STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
The DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
In clinical measurement comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Such investigations are often analysed inappropriately, notably by using correlation coefficients. The use of correlation is misleading. An alternative approach, based on graphical techniques and simple calculations, is described, together with the relation between this analysis and the assessment of repeatability.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and all parameters used for the proposition and cross-validation of the model confirmed their validity for the studied population (15, 37–39, 57)”
From Development and Cross-Validation of a Predictive Equation for Fat-Free Mass in Brazilian Adolescents by Bioelectrical Impedance (2022) · cites this paper as [39]
In the citing paper
“Each volume measurement was performed twice to determine reproducibility by the method of Bland and Altman (1986).”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as Bland and Altman (1986)
In the citing paper
“After a fairly extensive search of physics textbooks and the scientific literature over a number of years, no direct references to the technique have been discovered 1 .”
From Archimedes revisited: a faster, better, cheaper method of accurately measuring the volume of small objects (2005) · cites this paper as [1]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
In the citing paper
“However, paired t test and Bland-Altman plot are methods for analyzing agreement, and Pearson correlation coefficient is only a measure of correlation, and hence, they are nonideal measures of reliability.”
From A guideline of selecting and reporting intraclass correlation coefficients for reliability research (2016) · cites this paper as [6]
2 more claims about this paper
Create a free account to see every claim other papers make about this one.
Citations
Cites
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Self-citationStatistical Methods in Medical Research
(1972) · Postgraduate Medical Journal · doi:10.1136/pgmj.48.555.67-a · Cited by 526
RELATION BETWEEN INITIAL BLOOD PRESSURE AND ITS FALL WITH TREATMENT
Gill, J. S., Beevers, D. G., Zezulka, A. V., & Davies, P. (1985) · The Lancet · doi:10.1016/s0140-6736(85)91219-x · Cited by 156
The Nellcor N‐101 pulse oximeter: A clinical evaluation in anaesthesia and intensive care
TYTLER, J. A., & SEELEY, H. F. (1986) · Anaesthesia · doi:10.1111/j.1365-2044.1986.tb12793.x · Cited by 32
Comparison of the new miniature Wright peak flow meter with the standard Wright peak flow meter.
Oldham, H. G., Bevan, M. M., & McDermott, M. (1979) · Thorax · doi:10.1136/thx.34.6.807 · Cited by 28
Estimation of gestational age at birth. Comparison of two methods.
Serfontein, G. L., & Jaroszewicz, A. M. (1978) · Archives of Disease in Childhood · doi:10.1136/adc.53.6.509 · Cited by 23
D'Arbela, P. G., Silayan, Z. M., & Bland, M. (1986) · Heart · doi:10.1136/hrt.56.5.445 · Cited by 4
Cited by
Items 45,101–45,200 of 47,868
Incidence of Acute Appendicitis during Pregnancy and Outcome of Appendectomy in Ramathibodi Hospital
Arpornsujaritkun, N., & Suwanthanma, W. (2017) · Journal of the Medical Association of Thailand
Yang, C.-L., Simposon, L. A., & Eng, J. J. (2020) · JMIR Rehabilitation and Assistive Technologies · doi:10.2196/26044
Crabtree, D. R., Buosi, W., Fyfe, C., Horgan, G. W., Holst, J. J., & Johnstone, A. M. (2021) · Physiology & Behavior · doi:10.1016/j.physbeh.2021.113622
Comparison of Pilot Fatalities and Number of Pilot Medical Examinations
Canfield, D. V., Dubowski, K. M., Salazar, G. J., & Forster, E. M. (2016) · ROSA P
Silva-Aragón, A., & Quezada, M. E. (2019) · Revista Digital Internacional de Psicología y Ciencia Social · doi:10.22402/j.rdipycs.unam.5.2.2019.239.150-157
UNAM, F. I., Arias-García, B. D., & Quezada, M. E. (2018) · doi:10.22402/j.rdipycs.unam.4.2.2018.203.120-126
Dion, M.-O., Simonyan, D., Faure, P.-A., Pelet, S., May, O., Bonin, N., Carsen, S., Beaulé, P. E., Nault, M., & Belzile, É. L. (2021) · Revue de Chirurgie Orthopédique et Traumatologique · doi:10.1016/j.rcot.2021.09.034
Pugh, A. L., & Gill, A. C. (2021) · Scientific investigations report · doi:10.3133/sir20215096 · Cited by 1
Barbotte, F., Delord, M., & Pujol, N. (2021) · Revue de Chirurgie Orthopédique et Traumatologique · doi:10.1016/j.rcot.2021.10.010
Ryckman, K. K., Ramesh, A., Cho, H., Oltman, S. P., Rogers, E. E., Dagle, J. M., & Jelliffe‐Pawlowski, L. L. (2021) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2021.10.007
Evaluation of Reading Performance and Visual Acuity Tests
Baashen, M. (2020) · Aston Publications Explorer (Aston University) · doi:10.48780/publications.aston.ac.uk.00042451
Concordância interobservadores em um checklist de cuidados em terapia nutricional enteral
Corrêa, A. P. A., Silva, S. M. R., Oleques, C. C., de Sousa, G. P., Viegas, G. L., Anziliero, F., Oliveira, A. C. D. S., & Beghetto, M. G. (2021) · Acta Paulista de Enfermagem · doi:10.37689/acta-ape/2021ao001525
Exploring functional limitations in people with traumatic and nontraumatic neck pain
Stenneberg, M. S. (2021) · Research Publications (Maastricht University) · doi:10.26481/dis.20211123ms
Reliability of the Norwegian version of PainDETECT: a test-retest study
Myhre, A. M., & Strand, L. I. (2021) · European Journal of Physiotherapy · doi:10.1080/21679169.2021.2001567
Idrissi, M., Saeid, N., Rami, A., Mzibri, M. E., Akom, A. G. A., Kossou, J., Hamdouchi, A. E., Safsaf, N., Belakhel, L., Benkirane, H., et al. (2021) · E3S Web of Conferences · doi:10.1051/e3sconf/202131901066
PandAcuity in paediatrics: a novel clinical measure of visual function based on the panda illusion
Kelbsch, C., Spieth, B., Zrenner, E., Besch, D., & Straßer, T. (2021) · British Journal of Ophthalmology · doi:10.1136/bjophthalmol-2021-319935
Tumour size, new lesions and other markers as longitudinal endpoints in oncology studies
Lombard, A. (2021) · Research Explorer (The University of Manchester)
Bernhart, J. A., Turner‐McGrievy, G. M., Eustis, S. J., Wilson, M. J., Hutto, B., & Wilcox, S. (2021) · Journal of Public Health · doi:10.1007/s10389-021-01666-z
Minoda, Y., Sugama, R., Ohta, Y., Takemura, S., Yamamoto, N., Nakatsuchi, T., & Nakamura, H. (2021) · Scientific Reports · doi:10.1038/s41598-021-01420-1
Validity of Drug and Diagnosis Data in Pharmacoepidemiology
Ritchey, M. E., West, S. L., & Maldonado, G. (2021) · Wiley eBooks · doi:10.1002/9781119701101.ch13
Bril, S. I. (2021) · Utrecht University Repository (Utrecht University) · doi:10.33540/1001
Masmas, T. N., Madsen, H., Petersen, S. L., Ryder, L. P., Svejgaard, A., Alizadeh, M., & Vindeløv, L. L. (2005)
Jacbs, R. S. W., & BILLING., G. C. (2018)
A probabilistic segmentation method for IVUS images
Rivera, M., & Kakadiaris, I. A. (2010)
Clinical Tests and Measurements
Lai, K. W., & Supriyanto, E. (2012) · SpringerBriefs in applied sciences and technology · doi:10.1007/978-981-4021-96-8_4
Accuracy of single intravenous access iohexol GFR in children is hampered by marker contamination
Eide, T. T., Hufthammer, K. O., Brun, A., Brackman, D., Svarstad, E., & Tøndel, C. (2021) · Scientific Reports · doi:10.1038/s41598-021-02759-1
Araújo, N., Lopes-Conceição, L., Morais, S., Fontes, F., Dias, T., Cruz, V. T., Ruano, L., Pereira, S., & Lunet, N. (2021) · Supportive Care in Cancer · doi:10.1007/s00520-021-06702-y
Otero, C. (2014) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas)
Ho, J. T. (2007) · Adelaide Research & Scholarship (AR&S) (University of Adelaide)
TIME SERIES ANALYSIS OF CARDIOMETABOLIC PARAMETERS: RELIABILITY AND ENERGY DRINK RESPONSE
Nelson, M. T. (2013) · University of Minnesota Digital Conservancy (University of Minnesota)
Validation of an Inertial Sensor System for Quantifying Knee Function
Koslin, A. J. (2006) · TigerPrints (Clemson University) · Cited by 1
Basic studies on estimation of glomerular filtration rate in cattle
Murayama, I. (2014) · Japanese Journal of Large Animal Clinics · doi:10.4190/jjlac.5.47
Effect of processing on the distribution of cadmium in wheaten products
Swinden, L. B. (2006) · Victoria University Research Repository (Victoria University)
Henriksson, H. (2015) · Linköping University medical dissertations · doi:10.3384/diss.diva-117420
Bewertung und Beurteilung humanpharmakologischer Studien
de Mey, C. (1992) · Konzepte in der Humanpharmakologi · doi:10.1007/978-3-642-84729-5_32
Activity-related parenting practices and young people's physical activity
Edwardson, C. L. (2010) · OpenGrey (Institut de l'Information Scientifique et Technique) · Cited by 1
Exercise measurements and interventions for children with childhood cancer
van der Torre, P. (2010) · Utrecht University Repository (Utrecht University)
Efeito agudo do alongamento dinâmico no salto vertical: uma revisão sistemática e meta-análise
Marques, N. K. (2014) · Dialnet (Universidad de la Rioja)
Determinação da tensão interfacial do sistema Água - Tolueno - Acetona
Silva, T. F. C. V. (2009) · Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT)
Cardiovascular magnetic resonance of the right ventricle
Alpendurada, F. (2013) · Middlesex University Research Repository (Middlesex University Of London)
Baxter, M. L. (2014) · Otago University Research Archive (University of Otago)
HRV estimation from BP and coding strategy analysis of the cardiovascular regulation
Lucena, F. (2012) · Institutional Repositories DataBase (IRDB)
Nieuwerburgh, F. V. (2007) · Ghent University Academic Bibliography (Ghent University)
Gas exchange kinetics in elite runners.
Edwards, A. M. (1999) · ResearchOnline at James Cook University (James Cook University)
The Physiology and Bioenergetics of Ultraendurance Mountain Bike Racing
Metcalfe, J. (2011) · CLOK (University of Central Lancashire)
Vasconcellos, F. V. D. A. (2014) · Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT) · doi:10.34626/g502-5p10
Geneen, L. J. (2014) · Queen Margaret University Publications Repository (Queen Margaret University)
Non-invasive rodent cardiac output: comparison of a compact clinical monitor with echocardiography
Sturgess, D. J., Haluska, B. A., Jones, M., & Venkatesh, B. (2010)
Caciagli, N. (2021) · Encyclopedia of earth sciences series/Encyclopedia of earth sciences · doi:10.1007/978-3-030-26050-7_261-1
Maghraoui, A. E. (2022) · Springer eBooks · doi:10.1007/978-3-030-87950-1_11
Royston, P. J., & Sauerbrei, W. (2008) · Wiley series in probability and statistics · doi:10.1002/9780470770771.refs
Validating Quantitative Diagnostic Tests
(2008) · Springer eBooks · doi:10.1007/978-1-4020-9523-8_36
Zhang, H., Wang, C., Tu, X., Liu, Z., Xi, Y., Hu, T., Feng, J., & Zhang, Y. (2021) · doi:10.2196/preprints.34375
Annes, M. L. F., Tabajara, F. B., da Rosa, R. D., Mattiello, R., Alves, A. L. S., & Schwanke, C. H. A. (2021) · Revista de Nutrição · doi:10.1590/1678-9865202134e200203
Montgomery‐Downs, H. E., & Meltzer, L. J. (2012) · Humana Press eBooks · doi:10.1007/978-1-60761-725-9_13
Osteoarthritis: Getting the picture
Gielis, W. P. (2022) · Utrecht University Repository (Utrecht University) · doi:10.33540/1153
(1997) · Proceedings of The Nutrition Society · doi:10.1079/pns19970084
Delahunty, F., Johansson, R., & Arčan, M. (2019) · doi:10.2196/preprints.16918
Sanitsakul, H., Woratanarat, P., Jaovisidha, S., & Woratanarat, T. (2022) · Journal of Pediatric Orthopaedics B · doi:10.1097/bpb.0000000000000947
(1990) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.1990.28.3.129
A Computer-Aided Recognition Method of Heart Rate Deflection Point
Wang, S., Li, J., Lan, M., Gao, L. H., & Gao, X. (2021) · IEEE International Conference on Progress in Informatics and Computing · doi:10.1109/pic53636.2021.9687083
Diawara, F., Fumeron, C., Duflot, L., Tia, M. W., & Bobrie, G. (2022) · Néphrologie & Thérapeutique · doi:10.1016/j.nephro.2021.10.004
Ŝifrer, F., Drinovec, I., Thomson, G., Baker, L., Mawby, E., Leach, R. M., Jensen, P. J., Michagin, G., Andersen, P. K., Peláez, J., et al. (1996) · Intensive Care Medicine · doi:10.1007/bf03216432
Jones, A. W. (2006) · CRC Press eBooks · doi:10.1201/9781420003468.ch5
Oliveira, A. G. (2020) · Wiley eBooks · doi:10.1002/9781119584254.refs
Fayers, P. M., & Machin, D. (2015) · Wiley eBooks · doi:10.1002/9781118758991.refs
Riley, R. D., Tierney, J. F., & Stewart, L. A. (2021) · Wiley eBooks · doi:10.1002/9781119333784.refs2
Choudhary, P. K., & Nagaraja, H. N. (2017) · Wiley series in probability and statistics · doi:10.1002/9781118553282.refs
Islam, M. R., Islam, J. S., Zatzman, G. M., Mughal, M. A. H., & Rahman, M. S. (2015) · Wiley eBooks · doi:10.1002/9781119184201.refs
Abstracts of Original Communications (University College Cork, Republic of Ireland. 18–20 June 2003)
(2003) · Proceedings of The Nutrition Society · doi:10.1017/s0029665103000764
(2016) · Wiley eBooks · doi:10.1002/9781119159704.refs
- Segmenting with Shape and Appearance Priors
El-Baz, A. S., Gimel’farb, G., & Suri, J. S. (2015) · CRC Press eBooks · doi:10.1201/b19253-12
Campbell, M. J. (2005) · Wiley eBooks · doi:10.1002/0470011815.b2a08042
(1991) · Proceedings of The Nutrition Society · doi:10.1079/pns19910017
Larner, A. J. (2018) · Springer eBooks · doi:10.1007/978-3-319-75259-4_2
Methods (2): Statistical Methods
Larner, A. J. (2019) · Springer eBooks · doi:10.1007/978-3-030-17562-7_3
López-Navarro, C., Serrano-Valero, M., Fages-Caravaca, E. M., Martínez-Payá, J. J., del Baño-Aledo, M. E., & Ríos-Díaz, J. (2022) · Neurología · doi:10.1016/j.nrl.2021.10.008
(1996) · Allergy · doi:10.1111/j.1398-9995.1996.tb04788.x
McLaren, S. J., Evans, W., Galna, B., Portas, M., Weston, M., & Spears, I. R. (2022) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2022.111015
Quality Management and Accreditation
Mortimer, D., Björndahl, L., Barratt, C. L. R., Castilla, J. A., Menkveld, R., Kvist, U., Álvarez, J. G., & Haugen, T. B. (2022) · Cambridge University Press eBooks · doi:10.1017/9781009181648.012 · Cited by 2
Henderson, G. R. (2011) · Wiley eBooks · doi:10.1002/9781119975328.refs
Cunningham, G. J., Cocor, C., Smith, M. M., Young, A. A., Cass, B., & Moor, B. K. (2022) · Revue de Chirurgie Orthopédique et Traumatologique · doi:10.1016/j.rcot.2021.08.015
Mizumoto, K., Terao, H., & Kamei, M. (2021) · JAPANESE ORTHOPTIC JOURNAL · doi:10.4263/jorthoptic.50f109
Re, F. L., Sponga, S., Angelini, J., Nalli, C., Zucchetto, A., Livi, U., & Baraldo, M. (2022) · IntechOpen eBooks · doi:10.5772/intechopen.102412
(2004) · Anaesthesia · doi:10.1046/j.0003-2409.2002.02745.x
Passerini, K., Kurmanavicius, J., Burkhardt, T., Ederhof, L., & Balsyte, D. (2022) · Archives of Gynecology and Obstetrics · doi:10.1007/s00404-022-06489-1
Babin, E., Cahill, J., Hicks, T., & Allen, D. H. (2022) · Journal of Radiology Nursing · doi:10.1016/j.jradnu.2022.02.001
Can listening provide a useful clinical estimate of low vision reading parameters?
Latham, K., & Subhi, H. (2022) · Ophthalmic and Physiological Optics · doi:10.1111/opo.12966
Radeisen, M. F., & Lahrmann, K. (2022) · Tierärztliche Praxis Ausgabe G Großtiere / Nutztiere · doi:10.1055/a-1729-3882
Tanaka, K., Yamasaki, H., & Takashi, M. O. R. I. (2022) · THE JOURNAL OF JAPAN SOCIETY FOR CLINICAL ANESTHESIA · doi:10.2199/jjsca.42.36
Uslu, H., Altınel, M. G., & Kanra, A. Y. (2022) · Akdeniz Medical Journal · doi:10.53394/akd.1057375
Validation of the inverse method for the determination of the access flow with thermodilution
Tocino, M. L. S., González, A. L., González, S. V., Sánchez, M. I. G., Macías, M. G., Horna, A. G., Puente-González, A. S., Blanc, M. H., Furaz, K., & Sánchez-Tocino, H. (2022) · Nefrología (English Edition) · doi:10.1016/j.nefroe.2022.03.004
Assessing the Validity of a Kinematic Knee Sleeve in a Resistance-Trained Population
Toon, N., McMaster, S. A., Outram, T., & Faghy, M. A. (2022) · International Conference on the Challenges, Opportunities, Innovations and Applications in Electronic Textiles (E-Textiles) · doi:10.3390/engproc2022015010
Peña, B., Ochoa, M., Flores, O., Gil, A. I., Ecker, L., Cornejo, R., Lanata, C. F., Howard, L. M., & Grijalva, C. G. (2022) · Journal of Virological Methods · doi:10.1016/j.jviromet.2022.114522
Echocardiographic measurements of left atrial volume in patients with ischaemic heart disease
Backman, M., & Rosfors, S. (2022) · Clinical Physiology and Functional Imaging · doi:10.1111/cpf.12748
Metadata
- Authors
- Martin Bland, J., & Altman, D.
- Venue
- The Lancet
- Publisher
- Elsevier BV
- Published
- 1986-02
- Volume
- 327
- Issue
- 8476
- Pages
- 307-310
- ISSN
- 0099-5355 (ISSN-L)
- 0140-6736
- OpenAlex
- W2015795623
- Cited by
- 47,965
- References
- 7
- Citations by year
- 2026: 1541 · 2025: 1310 · 2024: 1377 · 2023: 1382 · 2022: 1513 · 2021: 1827 · 2020: 1878 · 2019: 1797 · 2018: 1725 · 2017: 1906 · 2016: 1876 · 2015: 1983 · 2014: 2081 · 2013: 2031 · 2012: 2082
- FWCI
- 149.89
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.elsevier.com/tdm/userlicense/1.0/ (tdm)
- Locations
- The Lancet (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Social Sciences › Decision Sciences › Statistics, Probability and Uncertainty › Reliability and Agreement in Measurement
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Mathematics › Statistics and Probability › Statistical Methods in Epidemiology
- Keywords
- agreement between measurement methods, method comparison, correlation coefficient limitations, repeatability, clinical measurement
- MeSH terms
- Diagnosis, Humans, Peak Expiratory Flow Rate, Statistics as Topic
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.