A high aspect ratio microelectrode array for mapping neural activity in vitro
Open accessAbstract
A novel high-aspect-ratio penetrating microelectrode array was designed and fabricated for the purpose of recording neural activity. The array allows two dimensional recording of 64 sites in vitro with high aspect ratio penetrating electrodes. Traditional surface electrode arrays, although easy to fabricate, do not penetrate to the viable tissue such as central hippocampus slices and thus have a lower signal/noise ratio and lower selectivity than a penetrating array. In the unfolded hippocampus preparation, the CA1-CA3 pyramidal cell layer in the whole unfolded rodent hippocampus preparation is encased by the alveus on one side and the Schaffer tract on the other and requires penetrating electrodes for high signal to noise ratio recording. An array of 64 electrode spikes, each with a target height of 200μm and diameter of 20μm, was fabricated in silicon on a transparent glass substrate. The impedance of the individual electrodes was measured to be approximately 1.5MΩ±497kΩ. The signal to noise ratio was measured and found to be 19.4±3dB compared to 3.9±0.8dB S/N for signals obtained with voltage sensitive dye RH414. A mouse unfolded hippocampus preparation was bathed in solution containing 50 micro-molar 4-amino pyridine and a complex two dimensional wave of activity was recorded using the array. These results indicate that this novel penetrating electrode array is able to obtain data superior to that of voltage sensitive dye techniques for broad field two-dimensional neuronal activity recording. When used with the unfolded hippocampus preparation, the combination forms a uniquely capable tool for imaging hippocampal network activity in the entire hippocampus.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Some neural interface devices have been fabricated with high electrode densities, but their designs have limited their use to specific central nervous system structures such as cortical grey matter or hippocampal slices.
“Some neural interface devices have been fabricated with high electrode densities; however, their designs have limited the use of these arrays to specific central nervous system structures, such as cortical grey matter (6.25 electrodes/mm 2 [12,13]; >150 electrodes/mm 2 [17,18]) or hippocampal slices (<11 electrodes/mm 2 [19]).”
From A new high-density (25 electrodes/mm<sup>2</sup>) penetrating microelectrode array for recording and stimulating sub-millimeter neuroanatomical structures (2013) · cites this paper as [19]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
A miniature microelectrode array to monitor the bioelectric activity of cultured cells
THOMASJR, C., SPRINGER, P., Loeb, G. E., Berwald‐Netter, Y., & Okun, L. M. (1972) · Experimental Cell Research · doi:10.1016/0014-4827(72)90481-8 · Cited by 580 · see citing works
Aghajanian, G. K., & Rasmussen, K. (1989) · Synapse · doi:10.1002/syn.890030406 · Cited by 445 · see citing works
Perreault, P., & Avoli, M. (1991) · Journal of Neurophysiology · doi:10.1152/jn.1991.65.4.771 · Cited by 330 · see citing works
The transverse hippocampal slice: a well-defined cortical structure maintainedin vitro
Skrede, K. K., & Westgaard, R. H. (1971) · Brain Research · doi:10.1016/0006-8993(71)90508-7 · Cited by 293 · see citing works
Heuschkel, M. O., Fejtl, M., Raggenbass, M., Bertrand, D., & Renaud, P. (2002) · Journal of Neuroscience Methods · doi:10.1016/s0165-0270(01)00514-3 · Cited by 291 · see citing works
Brain slice preparation: Hippocampus
Teyler, T. J. (1980) · Brain Research Bulletin · doi:10.1016/s0361-9230(80)80009-8 · Cited by 192 · see citing works
256-Channel Neural Recording and Delta Compression Microsystem With 3D Electrodes
Aziz, J. N. Y., Abdelhalim, K., Shulyzki, R., Genov, R., Bardakjian, B. L., Derchansky, M., Serletis, D., & Carlen, P. L. (2009) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2008.2010997 · Cited by 184 · see citing works
Desai, S. A. (2010) · Frontiers in Neuroengineering · doi:10.3389/fneng.2010.00005 · Cited by 124 · see citing works
Optimizing recording capabilities of the Utah Intracortical Electrode Array
Nordhausen, C. T., Rousche, P. J., & Normann, R. A. (1994) · Brain Research · doi:10.1016/0006-8993(94)91213-0 · Cited by 100 · see citing works
A fundamental oscillatory state of isolated rodent hippocampus
Wu, C., Shen, H., Luk, W. P., & Zhang, L. (2002) · The Journal of Physiology · doi:10.1113/jphysiol.2001.013441 · Cited by 92 · see citing works
Kloosterman, F., Peloquin, P., & Leung, L.-W. S. (2001) · Journal of Neurophysiology · doi:10.1152/jn.2001.86.5.2435 · Cited by 91 · see citing works
Surface-modified microelectrode array with flake nanostructure for neural recording and stimulation
Kim, J., Kang, G., Nam, Y., & Choi, Y. (2010) · Nanotechnology · doi:10.1088/0957-4484/21/8/085303 · Cited by 90 · see citing works
Derchansky, M., Rokni, D., Rick, J. T., et al. (2006) · Neurobiology of Disease · doi:10.1016/j.nbd.2006.03.014 · Cited by 80 · see citing works
BioMEA™: A versatile high-density 3D microelectrode array system using integrated electronics
Charvet, G., Rousseau, L., Billoint, O., Gharbi, S., Rostaing, J.-P., Joucla, S., Trevisiol, M., Bourgerette, A., Chauvet, P., Moulin, C., et al. (2010) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2010.01.001 · Cited by 73 · see citing works
An array of Pt-tip microelectrodes for extracellular monitoring of activity of brain slices.
Thiebaud, P., Beuret, C., Koudelka-Hep, M., Bove, M., Martinoia, S., Grattarola, M., Jahnsen, H., Rebaudo, R., Balestrino, M., Zimmer, J., et al. (1999) · Biosensors & bioelectronics · doi:10.1016/s0956-5663(98)00098-0 · Cited by 68 · see citing works
The adult rat hippocampal slice revisited with multi-electrode arrays
Steidl, E.-M., Neveu, E., Bertrand, D., & Buisson, B. (2006) · Brain Research · doi:10.1016/j.brainres.2006.04.034 · Cited by 66 · see citing works
Roughened polysilicon for low impedance microelectrodes in neural probes
Paik, S., Park, Y., & Cho, D. (2003) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/13/3/304 · Cited by 59 · see citing works
Nam, Y., Wheeler, B. C., & Heuschkel, M. (2006) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2006.01.014 · Cited by 45 · see citing works
A novel masking method for high aspect ratio penetrating microelectrode arrays
Bhandari, R., Negi, S., Rieth, L. W., Normann, R. A., & Solzbacher, F. (2009) · Journal of Micromechanics and Microengineering · doi:10.1088/0960-1317/19/3/035004 · Cited by 38 · see citing works
Orthogonal wave propagation of epileptiform activity in the planar mouse hippocampus in vitro
Kibler, A. B., & Durand, D. M. (2011) · Epilepsia · doi:10.1111/j.1528-1167.2011.03125.x · Cited by 32 · see citing works
Self-citationGhezzi, D., Menegon, A., Pedrocchi, A., Valtorta, F., & Ferrigno, G. (2008) · Journal of Neuroscience Methods · doi:10.1016/j.jneumeth.2008.08.003 · Cited by 27 · see citing works
Charvet, G., Billoint, O., Gharbi, S., Heuschkel, M. O., Georges, C., Kauffmann, T., Pellissier, A., Yvert, B., & Guillemaud, R. (2010) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2010.5626403 · Cited by 10 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Wark, H. A. C., Sharma, R., Mathews, K. S., Fernandez, E., Yoo, J., Christensen, B., Tresco, P., Rieth, L., Solzbacher, F., Normann, R. A., et al. (2013) · Journal of Neural Engineering · doi:10.1088/1741-2560/10/4/045003 · Cited by 148 · see citing works
Xiang, Z., Yen, S., Sheshadri, S., Wang, J., Lee, S., Liu, Y., Liao, L., Thakor, N. V., & Lee, C. (2016) · Advanced Materials · doi:10.1002/adma.201503423 · Cited by 116 · see citing works
- Authors
- Kibler, A. B., Jamieson, B. G., & Durand, D. M.
- Affiliations
- Case Western Reserve University (US)
- Goddard Space Flight Center (US)
- Corresponding author
- Durand, Dominique M.
- ORCID
- Durand, Dominique M.
- Venue
- Journal of Neuroscience Methods
- Publisher
- Elsevier BV
- Published
- 2012-03
- Volume
- 204
- Issue
- 2
- Pages
- 296-305
- ISSN
- 0165-0270 (ISSN-L)
- PMC
- PMC3273647
- OpenAlex
- W2017397002
- Cited by
- 25
- References
- 22
- Citations by year
- 2026: 1 · 2024: 2 · 2022: 2 · 2021: 2 · 2020: 2 · 2018: 1 · 2016: 2 · 2015: 4 · 2014: 5 · 2013: 3 · 2012: 1
- FWCI
- 1.19
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of Neuroscience Methods (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- Funders
- NIH (R01NS40785)
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Neural dynamics and brain function
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Keywords
- high aspect ratio electrodes, silicon micromachining, electrode impedance, signal-to-noise ratio, 4-aminopyridine, hippocampal network activity
- MeSH terms
- Action Potentials, Animals, Brain Mapping, Electric Stimulation, Electrophysiology, Microelectrodes, Microscopy, Electron, Scanning, Nerve Net, Neurons, 4-Aminopyridine, Potassium Channel Blockers, Tissue Culture Techniques, Mice, CA1 Region, Hippocampal, CA3 Region, Hippocampal, Signal-To-Noise Ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review