A new high-density (25 electrodes/mm2) penetrating microelectrode array for recording and stimulating sub-millimeter neuroanatomical structures
Open accessAbstract
OBJECTIVE: Among the currently available neural interface devices, there has been a need for a penetrating electrode array with a high electrode-count and high electrode-density (the number of electrodes/mm(2)) that can be used for electrophysiological studies of sub-millimeter neuroanatomical structures. We have developed such a penetrating microelectrode array with both a high electrode-density (25 electrodes/mm(2)) and high electrode-count (up to 96 electrodes) for small nervous system structures, based on the existing Utah Slanted Electrode Array (USEA). Such high electrode-density arrays are expected to provide greater access to nerve fibers than the conventionally spaced USEA especially in small diameter nerves. APPROACH: One concern for such high density microelectrode arrays is that they may cause a nerve crush-type injury upon implantation. We evaluated this possibility during acute (<10 h) in vivo experiments with electrode arrays implanted into small diameter peripheral nerves of anesthetized rats (sciatic nerve) and cats (pudendal nerve). MAIN RESULTS: Successful intrafascicular implantation and viable nerve function was demonstrated via microstimulation, single-unit recordings and histological analysis. Measurements of the electrode impedances and quantified electrode dimensions demonstrated fabrication quality. The results of these experiments show that such high density neural interfaces can be implanted acutely into neural tissue without causing a complete nerve crush injury, while mediating intrafascicular access to fibers in small diameter peripheral nerves. SIGNIFICANCE: This new penetrating microelectrode array has characteristics un-matched by other neural interface devices currently available for peripheral nervous system neurophysiological research.
AI snapshot
Written from the paper’s full text.
TL;DR
The paper reports a high-density Utah Slanted Electrode Array (HD-USEA) with 25 electrodes/mm² and up to 96 electrodes for penetrating small peripheral nerves. Acute rat sciatic and cat pudendal implants showed viable recording/stimulation and no complete nerve crush injury.
Key findings
- HD-USEAs were fabricated at 25 electrodes/mm² with up to 96 electrodes; final wired arrays used 4 or 5 × 10 electrodes with targeted lengths of 300–800 μm and 200 μm spacing. Across ten devices, average longest and shortest electrode lengths ranged from 884 ± 39 to 310 ± 60 μm.
- In acute rat sciatic nerve experiments, >25 of 48 wired electrodes recorded single- or multiunit activity in each preparation, with mean SNR >4. Stimulation via individual electrodes selectively evoked ankle flexion in 15 ± 6 electrodes and extension in 17 ± 8 electrodes, with selectivity at tip-to-tip separation as low as 208 μm.
- Histology from rat sciatic nerves (n = 2) and a feline pudendal nerve (n = 1) showed no complete nerve crush injury; damage was limited to electrode tracks, with little or no compression around electrode tips or shanks. One feline pudendal electrode was broken before implantation and did not penetrate the fascicular space.
- Tip de-insulation by oxygen plasma produced GSA of 4543 ± 526 μm² (n = 17 tips from one device), while excimer laser plus plasma produced 473 ± 365 μm² (n = 22 tips from one device; significantly lower, p = 1.06 × 10⁻²⁶). In vitro mean tip impedances were 440 ± 340 kΩ for laser-de-insulated and 179 ± 216 kΩ for plasma-de-insulated electrodes.
Methods
- Devices were fabricated from boron-doped silicon wafers using back-side dicing, glassing, metallization, front-side dicing, wet etching, Ti/IrOx tip metallization, parylene-C coating, and tip de-insulation by oxygen plasma etching or excimer laser ablation.
- In vitro characterization included 1 kHz, 10 mV impedance testing in saline, SEM imaging, and optical microscopy for electrode dimensions and geometric surface area.
- In vivo validation used acute (<10 h) implants in anesthetized Sprague-Dawley rats (sciatic nerve, 350–500 g) and cats (pudendal nerve), with pneumatic insertion, 1100-channel stimulation and EMG recording, Cerebus data acquisition, and serial semithin histology.
Limitations
- Stated: only acute (<10 h) implantation was studied; chronic damage, long-term efficacy, and biocompatibility were not assessed. Laser de-insulation was manual and less uniform, and further optimization is needed for higher-density arrays; foil-mask de-insulation still produced relatively large GSAs. Feline preparations were histology only, and lower stimulation voltages were not tested because of the stimulator’s lower limit.
- Evident: sample sizes were small (physiology n = 3 rats; histology n = 2 rat sciatic nerves and n = 1 cat pudendal nerve), and the absence of chronic data limits conclusions about long-term safety and function.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 38 cited sources checked · 0 supported · 0 flagged
- Findingown data
Measurements of electrode impedances and quantified electrode dimensions demonstrated fabrication quality.
“Measurements of the electrode impedances and quantified electrode dimensions demonstrated fabrication quality.”
Quote in the paperNot a substantial claim - Findingnone given
The authors developed a penetrating electrode array with a 25 electrodes/mm2 density, called the High-Density Utah Slanted Electrode Array (HD-USEA).
“We have recently developed a penetrating electrode array with a 25 electrodes/mm 2 density (called the High-Density Utah Slanted Electrode Array or HD-USEA).”
Quote in the paperSubstantial claim - Findingcitation
Physiological and histological evaluations showed no evidence of complete nerve trunk injury after acute (<10 h) implantation of HD-USEAs.
“Physiological and histological evaluations showed no evidence of complete nerve trunk injury [23] after acute (<10 h) implantation of HD-USEAs.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [23] Lee S K, 2000, J. Am. Acad. Orthop. Surg. source
Not checked in the free run.
- Findingown data
The average lengths of the longest and shortest electrodes measured from ten HD-USEAs ranged from 884 ± 39 to 310 ± 60 μm.
“The average lengths of the longest and shortest electrodes measured from these ten devices ranged from 884 ± 39 to 310 ± 60 μm, respectively (averages were calculated from all ten electrode rows).”
Quote in the paperSubstantial claim - Findingown data
Selective activation of flexor versus extensor muscles was achieved by delivering stimulation via two catercornered electrodes with an approximate tip-to-tip separation of 208 μm.
“This selective activation of flexor versus extensor muscles was achieved by delivering stimulation via two catercornered electrodes with an approximate tip-to-tip separation of 208 μm.”
Quote in the paperSubstantial claim
- Findingown data
In each rat experiment, more than 25 of the 48 wired electrodes were able to record either single-unit or multiunit waveforms.
“In each experiment, there were >25 electrodes out of the 48 wired electrodes that were able to record either single unit or multiunit waveforms.”
Quote in the paperSubstantial claim - Findingown data
In rat preparations, afferent neuronal activity could be recorded via individual HD-USEA electrodes when evoked by tactile and proprioceptive stimuli.
“In the same rat preparations, we were also able to record afferent neuronal activity via individual electrodes that was evoked by tactile and proprioceptive stimuli.”
Quote in the paperSubstantial claim - Findingown data
Electrical stimulation of rat sciatic nerve fibers delivered via individual HD-USEA electrodes could selectively evoke hind limb muscle twitches.
“We observed that electrical stimulation of rat sciatic nerve fibers delivered via individual HD-USEA electrodes could selectively evoke hind limb muscle twitches.”
Quote in the paperSubstantial claim - Findingown data
Delivering voltage-controlled stimulation via individual electrodes on the implanted HD-USEA evoked selective muscle twitches in the ankle flexor and extensor muscles.
“Delivering voltage-controlled stimulation via individual electrodes on the implanted HD-USEA evoked selective muscle twitches in the ankle flexor and extensor muscles.”
Quote in the paperSubstantial claim - Findingown data
The mean signal-to-noise ratio for all three rat preparations across all recording channels was greater than 4.
“The mean SNR for all three preparations across all recording channels was >4.”
Quote in the paperNot a substantial claim - Findingown data
The impedances of electrodes de-insulated via plasma etching (n = three arrays) were 440 ± 340 and 179 ± 216 (mean ± sd) kΩ.
“etching (n = three arrays) were 440 ± 340 and 179 ± 216 (mean ± sd) k , respectively.”
Quote in the paperNot a substantial claim - Findingnone given
The authors successfully fabricated 25 electrodes/mm2 penetrating microelectrode arrays with varying length electrodes designed to access nerve fibers at different depths in small peripheral nerves.
“We have successfully fabricated 25 electrodes/mm 2 penetrating microelectrode arrays with varying length electrodes designed to access nerve fibers that are found at different depths in small peripheral nerves.”
Quote in the paperSubstantial claim - Findingown data
In a feline pudendal nerve implanted with an HD-USEA, there was preservation of the intrafascicular anatomy in the region between electrode tracks.
“As with the sciatic nerve implant, there is preservation of the intrafascicular anatomy in the region between electrode tracks.”
Quote in the paperNot a substantial claim - Findingown data
The device de-insulated using the excimer laser technique had a large drop in mean tip impedance to 232 ± 150 kΩ when measured in vivo.
“The device de-insulated using the excimer laser technique had a large drop in mean tip impedance to 232 ± 150 k when measured in vivo.”
Quote in the paperSubstantial claim - Findingown data
Analysis of electrode shank widths at different locations along their lengths across ten electrodes on ten different devices demonstrated uniformity achieved during fabrication.
“Analysis of the widths of the electrode shanks at different locations along their lengths across ten electrodes on ten different devices demonstrates the uniformity that was achieved during fabrication of these devices.”
Quote in the paperNot a substantial claim - Findingcitation
Mean geometric surface areas for laser de-insulated electrodes were well within the safe limits established for penetrating microelectrodes (GSA 2000 μm2).
“The mean GSAs for the laser de-insulated electrodes were well within the safe limits established for penetrating microelectrodes (GSA 2000 μm 2 ) [29]; however, more work is needed to decrease the GSA of electrodes de-insulated using the foil mask technique.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [29] source
Not checked in the free run.
- Findingown data
Mean geometric surface areas obtained with laser ablation de-insulation (473 ± 365 μm2) were significantly lower than those from plasma etching (p = 1.06 × 10−26).
“The mean GSAs obtained with the laser ablation de-insulation (473 ± 365 μm 2 ) were significantly lower (alpha = 0.05; p = 1.06 × 10 -26 ; two-tailed student t-test); however, this technique produced much less uniform tip exposures (larger distribution around the mean) likely due to the manual nature of the process.”
- Also stated in: § 5. Conclusions · plain
“The average GSA of electrode tips de-insulated with the oxygen plasma etching (figure1(b)) and laser ablation (figure1(c)) methods were 4543 ± 526 μm 2 (mean ± sd; n = 17 tips from one device) and 473 ± 365 μm 2 (n = 22 tips from one device), respectively.”
Quote in the paperSubstantial claim - Also stated in: § 5. Conclusions · plain
- Findingown data
In all three rat preparations, selective contractions of ankle flexor and extensor muscles were evoked by voltage-controlled stimulation via individual HD-USEA electrodes.
“In all three rat preparations, we were able to evoke selective contractions of ankle flexor and extensor muscles, and importantly, such selectivity could be achieved in electrodes with a tip-totip separation as low as 208 μm between electrodes.”
Quote in the paperSubstantial claim - Findingown data
Histological analyses of HD-USEA implant sites showed the nerve axons and fascicles appear to manifest no global damage.
“Importantly, histological analyses of the HD-USEA implant sites showed the nerve axons and fascicles appear to manifest no global damage.”
Quote in the paperSubstantial claim - Findingown data
There was little compression of nerve fibers around the tips and shanks of the HD-USEA electrodes.
“Moreover, there was little compression of the nerve fibers around the tips and shanks of the HD-USEA electrodes, as had been previously noted for the conventionally spaced USEAs [1].”
- Also stated in: § 5. Conclusions · plain
“Notice there is no axonal compression around the tips (d)-(e) of the electrodes or around the shanks (f) of the electrodes.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [1] source
Not checked in the free run.
- Also stated in: § 5. Conclusions · plain
- Findingown data
Cross-sections of acutely implanted rat sciatic nerves (n = 2) and a feline pudendal nerve (n = 1) showed no signs of these types of nerve injuries outside the tracks of the penetrating electrodes.
“The cross-sections of acutely implanted rat sciatic nerves (n = 2) and a feline pudendal nerve (n = 1) showed no signs of these types of nerve injuries outside the tracks of the penetrating electrodes.”
Quote in the paperSubstantial claim - Findingown data
Selective activation of ankle flexor and extensor muscles could be achieved with electrodes having tip-to-tip separation as low as 208 μm.
“In the rat preparation, selective activation of ankle flexor and extensor muscles was achieved by delivering stimulation via two neighboring electrodes in adjacent rows (≈ tip-to-tip separation of 208 μm).”
- Also stated in: § 4.2. In vivo characterization of HD-USEAs · hedged
“such selectivity could be achieved in electrodes with a tip-totip separation as low as 208 μm between electrodes.”
Quote in the paperSubstantial claim - Also stated in: § 4.2. In vivo characterization of HD-USEAs · hedged
- Findingown data
The shortest HD-USEA electrodes were able to penetrate into fascicles.
“Figure 5 shows that, in fact, the shortest electrodes were able to do so.”
- Also stated in: § 5. Conclusions · plain
“Microscopic cross section of the nerve shows penetration of a row of shorter length electrodes into the top portion of the rat sciatic nerve fascicles (scale bar = 200 μm).”
Quote in the paperSubstantial claim - Also stated in: § 5. Conclusions · plain
- Findingown data
Tissue insult from HD-USEA implantation was limited to that produced by the penetrating electrode tips, similar to implantation of arrays four times less dense.
“Tissue insult was limited to that which was produced by the penetrating electrode tips, similar to implantation of arrays that were four times less dense.”
Quote in the paperSubstantial claim - Findingown data
In cat pudendal nerve implanted with HD-USEA, no evidence of complete crush injury or nerve compression was observed.
“Additionally, no evidence of complete crush injury or nerve compression is observed.”
Quote in the paperSubstantial claim - Findingown data
Single or multiunit waveforms were recorded from the 300 μm long electrodes in one animal.
“In addition, we were able to record either single or multiunit waveforms from the 300 μm long electrodes in one animal.”
Quote in the paperSubstantial claim - Findingown data
A microscopic cross section of the rat sciatic nerve shows penetration of shorter HD-USEA microelectrodes (<400 μm in length) with preservation of intrafascicular anatomy between electrode penetration tracks.
“A microscopic cross section of the 1.8 × 0.8 mm nerve shows penetration of the shorter HD-USEA microelectrodes (<400 μm in length) with preservation of the intrafascicular anatomy (epineurial, endoneurial and perineurial) between electrode penetration tracks into the nerve (figure5(c)).”
Quote in the paperSubstantial claim - Findingown data
The three electrode tracks that penetrated into the endoneurium extended approximately 450-500 μm from the edge of the epineurium to the electrode tips.
“The three tracks that penetrate into the endoneurium extend approximately 450-500 μm from the edge of the epineurium to the electrode tips (scale bar = 200 μm).”
Quote in the paperSubstantial claim - Findingown data
In cat pudendal nerve, fibers immediately surrounding the electrode track showed minimal signs of acute damage.
“While substantial loss of fibers in the electrode track is observed, fibers immediately surrounding the track show minimal signs of acute damage.”
Quote in the paperSubstantial claim - Findingown data
In cat pudendal nerve implanted with HD-USEA for <10 h, three of four electrodes penetrated the fascicular space.
“Low-resolution image showing penetration of four electrodes into the nerve, three of which penetrated the fascicular space.”
Quote in the paperSubstantial claim - Findingown data
The results reported herein illustrate that such a delicate architecture can be fabricated consistently.
“The results reported herein illustrate that such a delicate architecture can be fabricated consistently, and importantly, that implantation of such high-density electrode arrays does not cause nerve damage, at least on the acute time frame (<10 h) studied here.”
Quote in the paperSubstantial claim - Findingown data
Selective stimulation of flexor or extensor ankle muscles was achieved by delivering low amplitude (-2 V, steps of increasing pulse-width of 50 μs) stimulation to the 300 μm length electrodes.
“We were also able to selectively stimulate flexor or extensor ankle muscles by delivering low amplitude (-2 V, steps of increasing pulse-width of 50 μs) stimulation to the 300 μm length electrodes.”
Quote in the paperSubstantial claim - Findingown data
In cat pudendal nerve, the electrode penetrated through the epineurium and perineurium into the endoneurial space.
“Close up view of an electrode track from (a) (denoted by arrow) showing penetration of the electrode through the epineurium and perineurium into the endoneurial space (scale bar = 100 μm).”
Quote in the paperSubstantial claim - Findingown data
The desired geometric surface area of the exposed SIROF electrode tips was achieved by oxygen plasma etching and laser ablation techniques.
“The desired GSA of the exposed SIROF electrode tips was achieved by oxygen plasma etching and laser ablation techniques.”
Quote in the paperSubstantial claim - Premisecitation
Current peripheral nerve interfaces have electrodes that vary in number of contacts, separated by hundreds of micrometers or several millimeters.
“Current peripheral nerve interfaces have electrodes that vary in their number of contacts, which can be separated by hundreds of micrometers [2,4,5,[11][12][13] or several millimeters [3,7,14].”
Quote in the paperSubstantial claim0 of 4 sources checked - Premisecitation
Recent penetrating neural interface applications require a greater number of electrodes (>60) to achieve complex neuromodulation or neurophysiological goals.
“Recent penetrating neural interface applications are requiring a greater number of electrodes (>60) to achieve complex neuromodulation [9,10,16] or neurophysiological [8] goals.”
Quote in the paperSubstantial claim0 of 4 sources checked - Premisecitation
Many current peripheral nerve interface devices have a low number of electrodes (<10) implanted interfascicularly, intrafascicularly, or epineurally.
“Many of the devices have a low number of electrodes (<10) that are implanted interfascicularly (e.g.slowly penetrating interfascicular electrode, SPINE [6]), intrafascicularly (e.g.longitudinal intrafascicular electrode, LIFE [15]), or epineurally (e.g.flat interface nerve electrode, FINE [7]).”
Quote in the paperSubstantial claim0 of 3 sources checked - Premisecitation
Currently, there is one high electrode-density (6.25 electrodes/mm2), high electrode-count (96 electrodes) penetrating device available for neuromodulation of large structures (>4 mm) in the peripheral nervous system.
“Currently, there is one high electrode-density (6.25 electrodes/mm 2 ), high electrodecount (96 electrodes) penetrating device available for neuromodulation of large structures (> 4 mm) in the peripheral nervous system [1].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [1] source
Not checked in the free run.
- Premisecitation
Some neural interface devices have been fabricated with high electrode densities, but their designs have limited their use to specific central nervous system structures such as cortical grey matter or hippocampal slices.
“Some neural interface devices have been fabricated with high electrode densities; however, their designs have limited the use of these arrays to specific central nervous system structures, such as cortical grey matter (6.25 electrodes/mm 2 [12,13]; >150 electrodes/mm 2 [17,18]) or hippocampal slices (<11 electrodes/mm 2 [19]).”
Quote in the paperSubstantial claim0 of 5 sources checked - Premisecitation
Production of a defined geometric surface area (GSA) of the electrode tips was possible using either an oxygen plasma etching technique or excimer laser ablation.
“We have shown that the production of a defined geometric surface area (GSA) of the electrode tips was possible using either an oxygen plasma etching technique [1,21] or by excimer laser ablation [22].”
Quote in the paperSubstantial claim0 of 3 sources checked - Interpretationreasoning
As few as 6 electrodes of the conventionally spaced arrays might be functionally implanted into a 1 mm nerve.
“Thus, as few as 6 electrodes of the conventionally spaced arrays might be functionally implanted into a 1 mm nerve.”
Quote in the paperSubstantial claim - Interpretationreasoning
Implantation of a 5 × 10 conventional array (0.4 mm electrode spacing with lengths from 0.5 to 1.5 mm) into a 1 mm nerve would result in partial coverage with at most only two electrode rows penetrating into the nerve (20 electrodes).
“For example, implantation of a 5 × 10 conventional array (0.4 mm electrode spacing with lengths from 0.5 to 1.5 mm) into a 1 mm nerve would result in partial coverage with at most only two electrode rows penetrating into the nerve (20 electrodes).”
Quote in the paperSubstantial claim - Interpretationown data
The nerve remained viable for at least the duration of these experiments (<10 h).
“We were able to record spontaneous and stimulus driven neuronal units on >25 electrodes with an average SNR >4 in all animal preparations, indicating the nerve remained viable for at least the duration of these experiments (<10 h).”
Quote in the paperSubstantial claim - Premisecitation
The carbon debris layer is thinner than the parylene-C layer.
“The carbon debris layer is thinner than the parylene-C layer [22], and thus, we hypothesized that the etching of the carbon debris could be carried out in 2 min without the need of a foil mask.”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [22] source
Not checked in the free run.
- Interpretationown data
Implantation of such high-density electrode arrays does not cause nerve damage, at least on the acute time frame (<10 h) studied here.
“The results reported herein illustrate that such a delicate architecture can be fabricated consistently, and importantly, that implantation of such high-density electrode arrays does not cause nerve damage, at least on the acute time frame (<10 h) studied here.”
Quote in the paperSubstantial claim - Interpretationown data
The HD-USEA provides selective access to the fibers of small diameter peripheral nerves.
“Furthermore, the HD-USEA provides selective access to the fibers of small diameter peripheral nerves.”
Quote in the paperSubstantial claim
- May combine two claims. 20
- Classified as probably not a substantial claim. 12
- The quote only partly states this claim. 1
May combine two claims.
Findingnone givenThe HD-USEA grid used in these studies had graded electrode lengths ranging from 300 to 800 μm with neighboring electrodes separated by 200 μm.
“The grid of electrodes used in these studies had graded electrode lengths ranging from 300 to 800 μm with neighboring electrodes separated by 200 μm.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataSuccessful intrafascicular implantation and viable nerve function were demonstrated via microstimulation, single-unit recordings, and histological analysis.
“Successful intrafascicular implantation and viable nerve function was demonstrated via microstimulation, single-unit recordings and histological analysis.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn all three rat preparations, there were 15 ± 6 and 17 ± 8 electrodes that selectively evoked ankle flexion and extension, respectively.
“In all three rat preparations, there were 15 ± 6 and 17 ± 8 (mean ± sd) electrodes that selectively evoked ankle flexion and extension, respectively.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn each preparation, there were neural units that fired spontaneously, and different neural units could be selectively driven by tactile or proprioceptive stimuli.
“In each preparation, there were neural units that fired spontaneously, and different neural units that could be selectively driven by tactile or proprioceptive stimuli.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataHistological analysis of the implantation sites verified that multiple HD-USEA electrodes penetrated into individual fascicles (<500 μm in diameter) of both the rat sciatic and feline pudendal nerves with little or no evidence of nerve fiber compression around the electrode tips.
“Histological analysis of the implantation sites verified that multiple HD-USEA electrodes penetrated into the individual fascicles (<500 μm in diameter) of both the rat sciatic and feline pudendal nerves with little or no evidence of nerve fiber compression around the electrode tips.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataOne electrode was broken prior to implantation and did not penetrate into the fascicular space.
“One electrode was broken prior to implantation and does not penetrate into the fascicular space (track marked with an arrowhead; figure 6(a)).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingnone givenOne wafer yields approximately 20 arrays that do not have any broken electrodes.
“One wafer yields approximately 20 arrays that do not have any broken electrodes.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingnone givenUp to 96 electrodes could be wire-bonded for use in other nerve structures.
“This number of electrodes was chosen to best fit the dimensions of the peripheral nerves tested in this work, but up to 96 electrodes could be wire-bonded for use in other nerve structures.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataDe-insulating HD-USEAs using a foil mask and oxygen plasma etching resulted in uniform geometric surface areas (standard deviation 526 μm2) but relatively large areas (4543 μm2).
“De-insulating HD-USEAs using the foil mask and oxygen plasma etching resulted in uniform GSAs (small distribution with standard deviation of 526 μm 2 ), but relatively large areas (4543 μm 2 ).”
- Also stated in: § 5. Conclusions · plain
“The average GSA of electrode tips de-insulated with the oxygen plasma etching (figure1(b)) and laser ablation (figure1(c)) methods were 4543 ± 526 μm 2 (mean ± sd; n = 17 tips from one device) and 473 ± 365 μm 2 (n = 22 tips from one device), respectively.”
Quote in the paperSubstantial claim- Also stated in: § 5. Conclusions · plain
The quote only partly states this claim.
FindingreasoningLaser ablation de-insulation produced much less uniform tip exposures (larger distribution around the mean) than plasma etching.
“The mean GSAs obtained with the laser ablation de-insulation (473 ± 365 μm 2 ) were significantly lower (alpha = 0.05; p = 1.06 × 10 -26 ; two-tailed student t-test); however, this technique produced much less uniform tip exposures (larger distribution around the mean) likely due to the manual nature of the process.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataSelective and complete recruitment of rat hind limb muscles was possible with stimulation amplitudes of only -2 V, with 50% recruitment of muscle twitches seen in some trials for pulse-widths <100 μs.
“Selective and complete recruitment of rat hind limb muscles was possible by delivering stimulation amplitudes of only -2 V with 50% recruitment of muscle twitches seen in some trials for pulse-widths <100 μs.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn all animal preparations, spontaneous and stimulus driven neuronal units were recorded on >25 electrodes with an average SNR >4.
“We were able to record spontaneous and stimulus driven neuronal units on >25 electrodes with an average SNR >4 in all animal preparations, indicating the nerve remained viable for at least the duration of these experiments (<10 h).”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataIn rat sciatic nerve, electrode 21 recorded neural responses only during dorsiflexion, while electrode 14 recorded responses only during plantarflexion.
“Note that electrode 21 recorded neural responses only during dorsiflexion, while electrode 14 recorded responses only during plantarflexion.”
- Also stated in: § 5. Conclusions · plain
“968 waveforms show the summation of the responses from two units to dorsiflexion of the foot around the ankle joint recorded by electrode 21 (SNR = 5.9).”
- Also stated in: § 5. Conclusions · plain
“1416 waveforms show a unit recorded on electrode 14 that was driven by plantarflexion (SNR = 10.8).”
Quote in the paperSubstantial claim- Also stated in: § 5. Conclusions · plain
May combine two claims.
Findingown dataThe authors developed a novel, high electrode-density (25 electrodes/mm2), high electrode-count (48-96 electrodes) penetrating neural interface designed for stimulation and recording in sub-millimeter neuroanatomical structures, such as small diameter nerves (<2 mm).
“We have developed a novel, high electrode-density (25 electrodes/mm 2 ), high electrode-count (48-96 electrodes) penetrating neural interface, designed for use in stimulation and recording in sub-millimeter neuroanatomical structures, such as small diameter nerves (<2 mm) like the rat sciatic [37], feline pudendal [38], or feline/human cochlear nerves [46].”
Quote in the paperSubstantial claim0 of 3 sources checkedClassified as probably not a substantial claim.
Premisenone givenThere has been a need for a penetrating electrode array with high electrode-count and high electrode-density for electrophysiological studies of sub-millimeter neuroanatomical structures.
“Among the currently available neural interface devices, there has been a need for a penetrating electrode array with a high electrode-count and high electrode-density (the number of electrodes/mm 2 ) that can be used for electrophysiological studies of sub-millimeter neuroanatomical structures.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationHigh-count microelectrode arrays have proven useful in basic physiology research and for clinical applications such as restoration of function in people with nervous system injury or stroke.
“Highcount microelectrode arrays have proven useful in basic physiology research [8], and for clinical applications such as the restoration of function in people with nervous system injury [9] or stroke [10].”
Quote in the paperNot a substantial claim0 of 3 sources checkedProbably not a substantial claim: may not pass S4 (based).
Interpretationnone givenHigher density microelectrode arrays would improve access to closely spaced individual or sub-populations of neurons and/or nerve fibers in order to investigate mechanisms underlying more complex spatial neural interactions.
“Moreover, higher density microelectrode arrays would improve the access to closely spaced individual or sub-populations of neurons and/or nerve fibers in order to investigate the mechanisms underlying more complex spatial neural interactions.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
PremisereasoningPenetrating electrode arrays with higher densities have been needed to expand neurophysiology research and to investigate future clinical applications in sub-millimeter neuroanatomical structures, such as small diameter peripheral nerves or ganglia.
“However, penetrating electrode arrays with higher densities have been needed to expand neurophysiology research and to investigate future clinical applications in sub-millimeter neuroanatomical structures, such as small diameter peripheral nerves or ganglia (e.g. the 1 mm cat pudendal nerve).”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataThe architecture of the HD-USEA makes it a unique and useful addition to currently available neural interfaces because it provides unprecedented nerve fiber access in sub-millimeter structures outside the central nervous system.
“The architecture of this device makes it a unique and useful addition to the neural interfaces currently available, as it provides unprecedented nerve fiber access in sub-millimeter structures outside the central nervous system.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataHD-USEAs are safe and effective for acute neurophysiology studies, with chronic use not yet investigated.
“Collectively, these results show that HD-USEAs are safe and effective for acute neurophysiology studies (we have yet to investigate their use in chronic experimentation).”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataThe quality and reproducibility of HD-USEAs is comparable to that of the conventional arrays currently being manufactured.
“The quality and reproducibility of HD-USEAs is comparable to that of the conventional arrays currently being manufactured.”
Quote in the paperNot a substantial claimMay combine two claims.
InterpretationreasoningImplantation of a 5 × 10 HD-USEA would result in more complete coverage (from 0.2 to 1.0 mm) of a 1 mm nerve and with a high-count number of electrodes (48 electrodes).
“In contrast, implantation of a 5 × 10 HD-USEA would result in more complete coverage (from 0.2 to 1.0 mm) of the nerve and with a high-count number of electrodes (48 electrodes).”
Quote in the paperSubstantial claimMay combine two claims.
InterpretationcitationThe electrode proportions, impedances, number of functionally distinct electrodes, and SNR of HD-USEAs are comparable to the conventionally spaced Utah electrode arrays.
“The electrode proportions, impedances, number of functionally distinct electrodes and SNR of HD-USEAs are comparable to the conventionally spaced Utah electrode arrays [1,13].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [1] source
Not checked in the free run.
- cites [13] Campbell P K, 1990, Biomed. Sci. Instrum.
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenAll procedures were approved by the Institutional Animal Care and Use Committee of the University of Utah.
“All procedures were approved by the Institutional Animal Care and Use Committee of the University of Utah.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningThe less uniform tip exposures from laser ablation were likely due to the manual nature of the process.
“however, this technique produced much less uniform tip exposures (larger distribution around the mean) likely due to the manual nature of the process.”
Quote in the paperNot a substantial claimMay combine two claims.
PremisecitationThe conventionally spaced, lower density USEA can be implanted acutely (<1 day) and chronically (>6 months) with insult to the epineurium or perineurium limited to the areas penetrated by the microelectrodes.
“The conventionally spaced, lower density USEA can be implanted both acutely (<1 days) [1] and chronically (>6 months) [40,41] with insult to the epineurium or perinerium limited to the areas penetrated by the microelectrodes.”
Quote in the paperNot a substantial claim0 of 3 sources checkedProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductcitationLower voltages were not investigated because -2 V is the lowest limit of the 1100 channel stimulator used in these studies.
“Lower voltages were not investigated, as this is the lowest limit of the 1100 channel stimulator used in these studies [33].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [33] Hiatt S D, 2010, 15th Annu. Conf. Int. Functional Electrical Stimulation Society
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenThe laser ablation technique (laser spot-size) was optimized for lower-density (6.25 electrodes/mm2) microelectrode arrays, and more work is needed to optimize parameters for higher-density neural interfaces.
“The laser ablation technique (laser spot-size) was optimized for lower-density (6.25 electrodes/mm 2 ) microelectrodes arrays, and more work is needed to optimize the parameters for performing laser ablation on higher-density neural interfaces.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
PremisecitationThe plasma etching deinsulation process can be carried out on individual electrode arrays or, for microelectrode arrays with uniform electrode lengths, on the wafer scale.
“This process can be carried out on individual electrode arrays [13] or-for microelectrode arrays that have uniform electrode lengths-on the wafer scale [24].”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [13] Campbell P K, 1990, Biomed. Sci. Instrum.
Not checked in the free run.
- cites [24] source
Not checked in the free run.
The paper relies on this without support.
High-density penetrating microelectrode arrays may cause a nerve crush-type injury upon implantation.
“One concern for such high density microelectrode arrays is that they may cause a nerve crush-type injury upon implantation.”
“One of the main concerns regarding the use of penetrating microelectrode arrays that have such high electrode densities is that a nerve crush injury may result from their implantation.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The Utah Slanted Electrode Array (USEA) provides comprehensive access to multiple independent motor nerves.
“The Utah Slanted Electrode Array (USEA) was proposed to provide comprehensive access to multiple independent motor nerves27,28”
From A flexible three-dimensional electrode mesh: An enabling technology for wireless brain–computer interface prostheses (2016) · cites this paper as [28]
The length of the Utah Slanted Electrode Array was restricted by the bulk silicon thickness and the etching process.
“The length of USEA was restricted by the bulk silicon thickness and the etching process, while for the varying polymer microneedle array, the longest microneedles were fixed in the central area due to the droplet profile.”
From A flexible three-dimensional electrode mesh: An enabling technology for wireless brain–computer interface prostheses (2016) · cites this paper as [28]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Conclusion says implantation causes no nerve damage despite reported fiber loss
MajorThe conclusion states that implantation does not cause nerve damage, but the figure caption reports substantial fiber loss in an electrode track. The observations support a narrower claim about the absence of complete crush injury, not the absence of damage.
“implantation of such high-density electrode arrays does not cause nerve damage, at least on the acute time frame (<10 h) studied here.”
“While substantial loss of fibers in the electrode track is observed, fibers immediately surrounding the track show minimal signs of acute damage.”
The citation lists haven’t been fetched yet
References 48 · Cited by 148
Research works printed but not linked: 3
- Campbell P K, 1990, Biomed. Sci. Instrum., 26, 161
- Cogan S F Ehrlich J Plante T D Van Wagenen R, 2009, Conf. Proc. IEEE Eng. Med. Biol. Soc., 7147
- Hiatt S D, 2010, 15th Annu. Conf. Int. Functional Electrical Stimulation Society, 081
- Authors
- Wark, H. A. C., Sharma, R., Mathews, K. S., Fernandez, E., Yoo, J., Christensen, B., Tresco, P., Rieth, L., Solzbacher, F., Normann, R. A., & Tathireddy, P.
- Affiliations
- University of Utah (US)
- Corresponding author
- Normann, R A
- Venue
- Journal of Neural Engineering
- Publisher
- IOP Publishing
- Published
- 2013-05-31
- Volume
- 10
- Issue
- 4
- Pages
- 045003
- ISSN
- 1741-2552 (ISSN-L)
- 1741-2560
- OpenAlex
- W2107611005
- Semantic Scholar
- 07de9cafcc0a30db429173ee33d27b06109d7eef
- Cited by
- 148
- References
- 48
- Citations by year
- 2026: 2 · 2025: 4 · 2024: 13 · 2023: 7 · 2022: 6 · 2021: 14 · 2020: 14 · 2019: 12 · 2018: 14 · 2017: 18 · 2016: 14 · 2015: 15 · 2014: 12 · 2013: 3
- FWCI
- 7.20
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Journal of Neural Engineering (publishedVersion, cc-by-nc-sa)
- PubMed (publishedVersion)
- Funders
- National Science Foundation (1134545)
- Ministerio de Economía y Competitividad
- Division of Chemical, Bioengineering, Environmental, and Transport Systems
- Topics
- Life Sciences › Neuroscience › Cellular and Molecular Neuroscience › Neuroscience and Neural Engineering
- Life Sciences › Neuroscience › Cognitive Neuroscience › Undergraduate Neuroscience Education and Research
- Physical Sciences › Engineering › Biomedical Engineering › Analog and Mixed-Signal Circuit Design
- Keywords
- high-density electrode arrays, peripheral nerve recording, peripheral nerve stimulation, nerve crush injury, sciatic nerve, pudendal nerve, single-unit recording, microstimulation, electrode impedance
- MeSH terms
- Animals, Cats, Diagnostic Techniques, Neurological, Electric Stimulation, Electrodes, Implanted, Equipment Design, Microelectrodes, Peripheral Nerves, Electric Impedance, Rats, Sprague-Dawley, Equipment Failure Analysis, Nanotechnology, Microarray Analysis, Rats
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review