Simplified size adjusted dose reference levels for adult CT examinations: A regional study
Abstract
Purpose To investigate retrospective classification of adult patients into small, average, and large based on effective diameter (EDia) from localizer image of computed tomography (CT) scans and to develop regional diagnostic reference levels (DRLs) and achievable doses (AD).Method The patients falling within the mean ± standard deviation (SD) of EDia were classified as average; those below this range as small and above as large. The CTDI vol,dose-length-product (DLP) and size-specific dose estimates (SSDE) of all adult patients undergoing CT examinations in 8 CT facilities for 11 months (Dec. 2019 - Oct. 2020) were evaluated. The 75th and 50th percentile values were compared with national and international values.Results Of the total of 69,434 CT examinations, nearly 80% fell within average size. The 75th percentile values of CTDI vol and DLP for small patients for abdomen-pelvic exams were nearly half of average sized patients. Similarly, the 75th percentile values for large patients were nearly double. Similar findings were not found for chest exams. Analysis of image quality and dose factors such as noise, mean axial length, slice thickness, mean number of sequences, use of iterative reconstruction and tube current modulation (TCM) resulted in identification of opportunities for improvement and optimization of different CT facilities.Conclusions DRLs for adult patients were found to vary widely with patient size and thus establishing DRLs only for standard sized patient is not adequate. Simplified and intuitive methods for size classification was shown to provide meaningful information for optimization for patients outside the standard size adult.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In CT, the predominant DRL metrics are CTDIvol (mGy) and DLP (mGy·cm).
“In CT, the two primary metrics recommended for DRLs are the volume-weighted computed tomography dose index (CTDIvol) and the dose-length product (DLP).85”
From Computed Tomography Diagnostic Reference Levels Worldwide and in Jordan: Implications for Patient Dose Reduction (2026) · cites this paper as [85]
Cites
ICRP Publication 135: Diagnostic Reference Levels in Medical Imaging
Vañó, E., Miller, D. L., Martin, C. J., Rehani, M. M., Kang, K., Rosenstein, M., Ortiz-López, P., Mattsson, S., Padovani, R., & Rogers, A. (2017) · Annals of the ICRP · doi:10.1177/0146645317717209 · Cited by 1,043 · see citing works
Self-citationSodickson, A. D., Baeyens, P. F., Andriole, K. P., Prevedello, L. M. S., Nawfel, R. D., Hanson, R., & Khorasani, R. (2009) · Radiology · doi:10.1148/radiol.2511081296 · Cited by 993 · see citing works
Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations
Boone, J. M., Strauss, K. J., CODY, D. D., et al. (2011) · doi:10.37206/143 · Cited by 554 · see citing works
U.S. Diagnostic Reference Levels and Achievable Doses for 10 Adult CT Examinations
Kanal, K. M., Butler, P. F., Sengupta, D., Bhargavan, M., Coombs, L. P., & Morin, R. L. (2017) · Radiology · doi:10.1148/radiol.2017161911 · Cited by 397 · see citing works
Image quality in CT: From physical measurements to model observers
Verdun, F. R., Racine, D., Ott, J. G., et al. (2015) · Physica Medica · doi:10.1016/j.ejmp.2015.08.007 · Cited by 299 · see citing works
Recent and Upcoming Technological Developments in Computed Tomography
Lell, M., & Kachelrieß, M. (2019) · Investigative Radiology · doi:10.1097/rli.0000000000000601 · Cited by 254 · see citing works
Mettler, F. A., Mahesh, M., Bhargavan, M., Chambers, C. E., Elee, J., Frush, D. P., Miller, D. L., Royal, H. D., Milano, M. T., Spelic, D. C., et al. (2020) · Radiology · doi:10.1148/radiol.2020192256 · Cited by 236 · see citing works
Patients undergoing recurrent CT scans: assessing the magnitude
Rehani, M. M., Yang, K., Melick, E. R., Heil, J., Šalát, D., Sensakovic, W. F., & Liu, B. (2019) · European Radiology · doi:10.1007/s00330-019-06523-y · Cited by 177 · see citing works
Self-citationCancer Risks from CT Scans: Now We Have Data, What Next?
Brenner, D. J., & Hall, E. J. (2012) · Radiology · doi:10.1148/radiol.12121248 · Cited by 172 · see citing works
Tsapaki, V., Aldrich, J. E., Sharma, R., et al. (2006) · Radiology · doi:10.1148/radiol.2403050993 · Cited by 141 · see citing works
Brambilla, M., Vassileva, J., Kuchcińska, A., et al. (2019) · European Radiology · doi:10.1007/s00330-019-06528-7 · Cited by 135 · see citing works
Patient-specific Radiation Dose and Cancer Risk for Pediatric Chest CT
Li, X., Samei, E., Segars, W. P., Sturgeon, G. M., Colsher, J. G., & Frush, D. P. (2011) · Radiology · doi:10.1148/radiol.11101900 · Cited by 116 · see citing works
Establishing national diagnostic reference levels (DRLs) for computed tomography in Egypt
Salama, D. H., Vassileva, J., Mahdaly, G., Shawki, M., Salama, A., Gilley, D., & Rehani, M. M. (2017) · Physica Medica · doi:10.1016/j.ejmp.2017.05.050 · Cited by 98 · see citing works
Self-citationRehani, M. M., Melick, E. R., Alvi, R. M., et al. (2019) · European Radiology · doi:10.1007/s00330-019-06551-8 · Cited by 85 · see citing works
Self-citationRadiation risk issues in recurrent imaging
Brower, C. H., & Rehani, M. M. (2021) · British Journal of Radiology · doi:10.1259/bjr.20210389 · Cited by 84 · see citing works
Self-citationMacGregor, K., Li, I., Dowdell, T., et al. (2015) · Radiology · doi:10.1148/radiol.2015141520 · Cited by 81 · see citing works
Rehani, M. M., & Hauptmann, M. (2020) · Physica Medica · doi:10.1016/j.ejmp.2020.07.014 · Cited by 80 · see citing works
Self-citationVassileva, J., Rehani, M. M., Applegate, K. E., et al. (2012) · European Radiology · doi:10.1007/s00330-012-2639-3 · Cited by 69 · see citing works
Self-citationLimitations of diagnostic reference level (DRL) and introduction of acceptable quality dose (AQD)
Rehani, M. M. (2015) · The British Journal of Radiology · doi:10.1259/bjr.20140344 · Cited by 67 · see citing works
Self-citationA study to establish international diagnostic reference levels for paediatric computed tomography
Vassileva, J., Rehani, M. M., Kostova-Lefterova, D. Z., AlNaemi, H., AlSuwaidi, J. S., Arandjic, D., Bashier, E., Renha, S. K., El-Nachef, L., Aguilar, J. G., et al. (2015) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncv116 · Cited by 65 · see citing works
Self-citationOptimizing Radiation Doses for Computed Tomography Across Institutions
Demb, J., Chu, P., Nelson, T. R., et al. (2017) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2017.0445 · Cited by 51 · see citing works
Brat, H. G., Zanca, F., Montandon, S., et al. (2019) · European Radiology · doi:10.1007/s00330-019-06257-x · Cited by 48 · see citing works
Challenges in Radiation Protection of Patients for the 21st Century
Rehani, M. M. (2013) · American Journal of Roentgenology · doi:10.2214/ajr.12.10244 · Cited by 45 · see citing works
Self-citationRoch, P., Célier, D., Dessaud, C., et al. (2019) · European Radiology · doi:10.1007/s00330-019-06422-2 · Cited by 45 · see citing works
Tang, H., Yu, N., Jia, Y., et al. (2017) · British Journal of Radiology · doi:10.1259/bjr.20170521 · Cited by 44 · see citing works
Effect of reduction in tube current on reader confidence in paediatric computed tomography
Shah, R., Gupta, A., Rehani, M. M., Pandey, A. K., & Mukhopadhyay, S. (2005) · Clinical Radiology · doi:10.1016/j.crad.2004.08.011 · Cited by 36 · see citing works
Self-citationVassileva, J., & Rehani, M. M. (2015) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncv113 · Cited by 33 · see citing works
Self-citationUse of Multiphase CT Protocols in 18 Countries: Appropriateness and Radiation Doses
Rastogi, S., Singh, R., Borse, R., et al. (2020) · Canadian Association of Radiologists Journal · doi:10.1177/0846537119888390 · Cited by 33 · see citing works
CT is still not a low‐dose imaging modality
Rehani, M. M., Szczykutowicz, T. P., & Zaidi, H. (2019) · Medical Physics · doi:10.1002/mp.14000 · Cited by 30 · see citing works
Self-citationLi, X., Hirsch, J. A., Rehani, M. M., et al. (2019) · American Journal of Roentgenology · doi:10.2214/ajr.19.21804 · Cited by 29 · see citing works
Self-citationCT dose survey in adults: what sample size for what precision?
Taylor, S. E., Muylem, A. V., Howarth, N., et al. (2016) · European Radiology · doi:10.1007/s00330-016-4333-3 · Cited by 28 · see citing works
Patient doses in paediatric CT: feasibility of setting diagnostic reference levels
Järvinen, H., Merimaa, K., Seuri, R. M., Tyrväinen, E., Perhomaa, M., Savikurki-Heikkilä, P., Svedström, E., Žiliukas, J., & Lintrop, M. (2011) · Radiation Protection Dosimetry · doi:10.1093/rpd/ncr293 · Cited by 26 · see citing works
Bibbo, G., Brown, S. M., & Linke, R. J. (2016) · Journal of Medical Imaging and Radiation Oncology · doi:10.1111/1754-9485.12488 · Cited by 23 · see citing works
Boos, J., Thomas, C. K., Appel, E., et al. (2017) · Journal of Radiological Protection · doi:10.1088/1361-6498/aaa32c · Cited by 22 · see citing works
Is it possible to kill the radiation risk issue in computed tomography?
Kachelrieß, M., & Rehani, M. M. (2020) · Physica Medica · doi:10.1016/j.ejmp.2020.02.017 · Cited by 20 · see citing works
Self-citationMapping Institution-Specific Study Descriptions to RadLex Playbook Entries
Mabotuwana, T., Lee, M., Cohen-Solal, E. V., & Chang, P. J. (2014) · Journal of Imaging Informatics in Medicine · doi:10.1007/s10278-013-9663-y · Cited by 19 · see citing works
Appel, E., Kröpil, P., Bethge, O., et al. (2018) · Clinical Radiology · doi:10.1016/j.crad.2018.02.012 · Cited by 14 · see citing works
Diagnostic Reference Levels for Adult Patients in the United States
Butler, P. F., & Kanal, K. M. (2018) · Journal of the American College of Radiology · doi:10.1016/j.jacr.2017.12.012 · Cited by 11 · see citing works
Implementation of Size-Dependent Local Diagnostic Reference Levels for CT Angiography
Boere, H., Eijsvoogel, N. G., Sailer, A. M., Wildberger, J. E., de Haan, M. W., Das, M., & Jeukens, C. R. L. P. N. (2018) · American Journal of Roentgenology · doi:10.2214/ajr.17.18566 · Cited by 11 · see citing works
CT diagnostic reference levels: are they appropriately computed?
Vanaudenhove, T., Muylem, A. V., Howarth, N., et al. (2019) · European Radiology · doi:10.1007/s00330-019-06141-8 · Cited by 8 · see citing works
Effect of contrast agent administration on water equivalent diameter in CT
Viggiano, B. T., Rose, S., & Szczykutowicz, T. P. (2021) · Medical Physics · doi:10.1002/mp.14721 · Cited by 5 · see citing works
MacGregor, K., Concepcion, L., Dowdell, T., et al. (2020) · Canadian Association of Radiologists Journal · doi:10.1177/0846537120923020 · Cited by 3 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Relationships of radiation dose indices with body size indices in adult body computed tomography
Inoue, Y., Itoh, H., Nagahara, K., Hata, H., & Mitsui, K. (2023) · Tomography · doi:10.3390/tomography9040110 · Cited by 16 · see citing works
Oglat, A., Shdaifat, S., Sayah, M., Abukonna, A., & Khogali, W. (2026) · Drug, Healthcare and Patient Safety · doi:10.2147/dhps.s595721
“85 These parameters constitute the foundation of national and international dose audits owing to their practicality, repeatability, and accessibility from practically all contemporary CT scanners.”
- Authors
- Moghadam, N., Lecomte, R., Mercure, S., Rehani, M. M., & Nassiri, M. A.
- Affiliations
- Centre Hospitalier Universitaire de Sherbrooke (CA)
- Université de Sherbrooke (CA)
- Massachusetts General Hospital (US)
- Corresponding author
- Moghadam, Narjes
- Venue
- European Journal of Radiology
- Publisher
- Elsevier BV
- Published
- 2021-09
- Volume
- 142
- Pages
- 109861
- ISSN
- 0720-048X (ISSN-L)
- OpenAlex
- W3181651327
- Cited by
- 16
- References
- 42
- Citations by year
- 2026: 5 · 2024: 3 · 2023: 4 · 2022: 4
- FWCI
- 1.45
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- Locations
- European Journal of Radiology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Physical Sciences › Physics and Astronomy › Radiation › Advanced Radiotherapy Techniques
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Digital Radiography and Breast Imaging
- Keywords
- diagnostic reference levels, effective diameter, size-specific dose estimate, dose-length product, CT radiation dose, abdominal and pelvic CT, chest CT, iterative reconstruction, tube current modulation, dose optimization, image quality
- MeSH terms
- Abdomen, Adult, Humans, Radiation Dosage, Reference Values, Retrospective Studies, Tomography, X-Ray Computed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review