Effect of decentration and tilt on the image quality of aspheric intraocular lens designs in a model eye
The DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.
Abstract
PURPOSE: To determine the impact of decentration and tilt on the imaging quality of aspheric intraocular lens (IOL) designs in a schematic model eye. SETTING: Institute of Medical Physics, University of Erlangen-Nuremberg, Erlangen, Germany. METHOD: A model eye was used to calculate the impact of misalignment on the imaging quality of 6 IOL designs. The crystalline lens in the model eye was replaced with IOL designs with 22.0 diopters nominal refractive power, and the anterior chamber depth (ACD) was set to the estimated ACD value provided by the manufacturer. The retinal position was optimized for the best image quality. The IOLs were decentered up to +/-1.0 mm and tilted up to +/-5 degrees relative to the line of sight. At each position, the modulation transfer function was recorded with 3.0 mm and 4.5 mm pupil diameters. The results between the IOL designs and those of the phakic model eye were then compared. RESULTS: Aberration-correcting IOLs were very sensitive to decentration and tilt. However, the impact of misalignment depended on IOL design. Aberration-free IOLs showed less sensitivity within a wide range of displacement but provided better results than the spherical IOL. CONCLUSIONS: Overall, modern aspheric IOLs provided better imaging quality than conventional spherical IOL designs. Aberration-free IOLs were less sensitive to decentration and tilt than aberration-correcting IOLs but provided better image quality than spherical IOLs. Aberration-correcting IOLs have the potential to provide diffraction-limited imaging quality when perfectly aligned.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Handbuch der physiologischen Optik
Helmholtz, H. L. F. (2011) · DFG Viewer (DFG Viewer) · Cited by 1,947
The luminous efficiency of rays entering the eye pupil at different points
Stiles, W. S., & Crawford, B. H. (1933) · Proceedings of the Royal Society of London. Series B, Containing Papers of a Biological Character · doi:10.1098/rspb.1933.0020 · Cited by 689
Anatomically accurate, finite model eye for optical modeling
Liou, H.-L., & Brennan, N. A. (1997) · Journal of the Optical Society of America A · doi:10.1364/josaa.14.001684 · Cited by 612 · cited 3×
“96 16.27 1.336 Retina 12.00 0.00 d d l Z wavelength; n Z refractive index *Gradient index material modeled by the equations supplied by Liou and Brennan.11 The retinal curvature was taken from Atchison and Smith.14 G - VOL 35, JUNE 2009 value with 2 iris pupil diameters (3.0 mm and 4.5 mm) to address image quality for…”
MethodInfluentialA New Intraocular Lens Design to Reduce Spherical Aberration of Pseudophakic Eyes
Holladay, J. T., Piers, P. A., Koranyi, G., van der Mooren, M., & Norrby, N. E. S. (2002) · Journal of Refractive Surgery · doi:10.3928/1081-597x-20021101-04 · Cited by 560 · cited 1×
“andHolladay et al.(6) TheMTF values of the 2 aberration-free IOLs (MC6125AS and SofPort AO) showed less decentration dependency, and the MTF of the spherical IOL was independent of decentration.”
BackgroundAtchison, D. A. (2018) · Encyclopedia of Modern Optics · doi:10.1016/b978-0-12-803581-8.09773-3 · Cited by 484 · cited 3×
“The retinal curvature was taken from Atchison and Smith.(14)”
MethodInfluentialOptical performance of 3 intraocular lens designs in the presence of decentration
Altmann, G. E., Nichamin, L. D., Lane, S. S., & Pepose, J. S. (2005) · Journal of Cataract and Refractive Surgery · doi:10.1016/j.jcrs.2004.09.024 · Cited by 210 · cited 4×
“Our methods were similar to those of Altmann et al.(30); however, we used a different model eye and presented the MTF at 30 cpd against the amount of misalignment instead of showing complete MTF curves for discrete values of misalignment.”
MethodResultInfluentialTheoretical Eye Model with Aspherics*
Lotmar, W. (1971) · Journal of the Optical Society of America · doi:10.1364/josa.61.001522 · Cited by 195
Improved functional vision with a modified prolate intraocular lens
Packer, M., Fine, H. I., Hoffman, R. S., & Piers, P. A. (2004) · Journal of Cataract and Refractive Surgery · doi:10.1016/j.jcrs.2003.10.022 · Cited by 159 · cited 2×
“These results are also supported in a study by Packer et al.(38) However, no significant improvement in visual acuity with low and high contrast and a decrease in spherical aberration were found in 2 studies by Kasper et al.”
ResultDesign of aspheric intraocular lenses
Atchison, D. A. (1991) · Ophthalmic and Physiological Optics · doi:10.1111/j.1475-1313.1991.tb00213.x · Cited by 118
Kasper, T., Bühren, J., & Kohnen, T. (2006) · Journal of Cataract & Refractive Surgery · doi:10.1016/j.jcrs.2005.11.018 · Cited by 113 · cited 2×
“However, no significant improvement in visual acuity with low and high contrast and a decrease in spherical aberration were found in 2 studies by Kasper et al.(39,40) Assuming that the IOLs in those studieswerewell centered, the results agreewith the finding in our simulation that the Tecnis Z9000 IOL provided much be…”
Result
Cited by
Comparison of visual performance with an aspheric intraocular lens and a spherical intraocular lens
Thiagarajan, M., McClenaghan, R., & Anderson, D. F. (2011) · Journal of Cataract and Refractive Surgery · doi:10.1016/j.jcrs.2011.05.046 · Cited by 19
Metadata
- Authors
- Eppig, T., Scholz, K., Löffler, A., Meßner, A., & Langenbucher, A.
- Venue
- Journal of Cataract and Refractive Surgery
- Published
- 2009-06
- Volume
- 35
- Issue
- 6
- Pages
- 1091-1100
- ISSN
- 0146-2776 (ISSN-L)
- 0886-3350
- OpenAlex
- W2000419123
- Cited by
- 185
- References
- 12
- FWCI
- 8.87
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.