STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 44,501–44,600 of 47,868
Arnold, T. D. W., & Balogh, Z. J. (2012) · The Journal of Trauma: Injury, Infection, and Critical Care · doi:10.1097/ta.0b013e318250914a
Agreement among Orbscan II, Pentacam HR and IOLMaster
Domínguez‐Vicent, A., Águila‐Carrasco, A. J. D., Monsálvez-Romín, D., Sanchis-Jurado, V., & Pérez‐Vives, C. (2013) · Journal of Emmetropia: Journal of Cataract, Refractive and Corneal Surgery
M-line processing for real-time investigation of large artery mechanics
Morganti, T., Ricci, S., Guidi, F., Valle, A. D., & Tortoli, P. (2004) · IEEE International Ultrasonics Symposium · doi:10.1109/ultsym.2003.1293212
Manuscripts in the Sport and Exercise Sciences
Ramsbottom, R. (2015) · International Journal of Sports and Exercise Medicine · doi:10.23937/2469-5718/1510021
Burns, R. D., Hannon, J. C., Allen, B. M., & Brusseau, T. A. (2013) · International Journal of Sports Science
Melchart, D., Doerfler, W., Weidenhammer, W., & Eustachi, A. (2016) · European Journal of Integrative Medicine · doi:10.1016/j.eujim.2016.11.002
McDonough, P. G., Williams, M., Thompson, L., Thomlinson, M., Kessopoulou, E., & Barratt, C. L. R. (1990) · Fertility and Sterility · doi:10.1016/s0015-0282(16)54032-1
Hanson, L. L. M. (2016) · University of Minnesota Digital Conservancy (University of Minnesota)
Metrologia applicata alla fisioterapia: misurazioni, test e bilanci, concetti fondamentali
Piette, P. (2016) · EMC - Medicina Riabilitativa · doi:10.1016/s1283-078x(16)80472-1 · Cited by 1
Prediction of photoplethysmogram signals using sequential ELM
Mahuli, A. P., Krupa, B., & George, K. (2016) · International Joint Conference on Neural Networks (IJCNN) · doi:10.1109/ijcnn.2016.7727442
variation coefficients for log transformed variables
le Cessie, S. (2008)
Novel methods to forecast emergency department crowding.
Hoot, N. R. (2007) · Europe PMC (PubMed Central) · Cited by 1
Measuring agreement in comparison of CA125 determined by three different immunoassay systems
Qiaodi, Z., Xie, E., Yan, Z., Ling, Y., Li, J., & Xu, H. (2015)
Malik, N., Minniti, C. P., Maivelett, J., Koroulakis, A. I., Delaney, K., Buscetta, A. J., Axelrod, K. C., & Gorbach, A. M. (2012)
Prediction of changes bone mineral in early postmenopausal women
Vestergaard, P. (2000) · Cited by 1
MEMOIRE pour l'obtention du DES d'Anesthésie-Réanimation
Dibiane, P. C. (2011)
Barbé, C., Harran, N. X., & Goulle, F. (2016) · Revue Vétérinaire Clinique · doi:10.1016/j.anicom.2016.10.001
Warner, D. S., & Woolf, C. J. (2007)
Testing a Model to Monitor Training Effect in Distance Running
Nummela, A. T., & Vesterinen, V. (2013) · Cited by 1
Preventable Deaths at Acute Care Hospitals
Kobewka, D. M. (2016) · uO Research (University of Ottawa) · doi:10.20381/ruor-5300
Using FATIMA — A robot mannequin head — For validation of head tracking software
Basteris, A., Pedler, A. R., & Sterling, M. M. (2016) · doi:10.1109/bme-hust.2016.7782104
Baker, K. M. (2016) · STARS (University of Central Florida)
Halligan, P., McGee, S., & Bartram, C. I. (1994)
Jamaan, A., & Saleh, T. Y. (2016) · Discovery Research Portal (University of Dundee)
Montenij, L. J. (2016) · Utrecht University Repository (Utrecht University)
Mackie, P. (2016) · University of Bedfordshire Repository (University of Bedfordshire)
Kluge, F., & Eskofier, B. M. (2016) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2016.07.043
Vinall, J. F. (2015) · cIRcle (University of British Columbia) · doi:10.14288/1.0166783
A Randomized Ttrial of Intensive Lipid-lowering Therapy in Calcific Aortic Stenosis
Cowell, S. J., Newby, D. E., & Prescott, R. J. (2005) · ACC Current Journal Review · doi:10.1016/j.accreview.2005.08.217
Schafer, M. A., Goss, F. L., Robertson, R. J., Nagle, E. F., & Kim, K. (2008) · Medicine & Science in Sports & Exercise · doi:10.1249/01.mss.0000322435.54108.57
Henriksson, P. (2015) · Linköping University medical dissertations · doi:10.3384/diss.diva-117435
Fetal volume measurements in the first trimester of pregnancy with three-dimensional ultrasound
Smeets, N. (. (2012) · TU/e Research Portal · doi:10.6100/ir735605
New Applications for Real-Time Three-Dimensional Echocardiography
Anwar, A. M. (2008) · EUR Research Repository (Erasmus University Rotterdam)
Femoral nailing in adults: doctor and patient reported outcomes
Moumni, M. E. (2016)
White, C. A. (2004) · Enlighten: Theses (The University of Glasgow)
Countering default mode in Teaching Statistics
MacGillivray, H., & Hewson, P. (2017) · Teaching Statistics · doi:10.1111/test.12122
Validation of an acoustic rumination sensor for dairy cows
Hendriksen, K., Büscher, W., Hoppe, S., Hoffmanns, C., & Riswick, H. (2014)
Bora, K., & Borah, P. (2017) · Korean Journal of Family Medicine · doi:10.4082/kjfm.2017.38.1.47
[Rotavirus: a frequent cause of acute gastroenteritis in children].
Ørstavik, I., Haug, K. W., Halvorsen, S., & Ulstrup, J. C. (1977) · PubMed
Reliability and Validity of Patient Oriented Eczema Measure (POEM) Korean Version
Kim, Y.-H., Han, J., Jung, I. C., Kang, W., & Park, Y. (2016) · 대한한의학회지 · doi:10.13048/jkm.16044
Explorations of pelvic floor ultrasound
Weemhoff, M. (2013) · Research Publications (Maastricht University) · doi:10.26481/dis.20130705mw
A composite photographic image – could it replace a lateral cephalogram?
Cala, A., Noar, J. H., Petrie, A., & O’Neill, J. (2017) · Journal of Orthodontics · doi:10.1080/14653125.2016.1277316
Buskermolen, A. G. (2014) · TU/e Research Portal
Tauseef, H., Fahiem, M. A., & Farhan, S. (2017) · Journal of Testing and Evaluation · doi:10.1520/jte20160214
Non-invasive Cardiac Output of Children in Health and Disease: Respiratory Gas Techniques
Schneiderman, J. E. (2012) · TSpace (University of Toronto)
Stewart, J. V., & Webster, D. (1992) · Annals of Emergency Medicine · doi:10.1016/s0196-0644(05)81786-x
Alwashmi, M. F. (2015) · Memorial University Research Repository (Memorial University) · doi:10.13140/rg.2.2.16303.79520
Jung, J. H., Jung, J., Cho, K. H., Ryu, M. R., Bae, S. H., Moon, S. K., Kim, Y. H., Choe, B., & Suh, T. S. (2017) · Journal of the Korean Physical Society · doi:10.3938/jkps.70.308
Moncho, D., & María, D. (2016) · TDX (Tesis Doctorals en Xarxa)
Investigations Related to Dietary Sodium in Chronic Heart Failure
Arcand, J. (2012) · TSpace (University of Toronto) · Cited by 1
Accurate tumor size determination in breast cancer: The debate continues
Noushad, S. N. (2016) · Clinical Cancer Investigation Journal · doi:10.4103/2278-0513.200107
Niang, M. M., Kane, C. T., Diop‐Ndiaye, H., Dièye, A., & Mboup, S. (2017)
Yu, J.-S., & Oh, J. (2017) · Journal of Physical Therapy Science · doi:10.1589/jpts.29.250
Monitoring of vascular health in children at risk for atherosclerosis.
Harrington, J. (2014) · Adelaide Research & Scholarship (AR&S) (University of Adelaide)
EXAMINATION OF THE VISUAL ACUITY ON THE LCD OPTOTYPE WITHWHOLE-LINE AND INTERPOLATION METHOD
Zajíček, T., Veselý, P., & Synek, S. (2012)
Comparison of haemoglobin values measured at point of care with the standard laboratory value
Johnson, M. (2016) · SUNScholar (Stellenbosch University)
AN INVESTIGATION INTO THE SLEEPING PATTERNS OF YOUTH SOCCER PLAYERS DURING THE COMPETITIVE SEASON
Turner, C. (2016) · Liverpool John Moores University · doi:10.24377/ljmu.t.00004888
Toong, C. M.-L., Lee, M. W. M., Connor, S. J., & Ng, W. (2017) · Pathology · doi:10.1016/j.pathol.2017.01.005
Validation of the Polar S410 Heart Rate Monitor for Estimating Energy Expenditure in Women
Newlin, M. K., Wingo, J. E., Ganio, M. S., Trilk, J. L., & Cureton, K. J. (2006) · Medicine & Science in Sports & Exercise · doi:10.1249/00005768-200605001-02816
Inwoley, A., Adiko, A. C., Kabran, M., Affi, R., Diomande, A., Wozan, F., & Dembelé, B. (2017) · Journal of Immunobiology · doi:10.4172/2476-1966.1000118
Test person operated 2-Alternative Forced Choice Audiometry compared to traditional audiometry
Schmidt, J. H., Brandt, C., Christensen‐Dalsgaard, J., Andersen, T. D., Bælum, J., & Poulsen, T. (2009) · University of Southern Denmark Research Portal (University of Southern Denmark)
Ultrasound Monitoring in ART Have We Made Any Progress?
Vandekerckhove, F., & Gerris, J. (2017) · Cambridge University Press eBooks · doi:10.1017/9781107295391.009
Aldughmi, M. (2016) · KU ScholarWorks (The University of Kansas) · Cited by 1
Inertial Sensors for Kinematic Measurement and Activity Classification of Gait Post-Stroke
Laudanski, A. F. (2013) · QSpace (Queen's University Library)
Crizzle, A. M. (2011) · UWSpace (University of Waterloo) · Cited by 4
Biomarkers of Renal Function in Acute Coronary Syndromes
Åkerblom, A. (2013) · Uppsala University Publications (Uppsala University)
Test–retest reliability of contrast visual acuities in a clinical population
Sukha, A. Y., & RUBIN, A. M. (2017) · African vision and eye health · doi:10.4102/aveh.v76i1.376
Anterior Superior Iliac Spine to the Tibial Tuberosity Length
Mahapatro, S., Mohanty, S., Panigrahi, S. K., Ray, R. K., & Saraswat, S. (2017) · Global Pediatric Health · doi:10.1177/2333794x16687190
Eskola, M. J. (2008) · Tampere University Institutional Repository (Tampere University) · Cited by 2
Validity of the International Physical Activity Questionnaire in a rural Hispanic Population
Espinoza, N. (2012) · ResearchWorks at the University of Washington (University of Washington) · Cited by 1
Erythrocyte sedimentation rate (ESR): Comparison between VES-Matic cube 80 and the manual method
Sofia, M. D. (2016) · Biochemistry & Analytical Biochemistry
Ferdousi, F. (2015) · Journal of Diabetes & Metabolism · doi:10.4172/2155-6156.1000564
Variations in the sonographic measurement techniques of BI-RADS 3 breast masses
Francisco, J., Jales, R. M., Oliveira, A., Arguello, C. H. F., & Derchain, S. (2017) · Journal of Clinical Ultrasound · doi:10.1002/jcu.22475
Burn scar assessment: identification and testing of an objective multi-component device
Gankande, U. (2015) · UWA Profiles and Research Repository (UWA)
A Modern Statistical Approach to Quality Improvement in Health Care using Quantile Regression
Dalton, J. E. (2013) · OhioLink ETD Center (Ohio Library and Information Network)
Towards reliable diffusion MRI of the in vivo human heart
Deuster, V., & von Deuster, C. (2016) · Research Portal (King's College London)
Pharmacological and physiological manipulation of ocular accommodation
Randhawa, P. K. (2017) · Aston Publications Explorer (Aston University) · doi:10.48780/publications.aston.ac.uk.00030401
Neal, M. L., Trister, A. D., & Ahn, S. (2013)
Estilo de vida y sus relaciones con la osteoporosis y pérdida del equilibrio corporal
Pôrto, E. F., de Oliveira, F. C., Rocha, G., Bernardo, K. M. D. A., Kümpel, C., & de Castro, A. A. M. (2017) · Revista Família Ciclos de Vida e Saúde no Contexto Social · doi:10.18554/refacs.v5i2.2068
Chew, J. (2016) · The Sydney eScholarship Repository (The University of Sydney)
Pocket-size ultrasound, a new diagnostic tool in clinical practice
Mjølstad, O. C. (2013) · NORA - Norwegian Open Research Archives
Adaptation culturelle du Female Pelvic Floor Questionnaire (FPFQ) en langue française
Deparis, J., Bonniaud, V., Desseauve, D., Guilhot, J., Masanovic, M., de Tayrac, R., Fauconnier, A., & Fritel, X. (2017) · The French Journal of Urology · doi:10.1016/j.purol.2017.03.008
Scarred Uterus: Subsequent Pregnancy and Delivery
Baranov, A. (2017) · Lund University Publications (Lund University)
Health risk assessment of lead exposure to children in Blantyre, Malawi
Utembe, W. (2016) · Cited by 1
Soares, D. P. (2012) · Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT) · doi:10.34626/6am5-y448
Katsumata, Y., TAKEI, H., Hayashi, H., & Ichikawa, K. (2017) · Rigakuryoho Kagaku · doi:10.1589/rika.32.215
McLean, S. K. (2008) · UWA Profiles and Research Repository (UWA)
In vivo validation of the time domain velocity measurement technique if blood flow in human fetuses
Alcázar, J. L., & Laparte, C. (2017) · Revista de medicina de la Universidad de Navarra · doi:10.15581/021.6704
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.