A rapid smartphone-based lactate dehydrogenase test for neonatal diagnostics at the point of care

Open access
Halvorsen, Cecilia Pegelow, Olson, Linus, Araújo, Ana Catarina, Karlsson, Mathias, Nguyễn, Trang Thị, Khu, Dung T. K., Le, Ha T. T., Nguyễn, Hoa T. B. +2 more·2019·Scientific Reports

Abstract

Abstract There is a growing recognition of the importance of point-of-care tests (POCTs) for detecting critical neonatal illnesses to reduce the mortality rate in newborns, especially in low-income countries, which account for 98 percent of reported neonatal deaths. Lactate dehydrogenase (LDH) is a marker of cellular damage as a result of hypoxia-ischemia in affected organs. Here, we describe and test a POC LDH test direct from whole blood to provide early indication of serious illness in the neonate. The sample-in-result-out POC platform is specifically designed to meet the needs at resource-limited settings. Plasma is separated from whole blood on filter paper with dried-down reagents for colorimetric reaction, combined with software for analysis using a smartphone. The method was clinically tested in newborns in two different settings. In a clinical cohort of newborns of Stockholm (n = 62) and Hanoi (n = 26), the value of R using Pearson’s correlation test was 0.91 (p < 0.01) and the R2 = 0.83 between the two methods. The mean LDH (±SD) for the reference method vs. the POC-LDH was 551 (±280) U/L and 552 (±249) U/L respectively, indicating the clinical value of LDH values measured in minutes with the POC was comparable with standardized laboratory analyses.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1038/s41598-019-45606-0

Cites

21 of 21 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Equipment for point-of-care LDH analysis is now available.

    “Equipment for point-of-care LDH analysis is now available, 135”

    From Umbilical cord pH, blood gases, and lactate at birth: normal values, interpretation, and clinical utility. (2023) · cites this paper as [135]

    The Results (Pre-clinical phase) describe a working point-of-care LDH system. But the Discussion calls it an early, non-commercially available prototype intended as a possible future diagnostic tool. The paper therefore supports the existence of POC LDH equipment, not an unqualified implication that it is generally available. (LLM, conf=0.86)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Halvorsen, C. P., Olson, L., Araújo, A. C., Karlsson, M., Nguyễn, T. T., Khu, D. T. K., Le, H. T. T., Nguyễn, H. T. B., Winbladh, B., & Russom, A.
Affiliations
  • Karolinska Institutet (SE)
  • Uppsala University (SE)
  • Vietnam National Children's Hospital (VN)
  • Science for Life Laboratory (SE)
Corresponding author
Russom, Aman
Venue
Scientific Reports
Published
2019-06-26
Volume
9
Issue
1
Pages
9301-9301
ISSN
2045-2322 (ISSN-L)
OpenAlex
W2955079455
Cited by
19
References
21
Citations by year
2026: 3 · 2025: 3 · 2024: 2 · 2023: 1 · 2022: 6 · 2021: 1 · 2020: 2 · 2019: 1
FWCI
0.88
Type
journal article
Language
en
Open access
Locations
  • Scientific Reports (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • Uppsala University Publications (Uppsala University) (submittedVersion, other-oa)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
Funders
  • Swedish Foundation for International Cooperation in Research and Higher Education
  • Karolinska Institutet
  • Sällskapet Barnavård
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Biosensors and Analytical Detection
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
  • Health Sciences › Medicine › Pediatrics, Perinatology and Child Health › Neonatal Health and Biochemistry
Keywords
lactate dehydrogenase, point-of-care testing, smartphone-based diagnostics, dried blood spots, colorimetric assay, hypoxia-ischemia, neonatal mortality, resource-limited settings, neonates
MeSH terms
Point-of-Care Testing, Smartphone, Hypoxia, Calibration, Colorimetry, Fetal Blood, Humans, Infant, Newborn, Ischemia, L-Lactate Dehydrogenase, Prothrombin Time, Reference Standards, Reference Values, Software, Sweden, Reproducibility of Results, Neonatal Screening, Mobile Applications

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review