Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference?
Open accessAbstract
During blood pressure (BP) measurement, the recommended positioning of the cuff bladder center is directly above the brachial artery. We investigated the relevance of incorrect cuff positioning during (1) auscultatory measurement with an appropriate or improperly small cuff and (2) oscillometric measurement with a wide-range cuff designed to guarantee accurate measurements regardless of position. In subjects with wide BP and arm circumference ranges, (1) auscultatory BP was repeatedly measured with a properly positioned cuff (reference) and, simultaneously, with an identical cuff placed on the other arm in either a correct or an incorrect position (test). The measurements were performed with a properly sized (N=57) or an improperly small cuff (N=33). (2) Auscultatory measurements obtained with a properly positioned and sized cuff were compared with oscillometric measurements obtained with a specially designed wide-range cuff (Omron IntelliWrap) placed on the contralateral arm either in a correct or an incorrect position. Auscultatory BP measures were unaffected by incorrect positioning of a properly sized cuff, whereas with undercuffing, BP was overestimated with the cuff displaced by 90° laterally (systolic/diastolic BP differences: 4.9±4.6/4.0±4.6 mm Hg, P<0.01) or by 180° (3.9±5.4/4.2±5.1 mm Hg, P<0.01) in relation to the correct position. Incorrect placement of the oscillometric cuff had no significant effect on the accuracy of the measurements (difference with correct position <1.5 mm Hg). Incorrect cuff positioning introduces a systematic overestimation of auscultatory BP when the cuff is too small in relation to arm circumference but not when it is correctly sized. No systematic error was observed with oscillometric measurements obtained with a specially designed wide-range cuff.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The clinical application and accuracy of HBPM may be limited by conditions related to the individual, the procedure, the oscillometric technique, or cuff-related issues.
“the clinical application and the accuracy of HBPM may be limited by certain conditions related either to the individual [63–66], to the procedure itself [4,67], to the oscillometric technique for BP measurement or to cuff-related issues [4,65,68–74].”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Intraclass correlations: Uses in assessing rater reliability.
Shrout, P. E., & Fleiss, J. L. (1979) · Psychological Bulletin · doi:10.1037/0033-2909.86.2.420 · Cited by 23,433 · see citing works
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · see citing works
Forouzanfar, M. H., Afshin, A., Alexander, L. T., Anderson, H. R., Bhutta, Z. A., Biryukov, S., Brauer, M., Burnett, R., Cercy, K., Charlson, F. J., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(16)31679-8 · Cited by 7,698 · see citing works
2007 Guidelines for the Management of Arterial Hypertension
Mancia, G., Backer, G. D., Dominiczak, A. F., Cífková, R., Fagard, R. H., Germanò, G., Grassi, G. M., Heagerty, A. M., Kjeldsen, S. E., Laurent, S., et al. (2007) · Journal of Hypertension · doi:10.1097/hjh.0b013e3281fc975a · Cited by 5,818 · see citing works
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985 · see citing works
Forouzanfar, M. H., Alexander, L., Anderson, H. R., Bachman, V. F., Biryukov, S., Brauer, M., Burnett, R., Casey, D., Coates, M. M., Cohen, A., et al. (2015) · The Lancet · doi:10.1016/s0140-6736(15)00128-2 · Cited by 2,701 · see citing works
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051 · see citing works
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655 · see citing works
Self-citationEuropean Society of hypertension position paper on ambulatory blood pressure monitoring
O’Brien, E., Parati, G., Stergiou, G., Asmar, R., Beilin, L., Bilo, G., Clement, D., de la Sierra, A., de Leeuw, P., Dolan, E., et al. (2013) · Journal of Hypertension · doi:10.1097/hjh.0b013e328363e964 · Cited by 1,419 · see citing works
Self-citationParati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e328308da66 · Cited by 837 · see citing works
Self-citation2013 ESH/ESC Guidelines for the management of arterial hypertension
Mancia, G., Fagard, R. H., Narkiewicz, K., Redón, J., Zanchetti, A., Böhm, M., Christiaens, T. C., Cífková, R., Backer, G. D., Dominiczak, A. F., et al. (2013) · Blood Pressure · doi:10.3109/08037051.2013.812549 · Cited by 425 · see citing works
Reliability of reporting self-measured blood pressure values by hypertensive patients.
Mengden, T., Medina, R. M. H., Beltran, B., Álvarez, E., Kraft, K., & Vetter, H. (1998) · American Journal of Hypertension · doi:10.1016/s0895-7061(98)00241-6 · Cited by 296 · see citing works
Effect of Vertical Displacement of the Arm on Indirect Blood-Pressure Measurement
Mitchell, P. L., Parlin, R. W., & Blackburn, H. (1964) · New England Journal of Medicine · doi:10.1056/nejm196407092710204 · Cited by 125 · see citing works
Miscuffing: inappropriate blood pressure cuff application.
Manning, D. M., Kuchirka, C., & Kamiński, J. (1983) · Circulation · doi:10.1161/01.cir.68.4.763 · Cited by 108 · see citing works
Parati, G., Omboni, S., & Bilo, G. (2009) · HYPERTENSION · doi:10.1161/hypertensionaha.108.122853 · Cited by 90 · see citing works
Self-citationEffect of back support and stethoscope head on seated blood pressure determinations.
Cushman, W. C., Cooper, K., Horne, R. A., & Meydrech, E. F. (1990) · American Journal of Hypertension · doi:10.1093/ajh/3.3.240 · Cited by 75 · see citing works
The effect of cuff size on blood pressure measurement in adults.
Sprafka, J., Strickland, D., Gomez-Marin, O. W., & Prineas, R. J. (1991) · Epidemiology · doi:10.1097/00001648-199105000-00010 · Cited by 65 · see citing works
The effect of crossing legs on blood pressure: a randomized single-blind cross-over study.
Peters, G. L., Binder, S. K., & Campbell, N. R. C. (1999) · Blood Pressure Monitoring · doi:10.1097/00126097-199900420-00007 · Cited by 62 · see citing works
Rectangular cuffs may overestimate blood pressure in individuals with large conical arms
Palatini, P., Benetti, E., Fania, C., Malipiero, G., & Saladini, F. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32834f98a0 · Cited by 54 · see citing works
Accuracy of home blood pressure readings: monitors and operators
Stryker, T., Wilson, M. P., & Wilson, T. (2004) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000130224.18777.ce · Cited by 48 · see citing works
(In)accuracy of blood pressure measurement in 14 Italian hospitals
Manzoli, L., Simonetti, V., D'Errico, M., de Vito, C., Flacco, M., Forni, C., Torre, G. L., Liguori, G., Messina, G., Mezzetti, A., et al. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e3283577b20 · Cited by 44 · see citing works
Graves, J., Bailey, K., & Sheps, S. (2003) · Blood Pressure Monitoring · doi:10.1097/00126097-200312000-00001 · Cited by 38 · see citing works
A survey of validated automated home blood pressure monitors available for the Internet shopper
Graves, J. (2005) · Blood Pressure Monitoring · doi:10.1097/00126097-200504000-00009 · Cited by 38 · see citing works
Boivin, J., Tsou-Gaillet, T.-J., Fay, R., Dobre, D., Rossignol, P., & Zannad, F. (2011) · Journal of Hypertension · doi:10.1097/hjh.0b013e32834b7efb · Cited by 36 · see citing works
The unsupported arm: a cause of falsely raised blood pressure readings.
Silverberg, D., Shemesh, E., & Iaina, A. (1977) · British medical journal · doi:10.1136/bmj.2.6098.1331 · Cited by 34 · see citing works
Mourad, J., Lopez-Sublet, M., Bahous, S. A., Villeneuve, F., Jaboureck, O., Dourmap-Collas, C., Denolle, T., Fourcade, J., & Baguet, J. (2013) · American Journal of Hypertension · doi:10.1093/ajh/hpt084 · Cited by 27 · see citing works
2013 ESH/ESC GUIDELINES FOR THE MANAGEMENT OF ARTERIAL HYPERTENSION
Mancia, G., Fagard, R. H., Narkiewicz, K., Redón, J., Zanchetti, A., Böhm, M., Christiaens, T. C., Cífková, R., Backer, G. D., Dominiczak, A. F., et al. (2014) · Russian Journal of Cardiology · doi:10.15829/1560-4071-2014-1-7-94 · Cited by 26 · see citing works
Treatment of the Infertile Marriage
Rc, B. (1948) · BMJ · doi:10.1136/bmj.2.4584.851 · Cited by 13 · see citing works
Tahlawi, M. E., Abdelbaset, M., Gouda, M., & Hussein, I. (2015) · Hypertension Research · doi:10.1038/hr.2014.170 · Cited by 11 · see citing works
[Selection and maintenance of the blood pressure monitor for self-blood pressure monitoring (SBPM)].
Wiliński, J., Kusiak, A., Wiliński, B., Smolik, P., Klima, Ł., Chmielewska, J., & Czarnecka, D. (2013) · PubMed · Cited by 2 · see citing works
[Information deficits concerning blood pressure self-measurement].
Patyna, W., Börsch, B., Patyna, M., & Mitrović, V. (2004) · Deutsche Medizinische Wochenschrift · doi:10.1055/s-2004-835288 · Cited by 2 · see citing works
Takahashi, H., Yoshika, M., & Yokoi, T. (2011) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e328348b688
ABC of blood pressure measurement: technique.
O'Brien, E. T., & O'Malley, K. (1979) · BMJ · doi:10.1136/bmj.2.6196.982
Flagged first, then by citations
2 of the 35 references couldn’t be matched to a known paper.
Cited by
Sorted by citations
Parati, G., Stergiou, G. S., Bilo, G., Kollias, A., Pengo, M., Ochoa, J. E., Agarwal, R., Asayama, K., Asmar, R., Burnier, M., et al. (2021) · Journal of Hypertension · doi:10.1097/hjh.0000000000002922 · Cited by 178 · see citing works
Self-citationOzoemena, E. L., Iweama, C., Agbaje, O., Umoke, P. C., Ene, O. C., Ofili, P. C., Agu, B., Orisa, C. U., Agu, M., & Anthony, E. (2019) · Archives of Public Health · doi:10.1186/s13690-019-0349-x · Cited by 73 · see citing works
Al‐Qatatsheh, A., Morsi, Y., Zavabeti, A., Zolfagharian, A., Salim, N. V., Kouzani, A. Z., Mosadegh, B., & Gharaie, S. (2020) · Sensors · doi:10.3390/s20164484 · Cited by 68 · see citing works
Teong, C. Y., Huang, C.-C., & Sun, F.-J. (2020) · Scientific Reports · doi:10.1038/s41598-020-65711-9 · Cited by 29 · see citing works
Spitz, R., Bell, Z., Wong, V., Viana, R., Chatakondi, R., Abe, T., & Loenneke, J. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab64b8 · Cited by 24
Park, S.-H., & Park, Y.-S. (2019) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000412 · Cited by 23 · see citing works
A Survey on Blood Pressure Measurement Technologies: Addressing Potential Sources of Bias
Mousavi, S. S., Reyna, M. A., Clifford, G., & Sameni, R. (2023) · Italian National Conference on Sensors · doi:10.3390/s24061730 · Cited by 19 · see citing works
“2) Cuff position: The brachial artery is commonly used for BP measurement [56], [144].”
BackgroundBlood Pressure and Cardio-Metabolic Risk Profile in Young Saudi Males in a University Setting
El-Ashker, S., Pednekar, M., Narake, S., Albaker, W., & Al-Hariri, M. (2021) · Medicina · doi:10.3390/medicina57080755 · Cited by 16
Single-Pill, Triple Antihypertensive Therapy in Rural Sub-Saharan Africa: Preliminary Experience
Stroppa, C., Hunjan, I., Umulisa, A., Irebe, B., Parati, G., Bianchetti, M., Muvunyi, B., Ntaganda, E., Sinabubaraga, V., Radovanović, D., et al. (2024) · Cardiology and Therapy · doi:10.1007/s40119-024-00358-5 · Cited by 13
“…established over three sets of blood pressure measurements taken by trained healthcare personnel at three different visits on separate days, employing a clinically validated, automated, oscillometric device [8, 9] and carefully following international blood pressure measurement recommendations [7].”
Self-citationLi, Y., Li, F., Li, Y., Cui, X.-R., Li, J., Zhi, H., Wang, W.-D., Sun, Y.-Y., & Cui, W. (2020) · The Journal of Clinical Hypertension · doi:10.1111/jch.13902 · Cited by 11 · see citing works
Grover-Páez, F., Cardona-Muñoz, E., Cardona-Muller, D., Guzmán-Saldívar, V. H., de la Cerda, M. L. R., Jiménez-Cázarez, M., Totsuka-Sutto, S., Alanis-Sánchez, G., & Ramos-Becerra, C. (2017) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000290 · Cited by 8
Tamborini, A., & Gharib, M. (2024) · Journal of the American Heart Association : Cardiovascular and Cerebrovascular Disease · doi:10.1161/jaha.123.033290 · Cited by 7 · see citing works
Sarkar, S., & Pahuja, S. K. (2024) · Biomedical Materials & Devices · doi:10.1007/s44174-023-00146-0 · Cited by 7 · see citing works
Lessof, C., Cooper, R., Wong, A., Bendayan, R., Caleyachetty, R., Cheshire, H., Cosco, T. D., Elhakeem, A., Hansell, A. L., Kaushal, A., et al. (2023) · PLoS ONE · doi:10.1371/journal.pone.0289052 · Cited by 7 · see citing works
“Blood pressure is affected by multiple factors [10] including the participant talking, actively listening, being exposed to cold, ingesting alcohol, having a distended bladder, recent smoking [44] and also to measurement protocols such as arm position and cuff size [45].”
BackgroundRaine, A., Choy, O., Achenbach, T., & Liu, J.-H. (2023) · European Child and Adolescent Psychiatry · doi:10.1007/s00787-023-02247-z · Cited by 7 · see citing works
“Children’s resting heart rate was assessed with an Omron M3 Comfort (HEM-7134-E) Automatic Upper Arm Blood Pressure Monitor whose Intelli Wrap cuff technology obtains accurate results in any position around the upper arm [7].”
Yüksel, S., Altun-Uğraş, G., Altinok, N., & Demir, N. (2020) · Florence Nightingale journal of nursing · doi:10.5152/fnjn.2020.19119 · Cited by 7
“This finding supports the results from previous research emphasizing that MUAC measurement and use of a properly sized cuff is essential for accurate BPM (Bilo et al., 2017; Mishra et al., 2017; Mourad et al., 2013; Ringrose et al. 2015; Watson et al., 2011).”
BackgroundMethodResultInfluentialA Novel Wearable Device for Continuous Blood Pressure Monitoring Utilizing Strain Gauge Technology
McMurray, J. P., DeVries, A., Frazee, K., Sizemore, B., Branan, K., Jennings, R., & Coté, G. L. (2025) · Biosensors · doi:10.3390/bios15070413 · Cited by 4 · see citing works
“For example, blood pressure accuracy is reduced when the device is misaligned with the artery; if the subject talks, shifts positions, or contracts their muscles during the measurement; or if they have not rested for at least 5 to 25 min prior to taking the measurement [9–11].”
Joukar, F., Yeganeh, S., Naghipour, M., Hassanipour, S., Nikbakht, H.-A., & Mansour-ghanaei, F. (2020) · Medical devices · doi:10.2147/mder.s253638 · Cited by 4 · see citing works
Blood pressure screening in Mata Sector, a rural area of Rwanda
Hunjan, I., Umulisa, A., Parati, G., Bianchetti, M., Milani, G., Muvunyi, B., Ntaganda, E., Radovanović, D., Stroppa, C., Suter, P., et al. (2024) · Journal of Human Hypertension · doi:10.1038/s41371-024-00912-7 · Cited by 3 · see citing works
“This pre-formed, expanded coverage BP cuff in fl ates around the upper arm to fi t arms with a circumference of 22 – 42 cm and ensures accurate readings [8].”
Self-citationTakahashi, H., Yamashita, S., & Yakura, N. (2024) · Vascular Health and Risk Management · doi:10.2147/vhrm.s468382 · Cited by 2 · see citing works
Hou, G.-C., Wu, Y.-L., Wang, J.-Y., & Zhi, J. (2024) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000688 · Cited by 2 · see citing works
Resistant Hypertension and Cardiorenovascular Risk
Prkačin, I., Vrdoljak, P., Cavrić, G., Važanić, D., Pervan, P., & Adam, V. (2017) · BANTAO Journal · doi:10.1515/bj-2017-0002 · Cited by 2 · see citing works
New trends in BP home monitoring
Chetan, M., & Moldovan, C. (2017) · International Symposium on Advanced Topics in Electrical Engineering · doi:10.1109/atee.2017.7905140 · Cited by 2 · see citing works
“accuracy of the measurements we can see the recent appearance of the Intelli Wrap cuff [7], in comparison with standard cuff.”
BackgroundMethodInfluentialCuff Design for Home Blood Pressure Monitors
Palatini, P., Asmar, R., Bilo, G., & Parati, G. (2020) · Updates in Hypertension and Cardiovascular Protection · doi:10.1007/978-3-030-23065-4_2 · Cited by 2 · see citing works
Self-citationSeoane-Casqueiro, Ó., Vázquez-Agra, N., Cruces-Sande, A., López-Prado, E., López-Paz, J., Camafort, M., Casariego-Vales, E., Hermida-Ameijeiras, Á., & Pose-Reino, A. (2026) · Revista clínica española (Ed. impresa) · doi:10.1016/j.rce.2026.502523 · Cited by 1 · see citing works
Chen, Y.-T., Xu, Z.-K., Ma, Y., Chen, H.-S., Xu, H.-B., Ye, Y., Xue, F., Zhang, H.-G., Su, S., & Liu, X.-Y. (2026) · Digital Health · doi:10.1177/20552076261456613 · Cited by 1 · see citing works
Saladini, F., Albini, F., Guardo, A. D., Fra, G. P., Mos, L., Omboni, S., Predotti, P., Rabbia, F., Tocci, G., Parati, G., et al. (2025) · High Blood Pressure & Cardiovascular Prevention · doi:10.1007/s40292-025-00778-x · Cited by 1 · see citing works
“Despite the fundamental importance of a correct office BP measurement as the first line diagnostic method in hypertension, several previous studies identified important problems in applying a correct methodology of this assessment in real life settings [6 –13 ].”
Self-citationSelf-blame, posttraumatic stress, and cardiovascular reactivity among sexual trauma survivors.
Sinnott, S. M., Park, C. L., Pescatello, L., & Sharda, K. (2024) · Stress and Health · doi:10.1002/smi.3453 · Cited by 1 · see citing works
Jin, Y.-K., Bei, X., & Jin, Y. (2024) · Blood Pressure Monitoring · doi:10.1097/mbp.0000000000000716 · Cited by 1
Ringrose, J., & Sridar, S. (2020) · The Journal of Clinical Hypertension · doi:10.1111/jch.13900 · Cited by 1 · see citing works
Prkačin, I., Rudan, D., Svaguša, T., Bakoš, M., & Šalek, Z. (2018) · Blood Pressure · doi:10.5772/intechopen.73004 · Cited by 1 · see citing works
Balamurugan, N., Thaha, T. M. I., A., A., A., A. M., & C, E. (2025) · Indian Journal of Clinical Cardiology · doi:10.1177/26324636251361613
“24 In critically ill patients, especially in those with hypotension, BP measured by cuff devices may be more unreliable.”
Gregory, P. (2025) · Journal of Paramedic Practice · doi:10.12968/jpar.2025.0017
Biceps muscle injury associated with non-invasive automatic blood pressure monitoring
Nicol, R., Lin, I., & Prasad, L. (2023) · BMJ Case Reports · doi:10.1136/bcr-2023-256459
Ergin, E., Aygün, H., Şahan, S., & İsmailoğlu, E. G. (2023) · Ege Üniversitesi Hemşirelik Fakültesi Dergisi · doi:10.53490/egehemsire.1114708
R., S., Ganesh, A., Jaiganesh, K., & S, P. (2022) · International Journal of Life Science and Pharma Research · doi:10.22376/ijpbs/lpr.2022.12.6.sp25.l34-40
Naqvi, S., Potluri, P., Mandal, P., & Lewis, P. (2022) · Journal of the Textile Institute · doi:10.1080/00405000.2022.2124615
Agreement between Student Paramedics When Measuring Blood Pressure
Hutchins, A., Momo, A. T., Tran, L., Thyer, L., & Simpson, P. M. (2019) · Australasian Journal of Paramedicine · doi:10.33151/ajp.16.630
Firdaus, Ayu, F., Herisanty, W., Ainiyah, N., & Hasanah, N. (2021) · Indian Journal of Forensic Medicine & Toxicology · doi:10.37506/ijfmt.v15i3.15756
Impact on variation in the direction of cuff bladder on blood pressure readings
Grema, B., Mohammed, I., Michael, G., & Aliyu, I. (2019) · Journal of the Practice of Cardiovascular Sciences · doi:10.4103/jpcs.jpcs_36_19
“Journal of the Practice of Cardiovascular Sciences ¦ Volume 5 ¦ Issue 3 ¦ September-December 2019 175 Bilo et al.[14] reported that incorrect bladder cuff position did not have any significant impact on BP readings; so far, the bladder size was appropriate.”
Background
- Authors
- Bilo, G., Sala, O., Perego, C., Faini, A., Gao, L., Głuszewska, A., Ochoa, J. E., Pellegrini, D., Lonati, L. M., & Parati, G.
- Affiliations
- IRCCS Istituto Auxologico Italiano (IT)
- Peking University First Hospital (CN)
- Jagiellonian University (PL)
- Corresponding author
- Parati, Gianfranco
- ORCID
- Venue
- Hypertension Research
- Published
- 2017-01-12
- Volume
- 40
- Issue
- 6
- Pages
- 573-580
- ISSN
- 0916-9636 (ISSN-L)
- 1348-4214
- PMC
- PMC5506235
- OpenAlex
- W2575424177
- Cited by
- 43
- References
- 35
- Citations by year
- 2026: 3 · 2025: 4 · 2024: 9 · 2023: 5 · 2022: 3 · 2021: 3 · 2020: 6 · 2019: 6 · 2018: 1 · 2017: 3
- FWCI
- 3.39
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Hypertension Research (publishedVersion, cc-by-nc-sa)
- Europe PMC (PubMed Central) (publishedVersion, cc-by-nc-sa)
- PubMed Central (publishedVersion, cc-by-nc-sa)
- PubMed (publishedVersion)
- BOA (University of Milano-Bicocca) (submittedVersion, cc-by-nc-sa)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Blood Pressure and Hypertension Studies
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- blood pressure measurement, cuff size, auscultatory blood pressure, oscillometric blood pressure, brachial artery, arm circumference, measurement accuracy
- MeSH terms
- Adult, Aged, Blood Pressure Determination, Female, Humans, Male, Middle Aged, Sphygmomanometers, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review