Chemical modification of poly(vinyl chloride) resin using poly(ethylene glycol) to improve blood compatibility
Abstract
Poly(vinyl chloride) (PVC) was aminated by treating the resin with a concentrated aqueous solution of ethylenediamine. The aminated PVC was then reacted with hexamethylene diisocyanate to incorporate the isocyanate group onto the polymer backbone. The isocyanated PVC was further reacted with poly(ethylene glycol) (PEG) of molecular weight 600 Da. The modified polymer was characterized using infrared and X-ray photoelectron spectroscopy (XPS) and thermal analysis. Infrared and XPS spectra showed the incorporation of PEG onto PVC. The thermal stability of the modified polymer was found to be lowered by the incorporation of PEG. Contact angle measurements on the surface of polymer films cast from a tetrahydrofuran solution of the polymer demonstrated that the modified polymer gave rise to a significantly hydrophilic surface compared to unmodified PVC. The solid/water interfacial free energy of the modified surface was 3.9 ergs/cm(2) as opposed to 18.4 ergs/cm(2) for bare PVC surface. Static platelet adhesion studies using platelet-rich plasma showed significantly reduced platelet adhesion on the surface of the modified polymer compared to control PVC. The surface hydrophilicity of the films was remarkably retained even in the presence of up to 30 wt% concentration of the plasticizer di-(2-ethylhexyl phthalate). The study showed that bulk modification of PVC with PEG using appropriate chemistry can give rise to a polymer that possesses the anti-fouling property of PEG and such bulk modifications are less cumbersome compared to surface modifications on the finished product to impart anti-fouling properties to the PVC surface.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Balakrishnan et al. showed that bulk modification of PVC resin with PEG 600 greatly reduced solid/water interfacial free energy and platelet adhesion in in vitro PRP studies.
“Balakrishnan et al.47 later demonstrated success using bulk modification of PVC resin with PEG 600 that yielded greatly reduced solid/water interfacial free energy and platelet adhesion in in vitro PRP studies.”
From Recent advances in thromboresistant and antimicrobial polymers for biomedical applications: just say yes to nitric oxide (NO) (2016) · cites this paper as [47]
Cites
Poly(ethylene glycol) chemistry : biotechnical and biomedical applications
Harris, J. M. (1992) · Plenum Press eBooks · Cited by 1,426 · see citing works
Larm, O., Larsson, R., & Olsson, P. (1983) · Biomaterials Medical Devices and Artificial Organs · doi:10.3109/10731198309118804 · Cited by 438 · see citing works
Surface-immobilized polyethylene oxide for bacterial repellence
Desai, N. P., Hossainy, S., & Hubbell, J. A. (1992) · Biomaterials · doi:10.1016/0142-9612(92)90160-p · Cited by 195 · see citing works
Andrade, J. D., King, R. N., GREGONIS, D. E., et al. (1979) · Journal of Polymer Science Polymer Symposia · doi:10.1002/polc.5070660130 · Cited by 149 · see citing works
Polymers of Biological and Biomedical Significance
Shalaby, S. W., Ikada2, Y., Langer3, R., et al. (1993) · ACS symposium series · doi:10.1021/bk-1994-0540 · Cited by 119 · see citing works
Potential Hazards of Exposure to Di-(2-Ethylhexyl)-Phthalate in Babies
Latini, G. (2000) · Neonatology · doi:10.1159/000014278 · Cited by 99 · see citing works
Lakshmi, & Jayakrishnan (1998) · Artificial Organs · doi:10.1046/j.1525-1594.1998.06124.x · Cited by 85 · see citing works
Carraher1, C. E., & Tsuda2, M. (1980) · ACS symposium series · doi:10.1021/bk-1980-0121 · Cited by 66 · see citing works
Park, J. H., Cho, Y. W., Kwon, I. C., Jeong, S. Y., & Bae, Y. H. (2002) · Biomaterials · doi:10.1016/s0142-9612(02)00144-8 · Cited by 61 · see citing works
Jones, D. S., McGovern, J. G., Woolfson, A. D., & Gorman, S. P. (1997) · Biomaterials · doi:10.1016/s0142-9612(96)00170-6 · Cited by 51 · see citing works
Moore, J. (1973) · Springer eBooks · doi:10.1007/978-94-010-2678-9 · Cited by 45 · see citing works
Immobile plasticizer in flexible PVC
Jayakrishnan, A., & Lakshmi, S. (1998) · Nature · doi:10.1038/25275 · Cited by 44 · see citing works
Self-citationRadiation grafting of hydrophilic monomers on to plasticized poly (vinyl chloride) sheets
Krishnan, V. K., Jayakrishnan, A., & Francis, J. D. (1991) · Biomaterials · doi:10.1016/0142-9612(91)90147-3 · Cited by 41 · see citing works
Self-citationIshikawa, Y., Honda, K., Sasakawa, S., et al. (1983) · Vox Sanguinis · doi:10.1111/j.1423-0410.1983.tb04125.x · Cited by 40 · see citing works
Surface studies of albumin immobilized onto PE and PVC films
de Queiroz, A. A. A., Barrak, É. R., Gil, H. A. C., et al. (1997) · Journal of Biomaterials Science Polymer Edition · doi:10.1163/156856297x00489 · Cited by 39 · see citing works
Ferruti, P., Barbucci, R., Danzo, N., et al. (1982) · Biomaterials · doi:10.1016/0142-9612(82)90058-8 · Cited by 34 · see citing works
Surface-grafted heparinizable materials
Barbucci, R., Casini, G., Ferruti, P., et al. (1985) · Polymer · doi:10.1016/0032-3861(85)90310-6 · Cited by 32 · see citing works
Yu, J., Johansson, S., & Ljungh, Å. (1997) · Biomaterials · doi:10.1016/s0142-9612(96)00154-8 · Cited by 31 · see citing works
Biomaterials for Blood Tubing: The Application of Plasticised Poly(Vinyl Chloride)
Blass, C. R., Jones, C., & Courtney, J. M. (1992) · The International Journal of Artificial Organs · doi:10.1177/039139889201500402 · Cited by 28 · see citing works
Luca, C., Drǎgan, S., Bărboiu, V., et al. (1980) · Journal of Polymer Science Polymer Chemistry Edition · doi:10.1002/pol.1980.170180205 · Cited by 15 · see citing works
Studies of fibrinogen measurement in the CAP survey program.
Koepke, J. A., Pr, G., Filip, D. J., et al. (1975) · PubMed · Cited by 15 · see citing works
Drǎgan, S., Luca, C., Petrariu, I., et al. (1980) · Journal of Polymer Science Polymer Chemistry Edition · doi:10.1002/pol.1980.170180206 · Cited by 12 · see citing works
Thrombus Formation on Surfaces in Contact with Blood
Schultz, J. S., Lindenauer, S. M., & Penner, J. A. (1982) · Advances in chemistry series · doi:10.1021/ba-1982-0199.ch005 · Cited by 12 · see citing works
Drăgan, S., Petrartu, I., & Dima, M. O. (1980) · Journal of Polymer Science Polymer Chemistry Edition · doi:10.1002/pol.1980.170180726 · Cited by 7 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Poly(ethylene glycol) in Drug Delivery: Pros and Cons as Well as Potential Alternatives
Knop, K., Hoogenboom, R., Fischer, D., & Schubert, U. S. (2010) · Angewandte Chemie International Edition · doi:10.1002/anie.200902672 · Cited by 3,533 · see citing works
Immunological properties of engineered nanomaterials
Dobrovolskaia, M. A., & McNeil, S. E. (2007) · Nature Nanotechnology · doi:10.1038/nnano.2007.223 · Cited by 1,823 · see citing works
Wo, Y., Brisbois, E. J., Bartlett, R. H., & Meyerhoff, M. E. (2016) · Biomaterials Science · doi:10.1039/c6bm00271d · Cited by 251 · see citing works
Zhou, Z., & Meyerhoff, M. E. (2005) · Biomaterials · doi:10.1016/j.biomaterials.2005.04.046 · Cited by 115 · see citing works
Major, T. C., Brisbois, E. J., Jones, A. M., Zanetti, M. E., Annich, G. M., Bartlett, R. H., & Handa, H. (2014) · Biomaterials · doi:10.1016/j.biomaterials.2014.05.036 · Cited by 83 · see citing works
Effect of liquid plasticizers on crystallization of PCL in soft PVC/PCL/plasticizer blends
Xu, Y., Xiong, Y., & Guo, S. (2020) · Journal of Applied Polymer Science · doi:10.1002/app.48803 · Cited by 31 · see citing works
- Authors
- Balakrishnan, B., Kumar, D. S., Yoshida, Y., & Jayakrishnan, A.
- Affiliations
- Sree Chitra Thirunal Institute for Medical Sciences and Technology (IN)
- Toyo University (JP)
- Corresponding author
- Jayakrishnan, A.
- Venue
- Biomaterials
- Publisher
- Elsevier BV
- Published
- 2005-06
- Volume
- 26
- Issue
- 17
- Pages
- 3495-3502
- ISSN
- 0142-9612 (ISSN-L)
- OpenAlex
- W2026294008
- Cited by
- 157
- References
- 24
- Citations by year
- 2026: 3 · 2025: 4 · 2024: 3 · 2023: 5 · 2022: 2 · 2021: 4 · 2020: 11 · 2019: 6 · 2018: 5 · 2017: 6 · 2016: 14 · 2015: 15 · 2014: 9 · 2013: 7 · 2012: 6
- FWCI
- 3.56
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Biomaterials (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Materials Science › Surfaces, Coatings and Films › Polymer Surface Interaction Studies
- Physical Sciences › Materials Science › Polymers and Plastics › Polymer Science and PVC
- Physical Sciences › Materials Science › Biomaterials › biodegradable polymer synthesis and properties
- Keywords
- polyvinyl chloride, polyethylene glycol, hemocompatibility, amination, hexamethylene diisocyanate, surface modification, hydrophilicity, platelet adhesion, antifouling, DEHP, plasticizers, contact angle
- MeSH terms
- Biocompatible Materials, Blood, Blood Platelets, Cells, Cultured, Composite Resins, Humans, Materials Testing, Molecular Weight, Platelet Adhesiveness, Polyethylene Glycols, Polyvinyl Chloride, Surface Properties, Hydrophobic and Hydrophilic Interactions
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review