Physiological preparation for studying the response of subcutaneously implanted glucose and oxygen sensors
Abstract
A physiological preparation has been developed for studying the response of glucose end oxygen sensors chronically implanted in subcutaneous tissues. The preparation employs a chamber permanently mounted on the back of a rat that supports the growth of vascularized subcutaneous tissue around the sensors and is used in conjunction with chronic intravascular catheters for blood sampling and fluid infusion. A total of 26 glucose and oxygen sensors were implanted in nine chambers. At 10 days, the tissue surrounding the sensors was cellular, well vascularized and permeable. Glucose sensors responded to glucose infusions with a 10-15 minute lag.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The optical tissue window enables tissue monitoring and correlation with sensor performance.
“The optical tissue window enables tissue monitoring and correlation to the sensor performance.116–118”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [117]
Cites
IMPROVEMENTS IN EPOXY RESIN EMBEDDING METHODS
Luft, J. H. (1961) · The Journal of Cell Biology · doi:10.1083/jcb.9.2.409 · Cited by 10,243 · see citing works
Ondracek, G. (1972) · Journal of Nuclear Materials · doi:10.1016/0022-3115(72)90033-5 · Cited by 2,664 · see citing works
Schultz, J. S., Mansouri, S., & Goldstein, I. (1982) · Diabetes Care · doi:10.2337/diacare.5.3.245 · Cited by 305 · see citing works
Cahill, G. F., Etzwiler, D. D., & Freinkel, N. (1976) · New England Journal of Medicine · doi:10.1056/nejm197604292941811 · Cited by 302 · see citing works
A transparent access chamber for the rat dorsal skin fold
Papenfuss, H. D., Gross, J. F., Intaglietta, M., & Treese, F. A. (1979) · Microvascular Research · doi:10.1016/0026-2862(79)90039-6 · Cited by 282 · see citing works
Two-dimensional enzyme electrode sensor for glucose
Gough, D., Lucisano, J. Y., & Tse, P. H. S. (1985) · Analytical Chemistry · doi:10.1021/ac00289a042 · Cited by 223 · see citing works
Self-citationApplication of Chronic Intravascular Blood Glucose Sensor in Dogs
Armour, J. C., Lucisano, J. Y., Bd, M., & Gough, D. A. (1990) · Diabetes · doi:10.2337/diab.39.12.1519 · Cited by 175 · see citing works
Self-citationGranulocytes and no-reflow phenomenon in irreversible hemorrhagic shock.
Barroso-Aranda, J., Schmid‐Schönbein, G. W., Zweifach, B. W., & Engler, R. L. (1988) · Circulation Research · doi:10.1161/01.res.63.2.437 · Cited by 173 · see citing works
The Measurement of Oxygen Tension in Healing Tissue
Silver, I. A. (2015) · Progress in respiratory research/Progress in research in emphysema and chronic bronchitis/Progress in respiration research · doi:10.1159/000386224 · Cited by 120 · see citing works
Shichiri, M., Asakawa, N., Yamasaki, Y., Kawamori, R., & Abe, H. (1986) · Diabetes Care · doi:10.2337/diacare.9.3.298 · Cited by 96 · see citing works
Subcutaneous implantation of a ferrocene-mediated glucose sensor in pigs
Claremont, D. J., Sambrook, E., Penton, C., et al. (1986) · Diabetologia · doi:10.1007/bf00873223 · Cited by 94 · see citing works
How Fibroblasts and Giant Cells Encapsulate Implants: Considerations in Design of Glucose Sensors
Woodward, S. C. (1982) · Diabetes Care · doi:10.2337/diacare.5.3.278 · Cited by 89 · see citing works
Measurement of wound oxygen with implanted Silastic tube.
Ninikoski, J., & Tk, H. (1972) · PubMed · Cited by 73 · see citing works
Oxygen and carbon dioxide tensions in experimental wounds.
Ninikoski, J., Heughan, C., & Hunt, T. K. (1971) · PubMed · Cited by 68 · see citing works
Transient response of the two-dimensional glucose sensor
Lucisano, J. Y., & Gough, D. A. (1988) · Analytical Chemistry · doi:10.1021/ac00164a007 · Cited by 58 · see citing works
Self-citationA telemetry-instrumentation system for chronically implanted glucose and oxygen sensors
Bd, M., & Gough, D. (1988) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.4581 · Cited by 55 · see citing works
Self-citationIn vitro stability of an oxygen sensor
Lucisano, J. Y., Armour, J. C., & Gough, D. (1987) · Analytical Chemistry · doi:10.1021/ac00132a012 · Cited by 48 · see citing works
Self-citationA NEW METHOD OF DETERMINING TISSUE OXYGEN TENSION
al Hunt, T. K. E. (1964) · The Lancet · doi:10.1016/s0140-6736(64)91161-4 · Cited by 42 · see citing works
Measurement of wound oxygen with implanted Silastic tube
Niinikoski, J., & Hunt, T. K. (1972) · Plastic & Reconstructive Surgery · doi:10.1097/00006534-197207000-00047 · Cited by 38 · see citing works
A Miniature Glucose Sensor, Implantable in the Blood Stream
Kondo, T., Ito, K., Ohkura, K., et al. (1982) · Diabetes Care · doi:10.2337/diacare.5.3.218 · Cited by 23 · see citing works
Glucose Concentration at Possible Sensor Tissue Implant Sites
Wolfson, S. K., Tokarsky, J. M., Yao, S. J., & Krupper, M. A. (1982) · Diabetes Care · doi:10.2337/diacare.5.3.162 · Cited by 22 · see citing works
Short-term in vivo operation of a glucose sensor.
Gough, D. A., Armour, J. C., Jy, L., & Bd, M. (1986) · PubMed · Cited by 21 · see citing works
Self-citationMeasurement of Glucose Concentration in the Presence of Coreactants with a Platinum Electrode
Lerner, H., Soeldner, J. S., Colton, C. K., et al. (1982) · Diabetes Care · doi:10.2337/diacare.5.3.229 · Cited by 20 · see citing works
Issues related to in vitro operation of potentially implantable enzyme electrode glucose sensors.
Gough, D. A. (1988) · PubMed · Cited by 10 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Continuous glucose monitoring and metrics for clinical trials: an international consensus statement
Battelino, T., Alexander, C. M., Amiel, S. A., Arreaza-Rubin, G., Beck, R. W., Bergenstal, R. M., Buckingham, B. A., Carroll, J., Ceriello, A., Chow, E., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00319-9 · Cited by 831 · see citing works
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347 · see citing works
Koschwanez, H. E., & Reichert, W. M. (2007) · Biomaterials · doi:10.1016/j.biomaterials.2007.03.034 · Cited by 181 · see citing works
Ju, Y. M., Yu, B., West, L., Moussy, Y., & Moussy, F. (2010) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32400 · Cited by 52 · see citing works
Simulations of the Frequency Response of Implantable Glucose Sensors
Jablecki, M., & Gough, D. A. (2000) · Analytical Chemistry · doi:10.1021/ac991018z · Cited by 24 · see citing works
Self-citationIntravital microscopy evaluation of angiogenesis and its effects on glucose sensor performance
Koschwanez, H. E., Reichert, W. M., & Klitzman, B. (2010) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.32630 · Cited by 15 · see citing works
Percutaneous Window Chamber Method for Chronic Intravital Microscopy of Sensor—Tissue Interactions
Koschwanez, H. E., Klitzman, B., & Reichert, W. M. (2008) · Journal of Diabetes Science and Technology · doi:10.1177/193229680800200608 · Cited by 9 · see citing works
- Authors
- Ertefai, S., & Gough, D. A.
- Affiliations
- University of California San Diego (US)
- Corresponding author
- Gough, D.A.
- Venue
- Journal of Biomedical Engineering
- Publisher
- IOP Publishing
- Published
- 1989-09-01
- Volume
- 11
- Issue
- 5
- Pages
- 362-368
- ISSN
- 0141-5425 (ISSN-L)
- OpenAlex
- W2018992015
- Cited by
- 52
- References
- 24
- Citations by year
- 2024: 1 · 2022: 1 · 2019: 1 · 2018: 2 · 2015: 1 · 2013: 2 · 2012: 1
- FWCI
- 1.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- Locations
- Journal of Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Funders
- American Diabetes Association
- Lee Foundation
- Charles Lee Powell Foundation
- National Institutes of Health
- Topics
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrochemical sensors and biosensors
- Physical Sciences › Chemical Engineering › Bioengineering › Analytical Chemistry and Sensors
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Keywords
- glucose sensor, oxygen sensors, subcutaneous implantation, chronic implantation, sensor response, glucose infusion, rat model, intravascular catheters
- MeSH terms
- Animals, Blood Glucose, Body Temperature, Connective Tissue, Foreign-Body Reaction, Male, Monitoring, Physiologic, Oxygen, Rats, Inbred Strains, Biosensing Techniques, Prostheses and Implants, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review