 1865-4584
- OpenAlex
- W1992412192
- Cited by
- 315
- References
- 18
- FWCI
- 2.05
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.