A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration
Open accessThe DOI matches its Crossref record; 42 of the 43 references it cites were found in a registry or an index.
Abstract
We report a new method for the real-time quantitative analysis of sodium in human sweat, consolidating sweat collection and analysis in a single, integrated, wearable platform. This temporal data opens up new possibilities in the study of human physiology, broadly applicable from assessing high performance athletes to monitoring Cystic Fibrosis (CF) sufferers. Our compact Sodium Sensor Belt (SSB) consists of a sodium selective Ion Selective Electrode (ISE) integrated into a platform that can be interfaced with the human body during exercise. No skin cleaning regime or sweat storage technology is required as the sweat is continually wicked from the skin to a sensing surface and from there to a storage area via a fabric pump. Our results suggest that after an initial equilibration period, a steady-state sodium plateau concentration was reached. Atomic Absorption Spectroscopy (AAS) was used as a reference method, and this has confirmed the accuracy of the new continuous monitoring approach. The steady-state concentrations observed were found to fall within ranges previously found in the literature, which further validates the approach. Daily calibration repeatability (n = 4) was ±3.0% RSD and over a three month period reproducibility was ±12.1% RSD (n = 56). As a further application, we attempted to monitor the sweat of Cystic Fibrosis (CF) sufferers using the same device. We observed high sodium concentrations symptomatic of CF (∼60 mM Na+) for two CF patients, with no conclusive results for the remaining patients due to their limited exercising capability, and high viscosity/low volume of sweat produced.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“Previously reported sweat-based and other non-invasive biosensors either can only monitor a single analyte at a time or lack on-site signal processing circuitry and sensor calibration mechanisms for accurate analysis of the physiological state14–18.”
From Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis (2016) · cites this paper as [14]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Exercise and Fluid Replacement
(2007) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e31802ca597 · Cited by 1,497
Gibson, L. E., & Cooke, R. E. (1959) · Pediatrics · doi:10.1542/peds.23.3.545 · Cited by 1,434
Variations in Regional Sweat Composition in Normal Human Males
Patterson, M. J., Galloway, S. D. R., & Nimmo, M. A. (2000) · Experimental Physiology · doi:10.1111/j.1469-445x.2000.02058.x · Cited by 438
Shirreffs, S. M., & Maughan, R. J. (1997) · Journal of Applied Physiology · doi:10.1152/jappl.1997.82.1.336 · Cited by 251
Calixarene-based sensing agents
Diamond, D., & McKervey, M. A. (1996) · Chemical Society Reviews · doi:10.1039/cs9962500015 · Cited by 247
Self-citationBio-sensing textile based patch with integrated optical detection system for sweat monitoring
Morris, D., Coyle, S., Wu, Y., Lau, K. T., Wallace, G., & Diamond, D. (2009) · Sensors and Actuators B: Chemical · doi:10.1016/j.snb.2009.02.032 · Cited by 235
Self-citationSolid-contact polymeric membrane electrodes with detection limits in the subnanomolar range
Sutter, J., Radu, A., Peper, S., Bakker, E., & Pretsch, E. (2004) · Analytica Chimica Acta · doi:10.1016/j.aca.2004.07.016 · Cited by 211
Solid Contact Potentiometric Sensors for Trace Level Measurements
Chumbimuni-Torres, K. Y., Rubinova, N., Radu, A., Kubota, L. T., & Bakker, E. (2006) · Analytical Chemistry · doi:10.1021/ac050749y · Cited by 206
Fluids and hydration in prolonged endurance performance
von Duvillard, S. P., Braun, W. A., Markofski, M., Beneke, R., & Leithäuser, R. (2004) · Nutrition · doi:10.1016/j.nut.2004.04.011 · Cited by 186
Peer Reviewed: Calixarenes: Designer Ligands for Chemical Sensors
Diamond, D., & Nolan, K. (2001) · Analytical Chemistry · doi:10.1021/ac012376g · Cited by 172
Self-citation
Research works printed but not linked: 2
- Northall, 1995, British Journal of Biomedical Science, 52, 68
- Legrys, 1994, Archives of Pathology & Laboratory Medicine, 118, 865
Cited by
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087
“have been used to measure analytes of interest (Supplementary Table 1) 14–18 .”
MethodBariya, M., Nyein, H. Y. Y., & Javey, A. (2018) · Nature Electronics · doi:10.1038/s41928-018-0043-y · Cited by 1,486
Wearable sensors: modalities, challenges, and prospects
Heikenfeld, J., Jajack, A., Rogers, J., Gutruf, P., Tian, L., Pan, T., Li, R., Khine, M., Kim, J., Wang, J., et al. (2018) · Lab on a Chip · doi:10.1039/c7lc00914c · Cited by 1,434
Bio-Integrated Wearable Systems: A Comprehensive Review
Ray, T. R., Choi, J., Bandodkar, A. J., Krishnan, S., Gutruf, P., Tian, L., Ghaffari, R., & Rogers, J. A. (2019) · Chemical Reviews · doi:10.1021/acs.chemrev.8b00573 · Cited by 1,350
Wearable and flexible electronics for continuous molecular monitoring
Yang, Y., & Gao, W. (2019) · Chemical Society Reviews · doi:10.1039/c7cs00730b · Cited by 1,315
Metadata
- Authors
- Schazmann, B., Morris, D., Slater, C., Beirne, S., Fay, C., Reuveny, R., Moyna, N., & Diamond, D.
- Affiliations
- Dublin Institute of Technology (IE)
- Dublin City University (IE)
- Corresponding author
- Schazmann, Benjamin
- Venue
- Analytical Methods
- Publisher
- Royal Society of Chemistry (RSC)
- Published
- 2010
- Volume
- 2
- Issue
- 4
- Pages
- 342
- ISSN
- 1759-9660 (ISSN-L)
- 1759-9679
- OpenAlex
- W2144506268
- Cited by
- 286
- References
- 43
- Citations by year
- 2026: 13 · 2025: 20 · 2024: 23 · 2023: 28 · 2022: 18 · 2021: 28 · 2020: 27 · 2019: 29 · 2018: 20 · 2017: 25 · 2016: 18 · 2015: 13 · 2014: 12 · 2013: 7 · 2012: 3
- FWCI
- 2.85
- Type
- journal article
- Language
- en
- Open access
- green
- orcid.org
- Locations
- Analytical Methods (publishedVersion)
- Dublin City University Open Access Institutional Repository (Dublin City University) (submittedVersion)
- Research Online (University of Wollongong) (submittedVersion)
- Arrow@dit (Dublin Institute of Technology) (submittedVersion)
- Arrow@dit (Dublin Institute of Technology) (submittedVersion, cc-by-nc-sa)
- Arrow - TU Dublin (Technological University Dublin) (submittedVersion)
- Topics
- Physical Sciences › Chemistry › Analytical Chemistry › Analytical Chemistry and Sensors
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Keywords
- sweat sodium concentration, wearable electrochemical sensors, ion-selective electrodes, real-time physiological monitoring, cystic fibrosis, exercise physiology, sweat collection
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration. https://app.cochat.ai/papers/10.1039/b9ay00184k
- The paper
- Schazmann, B., Morris, D., Slater, C., Beirne, S., Fay, C., Reuveny, R., Moyna, N., & Diamond, D. (2010). A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration. Analytical Methods. https://doi.org/10.1039/b9ay00184k
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.