Adipose tissue thickness affects in vivo quantitative near-IR spectroscopy in human skeletal muscle
The DOI matches its Crossref record; 20 of the 20 references it cites were found in a registry or an index.
Abstract
The influence of adipose tissue thickness (ATT) on near-IR spectroscopy (NIRS) measurements in vivo was studied in the human flexor digitorum superficialis muscle at rest and during sustained isometric handgrip exercise. NIRS was used for the quantitative measurement of muscle O(2) consumption (mV.O(2)) and forearm blood flow (FBF) in 78 healthy subjects. Skinfold thickness ranged from 1.4 to 8.9 mm within the group. Resting mV.O(2) was 0.11+/-0.04 ml of O(2).min(-1).100 g(-1), and FBF was 1.28+/-0.82 ml.min(-1).100 ml(-1). There was a negative correlation (r=-0.70, P< or =0.01), indicating a decrease in mV.O(2) with increasing ATT. mV.O(2) in the 10 leanest subjects appeared to be twice as high as that in the 10 subjects with the highest ATT. A poor correlation (r=0.29, P< or =0.01) was found between ATT and FBF. The gender difference that we found for mV.O(2) was due to the difference in ATT between female and male subjects. No correlation was found between maximum voluntary contraction and mV.O(2), nor between maximum voluntary contraction and ATT, indicating that the contraction force did not confound our results. These results show that ATT has a substantial confounding influence on in vivo NIRS measurements, and that it is essential to incorporate this factor into future NIRS muscle studies in order to justify comparisons between different groups. To facilitate such comparisons, upper and lower boundaries for normal values of mV.O(2) and FBF in relation to ATT are prese…
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Owing to the confounding effects of ATT,36 all measurement sites with ATT greater than 60% of the emitter-detector distance (15.0 mm) were removed from the analysis.”
From Near-infrared spectroscopy-derived muscle oxygen saturation on a 0% to 100% scale: reliability and validity of the Moxy Monitor (2019) · cites this paper as [36]
In the citing paper
“Also, it is well-known that scattering and absorbance due to adipose tissue affect NIRS signals, reducing absorbance by underlying muscle tissue [24–26].”
From Oxygenation Threshold Derived from Near-Infrared Spectroscopy: Reliability and Its Relationship with the First Ventilatory Threshold (2016) · cites this paper as [25]
In the citing paper
“Adipose tissue thickness was calculated by dividing skinfold thickness by two [25], resembling subcutaneous fat and skin.”
From Oxygenation Threshold Derived from Near-Infrared Spectroscopy: Reliability and Its Relationship with the First Ventilatory Threshold (2016) · cites this paper as [25]
In the citing paper
“ATT is known to affect the amplitude of NIRS measurements [7,24–26,42].”
From Oxygenation Threshold Derived from Near-Infrared Spectroscopy: Reliability and Its Relationship with the First Ventilatory Threshold (2016) · cites this paper as [25]
Citations
Cites
Recovery from exercise-induced desaturation in the quadriceps muscles of elite competitive rowers
Chance, B., Dait, M. T., Zhang, C., Hamaoka, T., & Hagerman, F. (1992) · American Journal of Physiology-Cell Physiology · doi:10.1152/ajpcell.1992.262.3.c766 · Cited by 528
Duncan, A., Meek, J. H., Clemence, M., Elwell, C. E., Fallon, P., Tyszczuk, L., Cope, M., & Delpy, D. T. (1996) · Pediatric Research · doi:10.1203/00006450-199605000-00025 · Cited by 469
van der Zee, P., Cope, M., Arridge, S. R., Essenpreis, M., Potter, L. A., Edwards, A. D., Wyatt, J. S., McCormick, D. C., Roth, S. C., Reynolds, E. O. R., et al. (1992) · Advances in Experimental Medicine and Biology · doi:10.1007/978-1-4615-3404-4_17 · Cited by 427
Near infrared monitoring of human skeletal muscle oxygenation during forearm ischemia
Hampson, N. B., & Piantadosi, C. A. (1988) · Journal of Applied Physiology · doi:10.1152/jappl.1988.64.6.2449 · Cited by 300
Noninvasive measurement of human forearm oxygen consumption by near infrared spectroscopy
De Blasi, R. A., Cope, M., Elwell, C., Safoue, F., & Ferrari, M. (1993) · European Journal of Applied Physiology and Occupational Physiology · doi:10.1007/bf00377698 · Cited by 238
Spectral dependence of temporal point spread functions in human tissues
Essenpreis, M., Elwell, C. E., Cope, M., van der Zee, P., Arridge, S. R., & Delpy, D. T. (1993) · Applied Optics · doi:10.1364/ao.32.000418 · Cited by 195
Comparative analysis of NMR and NIRS measurements of intracellular P O 2 in human skeletal muscle
Tran, T.-K., Sailasuta, N., Kreutzer, U., Hurd, R., Chung, Y., Mole, P., Kuno, S., & Jue, T. (1999) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.1999.276.6.r1682 · Cited by 175
Postprandial Substrate Deposition in Human Forearm and Adipose Tissues in Vivo
Coppack, S. W., Fisher, R. M., Gibbons, G. F., Humphreys, S. M., McDonough, M. J., Potts, J. L., & Frayn, K. N. (1990) · Clinical Science · doi:10.1042/cs0790339 · Cited by 172
Myoglobin desaturation with exercise intensity in human gastrocnemius muscle
Molé, P. A., Chung, Y., Tran, T. K., Sailasuta, N., Hurd, R., & Jue, T. (1999) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.1999.277.1.r173 · Cited by 164
Noninvasive Quantitative Analysis of Blood Oxygenation in Rat Skeletal Muscle
Seiyama, A., Hazeki, O., & Tamura, M. (1988) · The Journal of Biochemistry · doi:10.1093/oxfordjournals.jbchem.a122285 · Cited by 149
Cited by
Items 401–424 of 424
Buzza, G., Lovell, G. P., Askew, C. D., & Solomon, C. (2020) · Charles Sturt University Research Output (CRO)
Townsend, D. K. (2007) · K-State Research Exchange (Kansas State University)
Likvern, E. K. (2018) · NVA (Nasjonalt vitenarkiv)
Correa, B., & Ricardo, F. (2016) · LA Referencia (Red Federada de Repositorios Institucionales de Publicaciones Científicas)
Deegan, E. G. (2019) · cIRcle (University of British Columbia) · doi:10.14288/1.0377236
Ward, A. T. (2016) · OhioLink ETD Center (Ohio Library and Information Network)
Kerhervé, H. A. (2015) · UniSC Research Bank (University of the Sunshine Coast)
Measurement of oxidative metabolism of the working human muscles by near-infrared spectroscopy
Yücetaş, A., Şayli, Ö., Karahan, M., & Akın, A. (2006) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.644594
Versteeg, N., Wellauer, V., Wittenwiler, S., Aerenhouts, D., Clarys, P., & Clijsen, R. (2024) · doi:10.3389/fphys.2024.1347196.s001
Application of TR-NIRS in Healthy and Diseased Children during Exercise
Chiu, A. S. (2017) · eScholarship (California Digital Library)
Signal Processing for Monitoring Cerebral Hemodynamics in Neonates
Dorado, A. C. (2013) · Lirias
Determination of Lipid Layer Thickness with Broadband Diffuse Optical Spectroscopy
Kwong, R. Y., Cerussi, A. E., Merritt, S. I., Ruth, J. R., & Tromberg, B. J. (2006) · Biomedical optics · doi:10.1364/bio.2006.me42
Furset, T. (2015) · NVA (Nasjonalt vitenarkiv)
Dommann, K., Marx, G., & Rossaint, R. (2015) · RWTH Publications (RWTH Aachen)
Dech, S. (2022) · publish.UP (University of Potsdam) · doi:10.25932/publishup-57126
Košir, M. (2024) · Repository of the University of Ljubljana (University of Ljubljana)
Editorial: Diffusive optics for medical imaging
Urban, B. E., & Morgado, M. (2026) · Frontiers in Photonics · doi:10.3389/fphot.2026.1842429
Repetitive strain injury : a novel focus on an ancient problem
Brunnekreef, J. J. (2012) · Radboud Repository (Radboud University)
Metadata
- Authors
- VAN BEEKVELT, M. C. P., BORGHUIS, M. S., VAN ENGELEN, B. G. M., WEVERS, R. A., & COLIER, W. N. J. M.
- Affiliations
- Neuromuscular Centre Nijmegen, Institute of Neurology, University Medical Centre Nijmegen, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands
- Radboud University Medical Center (NL)
- Venue
- Clinical Science
- Publisher
- Portland Press Ltd.
- Published
- 2001-05-29
- Volume
- 101
- Issue
- 1
- Pages
- 21-28
- ISSN
- 0143-5221 (ISSN-L)
- 1470-8736
- OpenAlex
- W2005647586
- Cited by
- 369
- References
- 20
- Citations by year
- 2026: 12 · 2025: 9 · 2024: 17 · 2023: 11 · 2022: 16 · 2021: 18 · 2020: 23 · 2019: 15 · 2018: 19 · 2017: 21 · 2016: 16 · 2015: 14 · 2014: 14 · 2013: 19 · 2012: 23
- FWCI
- 4.43
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Clinical Science (publishedVersion)
- PubMed (publishedVersion)
- Radboud Repository (Radboud University) (submittedVersion)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Optical Imaging and Spectroscopy Techniques
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
- Keywords
- adipose tissue thickness, near-infrared spectroscopy, muscle oxygen consumption, forearm blood flow, flexor digitorum superficialis, isometric handgrip exercise, skinfold thickness, sex differences, maximal voluntary contraction, muscle oxygenation
- MeSH terms
- Adipose Tissue, Adult, Female, Humans, Male, Oxygen Consumption, Regional Blood Flow, Sex Factors, Skinfold Thickness, Reproducibility of Results, Exercise, Normal Distribution, Muscle, Skeletal, Spectroscopy, Near-Infrared
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.