CoChat

Subcutaneous glucose monitoring by means of electrochemical sensors: fiction or reality?

Rebrin, K., Fischer, U., Hahn von Dorsche, H., von Woetke, T., Abel, P., Brunstein, E. · 1992 · Journal of Biomedical Engineering
Cited by 81

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

Amperometric glucose oxidase/hydrogen peroxide sensors were inserted subcutaneously into the neck of normal and diabetic dogs (n = 10), to elucidate the conditions for stable long-term functioning. Their output current was observed in parallel with measurements of plasma glucose concentrations and their function was checked by means of induced alterations in glycaemia. After between 14 and 96 h the experiments were terminated due to losses in the apparent sensitivity of implanted sensors and/or increasing oscillations following stable measurements. This was accompanied by an inflammatory reaction which was analysed on the basis of the clinical picture and histology. In most cases there was a bacterial ingrowth from the normal skin flora of dogs. The inflammatory exsudate contained only 23 +/- 17% of the simultaneous steady state plasma glucose concentration, which was significantly different from the glucose level in the fluid obtained from non-irritate subcutaneous tissue (95 +/- 12%, separate set of experiments). The in vitro calibration of sensors exhibited essentially comparable sensitivities before and after the in vivo application. No differences in reported findings related to the biomaterials used (polyurethane versus cellulose acetate), the presence of diabetes, the history of individual electrodes and the effective duration of a given experiment were discernible. We conclude that the functional bioinstability of subcutaneous glucose sensors is largely due to the inflammatory tissue reaction which alters the effective glucose concentration within the measuring compartment of the electrodes; these drawbacks may be overcome by further miniaturization including implantable telemetric devices allowing the closure of the skin.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1016/0141-5425(92)90033-h

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cited by

See all 81

Metadata

Authors
Rebrin, K., Fischer, U., Hahn von Dorsche, H., von Woetke, T., Abel, P., & Brunstein, E.
Venue
Journal of Biomedical Engineering
Published
1992-01-01
Volume
14
Issue
1
Pages
33-40
ISSN
0141-5425 (ISSN-L)
OpenAlex
W4318599507
Cited by
81
References
1
FWCI
7.93
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Subcutaneous glucose monitoring by means of electrochemical sensors: fiction or reality? https://app.cochat.ai/papers/10.1016/0141-5425(92)90033-h
The paper
Rebrin, K., Fischer, U., Hahn von Dorsche, H., von Woetke, T., Abel, P., & Brunstein, E. (1992). Subcutaneous glucose monitoring by means of electrochemical sensors: fiction or reality? Journal of Biomedical Engineering. https://doi.org/10.1016/0141-5425(92)90033-h

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.