Reliability of blood glucose monitoring by patients with diabetes mellitus
Abstract
Nineteen patients with insulin-dependent diabetes mellitus were evaluated for overall reliability of self-generated data from capillary blood glucose monitoring. For a period of 12 to 14 days, standard reflectance meters used by these subjects were replaced by meters internally modified with memory chips capable of storing all glucose readings by date and time. The subjects were not aware of this modification and were instructed to continue to test capillary blood glucose as they had been and to continue their practice of recording the meter readings in a logbook. To assess reliability of patient-generated data as recorded in the logbook, the addition, deletion, and alteration of test results were determined. A significantly lower (p less than 0.0001) mean blood glucose level was reported in the logbooks than recorded in the memory reflectance meters. Differences in logbooks and memory reflectance meters ranged from 0 to 109 mg/dl. Three fourths of the subjects had reported lower than actual mean blood glucose values. Under-reporting, or omission of memory reflectance meter readings, averaged 10 percent, whereas over-reporting or addition of phantom values averaged 40 percent. An average of 26 percent of the logbook entries were not identical to memory reflectance meter values determined at the corresponding time. Two thirds of the subjects had reported values in such a manner as to obscure hyper- and hypoglycemia, leading to misleading clinical impressions about the fluctuat…
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The development of high-reliability glucose monitoring devices requires new technologies and strategies to make sensors more affordable, non-invasive, and suitable for continuous monitoring.
“Today, the development of devices for glucose monitoring with high reliability need new technologies and strategies to make those sensors more affordable, non-invasive, and suitable for continuous monitoring of glycemic status 40-42.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [40]
Cites
Correlation of Glucose Regulation and Hemoglobin A Ic in Diabetes Mellitus
Koenig, R. J., Peterson, C. M., Jones, R. L., Saudek, C. D., Lehrman, M. A., & Cerami, A. (1976) · New England Journal of Medicine · doi:10.1056/nejm197608192950804 · Cited by 1,173 · see citing works
Hemoglobin Components in Patients with Diabetes Mellitus
Trivelli, L., Ranney, H. M., & Lai, H.-T. (1971) · New England Journal of Medicine · doi:10.1056/nejm197102182840703 · Cited by 841 · see citing works
Glycosylated Hemoglobins and Long-Term Blood Glucose Control in Diabetes Mellitus
Gabbay, K. H., Hasty, K. A., Breslow, J. L., Ellison, R. C., Bunn, H. F., & Gallop, P. M. (1977) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jcem-44-5-859 · Cited by 520 · see citing works
Relation of diabetic control to development of microvascular complications
Tchobroutsky, G. (1978) · Diabetologia · doi:10.1007/bf00421230 · Cited by 427 · see citing works
HOME MONITORING OF BLOOD-GLUCOSE
Sönksen, P. H., Judd, S. L., & Lowy, C. (1978) · The Lancet · doi:10.1016/s0140-6736(78)90854-1 · Cited by 341 · see citing works
SELF-MONITORING OF BLOOD-GLUCOSE
Walford, S., Gale, E. A. M., Allison, S. P., & Tattersall, R. B. (1978) · The Lancet · doi:10.1016/s0140-6736(78)90855-3 · Cited by 269 · see citing works
Jovanovič, L., Druzin, M. L., & Peterson, C. M. (1981) · The American Journal of Medicine · doi:10.1016/0002-9343(81)90301-6 · Cited by 268 · see citing works
Schiffrin, A., & Belmonte, M. M. (1982) · Diabetes Care · doi:10.2337/diacare.5.5.479 · Cited by 200 · see citing works
Special Comment: Meticulous Control of Diabetes: Benefits, Risks, and Precautions
Unger, R. H. (1982) · Diabetes · doi:10.2337/diab.31.6.479 · Cited by 134 · see citing works
complications of Diabetes Mellitus: Relationship to Metabolic Dysfunction
Skyler, J. S. (1979) · Diabetes Care · doi:10.2337/diacare.2.6.499 · Cited by 102 · see citing works
Patient self-monitoring of blood glucose and refinements of conventional insulin treatment
Tattersall, R. B., & Gale, E. A. (1981) · The American Journal of Medicine · doi:10.1016/0002-9343(81)90424-1 · Cited by 69 · see citing works
Sönksen, P. H., Judd, S., & Lowy, C. (1980) · Diabetes Care · doi:10.2337/diacare.3.1.100 · Cited by 42 · see citing works
Home blood sampling for plasma glucose assay in control of diabetes.
Howe-Davies, S., Holman, R. R., Phillips, M., & Turner, R. C. (1978) · BMJ · doi:10.1136/bmj.2.6137.596 · Cited by 41 · see citing works
Efficacy of home glucose monitoring in diabetic pregnancy
Varner, M. W. (1983) · The American Journal of Medicine · doi:10.1016/0002-9343(83)90438-2 · Cited by 30 · see citing works
Home monitoring of blood glucose by diabetic patients.
Ph, S. (1980) · PubMed · Cited by 17 · see citing works
Intravenous Home Blood Glucose Monitoring
Schade, D. S., Eaton, R. P., Mitchell, W. J., & Friedman, N. M. (1981) · Diabetes Care · doi:10.2337/diacare.4.3.420 · Cited by 9 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Ecological momentary assessment
Shiffman, S., Stone, A. A., & Hufford, M. R. (2008) · Annual Review of Clinical Psychology · doi:10.1146/annurev.clinpsy.3.022806.091415 · Cited by 6,454 · see citing works
Assessment of Diabetes-Related Distress
Polonsky, W. H., Anderson, B. J., Lohrer, P. A., Welch, G., Jacobson, A. M., Aponte, J. E., & Schwartz, C. E. (1995) · Diabetes Care · doi:10.2337/diacare.18.6.754 · Cited by 1,497 · see citing works
Response Burden and Questionnaire Length: Is Shorter Better? A Review and Meta-analysis
Rolstad, S., Adler, J., & Rydén, A. (2011) · Value in Health · doi:10.1016/j.jval.2011.06.003 · Cited by 958 · see citing works
Patient compliance with paper and electronic diaries
Stone, A. A., Shiffman, S., Schwartz, J. E., Broderick, J. E., & Hufford, M. R. (2003) · Controlled Clinical Trials · doi:10.1016/s0197-2456(02)00320-3 · Cited by 910 · see citing works
Recommendations for the use of home (self) and ambulatory blood pressure monitoring
Pickering, T. G. (1996) · American Journal of Hypertension · doi:10.1016/0895-7061(95)00341-x · Cited by 473 · see citing works
Adherence and Persistence with Glaucoma Therapy
Schwartz, G. F., & Quigley, H. A. (2008) · Survey of Ophthalmology · doi:10.1016/j.survophthal.2008.08.002 · Cited by 337 · see citing works
Review of hemoglobin A1c in the management of diabetes
Gallagher, E. J., Roith, D. L., & Bloomgarden, Z. T. (2009) · Journal of Diabetes · doi:10.1111/j.1753-0407.2009.00009.x · Cited by 303 · see citing works
Friedman, D., Hahn, S., Gelb, L., Tan, J., Shah, S. N., Kim, E. E., Zimmerman, T., & Quigley, H. (2008) · Ophthalmology (Rochester, Minn.) · doi:10.1016/j.ophtha.2007.11.023 · Cited by 177 · see citing works
Guerci, B., Drouin, P., Grangé, V, Bougnères, P., Fontaine, P., Kerlan, V, Passa, P., Thivolet, C., Vialettes, B., & Charbonnel, B. (2003) · Diabetes & Metabolism · doi:10.1016/s1262-3636(07)70073-3 · Cited by 287 · see citing works
Reliability of reporting self-measured blood pressure values by hypertensive patients.
Mengden, T., Medina, R. M. H., Beltran, B., Álvarez, E., Kraft, K., & Vetter, H. (1998) · American Journal of Hypertension · doi:10.1016/s0895-7061(98)00241-6 · Cited by 296 · see citing works
Salivary cortisol sampling compliance: comparison of patients and healthy volunteers
Broderick, J. E., Arnold, D., Kudielka, B. M., & Kirschbaum, C. (2003) · Psychoneuroendocrinology · doi:10.1016/s0306-4530(03)00093-3 · Cited by 261 · see citing works
Pullano, S. A., Greco, M., Bianco, M. G., Foti, D., Brunetti, A., & Fiorillo, A. S. (2022) · Theranostics · doi:10.7150/thno.64035 · Cited by 210 · see citing works
Piasecki, T. M., Hufford, M. R., Solhan, M. B., & Trull, T. J. (2007) · Psychological Assessment · doi:10.1037/1040-3590.19.1.25 · Cited by 196 · see citing works
Self-management of diabetes mellitus: A critical review.
Goodall, T. A., & Halford, W. K. (1991) · Health Psychology · doi:10.1037//0278-6133.10.1.1 · Cited by 184 · see citing works
Adherence to Diabetes Regimens: Empirical Status and Clinical Applications
Kurtz, S. M. (1990) · The Science of Diabetes Self-Management and Care · doi:10.1177/014572179001600112 · Cited by 152 · see citing works
Signaling does not adequately improve diary compliance
Broderick, J. E., Schwartz, J. E., Shiffman, S. S., Hufford, M. R., & Stone, A. A. (2003) · Annals of Behavioral Medicine · doi:10.1207/s15324796abm2602_06 · Cited by 151 · see citing works
Ambulatory Glucose Profile: Representation of Verified Self-Monitored Blood Glucose Data
Mazze, R. S., Lucido, D., Langer, O., Hartmann, K., & Rodbard, D. (1987) · Diabetes Care · doi:10.2337/diacare.10.1.111 · Cited by 144 · see citing works
Self-citationAging and Medical Adherence: The Use of Automatic Processes to Achieve Effortful Things.
Liu, L. L., & Park, D. C. (2004) · Psychology and Aging · doi:10.1037/0882-7974.19.2.318 · Cited by 140 · see citing works
Reliability of Self-reported Blood Pressure Measurements
Johnson, K. A., Partsch, D. J., Rippole, L. L., & McVey, D. M. (1999) · Archives of Internal Medicine · doi:10.1001/archinte.159.22.2689 · Cited by 126 · see citing works
Continuous Glucose Monitoring: A Brief Review for Primary Care Practitioners
Ajjan, R., Slattery, D., & Wright, E. E. (2019) · Advances in Therapy · doi:10.1007/s12325-019-0870-x · Cited by 126 · see citing works
Wing, R. R., Epstein, L. H., Nowalk, M. P., Scott, N., Koeske, R. D., & Hagg, S. A. (1986) · The American Journal of Medicine · doi:10.1016/0002-9343(86)90354-2 · Cited by 120 · see citing works
Allen, B., DeLong, E. R., & Feussner, J. R. (1990) · Diabetes Care · doi:10.2337/diacare.13.10.1044 · Cited by 119 · see citing works
Quantifying Data Quality for Clinical Trials Using Electronic Data Capture
Nahm, M., Pieper, C. F., & Cunningham, M. M. (2008) · PLoS ONE · doi:10.1371/journal.pone.0003049 · Cited by 112 · see citing works
Marrero, D. G., Vandagriff, J. L., Kronz, K., Fineberg, N. S., Golden, M. P., Gray, D. L., Orr, D. P., Wright, J. C., & Johnson, N. B. (1995) · The Science of Diabetes Self-Management and Care · doi:10.1177/014572179502100409 · Cited by 112 · see citing works
The great myth of office blood pressure measurement
Myers, M. G. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e3283577b05 · Cited by 108 · see citing works
Moskowitz, D. S., Russell, J. J., Sadikaj, G., & Sutton, R. (2009) · Canadian Psychology / Psychologie canadienne · doi:10.1037/a0016625 · Cited by 103 · see citing works
Palsson, O. S., Baggish, J. S., Turner, M. J., & Whitehead, W. E. (2012) · American Journal of Gastroenterology · doi:10.1038/ajg.2011.358 · Cited by 99 · see citing works
Depression and the Reporting of Diabetes Symptoms
Lustman, P. J., Clouse, R. E., & Carney, R. M. (1989) · The International Journal of Psychiatry in Medicine · doi:10.2190/lw52-jfkm-jchv-j67x · Cited by 93 · see citing works
Wojcicki, J. M., Ładyżyński, P., Krzymień, J., Jóźwicka, E., Blachowicz, J., Janczewska, E., Czajkowski, K., & Karnafel, W. (2001) · Diabetes Technology & Therapeutics · doi:10.1089/15209150152811207 · Cited by 90 · see citing works
Cox, D. J., Kovatchev, B., Kovatchev, B., Koev, D., Koeva, L., Dachev, S., Tcharaktchiev, D., Protopopova, A., Gonder-Frederick, L., & Clarke, W. (2004) · International Journal of Behavioral Medicine · doi:10.1207/s15327558ijbm1104_4 · Cited by 88 · see citing works
Self monitoring of glucose by people with diabetes: evidence based practice: Fig 1
Gallichan, M. (1997) · BMJ · doi:10.1136/bmj.314.7085.964 · Cited by 84 · see citing works
Patient-Centered Communication to Assess and Enhance Patient Adherence to Glaucoma Medication
Hahn, S. R. (2009) · Ophthalmology · doi:10.1016/j.ophtha.2009.06.023 · Cited by 81 · see citing works
Psychological stress and blood glucose regulation in Type I diabetic patients.
Halford, W. K., Cuddihy, S., & Mortimer, R. H. (1990) · Health Psychology · doi:10.1037//0278-6133.9.5.516 · Cited by 81 · see citing works
Kyne‐Grzebalski, D., Wood, L., Marshall, S. M., & Taylor, R. (1999) · Diabetic Medicine · doi:10.1046/j.1464-5491.1999.00131.x · Cited by 79 · see citing works
Tests of Glycemia in Diabetes Mellitus
Singer, D. E., Coley, C. M., Samet, J. H., & Nathan, D. M. (1989) · Annals of Internal Medicine · doi:10.7326/0003-4819-110-2-125 · Cited by 77 · see citing works
Diabetes in Western Australian children: descriptive epidemiology
Glatthaar, C., Whittall, D. E., Welborn, T. A., Gibson, M. J., Brooks, B. H., Ryan, M. M. P., & Byrne, G. C. (1988) · The Medical Journal of Australia · doi:10.5694/j.1326-5377.1988.tb112770.x · Cited by 76 · see citing works
Strowig, S. M., & Raskin, P. (1998) · Diabetes Care · doi:10.2337/diacare.21.10.1694 · Cited by 72 · see citing works
Automated office blood pressure measurement in primary care.
Myers, M. G., Kaczorowski, J. A., Dawes, M. G., & Godwin, M. S. (2014) · Europe PMC (PubMed Central) · Cited by 71 · see citing works
Compliance with blood glucose monitoring in children with type 1 diabetes mellitus
Wilson, D. P., & Endres, R. K. (1986) · The Journal of Pediatrics · doi:10.1016/s0022-3476(86)80955-6 · Cited by 68 · see citing works
Pernick, N. L., & Rodbard, D. (1986) · Diabetes Care · doi:10.2337/diacare.9.1.61 · Cited by 65 · see citing works
Improved Care of Patients with Diabetes through Telecommunicationsa
Shultz, E. K., Bauman, A. J., Hayward, M. A., & Holzman, R. S. (1992) · Annals of the New York Academy of Sciences · doi:10.1111/j.1749-6632.1992.tb26084.x · Cited by 65 · see citing works
Langer, O., & Mazze, R. S. (1986) · American Journal of Obstetrics and Gynecology · doi:10.1016/0002-9378(86)90293-0 · Cited by 62 · see citing works
Self-citationLehmann, E. D., & Deutsch, T. (1995) · Medical Informatics · doi:10.3109/14639239509024284 · Cited by 62 · see citing works
A novel pen-based Bluetooth-enabled insulin delivery system with insulin dose tracking and advice
Bailey, T. S., & Stone, J. Y. (2017) · Expert Opinion on Drug Delivery · doi:10.1080/17425247.2017.1313831 · Cited by 60 · see citing works
Hahn, S. R., Friedman, D. S., Quigley, H. A., Kotak, S. V., Kim, E., Onofrey, M., Eagan, C. A., & Mardekian, J. (2010) · Ophthalmology · doi:10.1016/j.ophtha.2009.11.026 · Cited by 57 · see citing works
Monitoring glycemic control: The cornerstone ofdiabetes care
LeRoith, D., & Smith, D. O. (2005) · Clinical Therapeutics · doi:10.1016/j.clinthera.2005.10.010 · Cited by 58 · see citing works
Reporting bias in self-measurement of blood pressure
Myers, M. G. (2001) · Blood Pressure Monitoring · doi:10.1097/00126097-200108000-00003 · Cited by 57 · see citing works
Guilfoyle, S. M., Crimmins, N. A., & Hood, K. K. (2011) · Pediatric Diabetes · doi:10.1111/j.1399-5448.2010.00735.x · Cited by 56 · see citing works
Doctor–Patient Communication in Glaucoma Care
Friedman, D. S., Hahn, S. R., Quigley, H. A., Kotak, S. V., Kim, E., Onofrey, M., Eagan, C. A., & Mardekian, J. (2009) · Ophthalmology · doi:10.1016/j.ophtha.2009.04.052 · Cited by 55 · see citing works
Salivary proteins associated with hyperglycemia in diabetes: a proteomic analysis
Bencharit, S., Baxter, S. S., Carlson, J., Byrd, W. C., Mayo, M. V., Border, M. B., Kohltfarber, H., Urrutia, E., Howard-Williams, E. L., Offenbacher, S., et al. (2013) · Molecular BioSystems · doi:10.1039/c3mb70196d · Cited by 53 · see citing works
Blood glucose monitoring by diabetic adolescents: Compliance and metabolic control.
Wysocki, T., Green, L., & Huxtable, K. (1989) · Health Psychology · doi:10.1037//0278-6133.8.3.267 · Cited by 49 · see citing works
Impact of Glucose Self-monitoring on Glycohemoglobin Values in a Veteran Population
Newman, W. P. (1990) · Archives of Internal Medicine · doi:10.1001/archinte.1990.00390130103015 · Cited by 49 · see citing works
Given, J. E., O'Kane, M., Bunting, B. P., & COATES, V. E. (2013) · Diabetic Medicine · doi:10.1111/dme.12130 · Cited by 49 · see citing works
Blackwell, M., & Wheeler, B. J. (2016) · Acta Diabetologica · doi:10.1007/s00592-016-0907-4 · Cited by 48 · see citing works
Assessment Methods for Patient-Reported Outcomes
Hufford, M. R., & Shiffman, S. S. (2003) · Disease Management & Health Outcomes · doi:10.2165/00115677-200311020-00002 · Cited by 48 · see citing works
Blood Glucose Monitoring Technology: Translating Data Into Practice
Hirsch, I. B. (2004) · Endocrine Practice · doi:10.4158/ep.10.1.67 · Cited by 47 · see citing works
Clinical Evaluation of Computer-Assisted Self-Monitoring of Blood Glucose System
Marrero, D. G., Kronz, K. K., Golden, M. P., Wright, J. C., Orr, D. P., & Fineberg, N. S. (1989) · Diabetes Care · doi:10.2337/diacare.12.5.345 · Cited by 47 · see citing works
Can Smart Pens Help Improve Diabetes Management?
Sy, S. L., Munshi, M. N., & Toschi, E. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820965600 · Cited by 44 · see citing works
Reporting bias: Achilles' heel of home blood pressure monitoring
Myers, M. G., & Stergiou, G. S. (2014) · Journal of the American Society of Hypertension · doi:10.1016/j.jash.2014.02.001 · Cited by 42 · see citing works
Methodological issues affecting the value of patient-reported outcomes data
Hufford, M. R., & Shiffman, S. M. (2002) · Expert Review of Pharmacoeconomics & Outcomes Research · doi:10.1586/14737167.2.2.119 · Cited by 41 · see citing works
Laboratory Test Analysis Near the Patient
Belsey, R. E. (1986) · JAMA · doi:10.1001/jama.1986.03370060089026 · Cited by 41 · see citing works
Poor adherence to home blood pressure measurement schedule
van der Hoeven, N. V., van den Born, B. H., Cammenga, M., & van Montfrans, G. V. (2009) · Journal of Hypertension · doi:10.1097/hjh.0b013e328319917e · Cited by 45 · see citing works
Adherence improvement in Dutch glaucoma patients: a randomized controlled trial
Beckers, H. J. M., Webers, C. A. B., Busch, M. J. W. M., Brink, H., Colen, T. P., Schouten, J., & The Dutch Adherence Study Group (2012) · Acta Ophthalmologica · doi:10.1111/j.1755-3768.2012.02571.x · Cited by 39 · see citing works
Home blood pressure teletransmission for better diagnosis and treatment
Pickering, T. G., Gerin, W., & Holland, J. K. (1999) · Current Hypertension Reports · doi:10.1007/s11906-996-0020-0 · Cited by 39 · see citing works
Impact of blood glucose monitoring on diabetic control: Obstacles and interventions
Wysocki, T. (1989) · Journal of Behavioral Medicine · doi:10.1007/bf00846550 · Cited by 39 · see citing works
Accuracy of Self-Monitored Blood Glucose in Type 2 Diabetes
Kazlauskaite, R., Soni, S., Evans, A. T., Graham, K., & Fisher, B. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0111 · Cited by 37 · see citing works
McGrady, A. V., & Horner, J. M. (1999) · Applied Psychophysiology and Biofeedback · doi:10.1023/a:1022851232058 · Cited by 36 · see citing works
Klein, C. E., Oboler, S. K., Prochazka, A. V., Oboler, S., Frank, M., Glugla, M., & Winters, S. (1993) · Journal of General Internal Medicine · doi:10.1007/bf02599711 · Cited by 36 · see citing works
Broderick, J. E., & Stone, A. A. (2006) · Psychological Methods · doi:10.1037/1082-989x.11.1.106 · Cited by 35 · see citing works
Evaluation of the quality of bedside monitoring of the blood glucose level in a teaching hospital.
Ting, C.-M., & Nanji, A. A. (1988) · Europe PMC (PubMed Central) · Cited by 35 · see citing works
Reliability of Reporting of Self-Monitoring of Blood Glucose in Pregnant Women
Kendrick, J. M., Wilson, C., Elder, R. F., & Springer, C. M. (2005) · Journal of Obstetric, Gynecologic & Neonatal Nursing · doi:10.1177/0884217505276306 · Cited by 31 · see citing works
Patient-Provider Interactions and Patient Outcomes
Sanson‐Fisher, R., Campbell, E. M., Redman, S., & Hennrikus, D. J. (1989) · The Science of Diabetes Self-Management and Care · doi:10.1177/014572178901500209 · Cited by 31 · see citing works
New Technologies to Improve Clinical Trials
Kobak, K. A., Greist, J. H., Jefferson, J. W., Katzelnick, D. J., & Mundt, J. C. (2001) · Journal of Clinical Psychopharmacology · doi:10.1097/00004714-200106000-00001 · Cited by 31 · see citing works
Haakens, K., Hanssen, K. F., Dahl‐Jørgensen, K., Vaaler, S., Aagenaes, Ö., & Mosand, R. (1990) · Journal of Internal Medicine · doi:10.1111/j.1365-2796.1990.tb00263.x · Cited by 30 · see citing works
Peterson, C. M., Jovanovic, L. G., & Chanoch, L. H. (1986) · The American Journal of Medicine · doi:10.1016/0002-9343(86)90184-1 · Cited by 30 · see citing works
Blood Glucose Pattern Management in Diabetes: Creating Order from Disorder
Choudhary, P., Genovese, S., & Reach, G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700618 · Cited by 28 · see citing works
Adolescent adherence in type 1 diabetes
Hoffman, R. P. (2002) · Comprehensive Therapy · doi:10.1007/s12019-002-0050-8 · Cited by 28 · see citing works
Zimmet, P., Gerstman, M., Raper, L., Cohen, M., Crosbie, C., Kuykendall, V., Michaels, D. W., & Hartmann, K. (1985) · Diabetes Research and Clinical Practice · doi:10.1016/s0168-8227(85)80029-2 · Cited by 26 · see citing works
Self-citationDiabetes screening, diagnosis, and therapy in pediatric patients with type 2 diabetes.
Rodbard, H. W. (2008) · PubMed Central · Cited by 24 · see citing works
Stenström, U., Göth, A., Carlsson, C., & Andersson, P. (2003) · Diabetes Research and Clinical Practice · doi:10.1016/s0168-8227(03)00018-4 · Cited by 25 · see citing works
Glycated Fibrinogen: A New Index of Short-Term Diabetic Control
Hammer, M. R., John, P. N., Flynn, M. D., Bellingham, A. J., & Leslie, R. D. (1989) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456328902600108 · Cited by 25 · see citing works
Parfitt, V. J., Clark, J. D. A., TURNER, G. M., & Hartog, M. (1993) · Diabetic Medicine · doi:10.1111/j.1464-5491.1993.tb00035.x · Cited by 25 · see citing works
Delamater, A. M., Davis, S. G., Bubb, J., Santiago, J. V., Smith, J. A., & White, N. H. (1989) · The Science of Diabetes Self-Management and Care · doi:10.1177/014572178901500115 · Cited by 25 · see citing works
Bastyr, E. J., Zhang, S., Mou, J., Hackett, A. P., Raymond, S. A., & Chang, A. M. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0407 · Cited by 24 · see citing works
Continuous Glucose Monitoring During Pregnancy
Chitayat, L., Zisser, H., & Jovanovič, L. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0136 · Cited by 24 · see citing works
Sevick, M. A., Levine, D. W., Burkart, J. M., Rocco, M. V., Keith, J. D., & Cohen, S. J. (1999) · Peritoneal Dialysis International · doi:10.1177/089686089901900105 · Cited by 24 · see citing works
Ładyżyński, P., Wojcicki, J. M., Krzymień, J., Blachowicz, J., Jóźwicka, E., Czajkowski, K., Janczewska, E., & Karnafel, W. (2001) · The International Journal of Artificial Organs · doi:10.1177/039139880102400308 · Cited by 23 · see citing works
The Quality of Self‐monitoring of Blood Glucose
Colagiuri, R., Colagiuri, S., Jones, S., & Moses, R. G. (1990) · Diabetic Medicine · doi:10.1111/j.1464-5491.1990.tb01495.x · Cited by 23 · see citing works
Pickering, T. G. (1988) · Hypertension · doi:10.1161/01.hyp.11.3_pt_2.ii96 · Cited by 21 · see citing works
Mazze, R. S., Simonson, G., Strock, E. S., Bergenstal, R. M., Idrogo, M., Ramirez, S., Hsu, S. I., & the International SDM Study Group (2001) · Practical Diabetes · doi:10.1002/pdi.240 · Cited by 22 · see citing works
Self-citationKalergis, M., Nadeau, J., Pacaud, D., Yared, Z., & Yale, J. (2006) · Canadian Journal of Diabetes · doi:10.1016/s1499-2671(06)03006-1 · Cited by 21 · see citing works
Dumais, K. M., Dias, N. R., Khurana, L., Gary, S. T., Witherspoon, B., Evans, C. J., & Dallabrida, S. M. (2019) · Patient · doi:10.1007/s40271-019-00376-9 · Cited by 21 · see citing works
Assessing ambulatory blood pressure in renal diseases: facts and concerns
Lurbe, E., & Redón, J. (1999) · Nephrology Dialysis Transplantation · doi:10.1093/ndt/14.11.2564 · Cited by 21 · see citing works
Computerized home telemetry of maternal blood pressure in hypertensive pregnancy
Dalton, K. J., Manning, K., Robarts, P. J., Dripps, J. H., & Currie, J. R. (1987) · International Journal of Bio-Medical Computing · doi:10.1016/0020-7101(87)90085-7 · Cited by 21 · see citing works
Use of Memory Meters to Measure Reliability of Self Blood Glucose Monitoring
Williams, C. D., Scobie, I. N., Till, S., Crane, R., Lowy, C., & Sönksen, P. H. (1988) · Diabetic Medicine · doi:10.1111/j.1464-5491.1988.tb01028.x · Cited by 21 · see citing works
Dorchy, H., & Roggemans, M.-P. (1997) · Diabetes Research and Clinical Practice · doi:10.1016/s0168-8227(97)00034-x · Cited by 21 · see citing works
Can Technology Improve Adherence to Long-Term Therapies?
Reach, G. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300313 · Cited by 20 · see citing works
Self-monitoring of the blood glucose level: potential sources of inaccuracy.
Rasaiah, B. (1985) · PubMed Central · Cited by 20 · see citing works
Modern Management of Insulin-dependent Diabetes Mellitus
Nathan, D. M. (1988) · Medical Clinics of North America · doi:10.1016/s0025-7125(16)30712-x · Cited by 20 · see citing works
Photo‐ and Electrochemical Dual‐Responsive Iridium Probe for Saccharide Detection
Carrod, A. J., Graglia, F., Male, L., Duff, C. S. L., Simpson, P. V., Elsherif, M., Ahmed, Z., Butt, H., Xu, G., Lo, K. K., et al. (2021) · Chemistry - A European Journal · doi:10.1002/chem.202103541 · Cited by 19 · see citing works
- Authors
- Mazze, R. S., Shamoon, H., Pasmantier, R., Lucido, D., Murphy, J., Hartmann, K., Kuykendall, V., & Lopatin, W.
- Corresponding author
- Mazze, Roger S.
- ORCID
- Shamoon, Harry
- Venue
- The American Journal of Medicine
- Publisher
- Elsevier BV
- Published
- 1984-08
- Volume
- 77
- Issue
- 2
- Pages
- 211-217
- ISSN
- 0002-9343 (ISSN-L)
- OpenAlex
- W2171957007
- Cited by
- 261
- References
- 16
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 1 · 2021: 6 · 2020: 7 · 2019: 7 · 2018: 5 · 2017: 3 · 2016: 5 · 2015: 5 · 2014: 10 · 2013: 5 · 2012: 3
- FWCI
- 9.46
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- The American Journal of Medicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Health Professions › Emergency Medical Services › Patient Safety and Medication Errors
- Keywords
- diabetes mellitus, type 1 diabetes, capillary blood glucose monitoring, self-monitoring of blood glucose, glycemic control, hypoglycemia, hyperglycemia
- MeSH terms
- Adolescent, Adult, Blood Glucose, Diabetes Mellitus, Type 1, Evaluation Studies as Topic, Female, Humans, Insulin, Male, Monitoring, Physiologic, Self Care
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review