Comparing peripheral limb and forehead vital sign monitoring in newborn infants at birth
Open accessAbstract
Background: To study the feasibility of measuring heart rate (HR) and oxygen saturation (SpO 2 ) on the forehead, during newborn transition at birth, and to compare these measurements with those obtained from the wrist. Methods: Vital signs were measured and compared between forehead-mounted reflectance (remittance) photoplethysmography sensor (fhPPG) and a wrist-mounted pulse oximeter sensor (wrPO), from 20 enrolled term newborns born via elective caesarean section, during the first 10 min of life. Results: From the datasets available ( n = 13), the median (IQR) sensor placement times for fhPPG, ECG and wrPO were 129 (70) s, 143 (68) s, and 159 (76) s, respectively, with data recorded for up to 10 min after birth. The success rate (percentage of total possible HR values reported once sited) of fhPPG (median = 100%) was higher compared to wrPO (median = 69%) during the first 6 min of life ( P < 0.005). Both devices exhibited good HR agreement with ECG, achieving >95% agreement by 3 (fhPPG) and 4 (wrPO) min. SpO 2 for fhPPG correlated with wrPO ( r = 0.88), but there were significant differences in SpO 2 between the two devices between 3 and 8 min ( P < 0.005), with less variance observed with fhPPG SpO 2 . Conclusion: In the period of newborn transition at birth in healthy term infants, forehead measurement of vital signs was feasible and exhibited greater HR accuracy and higher estimated SpO 2 values compared to wrist-sited pulse oximetry. Further investigation of forehead monitoring based on the potential benefits over peripheral monitoring is warranted. Impact: This study demonstrates the feasibility of continuously monitoring heart rate and oxygen saturation from an infant’s forehead in the delivery room immediately after birth. Significantly higher SpO 2 measurements were observed from the forehead than the wrist during the transition from foetal to newborn life. Continuous monitoring of vital signs from the forehead could become a valuable tool to improve the delivery of optimal care provided for newborns at birth.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
For infants who experience hypothermia or low body temperatures, distal measurement locations like the hand and foot often face limited perfusion to protect vital organs.
“In addition, for infants who experience hypothermia or low body temperatures, which is common for infants in the Ethiopian NICU, distal measurement locations like the hand and foot often face limited perfusion to protect vital organs8–11.”
From Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia (2025) · cites this paper as [11]
The Introduction says peripheral monitoring in newborns can be affected by poor tissue perfusion, and the Discussion says fingers, toes and wrists are prone to vasoconstriction and low perfusion. The paper does not establish that this occurs specifically in hypothermic infants to protect vital organs, or that hypothermia is common in Ethiopian NICUs; its newborn study involved healthy term infants at birth. (LLM, conf=0.85)
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Part 11: Neonatal resuscitation
Wyllie, J. P., Perlman, J. M., Kattwinkel, J., Atkins, D. L., Chameides, L., Goldsmith, J. P., Guinsburg, R., Hazinski, M. F., Morley, C. J., Richmond, S. W. J., et al. (2010) · Resuscitation · doi:10.1016/j.resuscitation.2010.08.029 · Cited by 602 · see citing works
Defining the Reference Range for Oxygen Saturation for Infants After Birth
Dawson, J. A., Kamlin, C. O. F., Vento, M., et al. (2010) · PEDIATRICS · doi:10.1542/peds.2009-1510 · Cited by 556 · see citing works
Part 11: Neonatal Resuscitation
Perlman, J. M., Wyllie, J. P., Kattwinkel, J., Atkins, D. L., Chameides, L., Goldsmith, J. P., Guinsburg, R., Hazinski, M. F., Morley, C. J., Richmond, S. W. J., et al. (2010) · Circulation · doi:10.1161/circulationaha.110.971127 · Cited by 551 · see citing works
Heart Rate Assessment Immediately after Birth
Phillipos, E., Solevåg, A. L., Pichler, G., Aziz, K., van Os, S., O’Reilly, M., Cheung, P., & Schmölzer, G. M. (2015) · Neonatology · doi:10.1159/000441940 · Cited by 531 · see citing works
Madar, J., Roehr, C. C., Ainsworth, S. B., Ersdal, H. L., Morley, C. J., Rüdiger, M., Skåre, C., Szczapa, T., Pas, A. B. T., Trevisanuto, D., et al. (2021) · Resuscitation · doi:10.1016/j.resuscitation.2021.02.014 · Cited by 521 · see citing works
Jubran, A. (2015) · Critical Care · doi:10.1186/s13054-015-0984-8 · Cited by 511 · see citing works
Vento, M., Moro, M., Escrig, R., et al. (2009) · PEDIATRICS · doi:10.1542/peds.2009-0434 · Cited by 470 · see citing works
Fetal Physiology and the Transition to Extrauterine Life
Morton, S. U., & Brodsky, D. D. (2016) · Clinics in Perinatology · doi:10.1016/j.clp.2016.04.001 · Cited by 339 · see citing works
Pulse Oximetry: Theory and Applications for Noninvasive Monitoring
Mendelson, Y. (1992) · Clinical Chemistry · doi:10.1093/clinchem/38.9.1601 · Cited by 320 · see citing works
Accuracy and reliability of fetal heart rate monitoring using maternal abdominal surface electrodes
Cohen, W. R., Ommani, S., Hassan, S., Mirza, F., Solomon, M., BROWN, R., Schifrin, B. S., Himsworth, J. M., & Hayes‐Gill, B. (2012) · Acta Obstetricia Et Gynecologica Scandinavica · doi:10.1111/j.1600-0412.2012.01533.x · Cited by 177 · see citing works
Self-citationPulse Oximetry Measures a Lower Heart Rate at Birth Compared with Electrocardiography
van Vonderen, J. J., Hooper, S. B., Kroese, J. K., Roest, A. A. W., Narayen, I. C., van Zwet, E. W., & Pas, A. B. T. (2014) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2014.09.015 · Cited by 150 · see citing works
Transitional Changes in Cerebral Blood Volume at Birth
Schwaberger, B., Pichler, G., Binder‐Heschl, C., et al. (2015) · Neonatology · doi:10.1159/000437347 · Cited by 136 · see citing works
Longmore, S. K., Lui, G. Y., Naik, G. R., Breen, P. P., Jalaludin, B. B., & Gargiulo, G. D. (2019) · Sensors · doi:10.3390/s19081874 · Cited by 133 · see citing works
Measuring Physiological Changes during the Transition to Life after Birth
van Vonderen, J. J., Roest, A. A. W., Siew, M. L., et al. (2014) · Neonatology · doi:10.1159/000356704 · Cited by 121 · see citing works
Seifi, S., Khatony, A., Moradi, G., Abdi, A., & Najafi, F. (2018) · BMC Nursing · doi:10.1186/s12912-018-0283-1 · Cited by 79 · see citing works
Schallom, L., Sona, C. S., McSweeney, M., & Mazuski, J. E. (2007) · Heart & Lung · doi:10.1016/j.hrtlng.2006.07.007 · Cited by 75 · see citing works
Mendelson, Y., & Pujary, C. J. (2004) · doi:10.1109/iembs.2003.1280775 · Cited by 74 · see citing works
Evaluation of a New Pulse Oximeter Sensor
Fernandez, M., Burns, K., Calhoun, B., George, S., Martin, B., & Weaver, C. (2007) · American Journal of Critical Care · doi:10.4037/ajcc2007.16.2.146 · Cited by 66 · see citing works
Grubb, M., Carpenter, J., Crowe, J. A., Teoh, J. Y., Marlow, N., Ward, C., Mann, C., Sharkey, D., & Hayes‐Gill, B. (2014) · Physiological Measurement · doi:10.1088/0967-3334/35/5/881 · Cited by 56 · see citing works
Self-citationForehead oximetry in critically ill patients: the case for a new monitoring site
Branson, R. D., & Mannheimer, P. D. (2004) · Respiratory Care Clinics · doi:10.1016/j.rcc.2004.04.003 · Cited by 48 · see citing works
Nesseler, N., Frenel, J., Launey, Y., et al. (2012) · Intensive Care Medicine · doi:10.1007/s00134-012-2659-0 · Cited by 45 · see citing works
Forehead SpO 2 monitoring compared to finger SpO 2 recording in emergency transport
Nuhr, M. J., Hoerauf, K. H., Joldzo, A., Frickey, N., Barker, R., Gorove, L., Puskas, T., & Kober, A. (2004) · Anaesthesia · doi:10.1111/j.1365-2044.2004.03673.x · Cited by 40 · see citing works
Oxygen saturation trends in preterm infants during the first 15 min after birth
Nuntnarumit, P., Rojnueangnit, K., & Tangnoo, A. (2009) · Journal of Perinatology · doi:10.1038/jp.2009.178 · Cited by 40 · see citing works
Accurate neonatal heart rate monitoring using a new wireless, cap mounted device
Henry, C., Shipley, L., Ward, C., et al. (2020) · Acta Paediatrica · doi:10.1111/apa.15303 · Cited by 40 · see citing works
Self-citationBerkenbosch, J. W., & Tobias, J. D. (2006) · Respiratory Care · doi:10.4187/respcare.06510726 · Cited by 38 · see citing works
A Comparison of Transmittance and Reflectance Pulse Oximetry During Vascular Surgery
Wax, D. B., Rubin, P., & Neustein, S. M. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3181bbc446 · Cited by 36 · see citing works
Iglesias, B., Rodríguez, M. J., Aleo, E., et al. (2015) · Anales de Pediatría · doi:10.1016/j.anpedi.2015.08.012 · Cited by 28 · see citing works
Oxygen saturation in healthy newborns; influence of birth weight and mode of delivery
Røsvik, A., Øymar, K. A. A., Kvaløy, J. T., et al. (2009) · Journal of Perinatal Medicine · doi:10.1515/jpm.2009.070 · Cited by 25 · see citing works
Iglesias, B., Rodríguez, M. J., Aleo, E., et al. (2016) · Anales de Pediatría (English Edition) · doi:10.1016/j.anpede.2015.08.017 · Cited by 23 · see citing works
Perfusion Changes at the Forehead Measured by Photoplethysmography during a Head-Down Tilt Protocol
Ysehak Abay, T., Shafqat, K., & Kyriacou, P. A. (2019) · Biosensors · doi:10.3390/bios9020071 · Cited by 22 · see citing works
Reflective oxygen saturation monitoring at hypothenar and its validation by human hypoxia experiment
Guo, T., Cao, Z., Zhang, Z., Li, D., & Yu, M. (2015) · BioMedical Engineering OnLine · doi:10.1186/s12938-015-0071-z · Cited by 20 · see citing works
Oxygen Saturation Levels during the First Minutes of Life in Healthy Term Neonates
Zübarioğlu, U., Uslu, S., Can, E., et al. (2011) · The Tohoku Journal of Experimental Medicine · doi:10.1620/tjem.224.273 · Cited by 18 · see citing works
Oxygen saturation of healthy term neonates during the first 30 minutes of life
Beşkardeş, A., Salihoğlu, Ö., Can, E., et al. (2012) · Pediatrics International · doi:10.1111/j.1442-200x.2012.03731.x · Cited by 14 · see citing works
Kelly, K. L., Carlson, A. R., Allison, T. G., & Johnson, B. D. (2019) · Heart & Lung · doi:10.1016/j.hrtlng.2019.10.012 · Cited by 14 · see citing works
Pulse Oximetry Using a Disposable Finger Sensor Placed on the Forehead in Hypoxic Patients
Smithline, H. A., Rudnitzky, N., Macomber, S., et al. (2009) · Journal of Emergency Medicine · doi:10.1016/j.jemermed.2009.03.019 · Cited by 9 · see citing works
How to use saturation monitoring in newborns
McVea, S., McGowan, M., & Rao, B. (2018) · Archives of Disease in Childhood Education & Practice · doi:10.1136/archdischild-2017-313499 · Cited by 6 · see citing works
Resuscitation Council UK: review of updated 2021 neonatal life support guideline
Pemberton, C., & Howarth, C. (2022) · Archives of Disease in Childhood Education & Practice · doi:10.1136/archdischild-2021-323277 · Cited by 6 · see citing works
Forehead monitoring of heart rate in neonatal intensive care
Stockwell, S. J., Kwok, T. C., Morgan, S. P., Sharkey, D., & Hayes‐Gill, B. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1127419 · Cited by 6 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 1
- [18]Chen, W., Ayoola, I., Oetomo, S. B. & Feijs, L. In 2010 Design, Automation & Test in Europe Conference & Exhibition (DATE 2010) 1530–1535 (Institute of Electrical and Electronic Engineers, 2010).
Cited by
Sorted by citations
Zhou, L., Aljiffry, A., Lee, Y. J., Matthews, J., Seitter, B., Soltis, I., Huang, Y., Maher, K., & Yeo, W.-H. (2025) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2025.117372 · Cited by 9 · see citing works
Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia
Zhou, L., Joseph, M., Lee, Y. J., Yadav, D., Nayak, L., Woodall, J., Matthews, J., Soltis, I., Mekuria, F. M., Yeshanew, K. H., et al. (2025) · npj Digital Medicine · doi:10.1038/s41746-025-01974-8 · Cited by 3 · see citing works
“In addition, for infants who experience hypothermia or low body temperatures, which is common for infants in the Ethiopian NICU, distal 8 – 11 .”
- Authors
- Swamy, S. K. N., Stockwell, S. J., Liu, C., Henry, C., Shipley, L., Ward, C., Mirahmadi, S., Correia, R., Morgan, S. P., Crowe, J. A., Sharkey, D., & Hayes-Gill, B. R.
- Affiliations
- University of Nottingham (GB)
- ORCID
- Venue
- Pediatric Research
- Published
- 2024-10-17
- Volume
- 99
- Issue
- 2
- Pages
- 598-603
- ISSN
- 0031-3998 (ISSN-L)
- 1530-0447
- PMC
- PMC12956579
- OpenAlex
- W4403510694
- Cited by
- 7
- References
- 38
- Citations by year
- 2026: 2 · 2025: 5
- FWCI
- 0.77
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Locations
- Pediatric Research (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Repository@Nottingham (University of Nottingham) (publishedVersion, other-oa)
- PubMed Central (publishedVersion, cc-by)
- Repository@Nottingham (University of Nottingham) (submittedVersion, cc-by)
- Funders
- UK Research and Innovation
- University of Nottingham (EP/R511730/1)
- Medical Research Council (MC_PC_15012)
- Engineering and Physical Sciences Research Council (EP/R513283/1, EP/K503101/1, EP/T517902/1, EP/R511730/1)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Keywords
- pulse oximetry, heart rate monitoring, oxygen saturation, transition at birth, term infants, sensor placement, non-invasive monitoring, delivery room care
- MeSH terms
- Oxygen Saturation, Electrocardiography, Feasibility Studies, Female, Forehead, Heart Rate, Humans, Infant, Newborn, Male, Monitoring, Physiologic, Oximetry, Wrist, Photoplethysmography, Vital Signs
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review