A robust algorithm for heart rate variability time series artefact correction using novel beat classification
Abstract
Purpose: Heart rate variability is a commonly used measurement to evaluate functioning of autonomic nervous system, psychophysiological stress, and exercise intensity and recovery. HRV measurements contain artefacts such as extra, missed or misaligned beat detections, which can produce significant distortion on HRV parameters. In this paper, a robust automatic method for artefact detection from HRV time series is proposed.Methods: The proposed detection method is based on time-varying thresholds estimated from distribution of successive RR-interval differences combined with a novel beat classification scheme. The method is validated using simulated extra, missed and misaligned beat detections as well as real artefacts such as atrial and ventricular ectopic beats.Results: The sensitivity of the algorithm to detect simulated missed/extra beats was 100%. The sensitivity to detect real atrial and ventricular ectopic beats was 96.96%, the corresponding specificity being 99.94%. The mean error in HRV parameters after correction was <2% for missed and extra beats as well as for misaligned beats generated with large displacement factors. Misaligned beats with smallest displacement factor were the most difficult to detect and resulted in largest HRV parameter errors after correction, largest errors being <8%.Conclusions: The HRV artefact correction algorithm presented in this study provided comparable specificity and better sensitivity to detect ectopic beats as compared to state-of-the-art algorithms. The proposed algorithm detects abnormal beats with high accuracy, is relatively easy to implement, and secures reliable HRV analysis by reducing the effect of possible artefacts to tolerable level.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Heart Rate Variability: Standards of Measurement, Physiological Interpretation, and Clinical Use
Electrophysiology, T. F. O. T. E. S. O. C. T. N. A. (1996) · Circulation · doi:10.1161/01.cir.93.5.1043 · Cited by 17,159
Goldberger, A. L., Amaral, L. A. N., Glass, L., Hausdorff, J. M., Ivanov, P. C., Mark, R. G., Mietus, J. E., Moody, G. B., Peng, C.-K., & Stanley, H. E. (2000) · Circulation · doi:10.1161/01.cir.101.23.e215 · Cited by 14,809
The impact of the MIT-BIH Arrhythmia Database
Moody, G. B., & Mark, R. G. (2001) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.932724 · Cited by 4,812
Thayer, J. F., Åhs, F., Fredrikson, M., Sollers, J. J., & Wager, T. D. (2012) · Neuroscience & Biobehavioral Reviews · doi:10.1016/j.neubiorev.2011.11.009 · Cited by 3,224
Heart rate variability: A review
Rajendra Acharya, U., Paul Joseph, K., Kannathal, N., Lim, C. M., & Suri, J. S. (2006) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-006-0119-0 · Cited by 2,820
Kubios HRV–heart rate variability analysis software.
Tarvainen, M. P., Niskanen, J.-P., Lipponen, J. A., Ranta-aho, P. O., & Karjalainen, P. A. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2013.07.024 · Cited by 2,720 · cited 1×
“The sampling rate of the ECG was 500Hz and QRS-complexes were detected using the Kubios HRV software [18].”
MethodSelf-citationDepression, Heart Rate Variability, and Acute Myocardial Infarction
Carney, R. M., Blumenthal, J. A., Stein, P. K., Watkins, L., Catellier, D., Berkman, L. F., Czajkowski, S. M., O’Connor, C., Stone, P. H., & Freedland, K. E. (2001) · Circulation · doi:10.1161/hc4201.097834 · Cited by 754
Schmidt, G., Malik, M., Barthel, P., Schneider, R., Ulm, K., Rolnitzky, L., Camm, A. J., Bigger, J. T., & Schömig, A. (1999) · The Lancet · doi:10.1016/s0140-6736(98)08428-1 · Cited by 716 · cited 1×
“Decreased HRV has been found to associate with many clinical scenarios such as mortality after myocardial infarction [1,2], congestive heart failure [3] and diabetic neuropathy [4].”
BackgroundSchmidt, F., Malík, Barthel, & Schneider, R. (1999) · The Lancet · Cited by 594
Ponikowski, P., Anker, S. D., Chua, T. P., Szelemej, R., Piepoli, M., Adamopoulos, S., Webb-Peploe, K., Harrington, D., Banasiak, W., Wrabec, K., et al. (1997) · The American Journal of Cardiology · doi:10.1016/s0002-9149(97)00215-4 · Cited by 493
Cited by
Schaffarczyk, M., Rogers, B., Reer, R., & Gronwald, T. (2022) · Sensors · doi:10.3390/s22176536 · Cited by 349
Moya-Ramon, M., Mateo-March, M., Peña-González, I., Zabala, M., & Javaloyes, A. (2022) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2022.106696 · Cited by 122
Rogers, B., Giles, D., Draper, N., Mourot, L., & Gronwald, T. (2021) · Sensors · doi:10.3390/s21030821 · Cited by 54
Rogers, B., Schaffarczyk, M., Clauß, M., Mourot, L., & Gronwald, T. (2022) · Sensors · doi:10.3390/s22052032 · Cited by 41
Flory, S., Guglielmini, S., Scholkmann, F., Marcar, V. L., & Wolf, M. (2023) · Scientific Reports · doi:10.1038/s41598-023-39083-9 · Cited by 18
Metadata
- Authors
- Lipponen, J. A., & Tarvainen, M. P.
- Venue
- Journal of Medical Engineering & Technology
- Publisher
- Informa UK Limited
- Published
- 2019-04-03
- Volume
- 43
- Issue
- 3
- Pages
- 173-181
- ISSN
- 0309-1902 (ISSN-L)
- 1464-522X
- OpenAlex
- W2957009284
- Cited by
- 364
- References
- 18
- FWCI
- 15.81
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: A robust algorithm for heart rate variability time series artefact correction using novel beat classification. https://app.cochat.ai/papers/10.1080/03091902.2019.1640306
- The paper
- Lipponen, J. A., & Tarvainen, M. P. (2019). A robust algorithm for heart rate variability time series artefact correction using novel beat classification. Journal of Medical Engineering & Technology. https://doi.org/10.1080/03091902.2019.1640306
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.