Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia
Open accessThe DOI matches its Crossref record; 21 of the 21 references it cites were found in a registry or an index; 8 of the 14 sources its claims cite were checked for support.
Abstract
The neonatal period is critical and stressful, particularly in low-resource settings where existing monitoring methods for newborns are sporadic and labor-intensive. There is a pressing need for an affordable, user-friendly device that offers continuous and effective monitoring. Here, we present a soft, all-in-one, wearable system designed for continuous wireless monitoring of neonatal health in low-resource settings. Developed using advanced packaging technologies, this system features a chest-mounted patch and a forehead-mounted pulse oximeter that transmit real-time data to a smartphone app. The integrated device measures clinical-grade parameters, including electrocardiograms, heart rate, respiration rate, temperature, and blood oxygen saturation, showing excellent correlation with hospital-grade tabletop devices and reduced motion artifacts. This pilot study demonstrates a significant improvement over current monitoring methods by providing continuous oversight, ensures critical events are not missed, and automates the process for enhanced neonatal care, resulting in 84% satisfaction by parents interested in using the system at home.
Overview
CoChat Summary
TL;DR
This paper reports a soft, all-in-one wearable system—a chest patch plus forehead pulse oximeter—for continuous neonatal vital-sign monitoring in low-resource settings, tested in an Ethiopian NICU pilot with 25 neonates. It shows feasibility, generally improved agreement when compared against ECG-derived reference values, and high parent willingness to use the system at home, while highlighting training and device-appearance concerns.
Key findings
- The system continuously measures ECG, heart rate, respiration rate, temperature, and SpO₂, transmitting wirelessly via Bluetooth Low Energy to a smartphone/tablet app; the chest patch runs 14 h per charge and the forehead oximeter 22 h, with a listed cost of <$60.
- In the TASH NICU pilot (25 neonates, 798 data entries, 592 h), ECG-derived reference comparisons showed mean differences of 4.8 bpm for heart rate (95% CI [−7.5, 17]), 2.6 breaths/min for respiration rate (95% CI [−4.7, 9.8]), and −0.04% for SpO₂ (95% CI [−5.6, 5.5]); manual-reference comparisons were wider: 12 bpm (95% CI [−49, 72]), −1.4 breaths/min (95% CI [−25, 22]), and −1.6% (95% CI [−9.7, 6.6]).
- Device characterization found PCB dry electrodes had an SNR of 25.35 dB versus 12.06 dB for gel electrodes; electrode connectors were limited to 30 mating cycles; adhesive maintained adhesion after 20 isopropyl-alcohol sanitization cycles; and a 180° bend test (n = 10,000) increased ground-plane resistance by only 14.34 milliohms.
- Parent survey results: 84% said they would use the biopatch at home, 89% stated it was user-friendly, 89% agreed technical support is needed, and 58% preferred the biopatch over manual measurements for future use.
Methods
- Device design/fabrication: flexible PCBs with elastomer encapsulation, reusable dry PCB electrodes, and a forehead PPG sensor; electronics included an nRF52832 BLE microcontroller, ADS1291 ECG ADC sampled at 250 Hz, TMP117 temperature sensor at 1 Hz, and MAX30102 PPG at 50 Hz. The app used offline-first edge processing, encrypted SQLite storage, cloud sync when connectivity was available, and NTP-based time synchronization.
- Human subject pilot: approved by the Ethiopian Food and Drug Administration and IRB; 25 neonates aged 0–7 days admitted to TASH NICU were recruited using Schmettow’s sample-size methodology. First 10 patients used manual reference measurements (auscultation, clip oximeter, manual respiration count, axillary thermometer); next 15 used reusable suction cup ECG monitoring for heart and respiration rate. Nurses logged values, and parents completed usability/acceptability surveys.
- Analysis: Bland-Altman comparisons for heart rate, respiration rate, and SpO₂; Pan-Tompkins QRS detection for ECG-derived heart rate; SpO₂ from red/IR AC/DC ratios; motion-quality index for ECG motion handling; and peel testing for adhesive strength (n = 20).
Limitations
- Stated: this was a pilot study with only 25 patients; future iterations should use a hybrid rigid/flexible design to ruggedize the reusable portion; parents were wary of device appearance and exposed electrical components; training was the biggest parent concern; and the lack of a continuous measurement reference made it difficult to assess SpO₂ accuracy against the TASH NICU reference method.
- Evident from the text: the study was single-center and short-term (median participation 31 h, range 5–84 h), with intermittent manual reference measurements and paper journal logging; some sensor validation tests used healthy adults rather than neonates, including the forehead oximeter breath-hold comparison (n = 15 adults) and electrode signal comparison on adult physiology.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
5 Valid · 1 Not valid · 3 Not checked · 75 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“In addition, for infants who experience hypothermia or low body temperatures, which is common for infants in the Ethiopian NICU, distal measurement locations like the hand and foot often face limited perfusion to protect vital organs8–11.”
§ Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia > Results > Design and structure of the chest and forehead devices
What the sources say
- Not checked — no full text of the sourcecites [8] Berkenbosch, J. W. & Tobias, J. D. “Comparison of a new forehead reflectance pulse oximeter sensor with a conventional digit sensor in pediatric patients. Respir. Care 51, 726 (2006). source
- Not checked — no full text of the sourcecites [9] MacLeod, D. B. et al. The desaturation response time of finger pulse oximeters during mild hypothermia. Anaesthesia 60, 65–71 (2005). source
- Could not verifycites [10] Rao, A. et al. Clinical study of continuous non-invasive blood pressure monitoring in neonates. Sensors 23, 3690 (2023). sourceExperimental — accuracy not yet measured
The Introduction and Discussion address neonatal blood-pressure monitoring, and the paper describes measurements at the wrist or foot, but none of these passages discusses reduced distal perfusion during hypothermia. Searches for hypothermia, low temperature, perfusion, and vasoconstriction found no passage supporting the claimed mechanism; page numbers were not available in our copy. (LLM, conf=0.74)
- Partly supportedcites [11] Swamy, S. K. N. et al. Comparing peripheral limb and forehead vital sign monitoring in newborn infants at birth. Pediatric Res. 2451, 1 (2024) sourceExperimental — accuracy not yet measured
The Introduction says peripheral monitoring in newborns can be affected by poor tissue perfusion, and the Discussion says fingers, toes and wrists are prone to vasoconstriction and low perfusion. The paper does not establish that this occurs specifically in hypothermic infants to protect vital organs, or that hypothermia is common in Ethiopian NICUs; its newborn study involved healthy term infants at birth. (LLM, conf=0.85)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Affordable, continuous physiological monitoring systems are a crucial tool for early deterioration detection and intervention of at-risk neonates, but few studies have targeted this challenge to improve the quality, access, and efficiency of care to reduce neonatal mortality in low-resource settings20,21.”
§ Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia > Discussion
What the sources say
- Partly supportedcites [20] Ginsburg, A. S. et al. Evaluation of non-invasive continuous physiological monitoring devices for neonates in Nairobi, Kenya: a research protocol. BMJ Open 10, e035184 (2020). sourceExperimental — accuracy not yet measured
The Introduction (page unavailable) describes the need for low-cost continuous monitoring devices suitable for neonates in low-resource settings and outlines a project evaluating them. It does not say that few studies have addressed this challenge or assess how much research exists. (LLM, conf=0.82)
- Partly supportedcites [21] Ginsburg, A. S. et al. Multiparameter continuous physiological monitoring technologies in neonates among health care providers and caregivers at a private tertiary hospital in Nairobi, Kenya: feasibility, usability, and acceptability study. J. Med. Internet Res. 23, e29755 (2021). sourceExperimental — accuracy not yet measured
The Introduction says evidence on the feasibility, appropriateness, and acceptability of neonatal health technologies has not been adequately investigated. The Discussion identifies a literature gap concerning barriers to and enablers of these technologies in resource-constrained settings. Neither passage establishes that few studies have targeted affordable continuous monitoring specifically. (LLM, conf=0.78)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
Every source the paper cites for this claim was checked and supports it.
In the paper
“Using Schmettow’s methodology for sample size estimation22, a total of 25 neonates (n = 798 data entries) were recruited to assess the first generation of the device.”
§ Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia > Methods > Human subject study
What the sources say
- Supportedcites [22] Schmettow, M. Sample size in usability studies. Commun. ACM 55, 64–70 (2012). source
The citation is attached to the use of Schmettow’s sample-size methodology, not to the citing authors’ recruitment count. In “the timing of control” (page unavailable), the paper describes the LNBzt model as supporting sample-size prediction and study-continuation decisions; it does not report the 25 neonates or 798 data entries. (LLM, conf=0.91)
plainn = 798- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Previous wearables have been designed to condense the complicated setups in NICU monitoring into miniaturized wearables for facilitated monitoring4–6; however, they encounter problems with device longevity with either single-use or short-term use before replacement.”
§ Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia > Discussion
What the sources say
- Not checked — no full text of the sourcecites [4] Zhou, L. et al. Soft imperceptible wearable electronics for at-home cardiovascular monitoring of infants with single ventricle heart disease. Biosens. Bioelectron. 278, 117372 (2025). source
- Partly supportedcites [5] Chung, H. U. et al. “Skin-interfaced biosensors for advanced wireless physiological monitoring in neonatal and pediatric intensive-care units. Nat. Med. 26, 418–429 (2020). sourceExperimental — accuracy not yet measured
The Abstract describes an earlier NICU wearable as mechanically fragile, which lends some support to a general longevity concern. It does not say that previous wearables were single-use or had to be replaced after short-term use. The operating lifetimes in Results > Device and System Design refer to this paper’s devices and their batteries, not replacement of earlier wearables; page numbers are unavailable in our copy. (LLM, conf=0.76)
- Could not verifycites [6] Chung, H. U. et al. “Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care. Science 363, eaau0780 (2019). sourceExperimental — accuracy not yet measured
The Abstract describes wireless, skin-like devices for NICU monitoring, but does not say that previous wearables were single-use or needed replacement after short-term use. The section on pilot studies says this paper’s devices can be sterilized for safe reuse; it does not establish a longevity problem for previous wearables. I searched the indexed copy for single-use, replacement, reuse, and device lifetime. (LLM, conf=0.72)
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“The total power consumption of the device is approximately 5 mW when paired to Bluetooth and has an idle consumption of 2 mW.”
§ Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia > Results > Design and structure of the chest and forehead devices
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
51 more claims
Create a free account to see all 84 claims and every source verdict.
- Probably not a substantial claim. 9
- It may state two separate claims as one. 21
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
“replacement electrodes are $1.95 (estimated from the cost of 120 PCBs for $124.34 from a 3rd party manufacturer)”
§ Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia > Results > Design and structure of the chest and forehead devices
Reported 1.95 · recomputed 1.036167
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia > Results > Real-time measurement study in the neonatal intensive care unit
“Results revealed that 48% of parents found the biopatch easy to use”
§ Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia > Results > Real-time measurement study in the neonatal intensive care unit
“0% reported “very difficult”; 11% found it difficult; the rest found it easy”
Citations
Cites
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855 · cited 1×
“Real-timeheart andrespirationratesarederivedfromtherawECGsignalusingthePan-TompkinsalgorithmforQRS-complexdetection 14 .”
Binodal, wireless epidermal electronic systems with in-sensor analytics for neonatal intensive care
Chung, H. U., Kim, B. H., Lee, J. Y., Lee, J., Xie, Z., Ibler, E. M., Lee, K., Banks, A., Jeong, J. Y., Kim, J., et al. (2019) · Science · doi:10.1126/science.aau0780 · Cited by 807 · cited 1×
“Previous wearables have been designed to condense the complicated setups in NICU monitoring into miniaturized wearables for facilitated monitoring 4 – 6 ; however, they encounter problems with device longevity with either single-use or short-term use before replacement.”
Self-citationPulse oximetry: Understanding its basic principles facilitates appreciation of its limitations
Chan, E. D., Chan, M. M., & Chan, M. M. (2013) · Respiratory Medicine · doi:10.1016/j.rmed.2013.02.004 · Cited by 710 · cited 6×
“The resulting ratio is paired with an experimentally derived calibration curve to estimate SpO2 15 – 18 .”
InfluentialChung, H. U., Rwei, A. Y., Hourlier-Fargette, A., Xu, S., Lee, K., Dunne, E. C., Xie, Z., Liu, C., Carlini, A., Kim, D. H., et al. (2020) · Nature Medicine · doi:10.1038/s41591-020-0792-9 · Cited by 482
Neonatal mortality in Ethiopia: trends and determinants
Mekonnen, Y., Tensou, B., Telake, D. S., Degefie, T., & Bekele, A. (2013) · BMC Public Health · doi:10.1186/1471-2458-13-483 · Cited by 234
The Light–Tissue Interaction of Pulse Oximetry
Mannheimer, P. D. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000269522.84942.54 · Cited by 194 · cited 1×
“Theforehead is an ideal blood oxygen measurement location for its stability and high vascular density 7 .”
Sample size in usability studies
Schmettow, M. (2012) · Communications of the ACM · doi:10.1145/2133806.2133824 · Cited by 123
The desaturation response time of finger pulse oximeters during mild hypothermia*
MacLeod, D. B., Cortinez, L. I., Keifer, J. C., Cameron, D., Wright, D. R., White, W. D., Moretti, E. W., Radulescu, L. R., & Somma, J. (2005) · Anaesthesia · doi:10.1111/j.1365-2044.2004.04033.x · Cited by 98
Calibration of Contactless Pulse Oximetry
Verkruysse, W., Bartula, M., Bresch, E., Rocque, M., Meftah, M., & Kirenko, I. (2017) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000001381 · Cited by 92
Respiratory sinus arrhythmia in new‐born infants.
Hathorn, M. K. (1987) · The Journal of Physiology · doi:10.1113/jphysiol.1987.sp016480 · Cited by 47
Research works printed but not linked: 1
- Aynalem, Y. A., Mekonen, H., Akalu, T. Y., Gebremichael, B. & Shiferaw, W. S. “Preterm Neonatal Mortality and its predictors in Tikur Anbessa Specialized Hospital, Addis Ababa, Ethiopia: a retrospective cohort study. Ethiopian J. health Sci. 31, 43–54 (2021).
Cited by
Motion artefact-mitigated bioelectronics using meta-topological hydrogel
Kim, K. R., & Yeo, W. (2026) · Nature Sensors · doi:10.1038/s44460-026-00062-y
Self-citation
Metadata
- Authors
- Zhou, L., Joseph, M., Lee, Y. J., Yadav, D., Nayak, L., Woodall, J., Matthews, J., Soltis, I., Mekuria, F. M., Yeshanew, K. H., Mamo, Y. K., Fekadu, A., Demtse, A., Gleason, R., & Yeo, W.-H.
- Affiliations
- Georgia Institute of Technology (US)
- Addis Ababa University (ET)
- Tikur Anbessa Hospital (ET)
- Brighton and Sussex Medical School (GB)
- Corresponding author
- Yeo, Woon-Hong
- ORCID
- Venue
- npj Digital Medicine
- Published
- 2025-09-25
- Volume
- 8
- Issue
- 1
- Pages
- 575-575
- ISSN
- 2398-6352 (ISSN-L)
- PMC
- PMC12462490
- OpenAlex
- W4414490313
- Semantic Scholar
- 3332ea1b2e261f8e18e3a62683f000eff7f459e0
- Cited by
- 3
- References
- 21
- Citations by year
- 2026: 3
- FWCI
- 0.72
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- npj Digital Medicine (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed (publishedVersion)
- PubMed Central (publishedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Imlay Foundation,United States
- Bill and Melinda Gates Foundation (INV- 006189)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Wireless Body Area Networks
- Keywords
- wearable sensors, nanomembranes, pulse oximetry, electrocardiogram, continuous vital sign monitoring, resource-limited settings, smartphone data transmission, motion artifact reduction, home-based neonatal care, Ethiopia
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia. https://app.cochat.ai/papers/10.1038/s41746-025-01974-8
- The paper
- Zhou, L., Joseph, M., Lee, Y. J., Yadav, D., Nayak, L., Woodall, J., Matthews, J., Soltis, I., Mekuria, F. M., Yeshanew, K. H., Mamo, Y. K., Fekadu, A., Demtse, A., Gleason, R., & Yeo, W.-H. (2025). Soft, all-in-one, nanomembrane wearable system for advancing neonatal health monitoring in Ethiopia. npj Digital Medicine. https://doi.org/10.1038/s41746-025-01974-8
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.