Measuring orientation of human body segments using miniature gyroscopes and accelerometers
Open accessAbstract
In the medical field, there is a need for small ambulatory sensor systems for measuring the kinematics of body segments. Current methods for ambulatory measurement of body orientation have limited accuracy when the body moves. The aim of the paper was to develop and validate a method for accurate measurement of the orientation of human body segments using an inertial measurement unit (IMU). An IMU containing three single-axis accelerometers and three single-axis micromachined gyroscopes was assembled in a rectangular box, sized 20 x 20 x 30 mm. The presented orientation estimation algorithm continuously corrected orientation estimates obtained by mathematical integration of the 3D angular velocity measured using the gyroscopes. The correction was performed using an inclination estimate continuously obtained using the signal of the 3D accelerometer. This reduces the integration drift that originates from errors in the angular velocity signal. In addition, the gyroscope offset was continuously recalibrated. The method was realised using a Kalman filter that took into account the spectra of the signals involved as well as a fluctuating gyroscope offset. The method was tested for movements of the pelvis, trunk and forearm. Although the problem of integration drift around the global vertical continuously increased in the order of 0.50 degrees s(-1), the inclination estimate was accurate within 3 degrees RMS. It was shown that the gyroscope offset could be estimated continuously during a trial. Using an initial offset error of 1 rad s(-1), after 2 min the off-set error was roughly 5% of the original offset error. Using the Kalman filter described, an accurate and robust system for ambulatory motion recording can be realised.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Accelerometer data is typically considered to be compound of linear acceleration, gravity, and white additive noise.
“Accelerometer data is typically considered to be compound of three factors: linear acceleration, gravity, and white additive noise [21].”
From Real-Time Estimation of Pathological Tremor Parameters from Gyroscope Data (2010) · cites this paper as [21]
Experimental — accuracy not yet measuredData from accelerometers are composed of a combination of linear acceleration, gravity, and additive noise.
“Data from accelerometers are composed of a combination of linear acceleration, gravity, and additive noise [16].”
From Neurological Tremor: Sensors, Signal Processing and Emerging Applications (2010) · cites this paper as [16]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
A New Approach to Linear Filtering and Prediction Problems
Kalman, R. E. (1960) · Journal of Basic Engineering · doi:10.1115/1.3662552 · Cited by 31,751 · see citing works
introduction to random signals and applied kalman filtering
Brown, R. G., & Hwang, P. Y. (1992) · Medical Entomology and Zoology · Cited by 3,384 · see citing works
Bouten, C. V. C., Koekkoek, K. K., Verduin, M., Kodde, R., & Janssen, J. J. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.554760 · Cited by 911 · see citing works
Inertial navigation systems for mobile robots
Barshan, B., & Durrant-Whyte, H. F. (1995) · IEEE Transactions on Robotics and Automation · doi:10.1109/70.388775 · Cited by 726 · see citing works
Mayagoitia, R. E., Nene, A. V., & Veltink, P. H. (2002) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00231-7 · Cited by 655 · see citing works
Self-citationEstimation of gait cycle characteristics by trunk accelerometry
Moe‐Nilssen, R., & Helbostad, J. L. (2003) · Journal of Biomechanics · doi:10.1016/s0021-9290(03)00233-1 · Cited by 641 · see citing works
A reliable gait phase detection system
Pappas, I. P. I., Popović, M. R., Keller, T., Dietz, V., & Morari, M. (2001) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/7333.928571 · Cited by 497 · see citing works
Detection of posture and motion by accelerometry: a validation study in ambulatory monitoring
Foerster, F., Smeja, M., & Fahrenberg, J. (1999) · Computers in Human Behavior · doi:10.1016/s0747-5632(99)00037-0 · Cited by 487 · see citing works
A practical gait analysis system using gyroscopes
Tong, K., & Granat, M. H. (1999) · Medical Engineering & Physics · doi:10.1016/s1350-4533(99)00030-2 · Cited by 460 · see citing works
Najafi, B., Aminian, K., Loew, F., Blanc, Y., & Robert, P. A. (2002) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2002.800763 · Cited by 435 · see citing works
A New Mathematical Formulation for Strapdown Inertial Navigation
Bortz, J. C. (1971) · IEEE Transactions on Aerospace and Electronic Systems · doi:10.1109/taes.1971.310252 · Cited by 417 · see citing works
Moe‐Nilssen, R. (1998) · Clinical Biomechanics · doi:10.1016/s0268-0033(98)00089-8 · Cited by 406 · see citing works
Inertial head-tracker sensor fusion by a complementary separate-bias Kalman filter
Foxlin, E. M. (1996) · doi:10.1109/vrais.1996.490527 · Cited by 366 · see citing works
Inclination measurement of human movement using a 3-D accelerometer with autocalibration
Luinge, H. J., & Veltink, P. H. (2004) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2003.822759 · Cited by 307 · see citing works
Self-citationA study of tremor in multiple sclerosis
Alusi, S. H. (2001) · Brain · doi:10.1093/brain/124.4.720 · Cited by 300 · see citing works
Dingwell, J. B., Cusumano, J. P., Sternad, D., & Cavanagh, P. R. (2000) · Journal of Biomechanics · doi:10.1016/s0021-9290(00)00092-0 · Cited by 270 · see citing works
Trunk accelerometry as a measure of balance control during quiet standing
Moe‐Nilssen, R., & Helbostad, J. L. (2002) · Gait & Posture · doi:10.1016/s0966-6362(01)00200-4 · Cited by 270 · see citing works
Detection of daily physical activities using a triaxial accelerometer
Mathie, M. J., Coster, A. C. F., Lovell, N. H., & Celler, B. G. (2003) · Medical & Biological Engineering & Computing · doi:10.1007/bf02348434 · Cited by 237 · see citing works
Uswatte, G., Miltner, W. H. R., Foo, B., Varma, M., Moran, S., & Taub, E. (2000) · Stroke · doi:10.1161/01.str.31.3.662 · Cited by 233 · see citing works
Moe‐Nilssen, R. (1998) · Clinical Biomechanics · doi:10.1016/s0268-0033(98)00090-4 · Cited by 221 · see citing works
Validity and reliability of triaxial accelerometers for inclinometry in posture analysis
Hansson, G., Asterland, P., Holmer, N.-G., & Skerfving, S. (2001) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345361 · Cited by 201 · see citing works
Miyazaki, S. (1997) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.605434 · Cited by 194 · see citing works
Detecting absolute human knee angle and angular velocity using accelerometers and rate gyroscopes
Williamson, R., & Andrews, B. J. (2001) · Medical & Biological Engineering & Computing · doi:10.1007/bf02345283 · Cited by 185 · see citing works
Automatic stance-swing phase detection from accelerometer data for peroneal nerve stimulation
Willemsen, A., Bloemhof, F., & Boom, H. B. K. (1990) · IEEE Transactions on Biomedical Engineering · doi:10.1109/10.64463 · Cited by 170 · see citing works
Procedure for effortless in-field calibration of three-axial rate gyro and accelerometers
Ferraris, F., Grimaldi, U., & Parvis, M. (1995) · PORTO@Iris · Cited by 163 · see citing works
Optimal strapdown attitude integration algorithms
Ignagni, M. (1990) · Journal of Guidance Control and Dynamics · doi:10.2514/3.20558 · Cited by 148 · see citing works
A triaxial accelerometer for measuring arm movements
Bernmark, E., & Wiktorin, C. (2002) · Applied Ergonomics · doi:10.1016/s0003-6870(02)00072-8 · Cited by 143 · see citing works
Bachmann, E. R. (2000) · Cited by 140 · see citing works
Real-time gait assessment utilizing a new way of accelerometry
Willemsen, A., van Alste, J. A., & Boom, H. B. K. (1990) · Journal of Biomechanics · doi:10.1016/0021-9290(90)90033-y · Cited by 129 · see citing works
Body position can be monitored in 3D using miniature accelerometers and earth-magnetic field sensors
Kemp, B., Janssen, A. J. M. W., & van der Kamp, B. (1998) · Electroencephalography and Clinical Neurophysiology/Electromyography and Motor Control · doi:10.1016/s0924-980x(98)00053-8 · Cited by 121 · see citing works
A method for inverse dynamic analysis using accelerometry
van den Bogert, A. J., Read, L., & Nigg, B. M. (1996) · Journal of Biomechanics · doi:10.1016/0021-9290(95)00155-7 · Cited by 99 · see citing works
Veltink, P. H., Slycke, P. J., Hemssems, J., Buschman, R., Bultstra, G., & Hermens, H. J. (2002) · Medical Engineering & Physics · doi:10.1016/s1350-4533(02)00041-3 · Cited by 99 · see citing works
Self-citationImproved strapdown coning algorithms
Jiang, Y. F., & Lin, Y. (1992) · IEEE Transactions on Aerospace and Electronic Systems · doi:10.1109/7.144574 · Cited by 87 · see citing works
An ambulatory dyskinesia monitor
Manson, A. J. (2000) · Journal of Neurology, Neurosurgery & Psychiatry · doi:10.1136/jnnp.68.2.196 · Cited by 79 · see citing works
Sensor systems for lower limb functional electircal stimulation (FES) control
Williamson, R., & Andrews, B. (2000) · Medical Engineering & Physics · doi:10.1016/s1350-4533(00)00038-2 · Cited by 41 · see citing works
Finite state control of functional electrical stimulation for the rehabilitation of gait
Sweeney, P. C., Lyons, G. M., & Veltink, P. H. (2000) · Medical & Biological Engineering & Computing · doi:10.1007/bf02344765 · Cited by 38 · see citing works
Self-citationVirtual artificial sensor technique for functional electrical stimulation
Tong, R. K., & Granat, M. H. (1998) · Medical Engineering & Physics · doi:10.1016/s1350-4533(98)00038-1 · Cited by 33 · see citing works
Design, fabrication and characterization of a highly symmetrical capacitive triaxial accelerometer
Lötters, J. C., Bomer, J. G., Verloop, A. J., Droog, E., Olthuis, W., Veltink, P. H., & Bergveld, P. (1998) · Sensors and Actuators A Physical · doi:10.1016/s0924-4247(98)00036-3 · Cited by 28 · see citing works
Self-citationInertial sensing in ambulatory back load estimation
Baten, C. T. M., Oosterhoff, P., Kingma, I., Veltink, P. H., & Hermens, H. J. (2002) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.1996.651834 · Cited by 18 · see citing works
Self-citationAssessment of inertial and gravitational inputs to the vestibular system
Baselli, G., Legnani, G., Franco, P., Brognoli, F., Marras, A., Quaranta, F., & Zappa, B. (2001) · Journal of Biomechanics · doi:10.1016/s0021-9290(01)00012-4 · Cited by 17 · see citing works
Flagged first, then by citations
Research works printed but not linked: 3
- BROWN, R. G., and HWANG, P. Y. C. (1997): 'Introduction to random signals and applied Kalmanfiltering' (John Wiley and Sons, 1997)
- FERRARIS, F., GRIMALDI, U., and PARVIS, M. (1995): 'Procedure for effortless in-field calibration of three-axis rate gyros and acceler- ometers', Sensors Mater., 7, pp. 311 330
- L~3TrERS, J., BOMER, J., VERLOOP, T., DROOG, E., OLTHUIS, W., VELTINK, P. H., and BERGVELD, P. (1998): 'In-use calibration pro- cedure for a triaxial accelerometer', Sensors Actuators A, Phys., 66, pp. 205 212 Medical & Biological Engineering & Computing 2005, Vol.
Cited by
Sorted by citations
Estimation of IMU and MARG orientation using a gradient descent algorithm
Madgwick, S., Harrison, A. J., & Vaidyanathan, R. (2011) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2011.5975346 · Cited by 2,108 · see citing works
IMU-Based Joint Angle Measurement for Gait Analysis
Seel, T., Raisch, J., & Schauer, T. (2014) · Sensors · doi:10.3390/s140406891 · Cited by 871 · see citing works
Accelerometry: A technique for quantifying movement patterns during walking
Kavanagh, J. J., & Menz, H. B. (2008) · Gait & Posture · doi:10.1016/j.gaitpost.2007.10.010 · Cited by 617 · see citing works
Activity identification using body-mounted sensors—a review of classification techniques
Preece, S. J., Goulermas, J. Y., Kenney, L. P. J., Howard, D., Meijer, K., & Crompton, R. H. (2009) · Physiological Measurement · doi:10.1088/0967-3334/30/4/r01 · Cited by 606 · see citing works
Camomilla, V., Bergamini, E., Fantozzi, S., & Vannozzi, G. (2018) · Sensors · doi:10.3390/s18030873 · Cited by 601 · see citing works
Roetenberg, D., Luinge, H. J., Baten, C. T. M., & Veltink, P. H. (2005) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2005.847353 · Cited by 588 · see citing works
Self-citationSmart wearable body sensors for patient self-assessment and monitoring
Appelboom, G., Camacho, E., Abraham, M. E., Bruce, S. S., Dumont, E. L., Zacharia, B. E., D’Amico, R. S., Slomian, J., Reginster, J., Bruyère, O., et al. (2014) · Archives of Public Health · doi:10.1186/2049-3258-72-28 · Cited by 449 · see citing works
Wearable soft sensing suit for human gait measurement
Mengüç, Y., Park, Y., Pei, H., Vogt, D. M., Aubin, P. M., Winchell, E., Fluke, L., Stirling, L., Wood, R. J., & Walsh, C. J. (2014) · The International Journal of Robotics Research · doi:10.1177/0278364914543793 · Cited by 420 · see citing works
The use of inertial sensors system for human motion analysis
Cuesta-Vargas, A. I., Galán-Mercant, A., & Williams, J. M. (2010) · Physical Therapy Reviews · doi:10.1179/1743288x11y.0000000006 · Cited by 369 · see citing works
Survey of Motion Tracking Methods Based on Inertial Sensors: A Focus on Upper Limb Human Motion
Filippeschi, A., Schmitz, N., Miezal, M., Bleser, G., Ruffaldi, E., & Stricker, D. (2017) · Sensors · doi:10.3390/s17061257 · Cited by 361 · see citing works
Ambulatory measurement of arm orientation
Luinge, H. J., Veltink, P. H., & Baten, C. T. M. (2006) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2005.11.011 · Cited by 356 · see citing works
Self-citationUsing Inertial Sensors for Position and Orientation Estimation
Kok, M., Hol, J. D., & Schön, T. B. (2017) · Foundations and Trends® in Signal Processing · doi:10.1561/2000000094 · Cited by 356 · see citing works
Full body gait analysis with Kinect
Gabel, M., Gilad-Bachrach, R., Renshaw, E., & Schuster, A. (2012) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2012.6346340 · Cited by 334 · see citing works
Inertial Measurement Units for Clinical Movement Analysis: Reliability and Concurrent Validity
Al-Amri, M., Nicholas, K., Button, K., Sparkes, V., Sheeran, L., & Davies, J. (2018) · Sensors · doi:10.3390/s18030719 · Cited by 318 · see citing works
Functional calibration procedure for 3D knee joint angle description using inertial sensors
Favre, J., Aïssaoui, R., Jolles, B. M., de Guise, J. A., & Aminian, K. (2009) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2009.06.025 · Cited by 302 · see citing works
Estimation of Ground Reaction Forces and Moments During Gait Using Only Inertial Motion Capture
Karatsidis, A., Bellusci, G., Schepers, H. T. A. M., de Zee, M., Andersen, M. S., & Veltink, P. H. (2016) · Sensors · doi:10.3390/s17010075 · Cited by 288 · see citing works
Self-citationInertial sensor-based knee flexion/extension angle estimation
Cooper, G., Sheret, I., McMillian, L., Siliverdis, K., Sha, N., Hodgins, D., Kenney, L. P. J., & Howard, D. (2009) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2009.08.004 · Cited by 281 · see citing works
An inertial and magnetic sensor based technique for joint angle measurement
O’Donovan, K. J., Kamnik, R., O'Keeffe, D. T., & Lyons, G. M. (2007) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2006.12.010 · Cited by 281 · see citing works
Bergamini, E., Ligorio, G., Summa, A., Vannozzi, G., Cappozzo, A., & Sabatini, A. M. (2014) · Sensors · doi:10.3390/s141018625 · Cited by 267 · see citing works
Shoulder and Elbow Joint Angle Tracking With Inertial Sensors
El‐Gohary, M., & McNames, J. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2208750 · Cited by 264 · see citing works
Development of a wearable sensor system for quantitative gait analysis
Liu, T., INOUE, Y., & SHIBATA, K. (2009) · Measurement · doi:10.1016/j.measurement.2009.02.002 · Cited by 262 · see citing works
Ergonomic analysis of construction worker's body postures using wearable mobile sensors
Nath, N. D., Akhavian, R., & Behzadan, A. H. (2017) · Applied Ergonomics · doi:10.1016/j.apergo.2017.02.007 · Cited by 237 · see citing works
Lee, J. K., Park, E. J., & Robinovitch, S. N. (2012) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2012.2187245 · Cited by 225 · see citing works
Prediction of ground reaction forces and moments during various activities of daily living
Fluit, R., Andersen, M. S., Kolk, S., Verdonschot, N. J. J., & Koopman, H. F. J. M. (2014) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2014.04.030 · Cited by 205 · see citing works
Gait posture estimation using wearable acceleration and gyro sensors
Takeda, R., TADANO, S., Akiko Natorigawa, A., Masahiro Todoh, A., & Yoshinari, S. (2009) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2009.07.016 · Cited by 197 · see citing works
Dejnabadi, H., Jolles, B. M., Casanova, E., Fua, P., & Aminian, K. (2006) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.873678 · Cited by 189 · see citing works
Ambulatory Assessment of Ankle and Foot Dynamics
Schepers, H. M., Koopman, H. F. J. M., & Veltink, P. H. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2006.889769 · Cited by 187 · see citing works
Self-citationLiu, X., Zhou, Z. G., Diao, W., Li, Z., & Zhang, K. (2015) · ACM SIGSAC Conference on Computer and Communications Security (CCS) · doi:10.1145/2810103.2813668 · Cited by 179 · see citing works
Accurate Attitude Estimation of a Moving Land Vehicle Using Low-Cost MEMS IMU Sensors
Ahmed, H., & Tahir, M. (2016) · IEEE Transactions on Intelligent Transportation Systems · doi:10.1109/tits.2016.2627536 · Cited by 177 · see citing works
Understanding Minds in Real-World Environments: Toward a Mobile Cognition Approach
Ladouce, S., Donaldson, D. I., Dudchenko, P. A., & Ietswaart, M. (2017) · Frontiers in Human Neuroscience · doi:10.3389/fnhum.2016.00694 · Cited by 169 · see citing works
Inertial and magnetic sensing of human motion
Roetenberg, D. (2006) · University of Twente Research Information · Cited by 166 · see citing works
Mobility assessment in older people: new possibilities and challenges
Zijlstra, W., & Aminian, K. (2007) · European Journal of Ageing · doi:10.1007/s10433-007-0041-9 · Cited by 153 · see citing works
Improved kinect-based spatiotemporal and kinematic treadmill gait assessment
Eltoukhy, M. A., Oh, J., Kuenze, C. M., & Signorile, J. F. (2016) · Gait & Posture · doi:10.1016/j.gaitpost.2016.10.001 · Cited by 148 · see citing works
An Algorithm for the In-Field Calibration of a MEMS IMU
Qureshi, U., & Golnaraghi, F. (2017) · IEEE Sensors Journal · doi:10.1109/jsen.2017.2751572 · Cited by 148 · see citing works
Complementary Observer for Body Segments Motion Capturing by Inertial and Magnetic Sensors
Fourati, H., Manamanni, N., Afilal, L., & Handrich, Y. (2012) · IEEE/ASME Transactions on Mechatronics · doi:10.1109/tmech.2012.2225151 · Cited by 145 · see citing works
Evaluation of Validity and Reliability of Inertial Measurement Unit-Based Gait Analysis Systems
Cho, Y.-S., Jang, S., Cho, J., Kim, M. J., Lee, H. D., Lee, S. Y., & Moon, S.-B. (2018) · Annals of Rehabilitation Medicine · doi:10.5535/arm.2018.42.6.872 · Cited by 140 · see citing works
Nazarahari, M., & Rouhani, H. (2020) · Information Fusion · doi:10.1016/j.inffus.2020.10.018 · Cited by 138 · see citing works
Neurological Tremor: Sensors, Signal Processing and Emerging Applications
Grimaldi, G., & Manto, M. (2010) · Sensors · doi:10.3390/s100201399 · Cited by 138 · see citing works
“Data from accelerometers are composed of a combination of linear acceleration, gravity, and additive noise [16].”
BackgroundMa, C., Wong, D. W., Lam, W., Wan, A., & Lee, W. C. C. (2016) · Sensors · doi:10.3390/s16040434 · Cited by 138 · see citing works
Bouvier, B., Duprey, S., Claudon, L., Dumas, R., & Savescu, A. (2015) · Sensors · doi:10.3390/s150818813 · Cited by 133 · see citing works
Won, S.-H. P., Golnaraghi, F., & Melek, W. (2008) · IEEE Transactions on Industrial Electronics · doi:10.1109/tie.2008.2010166 · Cited by 134 · see citing works
Benchmarking of a full-body inertial motion capture system for clinical gait analysis
Cloete, T., & Scheffer, C. (2008) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2008.4650232 · Cited by 131 · see citing works
Karatsidis, A., Jung, M., Schepers, H. M., Bellusci, G., de Zee, M., Veltink, P. H., & Andersen, M. S. (2019) · Medical Engineering & Physics · doi:10.1016/j.medengphy.2018.12.021 · Cited by 129 · see citing works
Self-citationFeasibility of using inertial sensors to assess human movement
Saber-Sheikh, K., Bryant, E. C., Glazzard, C., Hamel, A., & Lee, R. (2009) · Manual Therapy · doi:10.1016/j.math.2009.05.009 · Cited by 128 · see citing works
Monje, M. H. G., Foffani, G., Obeso, J. A., & Sánchez‐Ferro, Á. (2019) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-062117-121036 · Cited by 128 · see citing works
Countermovement jump performance assessment using a wearable 3D inertial measurement unit
Picerno, P., Camomilla, V., & Capranica, L. (2010) · Journal of Sports Sciences · doi:10.1080/02640414.2010.523089 · Cited by 125 · see citing works
Toward Automatic Activity Classification and Movement Assessment During a Sports Training Session
Ahmadi, A., Mitchell, E., Richter, C., Destelle, F., Gowing, M., O'Connor, N. E., & Moran, K. (2014) · IEEE Internet of Things Journal · doi:10.1109/jiot.2014.2377238 · Cited by 123 · see citing works
Peruzzi, A., Croce, U. D., & Cereatti, A. (2011) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2011.04.035 · Cited by 123 · see citing works
Jarchi, D., & Casson, A. J. (2016) · Data · doi:10.3390/data2010001 · Cited by 120 · see citing works
Johnson, W. R., Mian, A. S., Robinson, M. A., Verheul, J., Lloyd, D. G., & Alderson, J. A. (2020) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2020.3006158 · Cited by 126 · see citing works
Reducing Drifts in the Inertial Measurements of Wrist and Elbow Positions
Zhou, H., & Hu, H. (2009) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2009.2025065 · Cited by 120 · see citing works
Tian, Y., Wei, H., & Tan, J. (2013) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2012.2205706 · Cited by 119 · see citing works
Evaluation of instrumented shoes for ambulatory assessment of ground reaction forces
Liedtke, C. B., Fokkenrood, S. A. W., Menger, J. T., van der Kooij, H., & Veltink, P. H. (2006) · Gait & Posture · doi:10.1016/j.gaitpost.2006.07.017 · Cited by 117 · see citing works
Self-citationAn IMU-to-Body Alignment Method Applied to Human Gait Analysis
Vargas-Valencia, L., Elías, A., Rocón, E., Bastos-Filho, T., & Frizera, A. (2016) · Sensors · doi:10.3390/s16122090 · Cited by 116 · see citing works
Wearable Sensor-Based Rehabilitation Exercise Assessment for Knee Osteoarthritis
Chen, K.-H., Chen, P.-C., Liu, K.-C., & Chan, C.-T. (2015) · Sensors · doi:10.3390/s150204193 · Cited by 115 · see citing works
Human pose recovery using wireless inertial measurement units
Lin, J. F.-S., & Kulić, D. (2012) · Physiological Measurement · doi:10.1088/0967-3334/33/12/2099 · Cited by 112 · see citing works
Wearable Sensing for Solid Biomechanics
Wong, C., Zhang, Z., Lo, B. P. L., & Yang, G. (2015) · IEEE Sensors Journal · doi:10.1109/jsen.2015.2393883 · Cited by 110 · see citing works
Ambulatory human motion tracking by fusion of inertial and magnetic sensing with adaptive actuation
Schepers, H. M., Roetenberg, D., & Veltink, P. H. (2009) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-009-0562-9 · Cited by 110 · see citing works
Self-citationReal-Time Estimation of Pathological Tremor Parameters from Gyroscope Data
Gallego, J., Rocon, E., Roa, J. O., Moreno, J., & Pons, J. L. (2010) · Sensors · doi:10.3390/s100302129 · Cited by 107 · see citing works
Accuracy of inertial motion sensors in static, quasistatic, and complex dynamic motion
Godwin, A., Agnew, M., & Stevenson, J. (2009) · Journal of Biomechanical Engineering · doi:10.1115/1.4000109 · Cited by 107 · see citing works
Movement Primitive Segmentation for Human Motion Modeling: A Framework for Analysis
Lin, J. F.-S., Karg, M. E., & Kulić, D. (2016) · IEEE Transactions on Human-Machine Systems · doi:10.1109/thms.2015.2493536 · Cited by 107 · see citing works
Kitagawa, N., & Ogihara, N. (2016) · Gait & Posture · doi:10.1016/j.gaitpost.2016.01.014 · Cited by 106 · see citing works
Evaluation of clinical spasticity assessment in Cerebral palsy using inertial sensors
van den Noort, J. C., Scholtes, V. A., & Harlaar, J. (2009) · Gait & Posture · doi:10.1016/j.gaitpost.2009.05.011 · Cited by 106 · see citing works
A novel art of continuous noninvasive blood pressure measurement
Fortin, J., Rogge, D. E., Fellner, C., Flotzinger, D., Grond, J., Lerche, K., & Saugel, B. (2021) · Nature Communications · doi:10.1038/s41467-021-21271-8 · Cited by 104 · see citing works
Support vector machine for classification of walking conditions using miniature kinematic sensors
Lau, H.-Y., Tong, R. K., & Zhu, H. (2008) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-008-0327-x · Cited by 103 · see citing works
Callaway, A. J., Cobb, J. E., & Jones, I. (2009) · International Journal of Sports Science & Coaching · doi:10.1260/1747-9541.4.1.139 · Cited by 103 · see citing works
Wearable Stretch Sensors for Motion Measurement of the Wrist Joint Based on Dielectric Elastomers
Huang, B., Li, M., Mei, T., McCoul, D., Qin, S., Zhao, Z., & Zhao, J. (2017) · Sensors · doi:10.3390/s17122708 · Cited by 102 · see citing works
Novel approach to ambulatory assessment of human segmental orientation on a wearable sensor system
Liu, K., Liu, T., SHIBATA, K., INOUE, Y., & Zheng, R. (2009) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2009.08.008 · Cited by 102 · see citing works
Sensor Evaluation for Wearable Strain Gauges in Neurological Rehabilitation
Giorgino, T., Tormene, P., Lorussi, F., De Rossi, D., & Quaglini, S. (2009) · IEEE Transactions on Neural Systems and Rehabilitation Engineering · doi:10.1109/tnsre.2009.2019584 · Cited by 100 · see citing works
Development of a real-time three-dimensional spinal motion measurement system for clinical practice
Goodvin, C., Park, E. J., Huang, K. I., & Sakaki, K. D. R. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0132-3 · Cited by 100 · see citing works
Quantitative evaluation of movement using the timed up-and-go test
Higashi, Y., Yamakoshi, K.-I., Fujimoto, T., Sekine, M., & TAMURA, T. (2008) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2008.919494 · Cited by 99 · see citing works
Estimation of Upper-Limb Orientation Based on Accelerometer and Gyroscope Measurements
Hyde, R. A., Ketteringham, L. P., Neild, S. A., & Jones, R. J. S. (2008) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.912647 · Cited by 98 · see citing works
Estimating Movement Smoothness From Inertial Measurement Units
Melendez-Calderon, A., Shirota, C., & Balasubramanian, S. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2020.558771 · Cited by 95 · see citing works
Schall, M. C., Fethke, N. B., Chen, H., Oyama, S., & Douphrate, D. I. (2016) · Ergonomics · doi:10.1080/00140139.2015.1079335 · Cited by 96 · see citing works
Wearable Goniometer and Accelerometer Sensory Fusion for Knee Joint Angle Measurement in Daily Life
Tognetti, A., Lorussi, F., Carbonaro, N., & Rossi, D. D. (2015) · Sensors · doi:10.3390/s151128435 · Cited by 95 · see citing works
Shahabpoor, E., & Pavić, A. (2017) · Sensors · doi:10.3390/s17092085 · Cited by 94 · see citing works
Kalman filter for mobile-robot attitude estimation: Novel optimized and adaptive solutions
Odry, Á., Fuller, R. C., Rudas, I. J., & Odry, P. (2018) · Mechanical Systems and Signal Processing · doi:10.1016/j.ymssp.2018.03.053 · Cited by 94 · see citing works
Ubiquitous Human Upper-Limb Motion Estimation using Wearable Sensors
Zhang, Z., Wong, W., & Wu, J. (2011) · IEEE Transactions on Information Technology in Biomedicine · doi:10.1109/titb.2011.2159122 · Cited by 92 · see citing works
Robust Foot Clearance Estimation Based on the Integration of Foot-Mounted IMU Acceleration Data
Benoussaad, M., Sijobert, B., Mombaur, K., & Coste, C. (2015) · Sensors · doi:10.3390/s16010012 · Cited by 90 · see citing works
Use of Inertial Sensors for Ambulatory Assessment of Center-of-Mass Displacements During Walking
Floor-Westerdijk, M. J., Schepers, H. M., Veltink, P. H., van Asseldonk, E. H. F., & Buurke, J. H. (2012) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2012.2197211 · Cited by 89 · see citing works
Self-citationBergmann, J. H., Mayagoitia, R. E., & Smith, I. C. (2009) · Dynamic Medicine · doi:10.1186/1476-5918-8-3 · Cited by 88 · see citing works
Gait mode recognition and control for a portable-powered ankle-foot orthosis
Li, Y. D., & Hsiao‐Wecksler, E. T. (2013) · International Conference on Rehabilitation Robotics · doi:10.1109/icorr.2013.6650373 · Cited by 88 · see citing works
SVM-based posture identification with a single waist-located triaxial accelerometer
Rodríguez-Martín, D., Samà, A., Pérez‐López, C., Catalá, A., Cabestany, J., & Rodríguez‐Molinero, A. (2013) · Expert Systems with Applications · doi:10.1016/j.eswa.2013.07.028 · Cited by 88 · see citing works
Study of the motion artefacts of skin-mounted inertial sensors under different attachment conditions
Forner‐Cordero, A., Mateu-Arce, M., Forner‐Cordero, I., Alcántara, E., Moreno, J. C., & Pons, J. L. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/4/n01 · Cited by 88 · see citing works
Faber, G. S., Chang, C.-C., Kingma, I., Dennerlein, J. T., & van Dieën, J. H. (2015) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2015.11.042 · Cited by 88 · see citing works
A method to qualitatively assess arm use in stroke survivors in the home environment
Leuenberger, K., Gonzenbach, R. R., Wachter, S., Luft, A. R., & Gassert, R. (2016) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-016-1496-7 · Cited by 86 · see citing works
Drift Removal for Improving the Accuracy of Gait Parameters Using Wearable Sensor Systems
Takeda, R., Lisco, G., Fujisawa, T., Gastaldi, L., Tohyama, H., & Tadano, S. (2014) · Sensors · doi:10.3390/s141223230 · Cited by 84 · see citing works
Ambulatory estimation of foot placement during walking using inertial sensors
Schepers, H. M., van Asseldonk, E. H. F., Baten, C. T. M., & Veltink, P. H. (2010) · Journal of Biomechanics · doi:10.1016/j.jbiomech.2010.07.039 · Cited by 85 · see citing works
Self-citationLutz, J., Memmert, D., Raabe, D., Dornberger, R., & Donath, L. (2019) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17010059 · Cited by 85 · see citing works
Kinematics of Gait: New Method for Angle Estimation Based on Accelerometers
Djuric-Jovicic, M. D., Jovičić, N. S., & Popović, D. B. (2011) · Sensors · doi:10.3390/s111110571 · Cited by 84 · see citing works
Pacher, L., Châtellier, C., Vauzelle, R., & Fradet, L. (2020) · Sensors · doi:10.3390/s20113322 · Cited by 83 · see citing works
Konrath, J. M., Karatsidis, A., Schepers, H. M., Bellusci, G., de Zee, M., & Andersen, M. S. (2019) · Sensors · doi:10.3390/s19071681 · Cited by 81 · see citing works
Ha, T. H., Saber-Sheikh, K., Moore, A. P., & Jones, M. P. (2012) · Manual Therapy · doi:10.1016/j.math.2012.04.003 · Cited by 81 · see citing works
Minimum-Order Kalman Filter With Vector Selector for Accurate Estimation of Human Body Orientation
Lee, J. K., & Park, E. J. (2009) · IEEE Transactions on Robotics · doi:10.1109/tro.2009.2017146 · Cited by 81 · see citing works
Dynamic accuracy of inertial measurement units during simple pendulum motion
Brodie, M. A. D., Walmsley, A., & Page, W. (2008) · Computer Methods in Biomechanics & Biomedical Engineering · doi:10.1080/10255840802125526 · Cited by 77 · see citing works
Ahmadi, A., Mitchell, E., Destelle, F., Gowing, M., O'Connor, N. E., Richter, C., & Moran, K. (2014) · Wearable and Implantable Body Sensor Networks · doi:10.1109/bsn.2014.29 · Cited by 77 · see citing works
Braz, G. P., Russold, M., & Davis, G. M. (2009) · Neuromodulation Technology at the Neural Interface · doi:10.1111/j.1525-1403.2009.00213.x · Cited by 77 · see citing works
Joint Inertial Sensor Orientation Drift Reduction for Highly Dynamic Movements
Fasel, B., Spörri, J., Chardonnens, J., Kröll, J., Müller, E. A., & Aminian, K. (2017) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2017.2659758 · Cited by 75 · see citing works
Foo, J. Y. A., & Wilson, S. J. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-005-0008-y · Cited by 74 · see citing works
Deep Learning Identifies Digital Biomarkers for Self-Reported Parkinson's Disease
Zhang, H., Deng, K., Li, H., Albin, R. L., & Guan, Y. (2020) · Patterns · doi:10.1016/j.patter.2020.100042 · Cited by 74 · see citing works
- Authors
- Luinge, H. J., & Veltink, P. H.
- Affiliations
- University of Twente (NL)
- Corresponding author
- Veltink, P. H.
- ORCID
- Veltink, P. H.
- Venue
- Medical & Biological Engineering & Computing
- Published
- 2005-04
- Volume
- 43
- Issue
- 2
- Pages
- 273-282
- ISSN
- 0140-0118 (ISSN-L)
- 1741-0444
- OpenAlex
- W2020496717
- Cited by
- 647
- References
- 40
- Citations by year
- 2026: 5 · 2025: 10 · 2024: 13 · 2023: 20 · 2022: 30 · 2021: 21 · 2020: 31 · 2019: 39 · 2018: 34 · 2017: 44 · 2016: 43 · 2015: 46 · 2014: 47 · 2013: 42 · 2012: 38
- FWCI
- 40.88
- Type
- journal article
- Language
- en
- Open access
- green
- research.utwente.nl
- Licence
- www.springer.com/tdm (tdm)
- Locations
- Medical & Biological Engineering & Computing (publishedVersion)
- PubMed (publishedVersion)
- University of Twente Research Information (submittedVersion)
- University of Twente Research Information (submittedVersion, other-oa)
- Funders
- Stichting voor de Technische Wetenschappen
- Topics
- Physical Sciences › Engineering › Aerospace Engineering › Inertial Sensor and Navigation
- Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- inertial measurement unit, Kalman filter, body kinematics, MEMS gyroscope, accelerometer, pelvis, trunk, forearm
- MeSH terms
- Acceleration, Biomechanical Phenomena, Humans, Models, Biological, Movement, Orientation, Signal Processing, Computer-Assisted, Monitoring, Ambulatory
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review