A comparison of two hybrid closed-loop systems in adolescents and young adults with type 1 diabetes (FLAIR): a multicentre, randomised, crossover trial
Open accessThe DOI matches its Crossref record; 24 of the 24 references it cites were found in a registry or an index.
Abstract
Background Management of type 1 diabetes is challenging. We compared outcomes using a commercially available hybrid closed-loop system versus a new investigational system with features potentially useful for adolescents and young adults with type 1 diabetes.Methods In this multinational, randomised, crossover trial (Fuzzy Logic Automated Insulin Regulation [FLAIR]), individuals aged 14-29 years old, with a clinical diagnosis of type 1 diabetes with a duration of at least 1 year, using either an insulin pump or multiple daily insulin injections, and glycated haemoglobin (HbA 1c) levels of 7·0-11·0% (53-97 mmol/mol) were recruited from seven academic-based endocrinology practices, four in the USA, and one each in Germany, Israel, and Slovenia. After a run-in period to teach participants how to use the study pump and continuous glucose monitor, participants were randomly assigned (1:1) using a computer-generated sequence, with a permuted block design (block sizes of two and four), stratified by baseline HbA 1c and use of a personal MiniMed 670G system (Medtronic) at enrolment, to either use of a MiniMed 670G hybrid closed-loop system (670G) or the investigational advanced hybrid closed-loop system (Medtronic) for the first 12-week period, and then participants were crossed over with no washout period, to the other group for use for another 12 weeks. Masking was not possible due to the nature of the systems used. The coprimary outcomes, measured with continuous glucose monitoring, were proportion of time that glucose levels were above 180 mg/dL (>10·0 mmol/L) during 0600 h to 2359 h (ie, daytime), tested for superiority, and proportion of time that glucose levels were below 54 mg/dL (<3·0 mmol/L) calculated over a full 24-h period, tested for non-inferiority (non-inferiority margin 2%). Analysis was by intention to treat. Safety was assessed in all participants randomly assigned to treatment. This trial is registered with ClinicalTrials.gov, NCT03040414, and is now complete.Findings Between June 3 and Aug 22, 2019, 113 individuals were enrolled into the trial. Mean age was 19 years (SD 4) and 70 (62%) of 113 participants were female. Mean proportion of time with daytime glucose levels above 180 mg/dL (>10·0 mmol/L) was 42% (SD 13) at baseline, 37% (9) during use of the 670G system, and 34% (9) during use of the advanced hybrid closed-loop system (mean difference [advanced hybrid closed-loop system minus 670G system] -3·00% [95% CI -3·97 to -2·04]; p<0·0001). Mean 24-h proportion of time with glucose levels below 54 mg/dL (<3·0 mmol/L) was 0·46% (SD 0·42) at baseline, 0·50% (0·35) during use of the 670G system, and 0·46% (0·33) during use of the advanced hybrid closed-loop system (mean difference [advanced hybrid closed-loop system minus 670G system] -0·06% [95% CI -0·11 to -0·02]; p<0·0001 for non-inferiority). One severe hypoglycaemic event occurred in the advanced hybrid closed-loop system group, determined to be unrelated to study treatment, and none occurred in the 670G group.Interpretation Hyperglycaemia was reduced without increasing hypoglycaemia in adolescents and young adults with type 1 diabetes using the investigational advanced hybrid closed-loop system compared with the commercially available MiniMed 670G system. Testing an advanced hybrid closed-loop system in populations that are underserved due to socioeconomic factors and testing during pregnancy and in individuals with impaired awareness of hypoglycaemia would advance the effective use of this technology FUNDING: National Institute of Diabetes and Digestive and Kidney Diseases.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“Medtronic 780G, 113,114 CamAPS FX, 115,116 and Diabeloop 117 received CE mark for use in European countries;”
From Developing the UVA/Padova Type 1 Diabetes Simulator: Modeling, Validation, Refinements, and Utility (2023) · cites this paper as [114]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,237
Self-citationOn the adaptive control of the false discovery rate in multiple testing with independent statistics
Benjamini, Y., & Hochberg, Y. (2000) · Journal of Educational and Behavioral Statistics · doi:10.3102/10769986025001060 · Cited by 2,015
Six-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1 Diabetes
Brown, S. A., Kovatchev, B. P., Raghinaru, D., Lum, J. W., Buckingham, B. A., Kudva, Y. C., Laffel, L. M., Levy, C. J., Pinsker, J. E., Wadwa, R. P., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1907863 · Cited by 1,030
Self-citationValidation of Time in Range as an Outcome Measure for Diabetes Clinical Trials
Beck, R. W., Bergenstal, R. M., Riddlesworth, T. D., Kollman, C., Li, Z., Brown, A. S., & Close, K. L. (2019) · Diabetes Care · doi:10.2337/dc18-1444 · Cited by 793
Self-citationSafety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes
Bergenstal, R. M., Garg, S., Weinzimer, S. A., Buckingham, B. A., Bode, B. W., Tamborlane, W. V., & Kaufman, F. R. (2016) · JAMA · doi:10.1001/jama.2016.11708 · Cited by 644
Self-citationGarg, S. K., Weinzimer, S. A., Tamborlane, W. V., Buckingham, B. A., Bode, B. W., Bailey, T. S., Brazg, R. L., Ilany, J., Slover, R. H., Anderson, S. M., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0421 · Cited by 585
Self-citationA Randomized Trial of Closed-Loop Control in Children with Type 1 Diabetes
Breton, M. D., Kanapka, L. G., Beck, R. W., Ekhlaspour, L., Forlenza, G. P., Cengiz, E., Schoelwer, M., Ruedy, K. J., Jost, E., Carria, L., et al. (2020) · New England Journal of Medicine · doi:10.1056/nejmoa2004736 · Cited by 455
Self-citationTauschmann, M., Thabit, H., Bally, L., Allen, J. M., Hartnell, S., Wilinska, M. E., Ruan, Y., Sibayan, J., Kollman, C., Cheng, P., et al. (2018) · The Lancet · doi:10.1016/s0140-6736(18)31947-0 · Cited by 440
Self-citationNocturnal Glucose Control with an Artificial Pancreas at a Diabetes Camp
Phillip, M., Battelino, T., Atlas, E., Kordonouri, O., Bratina, N., Miller, S., Biester, T., Avbelj Stefanija, M., Muller, I., Nimri, R., et al. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1206881 · Cited by 435
Self-citation
Research works printed but not linked: 1
- Isganaitis E, Raghinaru D, Ambler-Osborn L, et al. Closed-loop control (CLC) in teens and young adults improves glycemic control: Results from the International Diabetes Closed-Loop (iDCL) Trial. J Diabetes Sci Technol 2020; 14(2): A53. [Google Scholar]
Cited by
Phillip, M., Nimri, R., Bergenstal, R. M., Barnard-Kelly, K., Danne, T., Hovorka, R., Kovatchev, B. P., Messer, L. H., Parkin, C. G., Ambler-Osborn, L., et al. (2023) · Endocrine Reviews · doi:10.1210/endrev/bnac022 · Cited by 443
Self-citation14. Children and Adolescents: Standards of Medical Care in Diabetes—2022
American Diabetes Association Professional Practice Committee (2022) · Diabetes Care · doi:10.2337/dc22-s014 · Cited by 367
Grunberger, G., Sherr, J., Allende, M., Blevins, T., Bode, B., Handelsman, Y., Hellman, R., Lajara, R., Roberts, V. L., Rodbard, D., et al. (2021) · Endocrine Practice · doi:10.1016/j.eprac.2021.04.008 · Cited by 288
New closed-loop insulin systems
Boughton, C. K., & Hovorka, R. (2021) · Diabetologia · doi:10.1007/s00125-021-05391-w · Cited by 270
Carlson, A. L., Sherr, J. L., Shulman, D. I., Garg, S. K., Pop-Busui, R., Bode, B. W., Lilenquist, D. R., Brazg, R. L., Kaiserman, K. B., Kipnes, M. S., et al. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0319 · Cited by 230
“Although Collyns et al. did not analyze AIT setting, a recent short longitudinal MiniMed 780G system study (n = 52 adolescents and adults, aged 43 – 12 years) showed that at the most aggressive system settings consisting of a 100 mg/dL glucose target and AIT of 2 h, there was an overall 97% of time spent in closed loo…”
BackgroundSelf-citationReal-World Performance of the MiniMed™ 780G System: First Report of Outcomes from 4120 Users
Silva, J. D., Lepore, G., Battelino, T., Arrieta, A., Castañeda, J., Grossman, B., Shin, J., & Cohen, O. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0203 · Cited by 229
“A relatively lower rate of mean targets achieved, as well as the proportion attaining goals, has been reported in two RCTs of 3–6 months duration.(21,22) However, a shorter RCT investigating the same AHCL system used in the FLAIR trial demonstrated greatest improvement in the TIR (+14.”
BackgroundSelf-citationRandomized Trial of Closed-Loop Control in Very Young Children with Type 1 Diabetes
Ware, J., Allen, J. M., Boughton, C. K., Wilinska, M. E., Hartnell, S., Thankamony, A., de Beaufort, C., Schierloh, U., Fröhlich-Reiterer, E., Mader, J. K., et al. (2022) · New England Journal of Medicine · doi:10.1056/nejmoa2111673 · Cited by 222
Self-citationChoudhary, P., Kolassa, R., Keuthage, W., Kroeger, J., Thivolet, C., Evans, M., Ré, R., de Portu, S., Vorrink, L., Shin, J., et al. (2022) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(22)00212-1 · Cited by 204
14. Children and Adolescents: Standards of Care in Diabetes—2023
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s014 · Cited by 200
Yang, K., Yang, X., Jin, C., Ding, S., Liu, T., Ma, B., Sun, H., Zhang, J., & Li, Y. (2024) · BMJ · doi:10.1136/bmj-2023-078432 · Cited by 172
14. Children and Adolescents: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s014 · Cited by 173
14. Children and Adolescents: Standards of Care in Diabetes—2024
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s014 · Cited by 165
Arrieta, A., Battelino, T., Scaramuzza, A. E., Da Silva, J., Castañeda, J., Cordero, T. L., Shin, J., & Cohen, O. (2022) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14714 · Cited by 154
Self-citationISPAD Clinical Practice Consensus Guidelines 2022: Diabetes technologies: Insulin delivery
Sherr, J. L., Schoelwer, M., Dos Santos, T. J., Reddy, L., Biester, T., Galderisi, A., van Dyk, J. C., Hilliard, M. E., Berget, C., & DiMeglio, L. A. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13421 · Cited by 152
Self-citationGlycaemic management in diabetes: old and new approaches
Ceriello, A., Prattichizzo, F., Phillip, M., Hirsch, I. B., Mathieu, C., & Battelino, T. (2022) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(21)00245-x · Cited by 139
Self-citationBoughton, C. K., Hartnell, S., Thabit, H., Mubita, W. M., Draxlbauer, K., Poettler, T., Wilinska, M. E., Hood, K. K., Mader, J. K., Narendran, P., et al. (2022) · The Lancet Healthy Longevity · doi:10.1016/s2666-7568(22)00005-8 · Cited by 127
Self-citationAbraham, M. B., de Bock, M., Smith, G. J., Dart, J., Fairchild, J. M., King, B. R., Ambler, G. R., Cameron, F. J., McAuley, S. A., Keech, A. C., et al. (2021) · JAMA Pediatrics · doi:10.1001/jamapediatrics.2021.3965 · Cited by 121
A Systematic Review of Commercial Hybrid Closed-Loop Automated Insulin Delivery Systems
Peacock, S., Frizelle, I., & Hussain, S. (2023) · Diabetes Therapy · doi:10.1007/s13300-023-01394-5 · Cited by 122
“The following endpoints were the main focus of the systematic review: the percentage of TIR; the change in TIR; HbA1c; and the incidence of severe hypoglycaemia, diabetic ketoacidosis (DKA), or serious adverse events.”
BackgroundISPAD Clinical Practice Consensus Guidelines 2022: Diabetes education in children and adolescents
Olinder, A. L., Deabreu, M., Greene, S. A., Haugstvedt, A., Lange, K., Majaliwa, E., Pais, V., Pélicand, J., Town, M., & Mahmud, F. H. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13418 · Cited by 168
Type 1 diabetes glycemic management: Insulin therapy, glucose monitoring, and automation
Perkins, B. A., Sherr, J. L., & Mathieu, C. (2021) · Science · doi:10.1126/science.abg4502 · Cited by 125
Closed-Loop Therapy and Preservation of C-Peptide Secretion in Type 1 Diabetes
Boughton, C. K., Allen, J. M., Ware, J., Wilinska, M. E., Hartnell, S., Thankamony, A., Randell, T., Ghatak, A., Besser, R. E. J., Elleri, D., et al. (2022) · New England Journal of Medicine · doi:10.1056/nejmoa2203496 · Cited by 108
Self-citationPredictors of time in target glucose range in real‐world users of the MiniMed 780G system
Castañeda, J., Mathieu, C., Aanstoot, H., Arrieta, A., Da Silva, J., Shin, J., & Cohen, O. (2022) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14807 · Cited by 106
Petrovski, G., Campbell, J., Pasha, M., Day, E., Hussain, K., Khalifa, A., & van den Heuvel, T. (2023) · Diabetes Care · doi:10.2337/dc22-1692 · Cited by 108
Artificial intelligence for diabetes care: current and future prospects
Sheng, B., Pushpanathan, K., Guan, Z., Lim, Q. H., Lim, Z. W., Yew, S. M. E., Goh, J. H. L., Bee, Y. M., Sabanayagam, C., Sevdalis, N., et al. (2024) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(24)00154-2 · Cited by 106
Ware, J., Boughton, C. K., Allen, J. M., Wilinska, M. E., Tauschmann, M., Denvir, L., Thankamony, A., Campbell, F. M., Wadwa, R. P., Buckingham, B. A., et al. (2022) · The Lancet Digital Health · doi:10.1016/s2589-7500(22)00020-6 · Cited by 101
Self-citationBiester, T., Berget, C., Boughton, C. K., Cudizio, L., Ekhlaspour, L., Hilliard, M. E., Reddy, L., Sap, S. N. U., Schoelwer, M. J., Sherr, J. L., et al. (2024) · Hormone Research in Paediatrics · doi:10.1159/000543034 · Cited by 131
Self-citationAbraham, M. B., Karges, B., Dovč, K., Naranjo, D. M., Arbeláez, A. M., Mbogo, J., Javelikar, G., Jones, T. W., & Mahmud, F. H. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13443 · Cited by 143
Current Advances of Artificial Pancreas Systems: A Comprehensive Review of the Clinical Evidence
Moon, S. J., Jung, I., & Park, C.-Y. (2021) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2021.0177 · Cited by 98
Crabtree, T. S. J., Griffin, T. P., Yap, Y. W., Narendran, P., Gallen, G., Furlong, N., Cranston, I., Chakera, A., Philbey, C., Karamat, M. A., et al. (2023) · Diabetes Care · doi:10.2337/dc23-0635 · Cited by 88
Beato-Víbora, P. I., Gallego-Gamero, F., Ambrojo-López, A., Gil-Poch, E., Martín-Romo, I., & Arroyo-Díez, F. J. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0037 · Cited by 87
Moon, J. S., Kang, S., Choi, J. H., Lee, K. A., Moon, J. H., Chon, S., Kim, D. J., Kim, H. J., Seo, J. A., Kim, M. K., et al. (2024) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2024.0249 · Cited by 90
Pasqua, M.-R., Tsoukas, M. A., Kobayati, A., Aboznadah, W., Jafar, A., & Haidar, A. (2025) · Nature Medicine · doi:10.1038/s41591-024-03463-z · Cited by 68
Wheeler, B. J., Collyns, O. J., Meier, R. A., Betts, Z. L., Frampton, C., Frewen, C. M., Galland, B. C., Hewapathirana, N. M., Jones, S., Chan, D. S., et al. (2021) · Acta Diabetologica · doi:10.1007/s00592-021-01789-5 · Cited by 77
Current Status and Emerging Options for Automated Insulin Delivery Systems
Forlenza, G. P., & Lal, R. A. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0514 · Cited by 67
Passanisi, S., Piona, C. A., Salzano, G., Marigliano, M., Bombaci, B., Morandi, A., Alibrandi, A., Maffeis, C., & Lombardo, F. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0373 · Cited by 69
Unannounced Meal Challenges Using an Advanced Hybrid Closed-Loop System
Shalit, R., Minsky, N. C., Laron-Hirsh, M., Cohen, O., Kurtz, N., Roy, A., Grosman, B., Benedetti, A., & Tirosh, A. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0139 · Cited by 75
Benhalima, K., Beunen, K., Siegelaar, S. E., Painter, R., Murphy, H. R., Feig, D. S., Donovan, L. E., Polsky, S., Buschur, E., Levy, C. J., et al. (2023) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(23)00116-x · Cited by 61
Self-citationPetrovski, G., Khalaf, F. A., Campbell, J., Day, E., Almajaly, D., Hussain, K., Pasha, M., Umer, F., Hamdan, M., & Khalifa, A. (2022) · BMC Endocrine Disorders · doi:10.1186/s12902-022-00996-7 · Cited by 74
Kariyawasam, D., Morin, C., Casteels, K., Le Tallec, C., Sfez, A., Godot, C., Huneker, E., Garrec, N., Benhamou, P.-Y., Polak, M., et al. (2022) · The Lancet Digital Health · doi:10.1016/s2589-7500(21)00271-5 · Cited by 57
The role of automated insulin delivery technology in diabetes
Boughton, C. K., & Hovorka, R. (2024) · Diabetologia · doi:10.1007/s00125-024-06165-w · Cited by 57
“…CA, USA) has been super-seded by the second generation MiniMed 780G, which has improved usability (fewer auto mode exits and alarms) and improved glucose management with automated bolus insulin corrections and a meal detection module in addition to the existing features of the MiniMed 670G [27].”
BackgroundHood, K. K., Laffel, L. M., Danne, T., Nimri, R., Weinzimer, S. A., Sibayan, J., Bailey, R. J., Schatz, D., Bratina, N., Bello, R., et al. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0153 · Cited by 57
Self-citationGrassi, B., Gómez, A. M., Calliari, L. E., Franco, D., Raggio, M., Riera, F., Castro, M., McVean, J., van den Heuvel, T., Arrieta, A., et al. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.15023 · Cited by 55
Beato-Víbora, P. I., Gallego-Gamero, F., Ambrojo-López, A., Gil-Poch, E., Martín-Romo, I., & Arroyo-Díez, F. J. (2021) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2021.108986 · Cited by 56
Landscape of Continuous Glucose Monitoring (CGM) and Integrated CGM: Accuracy Considerations
Bailey, T. S., & Alva, S. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0236 · Cited by 56
Vetrani, C., Calabrese, I., Cavagnuolo, L., Pacella, D., Napolano, E., Rienzo, S. D., Riccardi, G., Rivellese, A. A., Annuzzi, G., & Bozzetto, L. (2021) · Diabetologia · doi:10.1007/s00125-021-05587-0 · Cited by 57
Fuchs, J., & Hovorka, R. (2021) · Frontiers in Pediatrics · doi:10.3389/fped.2021.679484 · Cited by 61
Kovatchev, B. P., Singh, H., Mueller, L., & Gonder-Frederick, L. A. (2022) · Diabetes Care · doi:10.2337/dc22-1217 · Cited by 49
Barriers to Uptake of Insulin Technologies and Novel Solutions
PAULEY, M. E., Berget, C., Messer, L. H., & Forlenza, G. P. (2021) · Medical Devices Evidence and Research · doi:10.2147/mder.s312858 · Cited by 73
Berget, C., Akturk, H. K., Messer, L. H., Vigers, T., Pyle, L., Snell‐Bergeon, J., Driscoll, K. A., & Forlenza, G. P. (2021) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14441 · Cited by 46
Time-in-range for monitoring glucose control: Is it time for a change?
Bellido, V., Pinés-Corrales, P. J., Villar‐Taibo, R., & Ampudia‐Blasco, F. J. (2021) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2021.108917 · Cited by 52
Developing the UVA/Padova Type 1 Diabetes Simulator: Modeling, Validation, Refinements, and Utility
Cobelli, C., & Kovatchev, B. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231195081 · Cited by 46
Pemberton, J. S., Wilmot, E. G., Barnard‐Kelly, K., Leelarathna, L., Oliver, N., Randell, T., Taplin, C. E., Choudhary, P., & Adolfsson, P. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.14962 · Cited by 44
Automated insulin delivery systems: from early research to routine care of type 1 diabetes
Renard, É. (2022) · Acta Diabetologica · doi:10.1007/s00592-022-01929-5 · Cited by 58
Aiello, E. M., Deshpande, S., Özaslan, B., Wolkowicz, K. L., Dassau, E., Pinsker, J. E., & Doyle, F. J. (2021) · Current Opinion in Biomedical Engineering · doi:10.1016/j.cobme.2021.100312 · Cited by 49
Tornese, G., Buzzurro, F., Carletti, C. V., Faleschini, E., & Barbi, E. (2021) · Frontiers in Endocrinology · doi:10.3389/fendo.2021.766314 · Cited by 48
Closed-loop insulin delivery: update on the state of the field and emerging technologies
Ware, J., & Hovorka, R. (2022) · Expert Review of Medical Devices · doi:10.1080/17434440.2022.2142556 · Cited by 47
Emerging Diabetes Technologies: Continuous Glucose Monitors/Artificial Pancreases
ALMURASHI, A. M., Rodriguez, E., & Garg, S. K. (2023) · Journal of the Indian Institute of Science · doi:10.1007/s41745-022-00348-3 · Cited by 67
Berge, T. V. D., Remus, K., Biester, S., Reschke, F., Klusmeier, B., Adolph, K., Holtdirk, A., Thomas, A., Kordonouri, O., Danne, T., et al. (2022) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14706 · Cited by 47
Self-citationHaidar, A., Lovblom, L. E., Cardinez, N., Gouchie‐Provencher, N., Orszag, A., Tsoukas, M. A., Falappa, C. M., Jafar, A., Ghanbari, M., Eldelekli, D., et al. (2022) · Nature Medicine · doi:10.1038/s41591-022-01805-3 · Cited by 48
Petruzelkova, L., Neuman, V., Plachy, L., Kozak, M., Obermannova, B., Kolouskova, S., Pruhova, S., & Sumnik, Z. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0562 · Cited by 40
Barriers and Facilitators to Diabetes Device Adoption for People with Type 1 Diabetes
Tanenbaum, M. L., & COMMISSARIAT, P. V. (2022) · Current Diabetes Reports · doi:10.1007/s11892-022-01469-w · Cited by 50
Boucsein, A., Watson, A. S., Frewen, C. M., Sanders, O. J., Haszard, J. J., Jones, S. D., Milford-Hughes, P. J., de Bock, M. I., & Wheeler, B. J. (2023) · Diabetes Care · doi:10.2337/dc22-1971 · Cited by 40
Passanisi, S., Salzano, G., Bombaci, B., Minuto, N., Bassi, M., Bonfanti, R., Scialabba, F., Mozzillo, E., Candia, F. D., Monti, S. M., et al. (2024) · Diabetes Care · doi:10.2337/dc23-2311 · Cited by 48
Jendle, J., Buompensiere, M. I., Holm, A. L., de Portu, S., Malkin, S. J. P., & Cohen, O. (2021) · Diabetes Therapy · doi:10.1007/s13300-021-01157-0 · Cited by 40
Dovč, K., Bergford, S., Fröhlich-Reiterer, E. E., Zaharieva, D. P., Potočnik, N., Müller, A., Lenarčič, Ž., Calhoun, P. M., Fritsch, M., Sourij, H., et al. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0178 · Cited by 44
Self-citationIslam, M. T., Hasan, M. W., Hassan, M. F., & Çeşmeci, Ş. (2024) · Pharmaceutics · doi:10.3390/pharmaceutics16070944 · Cited by 56
Kovatchev, B. P., Castillo, A., Pryor, E. C., Kollar, L. L., Barnett, C. L., DeBoer, M. D., Brown, S. A., & for the NAP Study Team (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0469 · Cited by 46
Kang, S., Kang, S. M., Choi, J. H., Ko, S., Koo, B. K., Kwon, H.-S., Kim, M. K., Kim, S. Y., Kim, S., Kim, Y., et al. (2025) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2025.0469 · Cited by 85
Lee, M. H., Paldus, B., Vogrin, S., Morrison, D., Zaharieva, D. P., Lu, J., Jones, H. M., Netzer, E., Robinson, L., Grosman, B., et al. (2021) · Diabetes Care · doi:10.2337/dc21-0814 · Cited by 37
Tornese, G., Carletti, C. V., Giangreco, M., Nisticò, D., Faleschini, E., & Barbi, E. (2022) · Diabetes Care · doi:10.2337/dc21-2643 · Cited by 46
Lepore, G., Rossini, A., Bellante, R., Corsi, A., Scaranna, C., Dodesini, A. R., & Trevisan, R. (2022) · Acta Diabetologica · doi:10.1007/s00592-022-01937-5 · Cited by 40
Automated Insulin Delivery with SGLT2i Combination Therapy in Type 1 Diabetes
Garcia‐Tirado, J., Farhy, L. S., Nass, R. M., Kollar, L. L., Oliveri, M. C., Basu, R., Kovatchev, B. P., & Basu, A. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0542 · Cited by 39
Benhamou, P., Adenis, A., Lebbad, H., Tourki, Y., Heredia, M., Gehr, B., Franc, S., & Charpentier, G. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.15008 · Cited by 35
Putman, M. S., Norris, A. W., Hull, R. L., Rickels, M. R., Sussel, L., Blackman, S. M., Chan, C. L., Ode, K. L., Daley, T., Stecenko, A. A., et al. (2023) · Diabetes Care · doi:10.2337/dc23-0380 · Cited by 35
The changing landscape of automated insulin delivery in the management of type 1 diabetes
Lakshman, R., Boughton, C. K., & Hovorka, R. (2023) · Endocrine Connections · doi:10.1530/ec-23-0132 · Cited by 49
Passanisi, S., Lombardo, F., Mameli, C., Bombaci, B., Macedoni, M., Zuccotti, G., Dovč, K., Battelino, T., Salzano, G., & Delvecchio, M. (2023) · Diabetes Therapy · doi:10.1007/s13300-023-01501-6 · Cited by 46
Self-citationMCVEAN, J. J. F., & Miller, J. D. (2021) · Expert Review of Medical Devices · doi:10.1080/17434440.2021.1926984 · Cited by 41
A Comparison of Two Hybrid Closed-Loop Systems in Italian Children and Adults With Type 1 Diabetes
Bassi, M., Teliti, M., Lezzi, M., Iosca, A., Strati, M. F., Carmisciano, L., d’Annunzio, G., Minuto, N., & Maggi, D. C. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2021.802419 · Cited by 41
Arunachalum, S., Velado, K., Vigersky, R. A., & Cordero, T. L. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221088608 · Cited by 34
Seget, S., Jarosz-Chobot, P. A., Ochab, A., Polańska, J., Rusak, E., Witoszek, P., & Chobot, A. P. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2022.1036808 · Cited by 39
Henry, Z., Fimbel, S., Bendelac, N., Perge, K., & Thivolet, C. H. (2023) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15344 · Cited by 36
Time in Tight Range for Patients With Type 1 Diabetes: The Time Is Now, or Is It Too Soon?
Hamidi, V., & Pettus, J. (2024) · Diabetes Care · doi:10.2337/dci23-0092 · Cited by 38
Automated Insulin Delivery for Young People with Type 1 Diabetes and Elevated A1c
Boucsein, A., Zhou, Y., Michaels, V. R., Haszard, J. J., Jefferies, C., Wiltshire, E. J., Paul, R. G., Parry‐Strong, A., Pasha, M., Petrovski, G., et al. (2024) · NEJM Evidence · doi:10.1056/evidoa2400185 · Cited by 39
Schiaffini, R., Deodati, A., Nicoletti, M. C., Carducci, C., Ciampalini, P., Lorubbio, A., Matteoli, M. C., Pampanini, V., Patera, I. P., Rapini, N., et al. (2022) · Acta Diabetologica · doi:10.1007/s00592-022-01886-z · Cited by 37
Benhamou, P., Adenis, A., Tourki, Y., Pou, S., Madrolle, S., Franc, S., Kariyawasam, D., Beltrand, J., Klonoff, D. C., & Charpentier, G. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221128565 · Cited by 33
Burnside, M. J., Lewis, D. M., Crocket, H. R., Meier, R. A., Williman, J. A., Sanders, O. J., Jefferies, C. A., Faherty, A. M., Paul, R. G., Lever, C. S., et al. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0484 · Cited by 32
Tanenbaum, M. L., & COMMISSARIAT, P. V. (2023) · Endocrine Connections · doi:10.1530/ec-23-0193 · Cited by 35
Stahl‐Pehe, A., Mashhadi, N. S., Wirth, M., Schlesinger, S., Kuß, O., Holl, R. W., Bächle, C., Warz, K.-D., Bürger-Büsing, J., Spörkel, O., et al. (2025) · EClinicalMedicine · doi:10.1016/j.eclinm.2025.103190 · Cited by 42
Ma, R., Shao, R., An, X., Zhang, Q., & Sun, S. (2022) · Journal of Materials Chemistry B · doi:10.1039/d2tb00749e · Cited by 37
Artificial Intelligence Algorithms for Treatment of Diabetes
Rashid, M. M., Askari, M. R., Chen, C., Liang, Y., Shu, K., & Çınar, A. (2022) · Algorithms · doi:10.3390/a15090299 · Cited by 42
Bassi, M., Patti, L., Silvestrini, I., Strati, M. F., Ponzano, M., Minuto, N., & Maggi, D. C. (2023) · Frontiers in Endocrinology · doi:10.3389/fendo.2023.1099024 · Cited by 37
Electro-responsive silk fibroin microneedles for controlled release of insulin
Qi, Z., Tao, X., Tan, G., Tian, B., Zhang, L., Kundu, S. C., & Lu, S. (2023) · International Journal of Biological Macromolecules · doi:10.1016/j.ijbiomac.2023.124684 · Cited by 38
Jacobsen, L. M., Sherr, J. L., Considine, E. G., Chen, A. X., Peeling, S., Hulsmans, M., Charleer, S., Urazbayeva, M., Tosur, M., ALAMARIE, S. A., et al. (2023) · Communications Medicine · doi:10.1038/s43856-023-00358-x · Cited by 42
Self-citationCherubini, V., Rabbone, I., Berioli, M. G., Giorda, S., Presti, D. L., Maltoni, G., Mameli, C., Marigliano, M., Marino, M., Minuto, N., et al. (2021) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14491 · Cited by 30
The automated pancreas: A review of technologies and clinical practice
Biester, T., Tauschmann, M., Chobot, A. P., Kordonouri, O., Danne, T., Kapellen, T. M., & Dovč, K. (2021) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14576 · Cited by 34
Self-citationAsgharzadeh, A., Patel, M., Connock, M. J., Damery, S. L., Ghosh, I., Jordan, M., Freeman, K., Brown, A., Court, R. A., Baldwin, S., et al. (2024) · Health Technology Assessment · doi:10.3310/jypl3536 · Cited by 38
Elbarbary, N. S., & Ismail, E. A. R. (2022) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2022.110045 · Cited by 31
Bassi, M., Franzone, D., Dufour, F., Strati, M. F., Scalas, M., Tantari, G., Aloi, C., Salina, A., d’Annunzio, G., Maghnie, M., et al. (2023) · Life · doi:10.3390/life13030783 · Cited by 47
Lambadiari, V., Saltik, A. Z. O., de Portu, S., Buompensiere, M. I., Kountouri, A., Korakas, E., Sharland, H., & Cohen, O. (2021) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0443 · Cited by 30
Lee, T., Collett, C., Man, M., Hammond, M., Shepstone, L., Hartnell, S., Gurnell, E. M., Byrne, C., SCOTT, E. M., Lindsay, R. S., et al. (2022) · BMC Pregnancy and Childbirth · doi:10.1186/s12884-022-04543-z · Cited by 29
Self-citation
Metadata
- Authors
- Bergenstal, R. M., Nimri, R., Beck, R. W., Criego, A., Laffel, L., Schatz, D., Battelino, T., Danne, T., Weinzimer, S. A., Sibayan, J., Johnson, M. L., Bailey, R. J., Calhoun, P., Carlson, A., Isganaitis, E., Bello, R., Albanese-O'Neill, A., Dovc, K., Biester, T., Weyman, K., Hood, K., & Phillip, M.
- Affiliations
- HealthPartners (US)
- Schneider Children's Medical Center (IL)
- Jaeb Center for Health Research (US)
- Joslin Diabetes Center (US)
- University of Florida Health (US)
- University of Ljubljana (SI)
- Kinderkrankenhaus auf der Bult (DE)
- Yale University (US)
- Stanford Medicine (US)
- Corresponding author
- Bergenstal, Richard M
- ORCID
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 2021-01
- Volume
- 397
- Issue
- 10270
- Pages
- 208-219
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- PMC
- PMC9194961
- OpenAlex
- W3118986275
- Semantic Scholar
- 3204173da18358b9918a9af55bfce7ef67927766
- Cited by
- 344
- References
- 24
- Citations by year
- 2026: 32 · 2025: 45 · 2024: 58 · 2023: 74 · 2022: 74 · 2021: 60
- FWCI
- 49.65
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- www.elsevier.com/legal/tdmrep-license (tdm)
- doi.org/10.15223/policy-017 (stm-asf)
- doi.org/10.15223/policy-037 (stm-asf)
- doi.org/10.15223/policy-012 (stm-asf)
- doi.org/10.15223/policy-029 (stm-asf)
- doi.org/10.15223/policy-004 (stm-asf)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- PubMed Central (acceptedVersion)
- Funders
- National Institute of Diabetes and Digestive and Kidney Diseases (UC4 DK108611)
- National Institutes of Health
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Genetics › Diabetes and associated disorders
- Keywords
- type 1 diabetes, hybrid closed-loop system, insulin pump therapy, adolescents and young adults, glycemic control, hypoglycemia, hyperglycemia, continuous glucose monitoring, MiniMed 670G, crossover trial, non-inferiority, safety and tolerability
- MeSH terms
- Adult, Diabetes Mellitus, Type 1, Female, Germany, Humans, Hyperglycemia, Insulin, Insulin Infusion Systems, Israel, Male, United States, Blood Glucose Self-Monitoring, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.