Recommendations from the Heart Rhythm Society Task Force on Device Performance Policies and Guidelines
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Two documents from the Heart Rhythm Society have addressed the management of device and lead recalls.
“Two documents from the Heart Rhythm Society have addressed the management of device 72 and lead 73 recalls.”
From Managing with pacemakers and implantable cardioverter defibrillators (2013) · cites this paper as [72]
Both documents emphasized the importance of communication, not just between industry and physicians, but between industry and patients, resulting in the implementation of direct patient communication about important device and lead performance issues.
“Both of these documents emphasized the importance of communication, not just between industry and physicians, but between industry and patients, resulting in the implementation of direct patient communication about important device and lead performance issues.”
From Managing with pacemakers and implantable cardioverter defibrillators (2013) · cites this paper as [72]
Cites
Moss, A. J., Zaręba, W., Hall, W. J., Klein, H., Wilber, D. J., Cannom, D. S., Daubert, J. P., Higgins, S. L., Brown, M. W., & Andrews, M. L. (2002) · New England Journal of Medicine · doi:10.1056/nejmoa013474 · Cited by 6,885 · see citing works
Amiodarone or an Implantable Cardioverter–Defibrillator for Congestive Heart Failure
Bardy, G. H., Lee, K. L., Mark, D. B., Poole, J. E., Packer, D. L., Boineau, R., Domanski, M., Troutman, C., Anderson, J., Johnson, G., et al. (2005) · New England Journal of Medicine · doi:10.1056/nejmoa043399 · Cited by 6,552 · see citing works
Cardiac Resynchronization in Chronic Heart Failure
Abraham, W. T., Fisher, W. G., Smith, A. L., Delurgio, D. B., Leon, A. R., Loh, E., Kocovic, D. Z., Packer, M., Clavell, A. L., Hayes, D. L., et al. (2002) · New England Journal of Medicine · doi:10.1056/nejmoa013168 · Cited by 4,648 · see citing works
Moss, A. J., Hall, W. J., Cannom, D. S., Daubert, J. P., Higgins, S. L., Klein, H., Levine, J. H., Saksena, S., Waldo, A. L., Wilber, D., et al. (1996) · New England Journal of Medicine · doi:10.1056/nejm199612263352601 · Cited by 4,165 · see citing works
CARDIAC RESYNCHRONIZATION IN CHRONIC HEART FAILURE
Braham, W. I. T. A., Isher, W. E. G. F., Mith, L. S., Elurgio, D. A. B. D., Eon, A. N. R. L., Oh, E., Ocovic, D. U. Z. K., Acker, M. G., Lavell, L. C., Ayes, D. A. L. H., et al. (2002) · Cited by 3,808 · see citing works
The AVID Investigators (1997) · New England Journal of Medicine · doi:10.1056/nejm199711273372202 · Cited by 3,221 · see citing works
A Randomized Study of the Prevention of Sudden Death in Patients with Coronary Artery Disease
Buxton, A. E., Lee, K. L., Fisher, J. D., Josephson, M. E., Prystowsky, E. N., & Hafley, G. (1999) · New England Journal of Medicine · doi:10.1056/nejm199912163412503 · Cited by 2,489 · see citing works
Self-citationProphylactic Defibrillator Implantation in Patients with Nonischemic Dilated Cardiomyopathy
Kadish, A. H., Dyer, A. R., Daubert, J. P., Quigg, R. J., Estes, N. A. M., Anderson, K. P., Calkins, H. G., Hoch, D., Goldberger, J. J., Shalaby, A. A., et al. (2004) · New England Journal of Medicine · doi:10.1056/nejmoa033088 · Cited by 2,064 · see citing works
Canadian Implantable Defibrillator Study (CIDS)
Connolly, S. J., Gent, M., Roberts, R. S., Dorian, P., Roy, D., Sheldon, R. S., Mitchell, L. B., Green, M. S., Klein, G. J., & O’Brien, B. (2000) · Circulation · doi:10.1161/01.cir.101.11.1297 · Cited by 1,601 · see citing works
Prophylactic use of an implantable cardioverter-defibrillator after acute myocardial infarction
Hohnloser, S. H., Kuck, K. H., Dorian, P., Roberts, R. S., Hampton, J. R., Hatala, R., Fain, E., Gent, M., & Connolly, S. J. (2004) · New England Journal of Medicine · doi:10.1056/nejmoa041489 · Cited by 1,519 · see citing works
Kück, K., Cappato, R., Siebels, J., & Rüppel, R. (2000) · Circulation · doi:10.1161/01.cir.102.7.748 · Cited by 1,474 · see citing works
Gregoratos, G., Abrams, J., Epstein, A. E., Freedman, R. A., Hayes, D. L., Hlatky, M. A., Kerber, R. E., Naccarelli, G. V., Schoenfeld, M. H., Silka, M. J., et al. (2002) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(02)02528-7 · Cited by 1,272 · see citing works
Self-citationMeta-analysis of the implantable cardioverter defibrillator secondary prevention trials
Connolly, S. J. (2000) · European Heart Journal · doi:10.1053/euhj.2000.2476 · Cited by 1,132 · see citing works
Ventricular Pacing or Dual-Chamber Pacing for Sinus-Node Dysfunction
Lamas, G. A., Lee, K. L., Sweeney, M. O., Silverman, R., León, A. R., Yee, R. K., Marinchak, R. A., Flaker, G., SCHRON, E. B., Orav, E. J., et al. (2002) · New England Journal of Medicine · doi:10.1056/nejmoa013040 · Cited by 988 · see citing works
Self-citationGregoratos, G., Abrams, J., Epstein, A. E., Freedman, R. A., Hayes, D. L., Hlatky, M. A., Kerber, R. E., Naccarelli, G. V., Schoenfeld, M. H., Silka, M. J., et al. (2002) · Circulation · doi:10.1161/01.cir.0000035996.46455.09 · Cited by 905 · see citing works
Self-citationConnolly, S. J., Kerr, C. R., Gent, M., Roberts, R. S., Yusuf, S., Gillis, A. M., Sami, M. H., Talajic, M., Tang, A. S. L., Klein, G. J., et al. (2000) · New England Journal of Medicine · doi:10.1056/nejm200005113421902 · Cited by 775 · see citing works
Gregoratos, G., Cheitlin, M. D., Conill, A., Epstein, A. E., Fellows, C., Ferguson, T. B., Freedman, R. A., Hlatky, M. A., Naccarelli, G. V., Saksena, S. V., et al. (1998) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(98)00024-2 · Cited by 632 · see citing works
Cost-Effectiveness of Implantable Cardioverter–Defibrillators
Sanders, G. D., Hlatky, M. A., & Owens, D. K. (2005) · New England Journal of Medicine · doi:10.1056/nejmsa051989 · Cited by 543 · see citing works
Nielsen, J. C., Kristensen, L., Andersen, H., Mortensen, P. T., Pedersen, O. L., & Pedersen, A. K. (2003) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(03)00757-5 · Cited by 510 · see citing works
Prospective randomised trial of atrial versus ventricular pacing in sick-sinus syndrome
Andersen, H. R., Thuesen, L., Bagger, J. P., Vesterlund, T., & Thomsen, P. E. B. (1994) · The Lancet · doi:10.1016/s0140-6736(94)90347-6 · Cited by 490 · see citing works
Pacemaker and ICD Generator Malfunctions
Maisel, W. H., Moynahan, M., Zuckerman, B. D., Gross, T. P., Tovar, O. H., Tillman, D.-B., & Schultz, D. B. (2006) · JAMA · doi:10.1001/jama.295.16.1901 · Cited by 344 · see citing works
Self-citationSudden Cardiac Death: Exploring the Limits of Our Knowledge
Myerburg, R. J. (2001) · Journal of Cardiovascular Electrophysiology · doi:10.1046/j.1540-8167.2001.00369.x · Cited by 296 · see citing works
Gould, P. A. (2006) · JAMA · doi:10.1001/jama.295.16.1907 · Cited by 284 · see citing works
Intravascular Extraction of Problematic or Infected Permanent Pacemaker Leads: 1994–1996
BYRD, C. L., Wilkoff, B. L., Love, C. J., Sellers, T. D., TURK, K. T., Reeves, R. A., YOUNG, R., Crevey, B. J., Kutalek, S. P., Freedman, R. A., et al. (1999) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1999.tb00628.x · Cited by 279 · see citing works
Self-citationGregoratos, G., Cheitlin, M. D., Conill, A., Epstein, A. E., Fellows, C., Ferguson, T. B., Freedman, R. A., Hlatky, M. A., Naccarelli, G. V., Saksena, S. V., et al. (1998) · Circulation · doi:10.1161/01.cir.97.13.1325 · Cited by 273 · see citing works
Complications Related to Permanent Pacemaker Therapy
KIVINIEMI, M. S., PIRNES, M. A., ERÄNEN, H. J. K., Kettunen, R., & Hartikainen, J. E. K. (1999) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1999.tb00534.x · Cited by 259 · see citing works
Adverse Events With Transvenous Implantable Cardioverter-Defibrillators
Rosenqvist, M., Beyer, T., Block, M. S., den Dulk, K., Minten, J., & Lindemans, F. W. (1998) · Circulation · doi:10.1161/01.cir.98.7.663 · Cited by 231 · see citing works
Al‐Khatib, S. M., Lucas, F. L., Jollis, J. G., Malenka, D. J., & Wennberg, D. E. (2005) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2005.04.063 · Cited by 193 · see citing works
Pacemaker and ICD Generator Reliability
Maisel, W. H. (2006) · JAMA · doi:10.1001/jama.295.16.1929 · Cited by 170 · see citing works
Self-citationLate complications following permanent pacemaker implantation or elective unit replacement
Harcombe, A., NEWELL, S. A., Ludman, P., Wistow, T., Sharples, L. D., Schofield, P. M., Stone, D. L., Shapiro, L. M., Cole, T., & Petch, M. C. (1998) · Heart · doi:10.1136/hrt.80.3.240 · Cited by 162 · see citing works
Kron, J., Herre, J. M., Renfroe, E. G., Rizo-Patrón, C., Raitt, M. H., Halperin, B. D., Gold, M. R., Goldner, B. G., Wathen, M. S., Wilkoff, B. L., et al. (2001) · American Heart Journal · doi:10.1067/mhj.2001.111261 · Cited by 149 · see citing works
Self-citationGold, M. R., Peters, R. W., Johnson, J. W., & Shorofsky, S. R. (1996) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(96)00314-2 · Cited by 102 · see citing works
Pediatric pacemaker infections: Twenty years of experience
Cohen, M. I., Bush, D. M., Gaynor, J. W., Vetter, V. L., Tanel, R. E., & Rhodes, L. A. (2002) · Journal of Thoracic and Cardiovascular Surgery · doi:10.1067/mtc.2002.123643 · Cited by 78 · see citing works
Pacemaker Lead Infection in Young Patients
Klug, D., Vaksmann, G., Jarwe, M., WALLET, F., Francart, C., Kacet, S., & Rey, C. (2003) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2003.t01-1-00215.x · Cited by 77 · see citing works
Infections with Nonthoracotomy Implantable Cardioverter Defibrillators: Can These Be Prevented?
Smith, P. N., Vidaillet, H. J., Jr., Hayes, J. J., WETHINGTON, P. J., VOSHAGE‐STAHL, L., HULL, M. L., & Broste, S. K. (1998) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1998.tb01060.x · Cited by 71 · see citing works
Takahashi, T., Bhandari, A. K., Watanuki, M., Cannom, D. S., Sakurada, H., & Hiraoka, M. (2002) · Circulation Journal · doi:10.1253/circj.66.746 · Cited by 64 · see citing works
Management of Recalled Pacemakers and Implantable Cardioverter-Defibrillators
Amin, M. S., Matchar, D. B., Wood, M. A., & Ellenbogen, K. A. (2006) · JAMA · doi:10.1001/jama.296.4.412 · Cited by 62 · see citing works
Self-citationLead extraction for device related infections: a single-centre experience
Bracke, F. A. L. E. (2004) · EP Europace · doi:10.1016/j.eupc.2004.01.007 · Cited by 62 · see citing works
The National ICD Registry: Now and into the future
Hammill, S. C., Phurrough, S. E., & Brindis, R. G. (2006) · Heart Rhythm · doi:10.1016/j.hrthm.2006.01.019 · Cited by 54 · see citing works
Self-citationNonthoracotomy Implantable Defibrillator Lead Extraction:
Saad, E., Saliba, W. I., Schweikert, R. A., Al-Khadra, A. S., Abdul‐Karim, A., Niebauer, M. J., & Wilkoff, B. L. (2003) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2003.00300.x · Cited by 54 · see citing works
Self-citationKantharia, B. K., Padder, F. A., Pennington, J. C., Wilbur, S. L., Samuels, F. L., MAQUILAN, M., & Kutalek, S. P. (2000) · The American Journal of Cardiology · doi:10.1016/s0002-9149(99)00817-6 · Cited by 32 · see citing works
Performance of implantable Cardiac Rhythm Management Devices
SONG, S. L. (1994) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1994.tb02410.x · Cited by 30 · see citing works
The Bilitch Report: Performance of Implantable Cardiac Rhythm Management Devices
Furman, S., Parsonnet, V., Goldman, B. S., Denes, P., SONG, S. L., Maloney, J. D., Harthorne, J. W., & Griffin, J. C. (1989) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1989.tb02694.x · Cited by 29 · see citing works
Improving Patient Care by Reporting Problems with Medical Devices
White, G. G., Weick-Brady, M. D., & Gross, T. P. (1998) · Journal of Toxicology Clinical Toxicology · doi:10.3109/15563659809028066 · Cited by 26 · see citing works
Management of Intracardiac Device Recalls: A Consensus Conference
Goldman, B. S., Newman, D. M., ERASER, J., IRWIN, M. E., & CONFERENCE, T. P. O. T. C. (1996) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1996.tb04785.x · Cited by 23 · see citing works
Maloney, J. D., Hayes, D. L., & TIMMIS, G. C. (1993) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1993.tb00986.x · Cited by 22 · see citing works
Technology assessment of medical devices at the Center for Devices and Radiological Health.
Kessler, L. G., & Richter, K. (1998) · PubMed · Cited by 14 · see citing works
PRATT, T. R. H., Pulling, C. C., & Stanton, M. S. (2000) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2000.tb00916.x · Cited by 13 · see citing works
Postmortem Analysis and Retrieval of Implantable Pacemakers and Defibrillators
Kirkpatrick, J. N., Burke, M. C., & Knight, B. P. (2006) · New England Journal of Medicine · doi:10.1056/nejmc053279 · Cited by 12 · see citing works
Management of intracardiac device recalls: a consensus conference.
Bs, G., Newman, D., Fraser, J. S., & IRWIN, M. E. (1996) · PubMed · Cited by 6 · see citing works
Furman, S. (1994) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1994.tb02409.x · Cited by 5 · see citing works
A Proposed Solution to the Notification Problem
Hall, R. F. (2005) · Minnesota journal of law science & technology · doi:10.24926/15529541.3668 · Cited by 1 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities
Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.02.032 · Cited by 2,546 · see citing works
Self-citationACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities
Epstein, A. E., Dimarco, J., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Circulation · doi:10.1161/circualtionaha.108.189742 · Cited by 1,768 · see citing works
Self-citationEpstein, A. E., Dimarco, J., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2012) · Circulation · doi:10.1161/cir.0b013e318276ce9b · Cited by 1,495 · see citing works
Self-citationKusumoto, F. M., Schoenfeld, M. H., Wilkoff, B. L., Berul, C. I., Birgersdotter-Green, U. M., Carrillo, R. G., Cha, Y., Clancy, J. F., Deharo, J., Ellenbogen, K. A., et al. (2017) · Heart Rhythm · doi:10.1016/j.hrthm.2017.09.001 · Cited by 1,211 · see citing works
Self-citationTracy, C. M., Epstein, A. E., Darbar, D., Dimarco, J., Dunbar, S. B., Estes, N. A. M., Ferguson, T. B., Hammill, S. C., Karasik, P., Link, M. S., et al. (2012) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2012.11.007 · Cited by 738 · see citing works
Self-citationVarma, N., Epstein, A. E., Irimpen, A., Schweikert, R., & Love, C. (2010) · Circulation · doi:10.1161/circulationaha.110.937409 · Cited by 594 · see citing works
ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities
Epstein, A. E., Dimarco, J., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.04.014 · Cited by 550 · see citing works
Self-citationSecurity and Privacy for Implantable Medical Devices
Halperin, D., Heydt-Benjamin, T. S., Fu, K. E., Kohno, T., & Maisel, W. H. (2008) · IEEE Pervasive Computing · doi:10.1109/mprv.2008.16 · Cited by 468 · see citing works
Self-citationWilkoff, B. L., Auricchio, A., Brugada, J., Cowie, M., Ellenbogen, K. A., Gillis, A. M., Hayes, D. L., Howlett, J. G., Kautzner, J., Love, C. J., et al. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.04.013 · Cited by 364 · see citing works
Self-citationWilkoff, B. L., Auricchio, A., Brugada, J., Cowie, M., Ellenbogen, K. A., Gillis, A. M., Hayes, D. L., Howlett, J. G., Kautzner, J., Love, C. J., et al. (2008) · Europace · doi:10.1093/europace/eun122 · Cited by 231 · see citing works
Self-citationAutomatic Remote Monitoring of Implantable Cardioverter-Defibrillator Lead and Generator Performance
Varma, N., Michalski, J., Epstein, A. E., & Schweikert, R. A. (2010) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.110.951962 · Cited by 164 · see citing works
Implantable Cardioverter-Defibrillator Therapy Before Death
Westerdahl, A. K., Sjöblom, J., Mattiasson, A., Rosenqvist, M., & Frykman, V. (2013) · Circulation · doi:10.1161/circulationaha.113.002648 · Cited by 129 · see citing works
Ferrick, A. M., Raj, S. R., Deneke, T., Kojodjojo, P., López-Cabanillas, N., Abe, H., Bovéda, S., Chew, D. S., Choi, J., Dagres, N., et al. (2023) · Heart Rhythm · doi:10.1016/j.hrthm.2023.03.1525 · Cited by 122 · see citing works
Recommendations from the Heart Rhythm Society Task Force on Lead Performance Policies and Guidelines
Maisel, W. H., Hauser, R. G., Hammill, S. C., Hauser, R. G., Ellenbogen, K. A., Epstein, A. E., Hayes, D. L., Alpert, J. S., Berger, R. D., Curtis, A. B., et al. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.04.024 · Cited by 114 · see citing works
Self-citationBraunschweig, F., Boriani, G., Bauer, A., Hatala, R., Herrmann‐Lingen, C., Kautzner, J., Pedersen, S. S., Pehrson, S. M., Ricci, R. P., & Schalij, M. J. (2010) · EP Europace · doi:10.1093/europace/euq316 · Cited by 113 · see citing works
Telemedicine and cardiac implants: what is the benefit?
Varma, N., & Ricci, R. P. (2013) · European Heart Journal · doi:10.1093/eurheartj/ehs388 · Cited by 102 · see citing works
Jung, W., Rillig, A., Birkemeyer, R., Miljak, T., & Meyerfeldt, U. (2008) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-008-9311-5 · Cited by 96 · see citing works
Francis, G. S., Greenberg, B., Hsu, D. T., Jaski, B. E., Jessup, M. L., LeWinter, M. M., Pagani, F. D., Piña, I. L., Semigran, M. J., Walsh, M. N., et al. (2010) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.04.014 · Cited by 91 · see citing works
Ferrick, A. M., Raj, S. R., Deneke, T., Kojodjojo, P., López-Cabanillas, N., Abe, H., Bovéda, S., Chew, D. S., Choi, J., Dagres, N., et al. (2023) · EP Europace · doi:10.1093/europace/euad123 · Cited by 89 · see citing works
Lead Perforation: Incidence in Registries
Carlson, M. D., Freedman, R. A., & LEVINE, P. A. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00943.x · Cited by 84 · see citing works
Self-citationKrahn, A. D., Champagne, J., Healey, J. S., Cameron, D., Simpson, C. S., Thibault, B., Mangat, I., Tung, S., Sterns, L. D., BIRNIE, D. H., et al. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.01.029 · Cited by 82 · see citing works
Risks and Challenges of Implantable Cardioverter-Defibrillators in Young Adults
Sherrid, M. V., & Daubert, J. P. (2008) · Progress in Cardiovascular Diseases · doi:10.1016/j.pcad.2008.10.001 · Cited by 78 · see citing works
eHealth in cardiovascular medicine: A clinical update
Saner, H. E., & van der Velde, E. T. (2016) · European Journal of Preventive Cardiology · doi:10.1177/2047487316670256 · Cited by 73 · see citing works
Outcome of advisory implantable cardioverter-defibrillator replacement: One-year follow-up
Gould, P. A., Gula, L. J., Champagne, J., Healey, J. S., Cameron, D., Simpson, C., Thibault, B., Pintér, A., Tung, S., Sterns, L. D., et al. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.09.020 · Cited by 70 · see citing works
Kirkpatrick, J. N., Ghani, S. N., Burke, M. C., & Knight, B. P. (2007) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2007.00773.x · Cited by 68 · see citing works
Akar, J. G., Bao, H., Jones, P. W., Wang, Y., Varosy, P. D., Masoudi, F. A., Steín, K. M., Saxon, L. A., Normand, S. T., & Curtis, J. P. (2015) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.114.003030 · Cited by 67 · see citing works
Self-citationPerformance of the St. Jude Medical Riata Leads
Epstein, A. E., Baker, J. H., Beau, S. L., Deering, T. F., Greenberg, S. M., & Goldman, D. S. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.10.030 · Cited by 62 · see citing works
Proclemer, A., Ghidina, M., Gregori, D., Facchin, D., Rebellato, L., Fioretti, P., & Brignole, M. (2008) · EP Europace · doi:10.1093/europace/eun370 · Cited by 61 · see citing works
Kirkpatrick, J. N., Papini, C., Baman, T. S., Khota, K., Eagle, K. A., Verdino, R. J., & Caplan, A. L. (2010) · Heart Rhythm · doi:10.1016/j.hrthm.2010.04.027 · Cited by 60 · see citing works
Risk Factors for Implantable Defibrillator Lead Fracture in a Recalled and a Nonrecalled Lead
Morrison, T. B., Rea, R. F., Hodge, D. O., Crusan, D. J., Koestler, C., Asirvatham, S. J., BRADLEY, D., Shen, W. K., Munger, T. M., Hammill, S. C., et al. (2010) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2009.01683.x · Cited by 60 · see citing works
Self-citationGuidance on reuse of cardio-vascular catheters and devices in India: A consensus document
Kapoor, A., Vora, A. M., Nataraj, G., Mishra, S., Kerkar, P. G., & Manjunath, C. N. (2017) · Indian Heart Journal · doi:10.1016/j.ihj.2017.04.003 · Cited by 57 · see citing works
Complications Associated with Generator Replacement in Response to Device Advisories
Costea, A., Rardon, D. P., Padanilam, B. J., Fogel, R. I., & Prystowsky, E. N. (2007) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2007.01047.x · Cited by 54 · see citing works
Self-citationSudden Death Prevention With Implantable Devices
PASSMAN, R. S., & Kadish, A. H. (2007) · Circulation · doi:10.1161/circulationaha.106.655704 · Cited by 54 · see citing works
Baman, T. S., Kirkpatrick, J. N., Romero, J. M., Gakenheimer, L., Romero, A., Lange, D. C., Nosowsky, R., Fuller, K., Sison, E. O. D., Tangco, R. V., et al. (2010) · Circulation · doi:10.1161/circulationaha.110.970483 · Cited by 52 · see citing works
Preventing tomorrow's sudden cardiac death today
Al‐Khatib, S. M., Sanders, G. D., Carlson, M. D., Cicic, A., Curtis, A. B., Fonarow, G. C., Groeneveld, P. W., Hayes, D. L., Heidenreich, P. A., Mark, D. B., et al. (2008) · American Heart Journal · doi:10.1016/j.ahj.2008.05.027 · Cited by 47 · see citing works
Self-citationVarma, N., Pavri, B. B., Stambler, B. S., Michalski, J., & TRUST Investigators (2012) · EP Europace · doi:10.1093/europace/eus410 · Cited by 47 · see citing works
Contemporary Pacemaker and Defibrillator Device Therapy
Schoenfeld, M. H. (2007) · Circulation · doi:10.1161/circulationaha.106.618587 · Cited by 45 · see citing works
Implantable Cardioverter Defibrillator‐Related Complications in the Pediatric Population
Shah, M. J. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02389.x · Cited by 42 · see citing works
Undavia, M. B., Goldstein, N. E., Cohen, P., Sinthawanarong, K., SINGSON, M., Bhutani, D., Munson, T., Gomes, J. A., Fischer, A., & Mehta, D. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01204.x · Cited by 41 · see citing works
Proclemer, A., Ghidina, M., Gregori, D., Facchin, D., Rebellato, L., Zakja, E., Gulizia, M. M., & Esente, P. (2009) · EP Europace · doi:10.1093/europace/eup346 · Cited by 41 · see citing works
Benefits of the Implantable Cardioverter-Defibrillator
Epstein, A. E. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.06.035 · Cited by 40 · see citing works
Reuse of Explanted Permanent Pacemakers Donated by Funeral Homes
Kantharia, B. K., Patel, S., Kulkarni, G. V., Shah, A. N., Lokhandwala, Y. Y., Mascarenhas, E., & Mascarenhas, D. A. N. (2011) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2011.08.036 · Cited by 38 · see citing works
Psychological Reaction to Potential Malfunctioning of Implantable Defibrillators
van den Broek, K. C., Denollet, J., Nyklíček, I., & van der Voort, P. H. (2006) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2006.00468.x · Cited by 36 · see citing works
Managing with pacemakers and implantable cardioverter defibrillators
Lampert, R. (2013) · Circulation · doi:10.1161/circulationaha.113.001555 · Cited by 34 · see citing works
Self-citationClinical Inferences of Cardiovascular Implantable Electronic Device Analysis at Autopsy
Sinha, S. K., Crain, B. J., Flickinger, K., Calkins, H. G., Rickard, J., Cheng, A., Berger, R. D., Tomaselli, G. F., & Marine, J. E. (2016) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2016.06.052 · Cited by 33 · see citing works
Effective communication and ethical consent in decisions related to ICDs
Clark, A. M., Jaarsma, T., Strachan, P. H., Davidson, P. M., Jerke, M., Beattie, J. M., Duncan, A. S., Ski, C. F., & Thompson, D. R. (2011) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2011.101 · Cited by 32 · see citing works
Gakenheimer, L., Lange, D. C., Romero, J. M., Kirkpatrick, J. N., Sovitch, P., Oral, H., Eagle, K. A., & Baman, T. S. (2011) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-010-9534-0 · Cited by 32 · see citing works
Pedersen, S. S., Brouwers, C. J., & Versteeg, H. (2012) · Expert Review of Medical Devices · doi:10.1586/erd.12.31 · Cited by 31 · see citing works
Implantable cardioverter defibrillators: risks accompany the life-saving benefits
Atwater, B. D., & Daubert, J. P. (2012) · Heart · doi:10.1136/heartjnl-2012-301853 · Cited by 29 · see citing works
Pacemaker recycling: A notion whose time has come
Runge, M., Baman, T. S., Davis, S., Weatherwax, K. J., Goldman, E., Eagle, K. A., & Crawford, T. (2017) · World Journal of Cardiology · doi:10.4330/wjc.v9.i4.296 · Cited by 29 · see citing works
Postmarket surveillance of medical devices: current capabilities and future opportunities
Blake, K. (2013) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-013-9778-6 · Cited by 29 · see citing works
Cha, Y., El‐Chami, M. F., Liu, C. F., Andreychuk, L. J., Beaver, T. M., Bergen, K. M., Berul, C. I., Birgersdotter-Green, U. M., Breitenstein, A., Epstein, L. M., et al. (2026) · Heart Rhythm · doi:10.1016/j.hrthm.2026.04.015 · Cited by 26 · see citing works
Larsen, J. M., Hjortshøj, S. P., Nielsen, J. C., Johansen, J. B., Petersen, H. H., Haarbo, J., Johansen, M. B., & Thøgersen, A. M. (2015) · Heart Rhythm · doi:10.1016/j.hrthm.2015.11.034 · Cited by 25 · see citing works
Krahn, A. D., Simpson, C. S., Parkash, R., Yee, R. K., Champagne, J., Healey, J. S., Cameron, D., Thibault, B., Mangat, I., Tung, S., et al. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.01.012 · Cited by 24 · see citing works
Reliability and longevity of implantable defibrillators
Hauser, R. G., Casey, S. A., Gitter, C. B., Tang, C. Y., Abdelhadi, R. H., Gornick, C. C., Stanberry, L. I., & Sengupta, J. D. (2021) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-020-00920-w · Cited by 24 · see citing works
Self-citationLogani, S., Gottlieb, M., Verdino, R. J., Baman, T. S., Eagle, K. A., & Kirkpatrick, J. N. (2011) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2011.03032.x · Cited by 22 · see citing works
Varma, N. (2013) · EP Europace · doi:10.1093/europace/eut116 · Cited by 23 · see citing works
Francis, G. S., Greenberg, B., Hsu, D. T., Jaski, B. E., Jessup, M. L., LeWinter, M. M., Pagani, F. D., Piña, I. L., Semigran, M. J., Walsh, M. N., et al. (2010) · Circulation · doi:10.1161/cir.0b013e3181ecbd97 · Cited by 23 · see citing works
Ferrick, A. M., Raj, S. R., Deneke, T., Kojodjojo, P., López-Cabanillas, N., Abe, H., Bovéda, S., Chew, D. S., Choi, J., Dagres, N., et al. (2023) · Journal of Arrhythmia · doi:10.1002/joa3.12851 · Cited by 21 · see citing works
A Viewpoint on the Impact of Device Advisories on Patient‐Centered Outcomes
PEDERSEN, S. S., VAN DEN BERG, M., & THEUNS, D. A. M. J. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02432.x · Cited by 20 · see citing works
The Sprint Fidelis lead fracture story: What do we really know and where do we go from here?
Ellenbogen, K. A., Wood, M. A., & Swerdlow, C. D. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.08.001 · Cited by 19 · see citing works
Self-citationComplications of replacing implantable devices in response to advisories: A single center experience
Moore, J. W., Barrington, W., Bazaz, R. R., Jain, S. K., Němec, J., Ngwu, O., Schwartzman, D. A., Shalaby, A. A., & Saba, S. (2008) · International Journal of Cardiology · doi:10.1016/j.ijcard.2007.12.070 · Cited by 19 · see citing works
Maron, B. J., & Hauser, R. G. (2007) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2007.02.072 · Cited by 19 · see citing works
Self-citationPatient Evaluation of ICD Recall Communication Strategies:A Vignette Study
Stutts, L. A., CONTI, J. B., Aranda, J. M., Miles, W. M., Burkart, T. A., & Sears, S. F. (2007) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2007.00820.x · Cited by 18 · see citing works
Pedersen, S. S., Versteeg, H., Nielsen, J. C., Mortensen, P. T., & Johansen, J. B. (2011) · EP Europace · doi:10.1093/europace/eur157 · Cited by 17 · see citing works
Pron, G. E., Ieraci, L., & Kaulback, K. (2012) · Europe PMC (PubMed Central) · Cited by 17 · see citing works
Global disparities in arrhythmia care: Mind the gap
Sharif, Z. I., & Ptaszek, L. M. (2022) · Heart Rhythm O2 · doi:10.1016/j.hroo.2022.09.007 · Cited by 16 · see citing works
Use of Administrative Data to Monitor Cardiac Implantable Electronic Device Complications
Parkash, R., Sapp, J. L., Gardner, M., Gray, C., AbdelWahab, A. M., & Cox, J. L. (2018) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2018.10.018 · Cited by 16 · see citing works
To Replace or Not to Replace: A Systematic Approach to Respond to Device Advisories
Priori, S. G., Auricchio, A., Nisam, S., & YONG, P. (2008) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2008.01291.x · Cited by 15 · see citing works
Laskey, W. K., Awad, K., Lum, J. J., Skodacek, K., Zimmerman, B., Selzman, K. A., & Zuckerman, B. D. (2012) · American Journal of Therapeutics · doi:10.1097/mjt.0b013e3182512ca5 · Cited by 15 · see citing works
Preventing tomorrow's sudden cardiac death today
Sanders, G. D., Al‐Khatib, S. M., Berliner, E., Bigger, J. T., Buxton, A. E., Califf, R. M., Carlson, M. D., Curtis, A. B., Curtis, J. P., DOMANSKI, M. J., et al. (2007) · American Heart Journal · doi:10.1016/j.ahj.2007.03.002 · Cited by 14 · see citing works
Self-citationThe Fidelis Recall: How Much Pressure Can the ICD World Bear?
Cannom, D. S., & Fisher, J. D. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01170.x · Cited by 14 · see citing works
Ricci, R. P., Pignalberi, C., Magris, B., Aquilani, S., Altamura, V., Morichelli, L., Porfili, A., Quarta, L., Saputo, F. A., & Santini, M. (2011) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-011-9612-y · Cited by 13 · see citing works
Remote Surveillance of Implantable Cardiac Devices
Orlov, M. V., Szombathy, T., CHAUDHRY, G. M., & Haffajee, C. I. (2009) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2009.02412.x · Cited by 13 · see citing works
Proceedings of the ICD Lead Performance Conference
Maisel, W. H., & Hauser, R. G. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.05.034 · Cited by 13 · see citing works
Self-citationElgaard, A. F., Larsen, J. M., Nielsen, J. C., Johansen, J. B., Haarbo, J., Petersen, H. H., & Riahi, S. (2016) · EP Europace · doi:10.1093/europace/euw226 · Cited by 13 · see citing works
Outcomes Before and After the Recall of a Heart Failure Pacemaker
Sengupta, J. N., Storey, K. M., Casey, S. A., Trager, L. E., Buescher, M., Horning, M., Gornick, C. C., Abdelhadi, R. H., Tang, C. Y., Brill, S., et al. (2019) · JAMA Internal Medicine · doi:10.1001/jamainternmed.2019.5171 · Cited by 13 · see citing works
Self-citationRisk Assessment for Defibrillator Therapy
Anderson, K. P. (2007) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.05.038 · Cited by 13 · see citing works
Mittal, S., Movsowitz, C., & Varma, N. (2014) · Cardiology Clinics · doi:10.1016/j.ccl.2014.01.001 · Cited by 12 · see citing works
Ethical dilemmas in device treatment for advanced heart failure
Kirkpatrick, J. N., Fedson, S., & Verdino, R. J. (2007) · Current Opinion in Supportive and Palliative Care · doi:10.1097/spc.0b013e3282f317a1 · Cited by 12 · see citing works
Remote monitoring of implantable devices: The European experience
Morichelli, L., & Ricci, R. P. (2009) · Heart Rhythm · doi:10.1016/j.hrthm.2009.02.030 · Cited by 12 · see citing works
Sengupta, J. N., Kendig, A. C., Goormastic, M., Hwang, E.-S., Ching, E. A., Chung, R., Lindsay, B. D., TCHOU, P. J., Wilkoff, B. L., Niebauer, M. J., et al. (2012) · Heart Rhythm · doi:10.1016/j.hrthm.2012.07.002 · Cited by 12 · see citing works
Self-citationOttenberg, A. L., Mueller, L. A., & Mueller, P. S. (2013) · Heart & Lung · doi:10.1016/j.hrtlng.2012.09.007 · Cited by 12 · see citing works
Postmortem Interrogation of Cardiac Implantable Electronic Devices
Paratz, E. D., Block, T. J., Stub, D. A., Gerche, A. L., Kistler, P. M., Kalman, J. M., Strathmore, N., Mond, H. G., Woodford, N. W. F., Burke, M., et al. (2021) · JACC. Clinical electrophysiology · doi:10.1016/j.jacep.2021.10.011 · Cited by 11 · see citing works
Clinical outcomes in patients with implantable cardioverter defibrillators and Sprint Fidelis leads
Verlato, R., Facchin, D., Catanzariti, D., Molon, G., Zanotto, G., Morani, G., Brieda, M., Zanon, F., Delise, P., Leoni, L., et al. (2013) · Heart · doi:10.1136/heartjnl-2012-303259 · Cited by 11 · see citing works
Swindle, J. P., Burroughs, T. E., Schnitzler, M. A., & Hauptman, P. J. (2008) · American Heart Journal · doi:10.1016/j.ahj.2008.04.003 · Cited by 11 · see citing works
Small-diameter defibrillation electrodes: Can they take a licking and keep hearts ticking?
Swerdlow, C. D. (2007) · Heart Rhythm · doi:10.1016/j.hrthm.2007.05.001 · Cited by 10 · see citing works
Yıldız, B. S., Fιndıkoğlu, G., Alihanoğlu, Y. İ., Kılıç, İ. D., Evrengül, H., & Şenol, H. (2018) · Rehabilitation Research and Practice · doi:10.1155/2018/5689353 · Cited by 10 · see citing works
Liu, I.-F., Chang, S., Lo, L., Hu, Y., Tuan, T., Kong, C., Wu, T., Chiang, C., Chen, S., & Lin, Y. (2011) · International Journal of Biometeorology · doi:10.1007/s00484-010-0388-z · Cited by 9 · see citing works
Donation of explanted pacemakers for reuse in underserved nations
Stanyon, R. (2010) · Journal of Healthcare Risk Management · doi:10.1002/jhrm.20018 · Cited by 9 · see citing works
Schwartz, J., Blangy, H., Zinzius, P. Y., Freysz, L., Aliot, E., & Sadoul, N. (2010) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2010.02918.x · Cited by 9 · see citing works
Hauser, R. G., & Kallinen, L. M. (2013) · Heart Rhythm · doi:10.1016/j.hrthm.2013.07.001 · Cited by 9 · see citing works
Self-citationProblems with Implantable Cardiac Device Therapy
Kowalski, M., Huizar, J. F., Kaszala, K., & Wood, M. A. (2008) · Cardiology Clinics · doi:10.1016/j.ccl.2008.03.012 · Cited by 9 · see citing works
Sharma, A. D., Blank, A., Patel, P., & Steín, K. M. (2013) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-013-9781-y · Cited by 9 · see citing works
Clinical Judgment Versus Decision Analysis for Managing Device Advisories
Amin, M. S., Wood, M. A., Shepard, R. K., Kalahasty, G., & Ellenbogen, K. A. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01171.x · Cited by 9 · see citing works
Self-citationRemote Monitoring for Follow-up of Patients with Cardiac Implantable Electronic Devices
Ricci, R. P., Morichelli, L., & Varma, N. (2014) · Arrhythmia & Electrophysiology Review · doi:10.15420/aer.2014.3.2.123 · Cited by 8 · see citing works
Guidelines for Managing Pacemaker and Implantable Defibrillator Advisories
Kramer, D. B., & Maisel, W. H. (2011) · Clinical Cardiac Pacing, Defibrillation and Resynchronization Therapy · doi:10.1016/b978-1-4377-1616-0.00034-5 · Cited by 8 · see citing works
Self-citationImplantable Cardioverter‐Defibrillators: Is There Life after Death?
Heuverswyn, F. E. V., Timmers, L., Stroobandt, R. X., & Barold, S. S. (2012) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.12023 · Cited by 7 · see citing works
Theuns, D. A. M. J., & Jordaens, L. J. L. M. (2011) · Netherlands Heart Journal · doi:10.1007/s12471-011-0228-8 · Cited by 7 · see citing works
The “Guidant Affair”: 15 years later
Maron, B. J., & Estes, N. A. M. (2020) · Heart Rhythm · doi:10.1016/j.hrthm.2020.10.016 · Cited by 7 · see citing works
- Authors
- Carlson, M. D., Wilkoff, B. L., Maisel, W. H., Carlson, M. D., Ellenbogen, K. A., Saxon, L. A., Prystowsky, E. N., Alpert, J. S., Cain, M. E., Ching, E. A., Curtis, A. B., Davies, D. W., Hammill, S. C., Hauser, R. G., Lampert, R., & Zipes, D. P.
- Affiliations
- Cleveland Clinic (US)
- Beth Israel Deaconess Medical Center (US)
- Virginia Commonwealth University Medical Center (US)
- Keck Hospital of USC (US)
- American Heart Association (US)
- Washington University in St. Louis (US)
- University of South Florida (US)
- St Mary's Hospital (GB)
- Mayo Clinic (US)
- Yale University (US)
- American College of Cardiology (US)
- ORCID
- Venue
- Heart Rhythm
- Publisher
- Elsevier BV
- Published
- 2006-10
- Volume
- 3
- Issue
- 10
- Pages
- 1250-1273
- ISSN
- 1547-5271 (ISSN-L)
- OpenAlex
- W2119072430
- Cited by
- 177
- References
- 52
- Citations by year
- 2026: 4 · 2025: 4 · 2024: 2 · 2023: 5 · 2022: 3 · 2021: 6 · 2020: 2 · 2019: 2 · 2018: 2 · 2017: 6 · 2016: 9 · 2015: 2 · 2014: 4 · 2013: 17 · 2012: 18
- FWCI
- 14.28
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Heart Rhythm (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Health Sciences › Health Professions › Health Information Management › Healthcare Regulation
- Health Sciences › Health Professions › Medical Laboratory Technology › Quality and Safety in Healthcare
- Keywords
- cardiac implantable electronic devices, equipment failure, device performance, battery depletion, pacemaker leads, cardiac pacing, FDA regulation
- MeSH terms
- Arrhythmias, Cardiac, Cardiac Pacing, Artificial, Electric Countershock, Humans, Pacemaker, Artificial, Defibrillators, Implantable
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review