Preliminary results of an absorbable interference screw
The DOI matches its Crossref record; 22 of the 22 references it cites were found in a registry or an index.
Abstract
A randomized, prospective multicenter comparison was done of a bioabsorbable interference screw (Bioscrew; Linvatec Corp, Largo, FL) made from poly L‐lactic acid and a metal interference screw produced by the same company. Assignment was randomized by sealed envelopes. A total of 110 patients underwent arthroscopic patellar tendon autografts. A minimum 12 months follow‐up is available on 85 patients (mean 19 months, range 12 to 33) including 42 with Bioscrews and 43 with metal screws. There were 56 male and 29 female patients. The average age was 29 years (16 to 50 years). Tourniquet times and associated surgical findings were similar for the two groups. Postoperative Tegner and Lysholm scores were not statistically different between the two groups. KT tests at 1 year showed an average 20‐lb laxity of 1.8 mm for the Bioscrew and 1.2 mm for the metal screw groups. The average 1‐year KT maximum manual side‐to‐side difference was 1.6 mm for Bioscrews and 1.6 mm for metal screws. A pivot shift was absent in 83% of Bioscrews and 90% of metal screws at follow‐up. Six of 85 Bioscrews inserted (7%) broke on insertion (all were 7‐mm diameter screws at the femoral site). No additional fixation was required in four cases. In two, the broken screw was replaced. No lytic bone changes or tunnel widening were found with any Bioscrew. One metal screw had tibial tunnel widening. No statistical difference was found between the Bioscrew and the metal screw groups. Short‐term data support the conclusion that the Bioscrew is a reasonable alternative to metal interference screws.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Citations
Cites
Rating systems in the evaluation of knee ligament injuries
TEGNER, Y., & LYSHOLM, J. (1985) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-198509000-00007 · Cited by 4,613
Evaluation of knee ligament surgery results with special emphasis on use of a scoring scale
Lysholm, J., & Gillquist, J. (1982) · The American Journal of Sports Medicine · doi:10.1177/036354658201000306 · Cited by 2,775
Kurosaka, M., Yoshiya, S., & Andrish, J. T. (1987) · The American Journal of Sports Medicine · doi:10.1177/036354658701500306 · Cited by 745
Clancy, W. G., Nelson, D. A., Reider, B., & Narechania, R. G. (1982) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-198264030-00004 · Cited by 581
OʼBrien, S. J., Warren, R. F., Pavlov, H., Panariello, R., & Wickiewicz, T. L. (1991) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199173020-00016 · Cited by 501
Biodegradable bone repair materials. Synthetic polymers and ceramics.
Hollinger, J. O., & Battistone, G. C. (1986) · PubMed · Cited by 486
Absorbable implants for the fixation of fractures.
Böstman, O. M. (1991) · The Journal of Bone & Joint Surgery · doi:10.2106/00004623-199173010-00022 · Cited by 459
Biodegradable Bone Repair Materials: Synthetic Polymers and Ceramics
HOLLINGER, J. O., & BATTISTONE, G. C. (1986) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-198606000-00046 · Cited by 381
Osteolytic changes accompanying degradation of absorbable fracture fixation implants
Bostman, O. (1991) · The Journal of Bone and Joint Surgery. British volume · doi:10.1302/0301-620x.73b4.1649195 · Cited by 294
Knee rehabilitation after anterior cruciate ligament reconstruction and repair
Paulos, L., Noyes, F. R., Grood, E., & Butler, D. L. (1981) · The American Journal of Sports Medicine · doi:10.1177/036354658100900303 · Cited by 270
Self-citation
Research works printed but not linked: 6
- Brostrom L, 1968, Behandling av invetererad ruptur av framre korsbandet, Lakartidningen, 64, 4477
- Alm A, 1974, Reconstruction of the anterior cruciate ligament by using the medial third of the patellar ligament, Acta Chir Scand, 140, 289
- James P, 1989, Stable fixation of bone‐tendon‐bone grafts in ACL reconstruction, Am J Knee Surg, 2, 71
- Bach BR, 1989, Potential pitfalls of Kurosaka screw interference fixation for ACL surgery, Am J Knee Surg, 2, 76
- FChabot PChristel MVert.In vivo tests of bioresorbable bone plates in long lasting poly (L‐lactic acid)UPAC Symposium Prague Czechoslovakia Ed 1 Polymers in Medicine and Biology July 9–12 1987
- BFElrod.Arthroscopic shoulder stabilization with a bioabsorbable takPresented at the Arthroscopy Association of North America Annual Meeting. April 1–4 1993 · doi:10.1097/00132585-199300140-00005
Cited by
Farah, S., Anderson, D. G., & Langer, R. (2016) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2016.06.012 · Cited by 3,443
Synthetic biodegradable polymers as orthopedic devices
Middleton, J. C., & Tipton, A. J. (2000) · Biomaterials · doi:10.1016/s0142-9612(00)00101-0 · Cited by 2,741
“Barber [36] evaluated 85 patients in two groups that received either a metal or LPLA interference screw.”
BackgroundProperties and medical applications of polylactic acid: A review
Hamad, K., Kaseem, M., Yang, H. W., Deri, F., & Ko, Y. G. (2015) · Express Polymer Letters · doi:10.3144/expresspolymlett.2015.42 · Cited by 722
Clinical biocompatibility of biodegradable orthopaedic implants for internal fixation: a review
Böstman, O., & Pihlajamäki, H. (2000) · Biomaterials · doi:10.1016/s0142-9612(00)00129-0 · Cited by 509
Adverse Tissue Reactions to Bioabsorbable Fixation Devices
B??stman, O. M., & Pihlajam??ki, H. K. (2000) · Clinical Orthopaedics and Related Research · doi:10.1097/00003086-200002000-00026 · Cited by 476
Metadata
- Authors
- Barber, F. A., Elrod, B. F., McGuire, D. A., & Paulos, L. E.
- Affiliations
- Plano Orthopedic Sports Medicine & Spine Center (US)
- Southern Sports Medicine And Orthopaedic Center Nashville Tennessee U.S.A.
- Knee and Arthroscopic Surgery Anchorage Alaska U.S.A.
- Orthopedic Specialty Hospital (US)
- Corresponding author
- Barber, F.Alan
- Venue
- Arthroscopy
- Publisher
- Wiley
- Published
- 1995-10
- Volume
- 11
- Issue
- 5
- Pages
- 537-548
- ISSN
- 0749-8063 (ISSN-L)
- 1526-3231
- OpenAlex
- W1968503265
- Cited by
- 178
- References
- 22
- Citations by year
- 2026: 1 · 2025: 1 · 2024: 3 · 2023: 1 · 2022: 1 · 2021: 1 · 2020: 2 · 2019: 5 · 2018: 2 · 2017: 3 · 2016: 8 · 2015: 4 · 2014: 4 · 2013: 11 · 2012: 5
- FWCI
- 8.18
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- onlinelibrary.wiley.com/termsAndConditions#vor (vor)
- doi.wiley.com/10.1002/tdm_license_1.1 (tdm)
- Locations
- Arthroscopy The Journal of Arthroscopic and Related Surgery (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Surgery › Knee injuries and reconstruction techniques
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Bone fractures and treatments
- Health Sciences › Medicine › Orthopedics and Sports Medicine › Foot and Ankle Surgery
- Keywords
- bioabsorbable interference screw, poly(L-lactic acid), patellar tendon autograft, knee arthroscopy, screw fixation, pivot shift, knee laxity, KT-1000 arthrometry, screw breakage, tunnel widening
- MeSH terms
- Absorption, Adolescent, Adult, Arthroplasty, Bone Screws, Female, Follow-Up Studies, Humans, Knee Joint, Lactates, Male, Metals, Middle Aged, Polyesters, Polymers, Prospective Studies, Radiography, Tendons, Transplantation, Autologous, Anterior Cruciate Ligament, Lactic Acid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Preliminary results of an absorbable interference screw. https://app.cochat.ai/papers/10.1016/0749-8063(95)90129-9
- The paper
- Barber, F. A., Elrod, B. F., McGuire, D. A., & Paulos, L. E. (1995). Preliminary results of an absorbable interference screw. Arthroscopy. https://doi.org/10.1016/0749-8063(95)90129-9
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.