Skin photoplethysmography — a review
Abstract
The photoplethysmograph has been used for over 50 years but there are still misconceptions in how and what is the information obtained. A photoplethysmograph signal from any site on the skin can be separated into an oscillating (a.c.) and a steady-state (d.c.) component, their amplitudes dependent upon the structure and flow in the vascular bed. Many simple applications are available: pulse counters, using the a.c. component, skin colour and haemoglobin saturation meters, using the d.c. component. The d.c. component of the photoplethysmograph signal is a function of the blood flux beneath the device. A good emitter for use in a photoplethysmograph of skin blood flow is one in the frequency range 600-700 nm and the best signal for a.c. analysis is obtained from the finger pulp. The frequency range of the electronic circuitry should be from 0.01 to 15 Hz, then all the information in the signal can be extracted about the autonomic nervous system control of the cardiovascular system, particularly between 0.01 and 2 Hz. Comparative studies may be drawn between similar skin sites on a subject or between subjects if the afferent inputs to the brain stem are controlled or driven at a known frequency. These afferents, inputs, will modulate the efferents, outputs, which generate variations in the a.c. component of the detected photoplethysmograph signal.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Without this type of real-time, in-sensor processing (IR_Non in Fig. 6D), robust operation would be impossible: PPG signals would quickly drift outside of the narrow operating range of the ADC as a result of patient-to-patient variations in skin pigmentation and unavoidable, time-dependent fluctuations in optical scattering that result from micromotions relative to underlying blood vessels and sub”
From Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care (2019) · cites this paper as [28]
Citations
Cites
McDonald, D. A. (1963) · The American Journal of the Medical Sciences · doi:10.1097/00000441-196301000-00017 · Cited by 1,568
Analysis of Heart Rate Variability
SAYKRS, B. M. (1973) · Ergonomics · doi:10.1080/00140137308924479 · Cited by 957
In vivo evaluation of microcirculation by coherent light scattering
Stern, M. D. (1975) · Nature · doi:10.1038/254056a0 · Cited by 866
Arterial assessment by Doppler-shift ultrasound.
Gosling, R. G., & King, D. H. (1974) · PubMed · Cited by 667
LASER DOPPLER MEASUREMENT OF CUTANEOUS BLOOD FLOW
Holloway, G. A., & Watkins, D. (1977) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12507665 · Cited by 543
THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPH
Hertzman, A. B. (1938) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1938.124.2.328 · Cited by 460
Bini, G., Hagbarth, K. E., Hynninen, P., & Wallin, B. G. (1980) · The Journal of Physiology · doi:10.1113/jphysiol.1980.sp013413 · Cited by 421
A theoretical and experimental study of light absorption and scattering by in vivo skin
Dawson, J. B., Barker, D., Ellis, D. J., COTTERILL, J. A., Grassam, E., Fisher, G. W., & Feather, J. W. (1980) · Physics in Medicine and Biology · doi:10.1088/0031-9155/25/4/008 · Cited by 355
Burton, A. C. (1939) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1939.127.3.437 · Cited by 333
Sympathetic Control of Human Blood Vessels
(1953) · The Medical Journal of Australia · doi:10.5694/j.1326-5377.1953.tb84977.x · Cited by 259
Photoelectric Plethysmography of the Fingers and Toes in Man
Hertzman, A. B. (1937) · Experimental Biology and Medicine · doi:10.3181/00379727-37-9630 · Cited by 221
The Input Impedance of an Assembly of Randomly Branching Elastic Tubes
Taylor, M. G. (1966) · Biophysical Journal · doi:10.1016/s0006-3495(66)86638-9 · Cited by 202
Regional similarities and differences in thermoregulatory vaso‐ and sudomotor tone.
Bini, G., Hagbarth, K. E., Hynninen, P., & Wallin, B. G. (1980) · The Journal of Physiology · doi:10.1113/jphysiol.1980.sp013414 · Cited by 164
A photoelectric plethysmograph for the measurement of cutaneous blood flow
CHALLONER, A. V. J., & Ramsay, C. A. (1974) · Physics in Medicine and Biology · doi:10.1088/0031-9155/19/3/003 · Cited by 145
Marks, R., & Payne, P. A. (1981) · Springer eBooks · doi:10.1007/978-94-009-7310-7 · Cited by 137
Psychophysiology of cardiovascular control: models, methods and data
(1987) · Comparative Biochemistry and Physiology Part A Physiology · doi:10.1016/0300-9629(87)90307-0 · Cited by 134
Mayer waves history and methodology
Peňáz, J. (1978) · Medical Entomology and Zoology · Cited by 126
Sympathetic Control of Human Blood Vessels
(1953) · JAMA · doi:10.1001/jama.1953.02940210196022 · Cited by 122
Non invasive physiological measurements
(1985) · General Pharmacology The Vascular System · doi:10.1016/0306-3623(85)90184-3 · Cited by 119
Burton, A. C., & Taylor, R. M. (1940) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1940.129.3.565 · Cited by 103
The absence of vasoconstrictor reflexes in the forehead circulation. Effects of cold
Hertzman, A. B., & Roth, L. W. (1942) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1942.136.4.692 · Cited by 85
RELATION OF ELECTRICAL PROPERTIES OF SKIN TO STRUCTURE AND PHYSIOLOGIC STATE
Edelberg, R. (1977) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12507771 · Cited by 85
Respiratory control of heart rate
Ahmed, A., Harness, J. B., & Mearns, A. J. (1982) · European Journal of Applied Physiology · doi:10.1007/bf00952248 · Cited by 79
Self-citationOptical radiation transfer in the human skin and applications in in vivo remittance spectroscopy
Anderson, R. R., Hu, J., & Parrish, J. (1981) · Springer eBooks · doi:10.1007/978-94-009-7310-7_28 · Cited by 74
In vivo studies with an intravascular and intracardiac reflection oximeter
Enson, Y., Briscoe, W. A., Polanyi, M. L., & Cournand, A. (1962) · Journal of Applied Physiology · doi:10.1152/jappl.1962.17.3.552 · Cited by 68
What does photoplethysmography measure?
de Trefford, J., & Lafferty, K. (1984) · Medical & Biological Engineering & Computing · doi:10.1007/bf02447713 · Cited by 65
The interaction between heart rate and respiration : Part 1 - Experimental studies in man
Selman, A. (1982) · Medical Entomology and Zoology · Cited by 57
Applications of photoelectric plethysmography in peripheral vascular disease
Hertzman, A. B., & Dillon, J. B. (1940) · American Heart Journal · doi:10.1016/s0002-8703(40)90534-8 · Cited by 57
Photoplethysmography- fundamental aspects of the optical properties of blood in motion
Roberts, V. C. (1982) · Transactions of the Institute of Measurement and Control · doi:10.1177/014233128200400205 · Cited by 54
VASCULAR CHANGES ACCOMPAXVIXG WHITE DERMOGRAPHISM AND DELAYED BLANCH IN ATOPIC DERMATITIS
Ramsay, C. A. (1969) · British Journal of Dermatology · doi:10.1111/j.1365-2133.1969.tb15917.x · Cited by 50
Cutaneous blood flow in man studied by freely diffusible radioactive indicators.
Sejrsen, P. (1967) · PubMed · Cited by 43
A NEW BLOODLESS METHOD FOR CONTINUOUS RECORDING OF PERIPHERAL CIRCULATORY CHANGES
Molitor, H., & Kniazuk, M. (1936) · Journal of Pharmacology and Experimental Therapeutics · doi:10.1016/s0022-3565(25)09748-4 · Cited by 40
Hokanson, D. E., Mozersky, D. J., Sumner, D. S., McLeod, F. D., & Strandness, D. E. (1972) · Radiology · doi:10.1148/102.2.435 · Cited by 40
The vasomotor components in the vascular reactions in the finger to cold
Hertzman, A. B., & Roth, L. W. (1942) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1942.136.4.669 · Cited by 39
Kyriacou, P. A., & Allen, J. (2021) · Elsevier eBooks · doi:10.1016/c2020-0-00098-8 · Cited by 35
Hyman, C. B., & Winsor, T. (1961) · PubMed · Cited by 35
Lichtelektrische Plethysmogramme
Matthes, K. J., & Hauss, W. H. (1938) · Journal of Molecular Medicine · doi:10.1007/bf01778576 · Cited by 29
A method for clinical evaluation of peripheral circulation: photoelectric hemodensitometry.
ELDRUP-JORGENSEN, S., Schwartz, S. I., & Jd, W. (1966) · PubMed · Cited by 28
A photoelectric approach to the study of peripheral circulation
Weinman, J., & Manoach, M. (1962) · American Heart Journal · doi:10.1016/0002-8703(62)90198-9 · Cited by 28
Photoplethysmography: System Calibration and Light History Effects
Novelly, R. A., Perona, P. J., & Ax, A. F. (1973) · Psychophysiology · doi:10.1111/j.1469-8986.1973.tb01084.x · Cited by 26
Estimation of the nutrient skin blood flow using a segmented thermal clearance probe
Holti, G., & MITCHELL, K. W. (1978) · Clinical and Experimental Dermatology · doi:10.1111/j.1365-2230.1978.tb01483.x · Cited by 24
Cutaneous blood flow in man studied by freely diffusible radioactive indicators.
Sejrsen, P. (1967) · PubMed · Cited by 21
Cummings, E. G. (1969) · Journal of Investigative Dermatology · doi:10.1038/jid.1969.109 · Cited by 20
Non-invasive measurement of limb and digit blood flow
Porter, J. M., & Swain, I. (1986) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(86)90082-8 · Cited by 19
REACTIONS OF LARGE AND SMALL ARTERIES IN MAN TO VASOCONSTRICTOR STIMULI
Hertzman, A. B., & Dillon, J. B. (1940) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1940.130.1.56 · Cited by 19
Correlation between blood volume and opacity changes in the finger.
Hocherman, S. D., & Palti, Y. (1967) · Journal of Applied Physiology · doi:10.1152/jappl.1967.23.2.157 · Cited by 19
The reactions of the digital artery and minute pad arteries to local cold
Hertzman, A. B., & Roth, L. W. (1942) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1942.136.4.680 · Cited by 17
Measurement of skin blood flow
Payne, P. A. (1984) · Electronics and Power · doi:10.1049/ep.1984.0114 · Cited by 14
The copper-tellurite-copper thermocouple adapted as a skin thermometer.
Holti, G. (1955) · PubMed · Cited by 14
A thermal entrainment device for cardiovascular investigation
Hitchen, M., Harness, J. B., & Mearns, A. J. (1980) · Journal of Medical Engineering & Technology · doi:10.3109/03091908009161117 · Cited by 10
Self-citationA PLETHYSMOGRAPHIC STUDY OF THE HUMAN SKIN UNDER VARIOUS ENVIRONMENTAL CONDITIONS
KUMAZAWA, T., Kobayashi, M., & Takagi, K. (1964) · The Japanese Journal of Physiology · doi:10.2170/jjphysiol.14.354 · Cited by 9
A study of the peripheral blood flow in skin tubed flaps by photoconductive plethysmography.
BEN-HUR, N., Mahler, Y., Shulman, J., & Neuman, Z. (1966) · PubMed · Cited by 7
Properties and applications of piezo film transducers
Halvorsen, D. L. (1986) · Electronics and Power · doi:10.1049/ep.1986.0432 · Cited by 3
Cross correlation between blood pressure and intracranial pressure
Ahmed, A., Harness, J. B., Mason, J. M., & Mearns, A. J. (1981) · Clinical Physics and Physiological Measurement · doi:10.1088/0143-0815/2/2/004 · Cited by 3
Self-citationA NEW FORM OF PISTON RECORDER AND SOME OF THE CHANGES OF THE VOLUME OF THE FINGER WHICH IT RECORDS
Lombard, W. P., & Pillsbury, W. B. (1899) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1899.3.5.186 · Cited by 3
Thermal entrainment and human ovulation
de Trafford, J. C., Edwards, S., Lafferty, K., Potter, C., Cotton, L. T., & Roberts, V. C. (1984) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(84)90081-5 · Cited by 2
Research works printed but not linked: 8
- Burton, 1940, Direct measurement of thermal conductance of skin as ideal index of peripheral blood flow, Am. J. Physiol., 129, 389
- Challoner, 1979, Photoelectric plethysmography for estimating cutaneous blood flow, Vol. 1, 127
- Fairs, 1985, Optical methods for measuring blood flow in skin, Vol. 1, 11
- Hertzman, 1937, Observations on the finger volume pulse recorded photoelectrically, Am. J. Physiol., 119, 334
- Mearns, 1985, Forcing frequency testing, theory and practice with reference to skin blood flow, Vol. 1, 23
- Molitor, 1936, A new continuous recording of peripheral circulatory changes, J. Pharmacol. Exp. Ther., 57, 6
- Sejrsen, 1971, Measurement of cutaneous bloodflow by freely diffusible radioactive isotopes, Dan. Med. Bull., 18, 9
- Thune, 1970, Plethysmographic recordings of skin pulses with particular reference to the piezoelectric method, Acta Dermatovener., 50, 27
Cited by
Sorted by citations
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
“…processes in relation to PPG measurements (Hertzman and Randall 1948, Brown et al 1965, D’Agrosa and Hertzman 1967, Weinman 1967, Zweifler et al 1967, Challoner 1979, Ochoa and Ohara 1980, Nijboer et al 1981, Roberts 1982, Lindberg and Öberg 1993, de Trafford and Lafferty 1984, Kamal et al 1989).”
BackgroundMethodWearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912
“The principle of PPG has been reviewed previously [2–4], and is explained briefly here.”
BackgroundBinodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., et al. (2019) · Science · doi:10.1126/science.aau0780 · Cited by 807
Zhang, Z., Pi, Z., & Liu, B. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2359372 · Cited by 759
Photoplethysmography Revisited: From Contact to Noncontact, From Point to Imaging
Sun, Y., & Thakor, N. V. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2476337 · Cited by 589
“sue bed beneath the skin, which are due to the pulsatile nature of the circulatory system [13], [14].”
Metadata
- Authors
- Kamal, A. A. R., Harness, J. B., Irving, G., & Mearns, A. J.
- Affiliations
- University of Bradford (GB)
- Corresponding author
- Harness, J.B.
- Venue
- Computer Methods and Programs in Biomedicine
- Publisher
- Elsevier BV
- Published
- 1989-04
- Volume
- 28
- Issue
- 4
- Pages
- 257-269
- ISSN
- 0169-2607 (ISSN-L)
- OpenAlex
- W2054173774
- Cited by
- 328
- References
- 56
- Citations by year
- 2026: 1 · 2025: 13 · 2024: 16 · 2023: 21 · 2022: 31 · 2021: 15 · 2020: 19 · 2019: 26 · 2018: 15 · 2017: 22 · 2016: 14 · 2015: 17 · 2014: 13 · 2013: 8 · 2012: 12
- FWCI
- 0.25
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Computer Methods and Programs in Biomedicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- photoplethysmography, skin blood flow, pulse counting, hemoglobin oxygen saturation, finger pulp, vascular bed
- MeSH terms
- Equipment Design, Humans, Monitoring, Physiologic, Plethysmography, Regional Blood Flow, Skin
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review