Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices
Open accessAbstract
BACKGROUND: Patients who have pacemakers or defibrillators are often denied the opportunity to undergo magnetic resonance imaging (MRI) because of safety concerns, unless the devices meet certain criteria specified by the Food and Drug Administration (termed "MRI-conditional" devices). METHODS: We performed a prospective, nonrandomized study to assess the safety of MRI at a magnetic field strength of 1.5 Tesla in 1509 patients who had a pacemaker (58%) or an implantable cardioverter-defibrillator (42%) that was not considered to be MRI-conditional (termed a "legacy" device). Overall, the patients underwent 2103 thoracic and nonthoracic MRI examinations that were deemed to be clinically necessary. The pacing mode was changed to asynchronous mode for pacing-dependent patients and to demand mode for other patients. Tachyarrhythmia functions were disabled. Outcome assessments included adverse events and changes in the variables that indicate lead and generator function and interaction with surrounding tissue (device parameters). RESULTS: No long-term clinically significant adverse events were reported. In nine MRI examinations (0.4%; 95% confidence interval, 0.2 to 0.7), the patient's device reset to a backup mode. The reset was transient in eight of the nine examinations. In one case, a pacemaker with less than 1 month left of battery life reset to ventricular inhibited pacing and could not be reprogrammed; the device was subsequently replaced. The most common notable change in device parameters (>50% change from baseline) immediately after MRI was a decrease in P-wave amplitude, which occurred in 1% of the patients. At long-term follow-up (results of which were available for 63% of the patients), the most common notable changes from baseline were decreases in P-wave amplitude (in 4% of the patients), increases in atrial capture threshold (4%), increases in right ventricular capture threshold (4%), and increases in left ventricular capture threshold (3%). The observed changes in lead parameters were not clinically significant and did not require device revision or reprogramming. CONCLUSIONS: We evaluated the safety of MRI, performed with the use of a prespecified safety protocol, in 1509 patients who had a legacy pacemaker or a legacy implantable cardioverter-defibrillator system. No long-term clinically significant adverse events were reported. (Funded by Johns Hopkins University and the National Institutes of Health; ClinicalTrials.gov number, NCT01130896 .).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla
Martin, E. T., Coman, J. A., Shellock, F. G., Pulling, C. C., Fair, R., & Jenkins, K. (2004) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2003.12.016 · Cited by 422
Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator
Russo, R. J., Costa, H. S., Silva, P. D., Anderson, J. L., Arshad, A., Biederman, R. W. W., Boyle, N. G., Frabizzio, J. V., Birgersdotter-Green, U., Higgins, S. L., et al. (2017) · New England Journal of Medicine · doi:10.1056/nejmoa1603265 · Cited by 373
Nazarian, S., Roguin, A., Zviman, M. M., Lardo, A. C., Dickfeld, T. L., Calkins, H., Weiss, R. G., Berger, R. D., Bluemke, D. A., & Halperin, H. R. (2006) · Circulation · doi:10.1161/circulationaha.105.607655 · Cited by 356
Self-citationRoguin, A., Zviman, M. M., Meininger, G. R., Rodrigues, E. R., Dickfeld, T. M., Bluemke, D. A., Lardo, A., Berger, R. D., Calkins, H., & Halperin, H. R. (2004) · Circulation · doi:10.1161/01.cir.0000137121.28722.33 · Cited by 355
Self-citationNazarian, S., Hansford, R., Roguin, A., Goldsher, D., Zviman, M. M., Lardo, A. C., Caffo, B. S., Frick, K. D., Kraut, M. A., Kamel, I. R., et al. (2011) · Annals of Internal Medicine · doi:10.7326/0003-4819-155-7-201110040-00004 · Cited by 302
Self-citationWilkoff, B. L., Bello, D., Taborsky, M., Vymazal, J., Kanal, E., Heuer, H., Hecking, K., Johnson, W. B., Young, W., Ramza, B., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2010.10.002 · Cited by 266
Sommer, T., Naehle, C. P., Yang, A., Zeijlemaker, V., Hackenbroch, M., Schmiedel, A., Meyer, C., Strach, K., Skowasch, D., Vahlhaus, C., et al. (2006) · Circulation · doi:10.1161/circulationaha.105.597013 · Cited by 256
MR Imaging and Cardiac Pacemakers: In Vitro Evaluation and in Vivo Studies in 51 Patients at 0.5 T
Sommer, T., Vahlhaus, C., Lauck, G., Smekal, A. V., Reinke, M., Hofer, U., Block, W., Träber, F., Schneider, C., Gieseke, J., et al. (2000) · Radiology · doi:10.1148/radiology.215.3.r00jn08869 · Cited by 244
Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heating
Baker, K. B., Tkach, J. A., Nyenhuis, J. A., Phillips, M., Shellock, F. G., Gonzalez‐Martinez, J., & Rezai, A. R. (2004) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.20103 · Cited by 234
Safe Performance of Magnetic Resonance Imaging on Five Patients with Permanent Cardiac Pacemakers
Gimbel, J., JOHNSON, D., LEVINE, P. A., & Wilkoff, B. L. (1996) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1996.tb03387.x · Cited by 145
Randomized trial of pacemaker and lead system for safe scanning at 1.5 Tesla
Gimbel, J., Bello, D. B., Schmitt, M., Merkely, B. P., Schwitter, J., Hayes, D. L., Sommer, T., Schloss, E. J., Chang, Y., Willey, S., et al. (2013) · Heart Rhythm · doi:10.1016/j.hrthm.2013.01.022 · Cited by 134
Faris, O., & Shein, M. J. (2006) · Circulation · doi:10.1161/circulationaha.106.647800 · Cited by 131
Magnetic Resonance Imaging at 1.5-T in Patients With Implantable Cardioverter-Defibrillators
Naehle, C. P., Strach, K. A., Thomas, D., Meyer, C. H., Linhart, M., Bitaraf, S., Litt, H. I., SCHWAB, J. O., Schild, H. H., & Sommer, T. (2009) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2009.04.050 · Cited by 122
Gimbel, J., Kanal, E., Schwartz, K. M., & Wilkoff, B. L. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.09520.x · Cited by 103
Vahlhaus, C., Sommer, T., Lewalter, T., Schimpf, R., Schumacher, B. M., Jung, W., & Lüderitz, B. (2001) · Pacing and Clinical Electrophysiology · doi:10.1046/j.1460-9592.2001.00489.x · Cited by 102
Gimbel, J., & Kanal, E. (2004) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2004.02.009 · Cited by 98
Cohen, J. D., Costa, H. S., & Russo, R. J. (2012) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2012.07.030 · Cited by 95
How to perform magnetic resonance imaging on patients with implantable cardiac arrhythmia devices
Nazarian, S., & Halperin, H. R. (2008) · Heart Rhythm · doi:10.1016/j.hrthm.2008.10.021 · Cited by 89
Self-citationMollerus, M. E., Albin, G., Lipinski, M., & Lucca, J. (2010) · EP Europace · doi:10.1093/europace/euq092 · Cited by 83
Government Viewpoint: U.S. Food & Drug Administration: Pacemakers, ICDs and MRI
Faris, O., & Shein, M. J. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.50035.x · Cited by 67
Strategies for the Safe Magnetic Resonance Imaging of Pacemaker‐Dependent Patients
Gimbel, J., BAILEY, S. M., TCHOU, P. J., Ruggieri, P. M., & Wilkoff, B. L. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.00230.x · Cited by 65
Burke, P., Ghanbari, H., Alexander, P. B., Shaw, M. K., Daccarett, M., & Machado, C. (2010) · Journal of Interventional Cardiac Electrophysiology · doi:10.1007/s10840-009-9463-y · Cited by 61
Is Magnetic Resonance Imaging Safe in Cardiac Pacemaker Recipients?
del Ojo, J. L., Moya, F., Villalba, J., SANZ, O., Pavón, R., García, D. H., & Pastor, L. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.50033.x · Cited by 61
Mollerus, M. E., Albin, G., Lipinski, M., & LUCCA, J. (2008) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2008.01172.x · Cited by 59
Clinical Evaluation of the Safety of Repetitive Intraoperative Defibrillation Threshold Testing
Frame, R., Brodman, R., Furman, S., KIM, S. G., Roth, J. A., Ferrick, K. J., Hollinger, I., GROSS, J. N., & Fisher, J. D. (1992) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1992.tb03077.x · Cited by 59
Nazarian, S., Reynolds, M. R., Ryan, M. P., Wolff, S. D., Mollenkopf, S. A., & Turakhia, M. P. (2015) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.24971 · Cited by 56
Self-citationSafety of serial MRI in patients with implantable cardioverter defibrillators
Junttila, J. M., Fishman, J. E., Lopera, G. A., Pattany, P. M., Velazquez, D. L., Williams, A. R., Trachtenberg, B. H., Sanina, C., Mather, J., & Hare, J. M. (2011) · Heart · doi:10.1136/heartjnl-2011-300153 · Cited by 54
“Power-on resets” in cardiac implantable electronic devices during magnetic resonance imaging
Higgins, J. V., Sheldon, S. H., Watson, R. E., Dalzell, C. M., Acker, N. G., Cha, Y., Asirvatham, S. J., Kapa, S., Felmlee, J. P., & Friedman, P. A. (2014) · Heart Rhythm · doi:10.1016/j.hrthm.2014.10.039 · Cited by 53
Wollmann, C. G., Thudt, K., Kaiser, B., Salomonowitz, E. K., Mayr, H., & Globits, S. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-16-30 · Cited by 53
Forleo, G. B., Santini, L., Rocca, D. G. D., Romano, V. C., Papavasileiou, L. P., Magliano, G., Sgueglia, M., & Romeo, F. (2010) · Heart Rhythm · doi:10.1016/j.hrthm.2010.02.020 · Cited by 50
Industry Viewpoint: St. Jude Medical: Pacemakers, ICDs and MRI
LEVINE, P. A. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.50027.x · Cited by 39
DANILOVIC, D., & Ohm, O. (1999) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1999.tb00498.x · Cited by 39
Industry Viewpoint: Medtronic: Pacemakers, ICDs, and MRI
Stanton, M. S. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.50025.x · Cited by 38
Industry Viewpoint: Guidant: Pacemakers, ICDs, and MRI
Smith, J. M. (2005) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.2005.09591.x · Cited by 38
Camacho, J. C., Moreno, C. C., SHAH, A. D., Mittal, P., Mengistu, A., Lloyd, M. S., El‐Chami, M. F., Lerakis, S., & Saindane, A. M. (2016) · American Journal of Roentgenology · doi:10.2214/ajr.16.16033 · Cited by 32
Callans, D. J., Hook, B. G., & Marchlinski, F. E. (1993) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(93)90186-5 · Cited by 28
Changes in pacing lead impedance over time predict lead failure
Sharif, M., Wyse, D. G., Rothschild, J., & Gillis, A. M. (1998) · The American Journal of Cardiology · doi:10.1016/s0002-9149(98)00414-7 · Cited by 25
Glikson, M., Luria, D. M., Friedman, P. A., Trusty, J. M., Benderly, M., Hammill, S. C., & Stanton, M. S. (2000) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2000.tb00311.x · Cited by 25
Pacing Impedance Variability in Tined Steroid Eluting Leads
DANILOVIC, D., & Ohm, O. (1998) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1998.tb00205.x · Cited by 24
MRI with cardiac pacing devices – Safety in clinical practice
Kaasalainen, T. T., Pakarinen, S., Kivistö, S., Holmström, M., Hänninen, H., Peltonen, J. I., Lauerma, K., & Sipilä, O. (2014) · European Journal of Radiology · doi:10.1016/j.ejrad.2014.04.022 · Cited by 24
Safety of magnetic resonance scanning without monitoring of patients with pacemakers
Bertelsen, L., Petersen, H. H., Philbert, B. T., Svendsen, J. H., Thomsen, C. E., & Vejlstrup, N. G. (2016) · EP Europace · doi:10.1093/europace/euw066 · Cited by 23
1147-210 Implantable cardiac defibrillator interactions with magnetic resonance imaging at 1.5 tesia
Coman, J. A., Martin, E. T., Sandler, D. A., & Thomas, J. R. (2004) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(04)90585-2 · Cited by 14
Cited by
2023 ESC guidelines for the management of cardiomyopathies
Arbelo, E., Protonotarios, A., Gimeno, J. R., Arbustini, E., Barriales-Villa, R., Basso, C., Bezzina, C. R., Biagini, E., Blom, N. A., de Boer, R. A., et al. (2023) · European Heart Journal · doi:10.1093/eurheartj/ehad194 · Cited by 2,844
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2021) · European Heart Journal · doi:10.1093/eurheartj/ehab364 · Cited by 2,019
2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy
Glikson, M., Nielsen, J. C., Kronborg, M. B., Michowitz, Y., Auricchio, A., Barbash, I. M., Barrabés, J. A., Boriani, G., Braunschweig, F., Brignole, M., et al. (2022) · EP Europace · doi:10.1093/europace/euab232 · Cited by 914
Metadata
- Authors
- Nazarian, S., Hansford, R., Rahsepar, A. A., Weltin, V., McVeigh, D., Gucuk Ipek, E., Kwan, A., Berger, R. D., Calkins, H., Lardo, A. C., Kraut, M. A., Kamel, I. R., Zimmerman, S. L., & Halperin, H. R.
- Venue
- New England Journal of Medicine
- Publisher
- Massachusetts Medical Society
- Published
- 2017-12-28
- Volume
- 377
- Issue
- 26
- Pages
- 2555-2564
- ISSN
- 0028-4793 (ISSN-L)
- 1533-4406
- OpenAlex
- W2776345759
- Cited by
- 301
- References
- 42
- FWCI
- 24.64
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.