Skin photoplethysmography — a review
Abstract
The photoplethysmograph has been used for over 50 years but there are still misconceptions in how and what is the information obtained. A photoplethysmograph signal from any site on the skin can be separated into an oscillating (a.c.) and a steady-state (d.c.) component, their amplitudes dependent upon the structure and flow in the vascular bed. Many simple applications are available: pulse counters, using the a.c. component, skin colour and haemoglobin saturation meters, using the d.c. component. The d.c. component of the photoplethysmograph signal is a function of the blood flux beneath the device. A good emitter for use in a photoplethysmograph of skin blood flow is one in the frequency range 600-700 nm and the best signal for a.c. analysis is obtained from the finger pulp. The frequency range of the electronic circuitry should be from 0.01 to 15 Hz, then all the information in the signal can be extracted about the autonomic nervous system control of the cardiovascular system, particularly between 0.01 and 2 Hz. Comparative studies may be drawn between similar skin sites on a subject or between subjects if the afferent inputs to the brain stem are controlled or driven at a known frequency. These afferents, inputs, will modulate the efferents, outputs, which generate variations in the a.c. component of the detected photoplethysmograph signal.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Without this type of real-time, in-sensor processing (IR_Non in Fig. 6D), robust operation would be impossible: PPG signals would quickly drift outside of the narrow operating range of the ADC as a result of patient-to-patient variations in skin pigmentation and unavoidable, time-dependent fluctuations in optical scattering that result from micromotions relative to underlying blood vessels and sub”
From Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care (2019) · cites this paper as [28]
Citations
Cites
McDonald, D. A. (1963) · The American Journal of the Medical Sciences · doi:10.1097/00000441-196301000-00017 · Cited by 1,568
Analysis of Heart Rate Variability
SAYKRS, B. M. (1973) · Ergonomics · doi:10.1080/00140137308924479 · Cited by 957
In vivo evaluation of microcirculation by coherent light scattering
Stern, M. D. (1975) · Nature · doi:10.1038/254056a0 · Cited by 866
Arterial assessment by Doppler-shift ultrasound.
Gosling, R. G., & King, D. H. (1974) · PubMed · Cited by 667
LASER DOPPLER MEASUREMENT OF CUTANEOUS BLOOD FLOW
Holloway, G. A., & Watkins, D. (1977) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12507665 · Cited by 543
THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPH
Hertzman, A. B. (1938) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1938.124.2.328 · Cited by 460
Bini, G., Hagbarth, K. E., Hynninen, P., & Wallin, B. G. (1980) · The Journal of Physiology · doi:10.1113/jphysiol.1980.sp013413 · Cited by 421
A theoretical and experimental study of light absorption and scattering by in vivo skin
Dawson, J. B., Barker, D., Ellis, D. J., COTTERILL, J. A., Grassam, E., Fisher, G. W., & Feather, J. W. (1980) · Physics in Medicine and Biology · doi:10.1088/0031-9155/25/4/008 · Cited by 355
Burton, A. C. (1939) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1939.127.3.437 · Cited by 333
Sympathetic Control of Human Blood Vessels
(1953) · The Medical Journal of Australia · doi:10.5694/j.1326-5377.1953.tb84977.x · Cited by 259
Research works printed but not linked: 8
- Burton, 1940, Direct measurement of thermal conductance of skin as ideal index of peripheral blood flow, Am. J. Physiol., 129, 389
- Challoner, 1979, Photoelectric plethysmography for estimating cutaneous blood flow, Vol. 1, 127
- Fairs, 1985, Optical methods for measuring blood flow in skin, Vol. 1, 11
- Hertzman, 1937, Observations on the finger volume pulse recorded photoelectrically, Am. J. Physiol., 119, 334
- Mearns, 1985, Forcing frequency testing, theory and practice with reference to skin blood flow, Vol. 1, 23
- Molitor, 1936, A new continuous recording of peripheral circulatory changes, J. Pharmacol. Exp. Ther., 57, 6
- Sejrsen, 1971, Measurement of cutaneous bloodflow by freely diffusible radioactive isotopes, Dan. Med. Bull., 18, 9
- Thune, 1970, Plethysmographic recordings of skin pulses with particular reference to the piezoelectric method, Acta Dermatovener., 50, 27
Cited by
Sorted by citations
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967
“…processes in relation to PPG measurements (Hertzman and Randall 1948, Brown et al 1965, D’Agrosa and Hertzman 1967, Weinman 1967, Zweifler et al 1967, Challoner 1979, Ochoa and Ohara 1980, Nijboer et al 1981, Roberts 1982, Lindberg and Öberg 1993, de Trafford and Lafferty 1984, Kamal et al 1989).”
BackgroundMethodWearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912
“The principle of PPG has been reviewed previously [2–4], and is explained briefly here.”
BackgroundBinodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., et al. (2019) · Science · doi:10.1126/science.aau0780 · Cited by 807
Zhang, Z., Pi, Z., & Liu, B. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2014.2359372 · Cited by 759
Photoplethysmography Revisited: From Contact to Noncontact, From Point to Imaging
Sun, Y., & Thakor, N. V. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2476337 · Cited by 589
“sue bed beneath the skin, which are due to the pulsatile nature of the circulatory system [13], [14].”
Photoplethysmogram Analysis and Applications: An Integrative Review
Park, J., Seok, H. S., Kim, S.-S., & Shin, H. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2021.808451 · Cited by 514
“The volume of blood volume in the measurement site, arterial diameter, hemoglobin concentration, and hemoglobin direction according to the cardiac cycle are also major factors that affect the detected light intensity (De Trafford and Lafferty, 1984; Kamal et al., 1989; Lindberg and Oberg, 1993).”
BackgroundUtility of the Photoplethysmogram in Circulatory Monitoring
Reisner, A. T., Shaltis, P., McCombie, D. B., Asada, H. H., Warner, D. S., & Warner, M. A. A. (2008) · Anesthesiology · doi:10.1097/aln.0b013e31816c89e1 · Cited by 365
Contour analysis of the photoplethysmographic pulse measured at the finger
Millasseau, S., Ritter, J. M., Takazawa, K., & Chowienczyk, P. (2006) · Journal of Hypertension · doi:10.1097/01.hjh.0000239277.05068.87 · Cited by 351
Tulyakov, S., Alameda-Pineda, X., Ricci, E., Yin, L., Cohn, J. F., & Sebe, N. (2016) · IEEE Computer Society Conference on Computer Vision and Pattern Recognition/Proceedings - IEEE Computer Society Conference on Computer Vision and Pattern Recognition/Proceedings · doi:10.1109/cvpr.2016.263 · Cited by 351
Self-powered cardiovascular electronic devices and systems
Zheng, Q., Tang, Q., Wang, Z. L., & Li, Z. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-020-0426-4 · Cited by 348
Lü, G., Yang, F., Taylor, J. A., & Stein, J. F. (2009) · Journal of Medical Engineering & Technology · doi:10.3109/03091900903150998 · Cited by 352
Accurate Heart Rate Monitoring During Physical Exercises Using PPG
Temko, A. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2676243 · Cited by 301
“The PPG signals [1]–[3], [29], [30] come from PPG sensors which are embedded in these wearable devices.”
BackgroundHeart Rate Estimation From Wrist-Worn Photoplethysmography: A Review
Biswas, D., Simões‐Capela, N., Hoof, C. A. V., & Helleputte, N. V. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2914166 · Cited by 273
Current progress of photoplethysmography and SPO2 for health monitoring
TAMURA, T. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00097-w · Cited by 274
A new look at the essence of the imaging photoplethysmography
Kamshilin, A. A., Nippolainen, E., Sidorov, I. S., Васильев, П. В., Ерофеев, Н. П., Podolyan, N. P., & Владимирович, Р. Р. (2015) · Scientific Reports · doi:10.1038/srep10494 · Cited by 250
Effects of acupuncture on skin and muscle blood flow in healthy subjects
Sandberg, M., Lundeberg, T., Lindberg, L., & Gerdle, B. U. C. (2003) · European Journal of Applied Physiology · doi:10.1007/s00421-003-0825-3 · Cited by 209
“Photoplethysmography (PPG) is a non-invasive optical technique mostly used for measuring changes in skin blood flow (Challoner 1979; Kamal et al. 1989).”
MethodMaeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9505-0 · Cited by 203
“01 to 15 Hz [10, 11].”
Tanveer, M. S., & Hasan, M. K. (2019) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2019.02.028 · Cited by 200
“or reflected from the subject’s tissue [25].”
Age-related changes in peripheral pulse timing characteristics at the ears, fingers and toes
Allen, J., & Murray, A. (2002) · Journal of Human Hypertension · doi:10.1038/sj.jhh.1001478 · Cited by 179
Influence of skin type and wavelength on light wave reflectance
Fallow, B. A., Tarumi, T., & Tanaka, H. (2013) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-013-9436-7 · Cited by 162
“Normalized modulation level is calculated as the AC/DC component and represents the change in flow over the underlying constant state of flow or perfusion [1, 8].”
BackgroundMethodInfluentialMonitoring of respiratory and heart rates using a fibre-optic sensor
Lindberg, L., Ugnell, H., & Öberg, P. Å. (1992) · Medical & Biological Engineering & Computing · doi:10.1007/bf02457833 · Cited by 158
Monitoring of respiratory rate in postoperative care using a new photoplethysmographic technique
Nilsson, L. M., Johansson, A., & Kálmán, S. (2000) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1011424732717 · Cited by 157
“The intensity of the transmitted or re£ected light, detected by a photodetector, measures blood content and £ow in the target skin volume [14].”
BackgroundLi, K. H. C., White, F. A., Tipoe, T., Liu, T., Wong, M. C. S., Jesuthasan, A., Baranchuk, A. M., Tse, G., & Yan, B. P. (2018) · JMIR mhealth and uhealth · doi:10.2196/11606 · Cited by 172
Khan, E., Al Hossain, F., Uddin, S. Z., Alam, S. K., & Hasan, M. K. (2016) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2466075 · Cited by 153
Photoplethysmographic imaging of high spatial resolution
Kamshilin, A. A., Miridonov, S. V., Teplov, V., Saarenheimo, R., & Nippolainen, E. (2011) · Biomedical Optics Express · doi:10.1364/boe.2.000996 · Cited by 142
Monitoring vital signs over multiplexed radio by near-field coherent sensing
Hui, X., & Kan, E. C.-C. (2017) · Nature Electronics · doi:10.1038/s41928-017-0001-0 · Cited by 140
Monitoring of Cardiorespiratory Signal: Principles of Remote Measurements and Review of Methods
Al‐Naji, A., Gibson, K., Lee, S., & Chahl, J. (2017) · IEEE Access · doi:10.1109/access.2017.2735419 · Cited by 133
“Theoptical properties of the skin color changes due to cardiorespiratory activity can be described as photoplethysmographic (PPG) [59]–[61].”
BackgroundSensors, vision and networks: From video surveillance to activity recognition and health monitoring
Prati, A., Shan, C., & Wang, K. I. (2019) · Journal of Ambient Intelligence and Smart Environments · doi:10.3233/ais-180510 · Cited by 135
“or transmitted through, the body, which captures the volumetric flow of blood within the vessels [7,63].”
BackgroundRespiration Signals from Photoplethysmography
Nilsson, L. M. (2013) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31828098b2 · Cited by 129
Non-contact estimation of heart rate and oxygen saturation using ambient light
Bal, U. (2014) · Biomedical Optics Express · doi:10.1364/boe.6.000086 · Cited by 126
Sandberg, M., Larsson, B. K., Lindberg, L., & Gerdle, B. U. C. (2004) · European Journal of Pain · doi:10.1016/j.ejpain.2004.11.002 · Cited by 121
“Photoplethysmography (PPG) is mostly used for skin blood flow monitoring (Challoner, 1979; Kamal et al., 1989).”
MethodThomson, E. A., Nuss, K. J., Comstock, A., Reinwald, S., Blake, S., Pimentel, R. E., Tracy, B., & Li, K. (2019) · Journal of Sports Sciences · doi:10.1080/02640414.2018.1560644 · Cited by 114
Chowdhury, S. S., Hyder, R., Hafiz, M. S. B., & Haque, M. A. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2632201 · Cited by 102
Optical properties of blood in motion
Lindberg, L.-G., & Oberg, P. A. (1993) · Optical Engineering · doi:10.1117/12.60688 · Cited by 102
Reflectance Photoplethysmography as Noninvasive Monitoring of Tissue Blood Perfusion
Abay, T. Y., & Kyriacou, P. A. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2417863 · Cited by 96
Allen, J., & Murray, A. (2000) · Physiological Measurement · doi:10.1088/0967-3334/21/3/303 · Cited by 93
“The PPG signal is complex and has several components which can be classified as either low frequency (LF) or high frequency (HF) (Kamal et al 1989).”
BackgroundJohansson, A., Öberg, P. Å., & Sedin, G. (1999) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1009912831366 · Cited by 91
“The skin blood perfusion contains many time-varying components, such as the arterial pulse, vasomotion, and blood pressure- and temperature-regulatory waves [11, 16].”
BackgroundSandberg, M., Zhang, Q., STYF, J. R., Gerdle, B. U. C., & Lindberg, L. (2005) · Acta Physiologica Scandinavica · doi:10.1111/j.1365-201x.2005.01412.x · Cited by 91
“Variations in the photodetector signal are related to changes in blood flow and blood volume in the underlying tissue (Challoner 1979, Kamal et al. 1989).”
BackgroundAn Algorithm for Real-Time Pulse Waveform Segmentation and Artifact Detection in Photoplethysmograms
Fischer, C., Dömer, B., Wibmer, T., & Penzel, T. (2016) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2016.2518202 · Cited by 95
“The power spectrum of the PPG consists of frequency components up to 15 Hz with a dominant power range between 6 and 8 Hz [46], [47].”
BackgroundNilsson, L., Goscinski, T., Kalman, S., Lindberg, L. ‐., & Johansson, A. (2007) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2007.01375.x · Cited by 88
“As a result of scattering and absorption within the tissue, the intensity detected becomes sensitive to changes in tissue blood flow and volume at the measurement site (1, 2).”
BackgroundPankaj, Kumar, A., Komaragiri, R., & Kumar, M. (2021) · Archives of Computational Methods in Engineering · doi:10.1007/s11831-021-09597-4 · Cited by 84
Nabeel, P. M., Jayaraj, J., & Sivaprakasam, M. (2017) · Physiological Measurement · doi:10.1088/1361-6579/aa9550 · Cited by 80
Nilsson, L. M., Johansson, A., & Kálmán, S. (2005) · Acta Anaesthesiologica Scandinavica · doi:10.1111/j.1399-6576.2005.00721.x · Cited by 78
“This was surprising, as the PPG signal is believed to arise mainly from variations in vascular filling [1].”
BackgroundMethodSandberg, M., Sandberg, M., & Dahl, J. (2007) · Physical Therapy · doi:10.2522/ptj.20060178 · Cited by 77
Matsumura, K., Rolfe, P. J., Lee, J., & Yamakoshi, T. (2014) · PLoS ONE · doi:10.1371/journal.pone.0091205 · Cited by 72
“The frequency of 6 Hz was chosen because the main bandwidth of the PPG power spectrum is usually less than 5 Hz [26,27], including the HR signal around 1 Hz, so it is easy to separate motion artifact from other sources in the PPG power spectrum.”
MethodRajala, S., Lindholm, H., & Taipalus, T. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aac7ac · Cited by 70
“This type of phenomenon has been studied previously, e.g. by Kamal et al (1989).”
BackgroundMethodAssessment of cardiovascular function from multi-Gaussian fitting of a finger photoplethysmogram
Couceiro, R., Carvalho, P., Paiva, R. P., Henriques, J. M. O., Quintal, I., Antunes, M., Muehlsteff, J., Eickholt, C., Brinkmeyer, C., Kelm, M., et al. (2015) · Physiological Measurement · doi:10.1088/0967-3334/36/9/1801 · Cited by 70
“…components in the acquired signal, such as the interference of the power sources, are removed using a low-pass Butterworth filter with a 18 Hz cut-off frequency (Chan et al 2007), ensuring the preservation of the physiological relevant information (usually bellow 15 Hz (Kamal et al 1989)).”
MethodOptical Techniques in the Assessment of Peripheral Arterial Disease
Alnaeb, M. E., Alobaid, N., Seifalian, A. M., Mikhailidis, D. P., & Hamilton, G. (2006) · Current Vascular Pharmacology · doi:10.2174/157016107779317242 · Cited by 69
“Conversely, the DC component (commonly used for venous evaluation) of the signal varies slowly and reflects variations in the total blood volume of the examined tissue [19].”
BackgroundNew Photoplethysmographic Signal Analysis Algorithm for Arterial Stiffness Estimation
Pilt, K., Ferenets, R., Meigas, K., Lindberg, L., Temitski, K., & Viigimaa, M. (2013) · The Scientific World JOURNAL · doi:10.1155/2013/169035 · Cited by 71
“The light is emitted from the LED to the skin and a small fraction of light intensity changes is received by the PD, which are related to blood flow, blood volume, blood vessel wall movement, and the orientation of red blood cells in the underlying tissue [6, 7].”
BackgroundAl‐Qatatsheh, A., Morsi, Y., Zavabeti, A., Zolfagharian, A., Salim, N. V., Kouzani, A. Z., Mosadegh, B., & Gharaie, S. (2020) · Sensors · doi:10.3390/s20164484 · Cited by 71
Aldughayfiq, B., Ashfaq, F., Jhanjhi, N. Z., & Humayun, M. (2023) · Diagnostics · doi:10.3390/diagnostics13142442 · Cited by 75
“PPG signals reflect changes in blood flow and can provide valuable information about the cardiovascular system, including the presence of irregular heart rhythms like AF [40].”
BackgroundLimitations of Oximetry to Measure Heart Rate Variability Measures
Lü, G., & Yang, F. (2009) · Cardiovascular Engineering An International Journal · doi:10.1007/s10558-009-9082-3 · Cited by 65
Yokus, M. A., & Daniele, M. A.-A. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113249 · Cited by 62
Variability of Microcirculation Detected by Blood Pulsation Imaging
Kamshilin, A. A., Teplov, V., Nippolainen, E., Miridonov, S. V., & Giniatullin, R. A. (2013) · PLoS ONE · doi:10.1371/journal.pone.0057117 · Cited by 57
“The ratio of the AC-component to the mean level of the back-reflected light intensity represents the blood pulsation amplitude [16,17].”
BackgroundNon-invasive continuous estimation of blood flow changes in human patellar bone
Näslund, J., Pettersson, J., Lundeberg, T., Linnarsson, D., & Lindberg, L. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0070-0 · Cited by 57
“It has mostly been suggested that the AC component is related to pulsatile volume changes due to varying lumen of the vessel [6, 16, 23, 24].”
BackgroundMethodBlood flow measurements at different depths using photoplethysmography and laser Doppler techniques
Bergstrand, S., Lindberg, L. G., Ek, A., Lindén, M., & Lindgren, M. (2009) · Skin Research and Technology · doi:10.1111/j.1600-0846.2008.00337.x · Cited by 57
“Like LDF, PPG is an optical, non-invasive technique for measuring blood flow in tissue (20).”
MethodEffects of filtering on multi-site photoplethysmography pulse waveform characteristics
Allen, J., & Murray, A. (2005) · Computers in cardiology · doi:10.1109/cic.2004.1442980 · Cited by 56
“In this study pulses were measured from the right ear, thumb and great toe of 10 healthy adult subjects.”
MethodPee, B. V., Massie, F., Vits, S., Dreesen, P., Klerkx, S., Bijwadia, J. S., Verbraecken, J. A., & Bergmann, J. (2022) · SLEEP · doi:10.1093/sleep/zsac028 · Cited by 57
“In 1937, Hertzman published a paper titled “Photoelectric plethysmography of the Fingers and Toes in Man” [2], which would later be credited as the founding paper for the research into photoplethysmography (PPG) [3].”
BackgroundInfluentialXintarakou, A., Sousonis, V., Asvestas, D., Vardas, P. E., & Tzeis, S. E. (2022) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2022.853614 · Cited by 58
Facial pain increases nausea and headache during motion sickness in migraine sufferers
Drummond, P. D. (2003) · Brain · doi:10.1093/brain/awh061 · Cited by 50
“The pulsatile component of the photoplethysmograph signal re¯ects changes in the calibre of cutaneous arterioles and thus mirrors changes in cutaneous blood ¯ow (Hertzman et al., 1946; Drummond and Lance, 1981; Kamal et al., 1989).”
BackgroundEvaluation of the signal quality of wrist-based photoplethysmography
Pradhan, N., Rajan, S., & Adler, A. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab225a · Cited by 50
“Unfortunately, the PPG signal is highly sensitive to noise-corruption, especially due to motion (Allen 2007, Kamal et al 1989, Abdallah and Bolz 2011).”
BackgroundZanelli, S., Ammi, M., Hallab, M., & El‐Yacoubi, M. A. (2022) · Sensors · doi:10.3390/s22134890 · Cited by 57
“The light emitted from the diode is reflected (or transmitted) through the tissues and the arteries and arterioles and then detected by the photodetector [40].”
BackgroundDesign and Development of a Modular, Multichannel Photoplethysmography System
Budidha, K., Rybynok, V., & Kyriacou, P. A. (2018) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2018.2810643 · Cited by 49
“and a pulsatile component (AC), which is related to changing blood volume [2].”
BackgroundCutting-edge nano-LED technology
Mikulics, M., Mayer, J., & Hardtdegen, H. H. (2022) · Journal of Applied Physics · doi:10.1063/5.0087279 · Cited by 47
An optical device to measure blood components by a photoplethysmographic method
Kraitl, J., Ewald, H., & Gehring, H. (2005) · Journal of Optics A Pure and Applied Optics · doi:10.1088/1464-4258/7/6/010 · Cited by 46
Cooke, E. D., Steinberg, M. D., Pearson, R. M., Fleming, C., Sl, T., & Elusade, J. A. (1993) · Journal of the Royal Society of Medicine · doi:10.1177/014107689308601205 · Cited by 46
Zhao, D., Sun, Y., Wan, S., & Feng, W. (2016) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2016.12.005 · Cited by 45
Advanced thermal sensing techniques for characterizing the physical properties of skin
Madhvapathy, S. R., Arafa, H., Patel, M. J., Winograd, J., Kong, J., Zhu, J., Xu, S., & Rogers, J. A. (2022) · Applied Physics Reviews · doi:10.1063/5.0095157 · Cited by 48
Monitoring heart and respiratory rates at radial artery based on PPG
Wang, C., Li, Z., & Wei, X. (2013) · Optik · doi:10.1016/j.ijleo.2012.11.044 · Cited by 44
Ye, Y., He, W., Cheng, Y., Huang, W., & Zhang, Z. (2017) · Sensors · doi:10.3390/s17020385 · Cited by 43
Is Continuous Heart Rate Monitoring of Livestock a Dream or Is It Realistic? A Review
Nie, L., Berckmans, D., Wang, C., & Li, B. (2020) · Sensors · doi:10.3390/s20082291 · Cited by 46
“Its principle has been reviewed previously [27].”
Decreased Pulsatile Blood Flow in the Patella in Patellofemoral Pain Syndrome
Näslund, J., Waldén, M., & Lindberg, L. (2007) · The American Journal of Sports Medicine · doi:10.1177/0363546507303115 · Cited by 41
“Subjects (n) 33 15 18 22 13 9 Age (y) 36 (10) 33 (10) 38 (10) 33 (11) 30 (9) 37 (12) Height (cm) 175 (11) 164 (5) 184 (6) 175 (6) 171 (4) 181 (4) Weight (kg) 78 (16) 64 (8) 90 (12) 72 (10) 66 (7) 81 (8) Body mass index 26 (4) 24 (4) 27 (3) 23 (3) 23 (2) 25 (4)”
Heuristic algorithm for photoplethysmographic heart rate tracking during maximal exercise test
Silva, S. M. L. (2011) · Journal of Medical and Biological Engineering · doi:10.5405/jmbe.898 · Cited by 41
“es commonly encountered in clinical environments and during controlled or moderate motion [1-8].”
BackgroundBeurton, A., Gavelli, F., Teboul, J., Vita, N. D., & Monnet, X. (2020) · Critical Care Medicine · doi:10.1097/ccm.0000000000004768 · Cited by 40
“Provided that the degree of vasoconstriction is constant, PI may estimate stroke volume and cardiac output (17, 23).”
BackgroundQuality Assessment for the Photoplethysmogram (PPG)
Orphanidou, C. (2017) · SpringerBriefs in bioengineering · doi:10.1007/978-3-319-68415-4_3 · Cited by 39
Novel Signal Noise Reduction Method through Cluster Analysis, Applied to Photoplethysmography
Waugh, W., Allen, J., Wightman, J., Sims, A. J., & Beale, T. A. W. (2018) · Computational and Mathematical Methods in Medicine · doi:10.1155/2018/6812404 · Cited by 38
Subaşı, A. (2019) · Elsevier eBooks · doi:10.1016/b978-0-12-817444-9.00002-7 · Cited by 37
A Proximity Coupling RF Sensor for Wrist Pulse Detection Based on Injection-Locked PLL
Kim, B., Hong, Y., An, Y., Kim, S., Lee, H., Kim, S.-W., Hong, S.-B., Yun, G., & Yook, J. (2016) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2016.2549531 · Cited by 37
“Another method involves the use of a photoplethysmography (PPG) sensor [11]–[15]; until recently, this was one of the most actively studied methods for tracking physiological information.”
MethodLin, Q., Song, S., Wegberg, R. V., Sijbers, W., Biswas, D., Konijnenburg, M., Vanhoof, C., Tavernier, F., & van Helleputte, N. (2021) · IEEE Transactions on Biomedical Circuits and Systems · doi:10.1109/tbcas.2021.3130470 · Cited by 30
“Such an AC component represents microvascular blood volume change within each heartbeat [7].”
BackgroundAndreozzi, E., Sabbadini, R., Centracchio, J., Bifulco, P., Irace, A., Breglio, G., & Riccio, M. (2022) · Sensors · doi:10.3390/s22197566 · Cited by 38
Chiang, P.-Y., Chao, P. C.-P., Tarng, D., & Yang, C. (2017) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2017.2701780 · Cited by 35
Prabhakar, S. K., Rajaguru, H., & Lee, S. (2019) · IEEE Access · doi:10.1109/access.2019.2950220 · Cited by 35
Technologies for non-invasive physiological sensing: Status, challenges, and future horizons
Yu, Y., Jain, B., Anand, G., Heidarian, M., Lowe, A., & Kalra, A. M. (2023) · Biosensors and Bioelectronics X · doi:10.1016/j.biosx.2023.100420 · Cited by 38
Remote Photoplethysmography for Heart Rate and Blood Oxygenation Measurement: A Review
Pirzada, P., Wilde, A., & Harris‐Birtill, D. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3405414 · Cited by 38
“These vitals can also be monitored with contact-based pulse oximetry [23], [24], [25], discussed next.”
A Method for Respiration Rate Detection in Wrist PPG Signal Using Holo-Hilbert Spectrum
Chang, H., Hsu, C.-C., Chen, C.-Y., Lee, W., Hsu, H.-T., Shyu, K., Yeh, J., Lin, P.-J., & Lee, P. (2018) · IEEE Sensors Journal · doi:10.1109/jsen.2018.2855974 · Cited by 35
Experimental evaluation of two new sensors for respiratory rate monitoring
Vegfors, M., Ugnell, H., Hök, B., Öberg, P. Å., & Lennmarken, C. (1993) · Physiological Measurement · doi:10.1088/0967-3334/14/2/008 · Cited by 34
Estimation of phase distortions of the photoplethysmographic signal in digital IIR filtering
Lapitan, D. G., Рогаткин, Д. А., Molchanova, E. A., & Tarasov, A. P. (2024) · Scientific Reports · doi:10.1038/s41598-024-57297-3 · Cited by 34
Zhong, Y., Koklu, A., Villalva, D. R., Zhang, Y., Hernandez, L. H., Moser, M., Hallani, R. K., McCulloch, I. A., Baran, D., & Inal, S. (2022) · Advanced Functional Materials · doi:10.1002/adfm.202211479 · Cited by 33
Arildsson, M. L., Asker, C. L., Salerud, E. G., & Tomas Strömberg, A. (2000) · Microvascular Research · doi:10.1006/mvre.1999.2206 · Cited by 32
Non-contact Reflection Photoplethysmography towards Effective Human Physiological Monitoring
Shi, P. A., Azorin-Peris, V., Echiadis, A., Zheng, J., Zhu, Y., Cheang, P.-Y., & Hu, S. (2010) · Loughborough University Institutional Repository (Loughborough University) · Cited by 32
A New Probe for Ankle Systolic Pressure Measurement Using Photoplethysmography (PPG)
J�nsson, B., Laurent, C. M., Skau, T., & Lindberg, L. (2005) · Annals of Biomedical Engineering · doi:10.1007/s10439-005-8982-7 · Cited by 31
Ambiguity of mapping the relative phase of blood pulsations
Teplov, V., Nippolainen, E., Makarenko, A. A., Giniatullin, R. A., & Kamshilin, A. A. (2014) · Biomedical Optics Express · doi:10.1364/boe.5.003123 · Cited by 30
Non-invasive Measurement of Blood Components
Kraitl, J., Klinger, D., Fricke, D., Timm, U., & Ewald, H. (2012) · Smart sensors, measurement and instrumentation · doi:10.1007/978-3-642-32180-1_14 · Cited by 30
Advances in PPG Sensors Data Acquisition With Light-to-Digital Converters: A Review
Chettri, N., Aprile, A., Bonizzoni, E., & Malcovati, P. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3420170 · Cited by 29
“…variability of a PPG signal: temperature fluctuations [29], LED placement (reflection mode or transmission mode) [30], [31], skin pigmentation and composition, arterial diameter, concentration and direction of oxygenated and deoxygenated hemoglobin [32], [33], [34], and site of measurement [35].”
BackgroundUsing imaging photoplethysmography for heart rate estimation in non-human primates
Михайлович, У. А., Moeller, S., Kagan, I., Gail, A., Treue, S., & Wolf, F. (2018) · PLoS ONE · doi:10.1371/journal.pone.0202581 · Cited by 29
Islam, M. S., Rabbi, M. S.-E., Dobaie, A. M., & Hasan, M. K. (2017) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2017.05.010 · Cited by 28
Kumaran, B., & Watson, T. (2018) · Electromagnetic Biology and Medicine · doi:10.1080/15368378.2017.1422260 · Cited by 28
Photoplethysmography for non-invasivein vivomeasurement of bone hemodynamics
Mateus, J., & Hargens, A. R. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/6/1027 · Cited by 28
LED Based Sensor System for Non-Invasive Measurement of the Hemoglobin Concentration in Human Blood
Timm, U., Lewis, E., McGrath, D., Kraitl, J., & Ewald, H. (2009) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-540-92841-6_203 · Cited by 28
“[5] According to the assumption that red blood cells are mainly filled with water, the absorption coefficient of blood is similar to a solution consisting HHb, HbO2 and water (H2O) and the absorption of HHb and Hb02 is indistinguishable to the absorption of H2O above 1200nm, it is necessary to select a wavelength valu…”
BackgroundDevelopment of a new analgesic index using nasal photoplethysmography
Choi, B., Park, C., Lee, Y. H., Shin, H., Lee, S. H., Jeong, S., Noh, G., & Lee, B. (2018) · Anaesthesia · doi:10.1111/anae.14327 · Cited by 27
“The AC component reflects variations in the blood volume below the sensor; the DC component represents the optical character of the underlying tissue and venous blood content [28, 29].”
Background
Metadata
- Authors
- Kamal, A. A. R., Harness, J. B., Irving, G., & Mearns, A. J.
- Affiliations
- University of Bradford (GB)
- Corresponding author
- Harness, J.B.
- Venue
- Computer Methods and Programs in Biomedicine
- Publisher
- Elsevier BV
- Published
- 1989-04
- Volume
- 28
- Issue
- 4
- Pages
- 257-269
- ISSN
- 0169-2607 (ISSN-L)
- OpenAlex
- W2054173774
- Cited by
- 328
- References
- 56
- Citations by year
- 2026: 1 · 2025: 13 · 2024: 16 · 2023: 21 · 2022: 31 · 2021: 15 · 2020: 19 · 2019: 26 · 2018: 15 · 2017: 22 · 2016: 14 · 2015: 17 · 2014: 13 · 2013: 8 · 2012: 12
- FWCI
- 0.25
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Computer Methods and Programs in Biomedicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- photoplethysmography, skin blood flow, pulse counting, hemoglobin oxygen saturation, finger pulp, vascular bed
- MeSH terms
- Equipment Design, Humans, Monitoring, Physiologic, Plethysmography, Regional Blood Flow, Skin
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review