Digital PCR on a SlipChip
Open accessAbstract
This paper describes a SlipChip to perform digital PCR in a very simple and inexpensive format. The fluidic path for introducing the sample combined with the PCR mixture was formed using elongated wells in the two plates of the SlipChip designed to overlap during sample loading. This fluidic path was broken up by simple slipping of the two plates that removed the overlap among wells and brought each well in contact with a reservoir preloaded with oil to generate 1280 reaction compartments (2.6 nL each) simultaneously. After thermal cycling, end-point fluorescence intensity was used to detect the presence of nucleic acid. Digital PCR on the SlipChip was tested quantitatively by using Staphylococcus aureus genomic DNA. As the concentration of the template DNA in the reaction mixture was diluted, the fraction of positive wells decreased as expected from the statistical analysis. No cross-contamination was observed during the experiments. At the extremes of the dynamic range of digital PCR the standard confidence interval determined using a normal approximation of the binomial distribution is not satisfactory. Therefore, statistical analysis based on the score method was used to establish these confidence intervals. The SlipChip provides a simple strategy to count nucleic acids by using PCR. It may find applications in research applications such as single cell analysis, prenatal diagnostics, and point-of-care diagnostics. SlipChip would become valuable for diagnostics, including applications in resource-limited areas after integration with isothermal nucleic acid amplification technologies and visual readout.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
SlipChip is especially appropriate for integrated DNA-based data storage because all microdroplets can be moved simultaneously by sliding, so no valves, pumps, or tubes are required.
“As a microfluidic device for handling a large array of microdroplets, SlipChip is a platform that is especially appropriate for the development of an integrated DNA-based data storage system. All the microdroplets can be moved simultaneously on a SlipChip by simply sliding the device, so no valves, pumps, or tubes are required (54–56).”
From Electrochemical DNA synthesis and sequencing on a single electrode with scalability for integrated data storage (2021) · cites this paper as [56]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Loop-mediated isothermal amplification of DNA
Notomi, T. (2000) · Nucleic Acids Research · doi:10.1093/nar/28.12.e63 · Cited by 8,805 · see citing works
THE USE OF CONFIDENCE OR FIDUCIAL LIMITS ILLUSTRATED IN THE CASE OF THE BINOMIAL
Clopper, C. J., & Pearson, E. S. (1934) · Biometrika · doi:10.1093/biomet/26.4.404 · Cited by 5,106 · see citing works
Probable Inference, the Law of Succession, and Statistical Inference
Wilson, E. B. (1927) · Journal of the American Statistical Association · doi:10.1080/01621459.1927.10502953 · Cited by 4,476 · see citing works
Interval Estimation for a Binomial Proportion
Brown, L. D., Cai, T., & Dasgupta, A. (2001) · Statistical Science · doi:10.1214/ss/1009213286 · Cited by 3,622 · see citing works
Approximate is Better than “Exact” for Interval Estimation of Binomial Proportions
Agresti, A., & Coull, B. A. (1998) · The American Statistician · doi:10.1080/00031305.1998.10480550 · Cited by 3,223 · see citing works
DNA Detection Using Recombination Proteins
Piepenburg, O., Williams, C. H., Stemple, D. L., & Armes, N. (2006) · PLoS Biology · doi:10.1371/journal.pbio.0040204 · Cited by 2,503 · see citing works
Reactions in Droplets in Microfluidic Channels
Song, H., Chen, D. L., & Ismagilov, R. F. (2006) · Angewandte Chemie International Edition · doi:10.1002/anie.200601554 · Cited by 1,844 · see citing works
Self-citationVogelstein, B., & Kinzler, K. W. (1999) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.96.16.9236 · Cited by 1,770 · see citing works
Mutation detection and single-molecule counting using isothermal rolling-circle amplification
Lizardi, P. M., Huang, X. C., Zhu, Z., Bray‐Ward, P., Thomas, D. C., & Ward, D. C. (1998) · Nature Genetics · doi:10.1038/898 · Cited by 1,443 · see citing works
Nucleic acid sequence-based amplification
Compton, J. (1991) · Nature · doi:10.1038/350091a0 · Cited by 1,351 · see citing works
Detection of Staphylococcus aureus by polymerase chain reaction amplification of the nuc gene
Brakstad, O. G., Aasbakk, K., & Maeland, J. A. (1992) · Journal of Clinical Microbiology · doi:10.1128/jcm.30.7.1654-1660.1992 · Cited by 1,147 · see citing works
A Microfluidic System for Controlling Reaction Networks in Time
Song, H., Tice, J. D., & Ismagilov, R. F. (2003) · Angewandte Chemie International Edition · doi:10.1002/anie.200390203 · Cited by 1,083 · see citing works
Self-citationHelicase‐dependent isothermal DNA amplification
Vincent, M., Xu, Y., & Kong, H. (2004) · EMBO Reports · doi:10.1038/sj.embor.7400200 · Cited by 877 · see citing works
Strand displacement amplification—an isothermal, in vitro DNA amplification technique
Walker, G. T., Fraiser, M. S., Schram, J. L., Little, M. C., Nadeau, J. G., & Malinowski, D. P. (1992) · Nucleic Acids Research · doi:10.1093/nar/20.7.1691 · Cited by 857 · see citing works
Dressman, D. C., Yan, H., Traverso, G., Kinzler, K. W., & Vogelstein, B. (2003) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1133470100 · Cited by 845 · see citing works
Microfluidic Digital PCR Enables Multigene Analysis of Individual Environmental Bacteria
Ottesen, E. A., Hong, J. W., Quake, S. R., & Leadbetter, J. R. (2006) · Science · doi:10.1126/science.1131370 · Cited by 733 · see citing works
Requirements for high impact diagnostics in the developing world
Urdea, M. S., Penny, L., Olmsted, S. S., Giovanni, M. Y., Kaspar, P., Shepherd, A., Wilson, P. L., Dahl, C. A., Buchsbaum, S., Moeller, G., et al. (2006) · Nature · doi:10.1038/nature05448 · Cited by 589 · see citing works
Chandran, U. R., Ma, C., Dhir, R., Bisceglia, M., Lyons‐Weiler, M. A., Liang, W., Michalopoulos, G. K., Becich, M. J., & Monzon, F. A. (2007) · BMC Cancer · doi:10.1186/1471-2407-7-64 · Cited by 489 · see citing works
Microdroplet-based PCR enrichment for large-scale targeted sequencing
Tewhey, R., Warner, J. B., Nakano, M., Libby, B., Medkova, M., David, P. H., Kotsopoulos, S. K., Samuels, M. L., Hutchison, J. B., Larson, J. W., et al. (2009) · Nature Biotechnology · doi:10.1038/nbt.1583 · Cited by 454 · see citing works
Microfluidics Digital PCR Reveals a Higher than Expected Fraction of Fetal DNA in Maternal Plasma
Lun, F. M. F., Chiu, R. W. K., Chan, K. C. A., Leung, T. Y., Lau, T. K., & Lo, Y. M. D. (2008) · Clinical Chemistry · doi:10.1373/clinchem.2008.111385 · Cited by 435 · see citing works
Digital PCR for the molecular detection of fetal chromosomal aneuploidy
Lo, Y. M. D., Lun, F. M. F., Chan, K. C. A., Tsui, N. B. Y., Chong, K. C., Lau, T. K., Leung, T. Y., Zee, B. C.-Y., Cantor, C. R., & Chiu, R. W. K. (2007) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.0705765104 · Cited by 434 · see citing works
Blyth, C. R., & Still, H. A. (1983) · Journal of the American Statistical Association · doi:10.1080/01621459.1983.10477938 · Cited by 406 · see citing works
Yung, T., Chan, K. C. A., Mok, T. S. K., TONG, J. H. M., To, K., & Lo, Y. M. D. (2009) · Clinical Cancer Research · doi:10.1158/1078-0432.ccr-08-2622 · Cited by 381 · see citing works
Transcription factor profiling in individual hematopoietic progenitors by digital RT-PCR
Warren, L. A., Bryder, D., Weissman, I., & Quake, S. (2006) · Proceedings of the National Academy of Sciences of the United States of America · doi:10.1073/pnas.0608512103 · Cited by 373 · see citing works
On-Chip, Real-Time, Single-Copy Polymerase Chain Reaction in Picoliter Droplets
Beer, N. R., Hindson, B. J., Wheeler, E. K., Hall, S. B., Rose, K. A., Kennedy, I. M., & Colston, B. W. (2007) · Analytical Chemistry · doi:10.1021/ac701809w · Cited by 358 · see citing works
Microfluidic Single-Cell mRNA Isolation and Analysis
Marcus, J. S., Anderson, W. F., & Quake, S. R. (2006) · Analytical Chemistry · doi:10.1021/ac0519460 · Cited by 351 · see citing works
Du, W., Li, L., Nichols, K. P., & Ismagilov, R. F. (2009) · Lab on a Chip · doi:10.1039/b908978k · Cited by 351 · see citing works
Self-citationSensitive digital quantification of DNA methylation in clinical samples
Li, M., Chen, W.-D., Papadopoulos, N., Goodman, S. N., Bjerregaard, N. C., Laurberg, S., Levin, B., Juhl, H., Arber, N., Moinova, H., et al. (2009) · Nature Biotechnology · doi:10.1038/nbt.1559 · Cited by 341 · see citing works
Application of Statistics to Problems in Bacteriology
Halvorson, H. O., & Ziegler, N. R. (1933) · Journal of Bacteriology · doi:10.1128/jb.25.2.101-121.1933 · Cited by 309 · see citing works
Dube, S., Qin, J., & Ramakrishnan, R. (2008) · PLoS ONE · doi:10.1371/journal.pone.0002876 · Cited by 302 · see citing works
Real-Time PCR Technology for Cancer Diagnostics
Bernard, P. S., & Wittwer, C. T. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.8.1178 · Cited by 273 · see citing works
Mažutis, L., Fallah-Araghi, A., Miller, O. J., Baret, J., Frenz, L., Janoshazi, A. K., Taly, V., Miller, B. J., Hutchison, J. B., Link, D. R., et al. (2009) · Analytical Chemistry · doi:10.1021/ac900403z · Cited by 264 · see citing works
Confidence intervals for the binomial parameter: some new considerations
Reiczigel, J. (2003) · Statistics in Medicine · doi:10.1002/sim.1320 · Cited by 259 · see citing works
A microfluidic processor for gene expression profiling of single human embryonic stem cells
Zhong, J. F., Chen, Y., Marcus, J. S., Scherer, A., Quake, S. R., Taylor, C. R., & Weiner, L. P. (2007) · Lab on a Chip · doi:10.1039/b712116d · Cited by 231 · see citing works
High-Throughput Single Copy DNA Amplification and Cell Analysis in Engineered Nanoliter Droplets
Kumaresan, P., Yang, C. J., Cronier, S. A., Blazej, R. G., & Mathies, R. A. (2008) · Analytical Chemistry · doi:10.1021/ac800327d · Cited by 226 · see citing works
Bhat, S., Herrmann, J., Armishaw, P., Corbisier, P., & Emslie, K. R. (2009) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-009-2729-5 · Cited by 218 · see citing works
On-Chip Single-Copy Real-Time Reverse-Transcription PCR in Isolated Picoliter Droplets
Beer, N. R., Wheeler, E. K., Lee-Houghton, L., Watkins, N. N., Nasarabadi, S., Hebert, N. E., Leung, P., Arnold, D. W., Bailey, C. G., & Colston, B. W. (2008) · Analytical Chemistry · doi:10.1021/ac800048k · Cited by 209 · see citing works
Protein Crystallization Using Microfluidic Technologies Based on Valves, Droplets, and SlipChip
Li, L., & Ismagilov, R. F. (2010) · Annual Review of Biophysics · doi:10.1146/annurev.biophys.050708.133630 · Cited by 209 · see citing works
Self-citationDigital PCR provides sensitive and absolute calibration for high throughput sequencing
White, R. A., III, Blainey, P. C., Fan, H. C., & Quake, S. R. (2009) · BMC Genomics · doi:10.1186/1471-2164-10-116 · Cited by 196 · see citing works
Detection of Aneuploidy with Digital Polymerase Chain Reaction
Fan, H. C., & Quake, S. R. (2007) · Analytical Chemistry · doi:10.1021/ac0709394 · Cited by 179 · see citing works
Some remarks on confidence or fiducial limits
Sterne, T. E. (1954) · Biometrika · doi:10.1093/biomet/41.1-2.275 · Cited by 160 · see citing works
Aryan, E., Makvandi, M., Sheikh, A. F., Huygen, K., Bifani, P. J., Gargari, S. L. M., Fateh, A., Jelodar, A., Gouya, M. M., & Romano, M. (2009) · Microbiological Research · doi:10.1016/j.micres.2009.05.001 · Cited by 153 · see citing works
Integrated microfluidic tmRNA purification and real-time NASBA device for molecular diagnostics
Dimov, I. K., Garcia-Cordero, J. L., O’Grady, J., Poulsen, C., Viguier, C., Kent, L., Daly, P., Lincoln, B., Maher, M., O’Kennedy, R. J., et al. (2008) · Lab on a Chip · doi:10.1039/b812515e · Cited by 144 · see citing works
Quantitative analysis of gene expression in a single cell by qPCR
Taniguchi, K., Kajiyama, T., & Kambara, H. (2009) · Nature Methods · doi:10.1038/nmeth.1338 · Cited by 136 · see citing works
Isothermal DNA amplification in vitro: the helicase-dependent amplification system
Jeong, Y., Park, K., & Kim, D. (2009) · Cellular and Molecular Life Sciences · doi:10.1007/s00018-009-0094-3 · Cited by 132 · see citing works
Spinning Disk Platform for Microfluidic Digital Polymerase Chain Reaction
Sundberg, S. O., Wittwer, C. T., Gao, C., & Gale, B. K. (2010) · Analytical Chemistry · doi:10.1021/ac902398c · Cited by 130 · see citing works
Nanoliter Multiplex PCR Arrays on a SlipChip
Shen, F., Du, W., Davydova, E. K., Karymov, M. A., Pandey, J., & Ismagilov, R. F. (2010) · Analytical Chemistry · doi:10.1021/ac1007249 · Cited by 115 · see citing works
Self-citationMicrofluidic digital PCR enables rapid prenatal diagnosis of fetal aneuploidy
Fan, H. C., Blumenfeld, Y. J., El-Sayed, Y. Y., Chueh, J. T., & Quake, S. R. (2009) · American Journal of Obstetrics and Gynecology · doi:10.1016/j.ajog.2009.03.002 · Cited by 108 · see citing works
SlipChip for Immunoassays in Nanoliter Volumes
Liu, W., Chen, D. L., Du, W., Nichols, K. P., & Ismagilov, R. F. (2010) · Analytical Chemistry · doi:10.1021/ac100044c · Cited by 102 · see citing works
Self-citationVincent, M. E., Liu, W., Haney, E. B., & Ismagilov, R. F. (2010) · Chemical Society Reviews · doi:10.1039/b917851a · Cited by 102 · see citing works
Self-citationLi, L., Du, W., & Ismagilov, R. F. (2009) · Journal of the American Chemical Society · doi:10.1021/ja908558m · Cited by 95 · see citing works
Self-citationD-lactic acidemia and aciduria in pediatric and adult patients with short bowel syndrome
Bongaerts, G. P. A., Tolboom, J. J. M., Naber, A. H. J., Bakkeren, J. A. J. M., Severijnen, R. S. V. M., & Willems, H. L. (1995) · Clinical Chemistry · doi:10.1093/clinchem/41.1.107 · Cited by 89 · see citing works
Nasba Technology: Isothermal RNA Amplification in Qualitative and Quantitative Diagnostics
Romano, J., Williams, K. G., Shurtliff, R. N., et al. (1997) · Immunological Investigations · doi:10.3109/08820139709048912 · Cited by 84 · see citing works
Isothermal Amplification and Multimerization of DNA by Bst DNA Polymerase
Hafner, G. J., Yang, I., Wolter, L. C., et al. (2001) · BioTechniques · doi:10.2144/01304rr03 · Cited by 83 · see citing works
User-Loaded SlipChip for Equipment-Free Multiplexed Nanoliter-Scale Experiments
Li, L., Du, W., & Ismagilov, R. F. (2009) · Journal of the American Chemical Society · doi:10.1021/ja908555n · Cited by 73 · see citing works
Self-citationStedtfeld, R. D., Baushke, S. W., Tourlousse, D. M., et al. (2008) · Applied and Environmental Microbiology · doi:10.1128/aem.02743-07 · Cited by 68 · see citing works
Strand displacement amplification: a versatile tool for molecular diagnostics
Hellyer, T. J., & Nadeau, J. G. (2004) · Expert Review of Molecular Diagnostics · doi:10.1586/14737159.4.2.251 · Cited by 59 · see citing works
Li, L., Karymov, M. A., Nichols, K. P., & Ismagilov, R. F. (2010) · Langmuir · doi:10.1021/la101460z · Cited by 29 · see citing works
Self-citationDetecting and describing heterogeneity in meta-analysis
Hardy, R. J., & Thompson, S. G. (1998) · Statistics in Medicine · doi:10.1002/(sici)1097-0258(19980430)17:8
Flagged first, then by citations
Research works printed but not linked: 2
- [36]Clopper CJ Pearson ES Biometrika 1934 26 404 413
- [55]White RA Blainey PC Fan HC Quake SR BMC Genomics 2009 10 116 19298667 PMC2667538
Cited by
Sorted by citations
Genome structural variation discovery and genotyping
Alkan, C., Coe, B. P., & Eichler, E. E. (2011) · Nature Reviews Genetics · doi:10.1038/nrg2958 · Cited by 1,630 · see citing works
Shen, F., Davydova, E. K., Du, W., Kreutz, J. E., Piepenburg, O., & Ismagilov, R. F. (2011) · Analytical Chemistry · doi:10.1021/ac200247e · Cited by 240 · see citing works
Self-citationShen, F., Sun, B., Kreutz, J. E., Davydova, E. K., Du, W., Reddy, P. L., Joseph, L. J., & Ismagilov, R. F. (2011) · Journal of the American Chemical Society · doi:10.1021/ja2060116 · Cited by 231 · see citing works
Self-citationKreutz, J. E., Munson, T., Huynh, T., Shen, F., Du, W., & Ismagilov, R. F. (2011) · Analytical Chemistry · doi:10.1021/ac201658s · Cited by 148 · see citing works
Self-citationXu, C., Ma, B., Gao, Z., Dong, X., Zhao, C., & Liu, H. (2021) · Science Advances · doi:10.1126/sciadv.abk0100 · Cited by 79 · see citing works
- Authors
- Shen, F., Du, W., Kreutz, J. E., Fok, A., & Ismagilov, R. F.
- Affiliations
- University of Chicago (US)
- Corresponding author
- Ismagilov, Rustem F.
- Venue
- Lab on a Chip
- Publisher
- Royal Society of Chemistry (RSC)
- Published
- 2010
- Volume
- 10
- Issue
- 20
- Pages
- 2666
- ISSN
- 1473-0189 (ISSN-L)
- 1473-0197
- PMC
- PMC2948063
- OpenAlex
- W2150204651
- Cited by
- 295
- References
- 59
- Citations by year
- 2026: 14 · 2025: 14 · 2024: 16 · 2023: 14 · 2022: 19 · 2021: 16 · 2020: 24 · 2019: 23 · 2018: 15 · 2017: 17 · 2016: 21 · 2015: 18 · 2014: 20 · 2013: 21 · 2012: 19
- FWCI
- 12.11
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Lab on a Chip (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- CaltechAUTHORS (California Institute of Technology) (submittedVersion)
- Funders
- National Science Foundation
- National Institutes of Health (DP1OD003584)
- Office of Naval Research (N00014)
- National Institute of General Medical Sciences
- Division of Materials Research
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Microfluidic and Capillary Electrophoresis Applications
- Physical Sciences › Engineering › Biomedical Engineering › Biosensors and Analytical Detection
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Electrowetting and Microfluidic Technologies
- Keywords
- digital PCR, SlipChip, microfluidics, droplet formation, fluorescence detection, DNA quantification, scoring methods, confidence intervals, low-cost diagnostics, point-of-care testing, resource-limited settings
- MeSH terms
- Chromosome Mapping, DNA, Bacterial, Equipment Design, Microscopy, Fluorescence, Signal Processing, Computer-Assisted, Staphylococcus aureus, Polymerase Chain Reaction, Equipment Failure Analysis, Microfluidic Analytical Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review