Three dimensional inertial sensing of foot movements for automatic tuning of a two-channel implantable drop-foot stimulator
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
A three dimensional inertial sensing system for measuring foot movements during gait is proposed and tested. It can form the basis for an automated tuning system for a two-channel implantable drop-foot stimulator. The foot orientation and position during the swing phase of gait can be reconstructed on the basis of three-dimensional measurement of acceleration and angular velocity, using initial and final conditions during mid-stance. The foot movements during gait of one stroke person using the implanted two-channel stimulator were evaluated for several combinations of stimulation parameters for both channels. The reconstructed foot movements during gait in this person indicated that the channel stimulating the deep peroneal nerve contributes mainly to dorsiflexion and provides some reduction of inversion seen without stimulation, while the channel activating the superficial peroneal nerve mainly provides additional reduction of inversion. This agrees with anatomical knowledge about the function of the muscles activated by both branches of the peroneal nerve. The inertial sensor method is expected to be useful for the clinical evaluation of foot movements during gait supported by the two-channel drop-foot stimulator. Furthermore, it is expected to be applicable for the automated balancing of the two stimulation channels to ensure optimal support of gait.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Clinical use of the odstock dropped foot stimulator: Its effect on the speed and effort of walking
Taylor, P. N., Burridge, J. H., Dunkerley, A. L., Wood, D. E., Norton, J. A., Singleton, C., & Swain, I. D. (1999) · Archives of Physical Medicine and Rehabilitation · doi:10.1016/s0003-9993(99)90333-7 · Cited by 275
Cutaneous whole nerve recordings used for correction of footdrop in hemiplegic man
Haugland, M. K., & Sinkjaer, T. (1995) · IEEE Transactions on Rehabilitation Engineering · doi:10.1109/86.481970 · Cited by 168
Cited by
Assessment of Walking Features From Foot Inertial Sensing
Sabatini, A. M., Martelloni, C., Scapellato, S., & Cavallo, F. (2005) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2004.840727 · Cited by 650
Measuring orientation of human body segments using miniature gyroscopes and accelerometers
Luinge, H. J., & Veltink, P. H. (2005) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345966 · Cited by 647
Self-citationRoetenberg, D., Luinge, H. J., Baten, C. T. M., & Veltink, P. H. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.847353 · Cited by 588
Self-citationYun, X., & Bachmann, E. R. (2006) · IEEE Transactions on Robotics · doi:10.1109/tro.2006.886270 · Cited by 569
3D gait assessment in young and elderly subjects using foot-worn inertial sensors
Mariani, B., Hoskovec, C., Rochat, S., Büla, C., Penders, J., & Aminian, K. (2010) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2010.07.003 · Cited by 370
Metadata
- Authors
- Veltink, P. H., Slycke, P., Hemssems, J., Buschman, R., Bultstra, G., & Hermens, H.
- Venue
- Medical Engineering & Physics
- Publisher
- IOP Publishing
- Published
- 2002-06-07
- Volume
- 25
- Issue
- 1
- Pages
- 21-28
- ISSN
- 1350-4533 (ISSN-L)
- 1873-4030
- OpenAlex
- W2105506122
- Cited by
- 99
- References
- 2
- FWCI
- 4.31
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Three dimensional inertial sensing of foot movements for automatic tuning of a two-channel implantable drop-foot stimulator. https://app.cochat.ai/papers/10.1016/s1350-4533(02)00041-3
- The paper
- Veltink, P. H., Slycke, P., Hemssems, J., Buschman, R., Bultstra, G., & Hermens, H. (2003). Three dimensional inertial sensing of foot movements for automatic tuning of a two-channel implantable drop-foot stimulator. Medical Engineering & Physics. https://doi.org/10.1016/s1350-4533(02)00041-3
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.