Validity of the Polar H10 Sensor for Heart Rate Variability Analysis during Resting State and Incremental Exercise in Recreational Men and Women
Open accessAbstract
Heart rate variability (HRV) is frequently applied in sport-specific settings. The rising use of freely accessible applications for its recording requires validation processes to ensure accurate data. It is the aim of this study to compare the HRV data obtained by the Polar H10 sensor chest strap device and an electrocardiogram (ECG) with the focus on RR intervals and short-term scaling exponent alpha 1 of Detrended Fluctuation Analysis (DFA a1) as non-linear metric of HRV analysis. A group of 25 participants performed an exhaustive cycling ramp with measurements of HRV with both recording systems. Average time between heartbeats (RR), heart rate (HR) and DFA a1 were recorded before (PRE), during, and after (POST) the exercise test. High correlations were found for the resting conditions (PRE: r = 0.95, rc = 0.95, ICC3,1 = 0.95, POST: r = 0.86, rc = 0.84, ICC3,1 = 0.85) and for the incremental exercise (r > 0.93, rc > 0.93, ICC3,1 > 0.93). While PRE and POST comparisons revealed no differences, significant bias could be found during the exercise test for all variables (p < 0.001). For RR and HR, bias and limits of agreement (LoA) in the Bland–Altman analysis were minimal (RR: bias of 0.7 to 0.4 ms with LoA of 4.3 to −2.8 ms during low intensity and 1.3 to −0.5 ms during high intensity, HR: bias of −0.1 to −0.2 ms with LoA of 0.3 to −0.5 ms during low intensity and 0.4 to −0.7 ms during high intensity). DFA a1 showed wider bias and LoAs (bias of 0.9 to 8.6% with LoA of 11.6 to −9.9% during low intensity and 58.1 to −40.9% during high intensity). Linear HRV measurements derived from the Polar H10 chest strap device show strong agreement and small bias compared with ECG recordings and can be recommended for practitioners. However, with respect to DFA a1, values in the uncorrelated range and during higher exercise intensities tend to elicit higher bias and wider LoA.
AI snapshot
Written from the paper’s full text.
TL;DR
In 25 recreational men and women, the Polar H10 chest strap with Elite HRV showed near-perfect agreement with a 12-channel ECG for RR and HR at rest and during incremental cycling, with small bias despite significant exercise differences. DFA a1 agreement was high but weaker, with wider bias and limits of agreement, especially at higher exercise intensities.
Key findings
- Linear HRV measures showed near-perfect agreement: PRE/POST RR and HR had r = 1.00, rc = 1.00, ICC3,1 = 1.00, and incremental exercise also had r = 1.00, rc = 1.00, ICC3,1 = 1.00. Resting PRE/POST comparisons showed no significant differences, while exercise comparisons showed significant bias for all variables (p < 0.001).
- DFA a1 correlations were high but comparatively weaker: PRE/POST r > 0.86, rc > 0.84, ICC3,1 > 0.85; incremental exercise r > 0.93, rc > 0.93, ICC3,1 > 0.93, with significant differences during exercise (p < 0.001).
- DFA a1 bias and limits of agreement were wider than for linear measures: PRE bias −2.1%, POST bias 2.5%; during exercise, low intensity bias 0.9% with LoA 11.6 to −9.9%, and high intensity bias 8.6% with LoA 58.1 to −40.9%.
- DFA a1 index values were approximately 6% and 7.5% higher for Polar H10 at fixed values of 0.75 and 0.5 compared with ECG, so threshold boundaries were crossed at higher HRs when using Polar H10.
Methods
- Cross-sectional validation study: 25 participants (men n = 14, age 40 ± 14 years; women n = 11, age 34 ± 10 years) performed an exhaustive cycling ramp with simultaneous Polar H10/Elite HRV and 12-channel ECG recordings before (PRE), during, and after (POST) exercise.
- RR, HR, and DFA a1 were analyzed in Kubios HRV Premium 3.5.0; resting analyses used a 2-min window, exercise used a 2-min moving window with 20-s grid, and datasets with artefacts >5% were excluded.
- Agreement was assessed with linear regression, Pearson’s r, Lin’s concordance correlation coefficient, ICC3,1, R², SEE, Bland–Altman LoA, paired t-test or Wilcoxon signed ranks test, and Cohen’s d, with significance at p ≤ 0.05.
Limitations
- Stated by the paper: the study did not group samples by individual personal characteristics; future studies could examine whether body composition, gender, age, or ventricular characteristics influence outcomes.
- Stated by the paper: the Polar H10 stores RR intervals but not ECG waveforms, so excessive artefacts cannot be identified or corrected, and noise cannot be differentiated from cardiac arrhythmia; five participants were excluded from exercise analysis and two from resting POST measurement due to artefacts >5%.
- Stated by the paper: raw data were analyzed with Kubios HRV Premium using automatic artefact correction, and the free standard Kubios version with threshold-based correction could yield slightly different results; the text also makes evident that device and lead placement differences prevent attributing discrepancies to a single variable.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 18 cited sources checked · 0 supported · 0 flagged
- Findingreasoning
The amount of deviation was variable, with some participants showing very close agreement between devices.
“However, the amount of deviation was variable, with some participants showing very close agreement between devices.”
Quote in the paperNot a substantial claim - Findingreasoning
Both threshold boundaries were crossed at higher HRs when using the Polar H10.
“Both threshold boundaries were crossed at higher HRs when using the Polar H10”
Quote in the paperSubstantial claim - Findingown data
In POST, RR bias was 0.2 to 0.0 ms with upper and lower LoA of 1.6 to 0.5 ms and −1.2 to −0.6 ms.
“and a bias of 0.2 to 0.0 ms with upper and lower LoA of 1.6 to 0.5 ms and −1.2 to −0.6 ms in POST”
Quote in the paperNot a substantial claim - Findingown data
During resting state conditions PRE and POST, the linear parameters RR and HR showed nearly perfect correlations (r = 1.00, rc = 1.00, ICC3,1 = 1.00).
“The comparisons during resting state conditions PRE and POST showed nearly perfect correlations for the linear parameters RR and HR (r = 1.00, rc = 1.00, ICC3,1 = 1.00).”
Quote in the paperSubstantial claim - Findingown data
DFA a1 bias in PRE was −2.1% (upper LoA 14.7 to 16.4%, lower LoA −18.9 to −20.5%).
“With respect to DFA a1 there was bias of −2.1% (upper LoA of 14.7 to 16.4% and lower LoA of −18.9 to −20.5%)”
Quote in the paperNot a substantial claim
- Findingown data
Bland-Altman analysis revealed a mean difference for RR in PRE of 0.2 to 0.1 ms with upper LoA of 2.6 to 1.3 ms and lower LoA of −2.2 to −1.1 ms.
“Bland–Altman analysis revealed a mean difference of 0.2 to 0.1 ms for RR with upper LoA of 2.6 to 1.3 ms and lower LoA of −2.2 to −1.1 ms in PRE”
Quote in the paperSubstantial claim - Findingown data
Index values tended to be approximately 6% and 7.5% higher at the fixed values of 0.75 and 0.5 for Polar H10 compared to ECG.
“Index values tended to be approximately 6% and 7.5% higher at the fixed values of 0.75 and 0.5 for Polar H10 in comparison to the ECG (Figure 5C)”
Quote in the paperSubstantial claim - Findingown data
There were no significant differences for PRE and POST conditions and the three variables of interest (p > 0.05).
“There were no significant differences for both conditions and the three variables of interest (p > 0.05) (Table 1).”
Quote in the paperSubstantial claim - Findingown data
DFA a1 bias in POST was 2.5% (upper LoA 15.6 to 23.7%, lower LoA −10.2 to −19.1%).
“and 2.5% for POST (upper LoA of 15.6 to 23.7% and lower LoA of −10.2 to −19.1%)”
Quote in the paperNot a substantial claim - Premisecitation
A more recent study found no differences in the results of the Polar H10 compared to ECG when Kubios HRV software was used to obtain HRV data during resting condition in supine and seated positions.
“There were no differences observed in the results of the Polar H10 in comparison to the ECG when Kubios HRV software was used to obtain HRV data.”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [10] Zabala, 2022, Validity and Reliability of Different Smartphones Applications to Measure HRV during Short and Ultra-Short Measurements in Elite Athletes, Comput. Methods Programs Biomed. source
Not checked in the free run.
- Premisecitation
The Polar H10 chest strap device has already proven validity to assess RR intervals correctly during rest and physical exercise conditions.
“A frequently used chest strap device is the Polar H10 (Polar Electro Oy, Kempele, Finland) which has already proven validity to assess RR intervals correctly during rest and physical exercise conditions [13].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [13] Schweizer, 2019, RR Interval Signal Quality of a Heart Rate Monitor and an ECG Holter at Rest and during Exercise, Eur. J. Appl. Physiol. source
Not checked in the free run.
- Interpretationreasoning
Even though the study attempted to exclude recordings exceeding 5% artifact, it cannot be guaranteed that sporadic DFA a1 variation did not occur.
“Even though this study attempted to exclude recordings exceeding 5% artifact, it cannot be guaranteed that sporadic DFA a1 variation did not occur.”
Quote in the paperSubstantial claim - Interpretationreasoning
It is impossible to attribute a single variable for the data discrepancy.
“therefore it is impossible to attribute a single variable for the data discrepancy.”
Quote in the paperNot a substantial claim - Interpretationcitation
Artifact correction can lead to variable results in terms of DFA a1 bias during incremental tests.
“In terms of DFA a1 bias during the incremental tests, artifact correction can lead to variable results [24].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [24] Rogers, B., Giles, D., Draper, N., Mourot, L., and Gronwald, T. (2021). Influence of Artefact Correction and Recording Device Type on the Practical Application of a Non-Linear Heart Rate Variability Biomarker for Aerobic Threshold Determination. Sensors, 21. source
Not checked in the free run.
- Interpretationreasoning
It is impossible to differentiate whether abundant artifacts originate from noise or due to cardiac arrhythmia.
“In this context, it is also impossible to differentiate whether abundant artifacts originate from noise or due to cardiac arrhythmia [30].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [30] Rogers, B., Schaffarczyk, M., Clauß, M., Mourot, L., and Gronwald, T. (2022). The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study. Sensors, 22. source
Not checked in the free run.
- Interpretationown data
With respect to DFA a1, values in the uncorrelated range and during higher exercise intensities tend to elicit higher bias and wider LoA.
“However, with respect to DFA a1, values in the uncorrelated range and during higher exercise intensities tend to elicit higher bias and wider LoA.”
- Also stated in: § 5. Conclusions · hedged
“However, with respect to DFA a1, values in the uncorrelated range and during higher exercise intensities tend to elicit higher bias and wider LoA.”
Quote in the paperSubstantial claim - Also stated in: § 5. Conclusions · hedged
- Interpretationown data
Linear HRV measurements derived from the Polar H10 sensor chest strap device recorded with the Elite HRV app correspond closely with measurements taken with a reference ECG in terms of RR intervals and HR.
“Linear HRV measurements derived from the Polar H10 sensor chest strap device recorded with the Elite HRV app correspond closely with measurements taken with a reference ECG in terms of RR intervals and HR.”
Quote in the paperSubstantial claim - Interpretationreasoning
Because the Polar H10 only stores RR intervals and not ECG waveform recordings, it is not possible to identify or potentially correct signals with excessive artefacts leading to data loss.
“Since the Polar H10 only stores RR intervals and not ECG waveform recordings, it is not possible to identify or potentially correct signals with excessive artefacts leading to a data loss.”
Quote in the paperNot a substantial claim- Rests on an assumption: The Polar H10 only stores RR intervals and not ECG waveform recordings.
- May combine two claims. 11
- Classified as probably not a substantial claim. 22
May combine two claims.
Findingown dataDFA a1 during incremental exercise showed wider bias (0.9 to 8.6%) and LoAs of 11.6 to −9.9% during low intensity and 58.1 to −40.9% during high intensity.
“DFA a1 showed wider bias (0.9 to 8.6%) and LoAs of 11.6 to −9.9% during low intensity and 58.1 to −40.9% during high intensity.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataAn example female participant's data during the incremental cycling ramp was excluded from validation analysis because artifact exceeded >5%, and on closer examination of the parallel recorded ECG these artifacts were classified as frequent atrial premature complexes (APCs).
“Figure 7 illustrates this with an example of a female participant from the present study whose data during the incremental cycling ramp was excluded from validation analysis since artifact exceeded >5%. On closer examination of the parallel recorded ECG, these “artifacts” were classified as frequent atrial premature complexes (APCs)”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataDuring resting state conditions PRE and POST, DFA a1 correlations were high but comparatively weaker (r > 0.86, rc > 0.84, ICC3,1 > 0.85).
“Although correlations were also high for DFA a1 (r > 0.86, rc > 0.84, ICC3,1 > 0.85), they were comparatively weaker”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataBland-Altman plot for RR during incremental exercise showed a mean difference of 0.7 to 0.4 ms with upper and lower LoA of 4.3 to −2.8 ms during low intensity and 1.3 to −0.5 ms during high intensity exercise.
“Bland–Altman plot for RR showed a mean difference of 0.7 to 0.4 ms with upper and lower LoA of 4.3 to −2.8 ms during low intensity and 1.3 to −0.5 ms during high intensity exercise.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataDuring incremental exercise, high correlations were found for linear parameters (r = 1.00, rc = 1.00, ICC3,1 = 1.00) and DFA a1 (r > 0.93, rc > 0.93, ICC3,1 > 0.93), although with a statistically significant difference (p < 0.001).
“High correlations could be found during the incremental exercise test for the linear parameters (r = 1.00. rc = 1.00, ICC3,1 = 1.00), and DFA a1 (r > 0.93. rc > 0.93, ICC3,1 > 0.93), although with a statistically significant difference (p < 0.001)”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataThe effects of high degrees of artifact correction are demonstrated, showing either data loss or HRV bias.
“The effects of high degrees of artifact correction are demonstrated, showing either data loss or HRV bias.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Findingown dataFive participants were excluded from exercise analysis and two participants from resting POST measurement due to artefacts >5%.
“Five participants were excluded from exercise analysis and two participants from resting POST measurement due to artefacts >5%.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
PremisecitationRecent studies show the potential of DFA a1 to demarcate physiological threshold boundaries by fixed DFA a1 values at the aerobic (~0.75) and anaerobic threshold (~0.5).
“recent studies show the potential of this parameter to demarcate physiological threshold boundaries by means of fixed DFA a1 values at the aerobic (~0.75) and anaerobic threshold (~0.5) for usage in performance monitoring and exercise intensity distribution in endurance sports [16,17].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [16] Rogers, B., Giles, D., Draper, N., Mourot, L., and Gronwald, T. (2021). Detection of the Anaerobic Threshold in Endurance Sports: Validation of a New Method Using Correlation Properties of Heart Rate Variability. J. Funct. Morphol. Kinesiol., 6. source
Not checked in the free run.
- cites [17] Mateo-March, M., Moya-Ramón, M., Javaloyes, A., Sánchez-Muñoz, C., and Clemente-Suárez, V.J. (2022). Validity of Detrended Fluctuation Analysis of Heart Rate Variability to Determine Intensity Thresholds in Professional Cyclists. Eur. J. Sport Sci., 1–20. source
Not checked in the free run.
Classified as probably not a substantial claim.
PremisecitationThe gold standard measure for quantification of RR intervals and assessment of HRV is the electrocardiogram (ECG).
“The gold standard measure for the quantification of RR intervals and assessment of HRV is the electrocardiogram (ECG) [7].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [7] Sammito, 2016, Reference Values for Time- and Frequency-Domain Heart Rate Variability Measures, Heart Rhythm. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationIn a previous report, the Polar H10 had substantially better signal quality during high-intensity activities compared to a 3-lead ECG Holter monitor.
“In this report, there was a substantially better signal quality during high-intensity activities for the Polar H10 in comparison to a 3-lead ECG Holter monitor”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [13] Schweizer, 2019, RR Interval Signal Quality of a Heart Rate Monitor and an ECG Holter at Rest and during Exercise, Eur. J. Appl. Physiol. source
Not checked in the free run.
Probably not a substantial claim: may not pass S2 (checkable), S3 (load-bearing).
PremisecitationMobile systems and chest straps offer superior practicability with respect to cost, ease of use, portability and interpretation.
“However, it has become common practice to use mobile systems (e.g., apps, wearables) and chest straps for data recording, storage, analysis and/or export which offers superior practicability with respect to cost, ease to use, portability and interpretation [8,9,10].”
Quote in the paperNot a substantial claim0 of 3 sources checked- cites [8] Perrotta, 2017, Validity of the Elite HRV Smartphone Application for Examining Heart Rate Variability in a Field-Based Setting, J. Strength Cond. Res. source
Not checked in the free run.
- cites [9] Dobbs, 2019, The Accuracy of Acquiring Heart Rate Variability from Portable Devices: A Systematic Review and Meta-Analysis, Sports Med. source
Not checked in the free run.
- cites [10] Zabala, 2022, Validity and Reliability of Different Smartphones Applications to Measure HRV during Short and Ultra-Short Measurements in Elite Athletes, Comput. Methods Programs Biomed. source
Not checked in the free run.
Classified as probably not a substantial claim. May combine two claims.
PremisecitationDFA a1 has shown suitability to describe complex cardiac autonomic regulation during various exercise intensities, modalities, and environmental conditions while possessing a wide dynamic range throughout all intensity zones.
“The non-linear short-term scaling exponent alpha 1 of Detrended Fluctuation Analysis (DFA a1) has shown its suitability to describe the complex cardiac autonomic regulation during various exercise intensities, modalities, and environmental conditions while possessing a wide dynamic range throughout all intensity zones in contrast to standard linear parameters [14,15].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [14] Gronwald, 2020, Correlation Properties of Heart Rate Variability during Endurance Exercise: A Systematic Review, Ann. Noninvasive Electrocardiol. source
Not checked in the free run.
- cites [15] Rogers, 2022, Fractal Correlation Properties of Heart Rate Variability as a Biomarker for Intensity Distribution and Training Prescription in Endurance Exercise: An Update, Front. Physiol. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationExercise-associated DFA a1 dynamics may give insights about physiological status in terms of monitoring fatigue during low-intensity exercise.
“Additionally, exercise associated DFA a1 dynamics may also give insights about physiological status in terms of monitoring of fatigue during low-intensity exercise [18].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [18] Rogers, 2021, Fractal Correlation Properties of Heart Rate Variability as a Biomarker of Endurance Exercise Fatigue in Ultramarathon Runners, Physiol. Rep. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationcitationThe basic, cost-free "standard" version of Kubios HRV with an alternate threshold-based artifact correction method could lead to slightly different results than presented.
“The basic, cost-free “standard” version of Kubios HRV with an alternate threshold-based artifact correction method could lead to slightly different results than presented [24].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [24] Rogers, B., Giles, D., Draper, N., Mourot, L., and Gronwald, T. (2021). Influence of Artefact Correction and Recording Device Type on the Practical Application of a Non-Linear Heart Rate Variability Biomarker for Aerobic Threshold Determination. Sensors, 21. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
InterpretationreasoningIt is plausible that some participants had either different or suboptimal signal strength using a chest strap device as opposed to ECG lead 2.
“Hence, it is certainly plausible that some participants had either different or sub optimal signal strength using a chest strap device as opposed to ECG lead 2.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenIt can be considered as a possible limitation that this study did not group the samples by their individual personal characteristics.
“It can be considered as a possible limitation that this study did not group the samples by their individual personal characteristics.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductcitationData sets with artefacts >5% were excluded from analysis based on [24].
“Data sets with artefacts >5% were excluded from analysis based on [24].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [24] Rogers, B., Giles, D., Draper, N., Mourot, L., and Gronwald, T. (2021). Influence of Artefact Correction and Recording Device Type on the Practical Application of a Non-Linear Heart Rate Variability Biomarker for Aerobic Threshold Determination. Sensors, 21. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenPrior to testing, the gas analyzers were calibrated according to the manufacturer's instructions.
“Prior to testing, the gas analyzers were calibrated according to the manufacturer’s instructions.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing), S4 (based).
Interpretationnone givenThe potential effect of lead placement (chest strap electrodes vs. ECG lead 2) is notable in terms of DFA a1 determination.
“The potential effect of lead placement (chest strap electrodes vs. ECG lead 2) is also notable in terms of DFA a1 determination.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenParticipants were asked to abstain from caffeine, alcohol, tobacco, and vigorous exercise 24 h before testing.
“Participants were asked to abstain from caffeine, alcohol, tobacco, and vigorous exercise 24 h before testing”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S2 (checkable).
InterpretationreasoningCommercial use of this mobile-based HRV recording setup for monitoring HRV data during resting and endurance exercise conditions could be justified.
“Nevertheless, since this mobile-based HRV recording setup displays superior practicability with generally comparable results, its commercial use for the monitoring of HRV data during resting and endurance exercise conditions could be justified.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataLinear HRV measurements derived from the Polar H10 chest strap device show strong agreement and small bias compared with ECG recordings and can be recommended for practitioners.
“Linear HRV measurements derived from the Polar H10 chest strap device show strong agreement and small bias compared with ECG recordings and can be recommended for practitioners.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S4 (based).
InterpretationreasoningThe higher DFA a1 bias and wider LoA may partly be related to expected differences in non-linear HRV associated with ECG lead placement.
“This may partly be related to the expected differences in non-linear HRV associated with ECG lead placement.”
Quote in the paperNot a substantial claimClassified as probably not a substantial claim.
Study design and conductcitationFor the 12-channel ECG, lead 2 was used as the comparable lead to the chest strap device.
“For the 12-channel ECG, lead 2 was used as the comparable lead to the chest strap device [21].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [21] Jeyhani, V., Mantysalo, M., Noponen, K., Seppanen, T., and Vehkaoja, A. (2019, January 23–27). Effect of Different ECG Leads on Estimated R–R Intervals and Heart Rate Variability Parameters. Proceedings of the 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
Study design and conductnone givenParticipants provided written informed consent.
“provided written informed consent.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenEntry criteria included adults (>18 years) of either sex and of any fitness level without previous medical history, current medications or recent illness.
“Entry criteria included adults (>18 years) of either sex and of any fitness level without previous medical history, current medications or recent illness.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenEthical approval for the study was obtained by the University of Hamburg, Department of Psychology and Movement Science, Germany (reference no.: 2021_400) and is in line with the principles of the Declaration of Helsinki.
“Ethical approval for the study was obtained by the University of Hamburg, Department of Psychology and Movement Science, Germany (reference no.: 2021_400) and is in line with the principles of the Declaration of Helsinki.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
InterpretationcitationThe degree of bias may simply relate to the discordance between actual ECG waveform signal strength and morphology.
“The degree of bias may simply relate to the discordance between actual ECG waveform signal strength and morphology [21].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [21] Jeyhani, V., Mantysalo, M., Noponen, K., Seppanen, T., and Vehkaoja, A. (2019, January 23–27). Effect of Different ECG Leads on Estimated R–R Intervals and Heart Rate Variability Parameters. Proceedings of the 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, Germany. source
Not checked in the free run.
The paper relies on this without support.
To date no validation has been conducted for non-linear measures of HRV.
“In addition to the requirement to perform population-based validations, to date no validation has been conducted for non-linear measures of HRV.”
The paper relies on this without support.
Researchers and recreational athletes frequently use the Polar H10 device for data collection and rely on appropriate measurements.
“since researchers and recreational athletes frequently use this device for data collection and therefore rely on appropriate measurements.”
The paper relies on this without support.
The Polar H10 only stores RR intervals and not ECG waveform recordings.
“Since the Polar H10 only stores RR intervals and not ECG waveform recordings”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
ECG and chest strap are two modalities that are known to correlate well.
“As a gold standard comparison for accuracy, either an ECG (n = 7) or a chest strap (n = 3) was used, two modalities that are known to correlate well [13].”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [13]
As a gold standard comparison for accuracy, either an ECG (n = 7) or a chest strap (n = 3) was used.
“As a gold standard comparison for accuracy, either an ECG (n = 7) or a chest strap (n = 3) was used, two modalities that are known to correlate well [13].”
From Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review (2023) · cites this paper as [13]
Cites
Measuring agreement in method comparison studies
Bland, J. M., & Altman, D. G. (1999) · Statistical Methods in Medical Research · doi:10.1177/096228029900800204 · Cited by 9,619 · see citing works
A concordance correlation coefficient to evaluate reproducibility.
Lin, L. I.-K. (1989) · Biometrics · doi:10.2307/2532051 · Cited by 8,785 · see citing works
A Real-Time QRS Detection Algorithm
Pan, J., & Tompkins, W. J. (1985) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.1985.325532 · Cited by 7,855 · see citing works
Quantification of scaling exponents and crossover phenomena in nonstationary heartbeat time series
Peng, C., Havlin, S., Stanley, E., & Goldberger, A. L. (1995) · Chaos An Interdisciplinary Journal of Nonlinear Science · doi:10.1063/1.166141 · Cited by 3,865 · see citing works
Age-predicted maximal heart rate revisited
Tanaka, H., Monahan, K. D., & Seals, D. R. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(00)01054-8 · Cited by 3,775 · cited 1× · see citing works
“The following criteria served for the assumption of exhaustion: (A) heart rate >90% of the maximum predicted heart rate (prediction model according to [20]: 208 − (0.”
BackgroundKubios HRV–heart rate variability analysis software.
Tarvainen, M. P., Niskanen, J.-P., Lipponen, J. A., Ranta-aho, P. O., & Karjalainen, P. A. (2014) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2013.07.024 · Cited by 2,720 · cited 4× · see citing works
“0 (Biosignal Analysis and Medical Imaging Group, Department of Physics, University of Kuopio, Kuopio, Finland, [19]) was conducted.”
BackgroundMethodInfluentialEstimates of Location Based on Rank Tests
Hodges, J. L., & Lehmann, E. L. (1963) · The Annals of Mathematical Statistics · doi:10.1214/aoms/1177704172 · Cited by 1,276 · see citing works
Biostatistics 104: correlational analysis.
Chan, Y. H. (2003) · PubMed · Cited by 1,223 · see citing works
Monitoring training status with HR measures: do all roads lead to Rome?
Buchheit, M. (2014) · Frontiers in Physiology · doi:10.3389/fphys.2014.00073 · Cited by 909 · see citing works
Cardiac Parasympathetic Reactivation Following Exercise: Implications for Training Prescription
Stanley, J., Peake, J. M., & Buchheit, M. (2013) · Sports Medicine · doi:10.1007/s40279-013-0083-4 · Cited by 551 · see citing works
RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise
Gilgen-Ammann, R., Schweizer, T., & Wyss, T. (2019) · European Journal of Applied Physiology · doi:10.1007/s00421-019-04142-5 · Cited by 438 · see citing works
Lipponen, J. A., & Tarvainen, M. P. (2019) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2019.1640306 · Cited by 374 · see citing works
Weippert, M., Kumar, M., Kreuzfeld, S., Arndt, D., Rieger, A., & Stoll, R. (2010) · European Journal of Applied Physiology · doi:10.1007/s00421-010-1415-9 · Cited by 302 · see citing works
Dobbs, W. C., Fedewa, M. V., MacDonald, H. V., Holmes, C. J., Cicone, Z. S., Plews, D. J., & Esco, M. R. (2019) · Sports Medicine · doi:10.1007/s40279-019-01061-5 · Cited by 194 · see citing works
Normal electrocardiographic findings: recognising physiological adaptations in athletes
Drezner, J. A., Fischbach, P., Froelicher, V. F., Marek, J. C., Pelliccia, A., Prutkin, J. M., Schmied, C. M., Sharma, S., Wilson, M. G., Ackerman, M. J., et al. (2013) · British Journal of Sports Medicine · doi:10.1136/bjsports-2012-092068 · Cited by 185 · see citing works
Perrotta, A. S., Jeklin, A. T., Hives, B. A., Meanwell, L. E., & Warburton, D. E. R. (2017) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000001841 · Cited by 185 · see citing works
Singh, N., Moneghetti, K. J., Christle, J. W., Hadley, D., Plews, D. J., & Froelicher, V. F. (2018) · Arrhythmia & Electrophysiology Review · doi:10.15420/aer.2018.27.2 · Cited by 140 · see citing works
Reference values for time- and frequency-domain heart rate variability measures
Sammito, S., & Böckelmann, I. (2016) · Heart Rhythm · doi:10.1016/j.hrthm.2016.02.006 · Cited by 131 · see citing works
Individual Endurance Training Prescription with Heart Rate Variability
Vesterinen, V., Nummela, A. T., Heikura, I. A., Laine, T., Hynynen, E., Botella, J., & Häkkinen, K. (2016) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000000910 · Cited by 125 · see citing works
Moya-Ramon, M., Mateo-March, M., Peña-González, I., Zabala, M., & Javaloyes, A. (2022) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2022.106696 · Cited by 122 · see citing works
Mühlen, J. M., Stang, J. S., Skovgaard, E. L., Júdice, P. B., Molina‐García, P., Johnston, W., Sardinha, L. B., Ortega, F. B., Caulfield, B., Bloch, W., et al. (2021) · British Journal of Sports Medicine · doi:10.1136/bjsports-2020-103148 · Cited by 118 · see citing works
Correlation properties of heart rate variability during endurance exercise: A systematic review
Gronwald, T., & Hoos, O. (2019) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12697 · Cited by 101 · see citing works
Self-citationStone, J. D., Ulman, H. K., Tran, K., Thompson, A. G., Halter, M. D., Ramadan, J. H., Stephenson, M., Finomore, V. S., Galster, S. M., Rezai, A. R., et al. (2021) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2021.585870 · Cited by 95 · see citing works
Main artifacts in electrocardiography
Pérez‐Riera, A. R., Barbosa‐Barros, R., Raimundo, R. D., & de Abreu, L. C. (2017) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12494 · Cited by 84 · see citing works
Fragmented QRS Complex: A Novel Marker of Cardiovascular Disease
Chatterjee, S., & Changawala, N. (2010) · Clinical Cardiology · doi:10.1002/clc.20709 · Cited by 75 · see citing works
Plews, D. J., Laursen, P. B., & Buchheit, M. (2016) · International Journal of Sports Physiology and Performance · doi:10.1123/ijspp.2016-0343 · Cited by 71 · see citing works
Rogers, B., & Gronwald, T. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.879071 · Cited by 63 · see citing works
Self-citationHernández‐Vicente, A., Hernando, D., Marín-Puyalto, J., Vicente‐Rodríguez, G., Garatachea, N., Pueyo, E., & Bailón, R. (2021) · Sensors · doi:10.3390/s21030902 · Cited by 57 · see citing works
Rogers, B., Giles, D., Draper, N., Mourot, L., & Gronwald, T. (2021) · Sensors · doi:10.3390/s21030821 · Cited by 54 · see citing works
Self-citationRogers, B., Giles, D., Draper, N., Mourot, L., & Gronwald, T. (2021) · Journal of Functional Morphology and Kinesiology · doi:10.3390/jfmk6020038 · Cited by 54 · see citing works
Self-citationMateo‐March, M., Moya‐Ramón, M., Javaloyes, A., Sánchez‐Muñoz, C., & Clemente‐Suárez, V. J. (2022) · European Journal of Sport Science · doi:10.1080/17461391.2022.2047228 · Cited by 54 · see citing works
An Analysis of the Effects of Noisy Electrocardiogram Signal on Heartbeat Detection Performance
Apandi, Z. F. M., Ikeura, R., Hayakawa, S., & Tsutsumi, S. (2020) · Bioengineering · doi:10.3390/bioengineering7020053 · Cited by 52 · see citing works
Rogers, B., Mourot, L., Doucende, G., & Gronwald, T. (2021) · Physiological Reports · doi:10.14814/phy2.14956 · Cited by 42 · see citing works
Self-citationRogers, B., Schaffarczyk, M., Clauß, M., Mourot, L., & Gronwald, T. (2022) · Sensors · doi:10.3390/s22052032 · Cited by 41 · see citing works
Self-citationEffect of Different ECG Leads on Estimated R–R Intervals and Heart Rate Variability Parameters
Jeyhani, V., Mäntysalo, M., Noponen, K., Seppänen, T., & Vehkaoja, A. T. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8857954 · Cited by 20 · see citing works
Im, S. I., Park, D. H., Kim, B., Cho, K. I., Kim, H. S., & Heo, J. H. (2018) · IJC Heart & Vasculature · doi:10.1016/j.ijcha.2018.05.002 · Cited by 15 · see citing works
Effects of R-R time series accuracy on heart rate variability indexes
Cassirame, J., Chevrolat, S., & Mourot, L. (2019) · Movement & Sport Sciences - Science & Motricité · doi:10.1051/sm/2019006 · Cited by 10 · see citing works
Flagged first, then by citations
1 of the 38 references couldn’t be matched to a known paper.
Cited by
Sorted by citations
Koerber, D., Khan, S., Shamsheri, T., Kirubarajan, A., & Mehta, S. (2023) · Journal of Racial and Ethnic Health Disparities · doi:10.1007/s40615-022-01446-9 · Cited by 91 · see citing works
A Systematic Review on the Use of Consumer-Based ECG Wearables on Cardiac Health Monitoring
Wang, R., Veera, S. C. M., Asan, O., & Liao, T. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3456028 · Cited by 42 · see citing works
Challenges and standardisation strategies for sensor-based data collection for digital phenotyping
Alam, N. B., Surani, M., Das, C. K., Giacco, D., Singh, S. P., & Jilka, S. (2025) · Communications Medicine · doi:10.1038/s43856-025-01013-3 · Cited by 41 · see citing works
Are Activity Wrist-Worn Devices Accurate for Determining Heart Rate during Intense Exercise?
Martín-Escudero, P., Cabanas, A. M., Dotor-Castilla, M. L., Galindo-Canales, M., Miguel‐Tobal, F., Fernández, C., Fuentes, M. E., & Giannetti, R. (2023) · Bioengineering · doi:10.3390/bioengineering10020254 · Cited by 40 · see citing works
Romagnoli, S., Ripanti, F., Morettini, M., Burattini, L., & Sbrollini, A. (2023) · Sensors · doi:10.3390/s23063350 · Cited by 37 · see citing works
Rehman, R. Z. U., Chatterjee, M., Manyakov, N. V., Daans, M., Jackson, A., O’Brisky, A., Telesky, T., Smets, S., Berghmans, P.-J., Yang, D., et al. (2024) · Sensors · doi:10.3390/s24216826 · Cited by 36 · see citing works
Validity and Efficacy of the Elite HRV Smartphone Application during Slow-Paced Breathing
Vondrasek, J. D., Riemann, B. L., Grosicki, G. J., & Flatt, A. A. (2023) · Sensors · doi:10.3390/s23239496 · Cited by 35 · see citing works
O’Grady, B., Lambe, R., Baldwin, M., Acheson, T., & Doherty, C. (2024) · Sensors · doi:10.3390/s24196220 · Cited by 31 · see citing works
Zheng, X. H., Liu, Z., Liu, J., Hu, C., Du, Y., Li, J., Pan, Z., & Ding, K. (2025) · ACS Applied Materials & Interfaces · doi:10.1021/acsami.4c22895 · Cited by 29 · see citing works
Occupant-centered indoor environmental quality management: Physiological response measuring methods
Kong, M., An, J., Jung, D., & Hong, T. (2023) · Building and Environment · doi:10.1016/j.buildenv.2023.110661 · Cited by 27 · see citing works
Schweizer, T., & Gilgen-Ammann, R. (2025) · JMIR Cardio · doi:10.2196/67110 · Cited by 25 · see citing works
Ali, A. E., Ney, R., van Berlo, Z. M. C., & Cesar, P. S. (2023) · IEEE Transactions on Visualization and Computer Graphics · doi:10.1109/tvcg.2023.3320228 · Cited by 23 · see citing works
Xie, H., Mao, X., & Wang, Z. (2024) · Frontiers in Physiology · doi:10.3389/fphys.2024.1451464 · Cited by 23 · see citing works
Validity of heart rate measurements in wrist-based monitors across skin tones during exercise
Hung, S. H., Serwa, K., Rosenthal, G., & Eng, J. J. (2025) · PLOS ONE · doi:10.1371/journal.pone.0318724 · Cited by 21 · see citing works
Awada, M., Becerik-Gerber, B., Lucas, G. M., & Roll, S. C. (2023) · Italian National Conference on Sensors · doi:10.3390/s23218694 · Cited by 17 · see citing works
Effects of Vibroacoustic Stimulation on Psychological, Physiological, and Cognitive Stress
Fooks, C., & Niebuhr, O. (2024) · Sensors · doi:10.3390/s24185924 · Cited by 21 · see citing works
Topalidis, P. I., Heib, D. P. J., Baron, S., Eigl, E., Hinterberger, A., & Schabus, M. (2023) · Sensors · doi:10.3390/s23052390 · Cited by 21 · see citing works
Coste, A., Millour, G., & Hausswirth, C. (2025) · Sensors · doi:10.3390/s25185745 · Cited by 20 · see citing works
Pilot turning behavior cognitive load analysis in simulated flight
Zhou, W., Yu, P.-P., Wu, L.-H., Cao, Y.-F., Zhou, Y., & Yuan, J. (2024) · Frontiers in Neuroscience · doi:10.3389/fnins.2024.1450416 · Cited by 19 · see citing works
From Pulses to Sleep Stages: Towards Optimized Sleep Classification Using Heart-Rate Variability
Topalidis, P. I., Baron, S., Heib, D. P. J., Eigl, E., Hinterberger, A., & Schabus, M. (2023) · Sensors · doi:10.3390/s23229077 · Cited by 19 · see citing works
Peak Running, Mechanical, and Physiological Demands of Professional Men’s Field Hockey Matches
Lin, L., Ji, X., Zhang, L., Weng, H., & Wang, X. (2023) · Journal of Human Kinetics · doi:10.5114/jhk/161551 · Cited by 16 · see citing works
Fleitas‐Paniagua, P. R., Marinari, G., Rasica, L., Rogers, B., & Murias, J. M. (2024) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000003412 · Cited by 16 · see citing works
Self-citationPotts, D., Broad, Z., Sehgal, T., Hartley, J. P., O'Neill, E. J., Jicol, C., Clarke, C., & Lutteroth, C. (2024) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3613904.3642611 · Cited by 17 · see citing works
Assessing Stress Level Scores Against Wearables‐Driven Physiological Measurements
Rosenbach, H., Itzkovitch, A., Gidron, Y., & Schönberg, T. (2025) · Stress and Health · doi:10.1002/smi.70125 · Cited by 16 · see citing works
Comparison of Slate Safety Wearable Device to Ingestible Pill and Wearable Heart Rate Monitor
Callihan, M., Cole, H. S., Stokley, H., Gunter, J., Clamp, K., Martin, A., & Doherty, H. (2023) · Sensors · doi:10.3390/s23020877 · Cited by 15 · see citing works
Kanniainen, M., Pukkila, T., Kuisma, J., Molkkari, M., Lajunen, K., & Räsänen, E. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1299104 · Cited by 14 · see citing works
Xu, R., Cao, S. C., Barnett‐Cowan, M., Bülbül, G. B., Irving, E. L., Niechwiej‐Szwedo, E., & Kearns, S. K. (2025) · MethodsX · doi:10.1016/j.mex.2025.103589 · Cited by 13 · see citing works
Howie, E. E., Harari, R., DIAS, R. D., Wigmore, S. J., Skipworth, R. J. E., & Yule, S. (2024) · Journal of Surgical Research · doi:10.1016/j.jss.2024.07.009 · Cited by 13 · see citing works
Hannon, J., O’Hagan, A. R., Lambe, R., O’Grady, B., & Doherty, C. (2025) · Sensors · doi:10.3390/s25144415 · Cited by 12 · see citing works
Ladriñán-Maestro, A., Sánchez-Infante, J., Martín-Vera, D., & Sánchez‐Sierra, A. (2024) · Frontiers in Physiology · doi:10.3389/fphys.2024.1457019 · Cited by 12 · see citing works
Yoshikawa, H., Adachi, Y., Baba, A., Takikawa, C., Yamaguchi, Y., Nakai, W., & Sudo, D. (2025) · PLOS mental health. · doi:10.1371/journal.pmen.0000240 · Cited by 12 · see citing works
Schwalm, L. C., Fohrmann, D., Schaffarczyk, M., HERRMANN, A., Gronwald, T., & Hollander, K. (2025) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0000000000003823 · Cited by 11 · see citing works
Self-citationMcLean, M. K., Weaver, R. G., Lane‐Cordova, A. D., Smith, M. T., Parker, H., Stone, B., McAninch, J., Matolak, D. W., Burkart, S. A., Chandrashekhar, M. V. S., et al. (2023) · Sensors · doi:10.3390/s23073429 · Cited by 11 · see citing works
Gronwald, T., Horn, L., Schaffarczyk, M., & Hoos, O. (2024) · European Journal of Sport Science · doi:10.1002/ejsc.12175 · Cited by 11 · see citing works
Self-citationCeugniez, M., Devanne, H., & Hermand, É. (2024) · JMIR mhealth and uhealth · doi:10.2196/54871 · Cited by 11 · see citing works
Messaoud, I. B., & Thamsuwan, O. (2025) · Computers · doi:10.3390/computers14020045 · Cited by 11 · see citing works
Yi, W., Zong, H., Antwi‐Afari, M. F., & Chan, A. P. C. (2025) · Building and Environment · doi:10.1016/j.buildenv.2025.112808 · Cited by 11 · see citing works
Developing a Fatigue Model for Construction Workers: An Interpretable Machine Learning Approach
Zong, H., Yi, W., Chan, A. P. C., Yang, H., Wu, P., & Xiao, B. (2025) · Journal of Management in Engineering · doi:10.1061/jmenea.meeng-6662 · Cited by 11 · see citing works
Moderate cardiac vagal tone is associated with more cooperation in children
Mastromatteo, L. Y., Girardi, P., Miller, J. G., & Scrimin, S. (2024) · International Journal of Psychophysiology · doi:10.1016/j.ijpsycho.2024.112371 · Cited by 11 · see citing works
Byrne, A., & Arnott, G. (2024) · Applied Animal Behaviour Science · doi:10.1016/j.applanim.2024.106403 · Cited by 10 · see citing works
Martínez-Martín, V., Verdejo-Herrero, J., Rey, R. R.-D., García‐González, J., Requena, M., & Alarcón, R. (2024) · Healthcare · doi:10.3390/healthcare12232424 · Cited by 10 · see citing works
Loro, F. L., Martins, R., Ferreira, J. B., de Araújo, C. L. P., Prade, L. R., Both, C. B., Nobre, J. C., Monteiro, M. B., & Lago, P. D. (2024) · JMIR Biomedical Engineering · doi:10.2196/57373 · Cited by 10 · see citing works
Pukkila, T., Molkkari, M., Hernesniemi, J. A., Kanniainen, M., & Räsänen, E. (2025) · Heart Rhythm O2 · doi:10.1016/j.hroo.2025.01.014 · Cited by 10 · see citing works
Bridging childhood to adulthood: the impact of early life stress on acute stress responses
Huang, Z., Bai, H., Yang, Z., Zhang, J., Wang, P., Wang, X., & Zhang, L. (2024) · Frontiers in Psychiatry · doi:10.3389/fpsyt.2024.1391653 · Cited by 10 · see citing works
Chapman, C. L., Schafer, E. A., Potter, A. W., Lavoie, E. M., Roberts, B. M., Castellani, J. W., Friedl, K. E., & Looney, D. P. (2024) · BMJ Military Health · doi:10.1136/military-2024-002687 · Cited by 10 · see citing works
Im, B., Keum, J., Kim, T., Lee, K.-I., & Koo, K.-I. (2023) · Applied Sciences · doi:10.3390/app14010004 · Cited by 9 · see citing works
Liu, Y., BARKER, A. R., Li, M., Gu, Y., & Williams, C. A. (2025) · Journal of Sports Sciences · doi:10.1080/02640414.2025.2491163 · Cited by 9 · see citing works
RetroSketch: A Retrospective Method for Measuring Emotions and Presence in Virtual Reality
Potts, D., Gada, M., Gupta, A., Goel, K., Krzok, K. P., Pate, G., Hartley, J. P., Weston-Arnold, M., Aylott, J., Clarke, C., et al. (2025) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/3706598.3713957 · Cited by 9 · see citing works
Inspiratory muscle warm up improves 400 m performance in elite male runners
Yılmaz, Y. A., Tosun, M. İ., Demirkan, E., Can, S., Özkan, A., Arıcı, M., Kutlu, M., Ayrancı, M., Marković, M., Arıcı, İ. E., et al. (2025) · Scientific Reports · doi:10.1038/s41598-025-14797-0 · Cited by 9 · see citing works
Touchless Heart Rate Monitoring from an Unmanned Aerial Vehicle Using Videoplethysmography
Pająk, A., Przybyło, J., & Augustyniak, P. (2023) · Sensors · doi:10.3390/s23167297 · Cited by 9 · see citing works
Maza, A., Goizueta, S., & Lloréns, R. (2024) · IEEE Sensors Journal · doi:10.1109/jsen.2024.3359301 · Cited by 9 · see citing works
Breunissen, E. H. W., Groenveld, T. D., Garms, L. M., Bonnes, J., van Goor, H., & Damman, P. (2024) · Open Heart · doi:10.1136/openhrt-2024-002628 · Cited by 9 · see citing works
Brandl, A., Engel, R., Egner, C., Schleip, R., & Schubert, C. (2024) · Journal of Medical Case Reports · doi:10.1186/s13256-024-04935-z · Cited by 9 · see citing works
Fennell, C. R. J., Mauger, A. R., & Hopker, J. (2023) · European Journal of Applied Physiology · doi:10.1007/s00421-023-05373-3 · Cited by 8 · see citing works
Wong, S. S. S., Liu, T. W., & Ng, S. S. M. (2024) · Heliyon · doi:10.1016/j.heliyon.2024.e35883 · Cited by 8 · see citing works
Well-being as a performance pillar: a holistic approach for monitoring tennis players
Michel, M.-F., Girard, O., Guillard, V., & Brechbuhl, C. (2023) · Frontiers in Sports and Active Living · doi:10.3389/fspor.2023.1259821 · Cited by 8 · see citing works
Jallouli, S., Jallouli, D., Damak, M., Sakka, S., Ghroubi, S., Mhiri, C., Driss, T., de Marco, G., Ayadi, F. M., & Hammouda, O. (2024) · Metabolic Brain Disease · doi:10.1007/s11011-024-01428-2 · Cited by 8 · see citing works
Impact of COVID-19 on heart rate variability in post-COVID individuals compared to a control group
Santos‐de‐Araújo, A. D., Dibai, D. B., Marinho, R. S., Dourado, I. M., de Almeida, L. V., dos Santos, S. D. S., Phillips, S. A., & Borghi‐Silva, A. (2024) · Scientific Reports · doi:10.1038/s41598-024-82411-w · Cited by 8 · see citing works
Activity Type Effects Signal Quality in Electrocardiogram Devices
Lindsey, B., Snyder, S. J., Zhou, Y., Shim, J. K., Hahn, J., Evans, W., & Martin, J. R. (2025) · Sensors · doi:10.3390/s25165186 · Cited by 7 · see citing works
Gellisch, M., & Burr, B. (2025) · Frontiers in Digital Health · doi:10.3389/fdgth.2025.1670464 · Cited by 7 · see citing works
Validity of the Polar H10 for Continuous Measures of Heart Rate and Heart Rate Synchrony Analysis
Chung, V., Chopin, L., Karadayi, J., & Grèzes, J. (2026) · Sensors · doi:10.3390/s26030855 · Cited by 7 · see citing works
Ammann, L., & Chmura, P. (2023) · Frontiers in Physiology · doi:10.3389/fphys.2023.1212573 · Cited by 7 · see citing works
Akgül, M. Ş., Tohănean, D. I., Akçay, N., Păun, L. I., Çetin, T., Uysal, H. Ş., & Alexe, C. I. (2024) · Applied Sciences · doi:10.3390/app142311269 · Cited by 7 · see citing works
Natural light control to improve awakening quality
Wang, X., Gu, Y., Yuan, J., & Matsushita, D. (2025) · Building and Environment · doi:10.1016/j.buildenv.2025.112733 · Cited by 7 · see citing works
Arney, D. E., Zhang, Y., Kennedy-Metz, L. R., DIAS, R. D., Goldman, J. M., & Zenati, M. A. (2023) · Sensors · doi:10.3390/s23083890 · Cited by 7 · see citing works
Fuentes-Barría, H., Aguilera-Eguía, R. A., Maureira-Sánchez, J., Alarcón-Rivera, M., Garrido-Osorio, V., López-Soto, O. P., Aristizábal-Hoyos, J. A., Angarita, L., Rojas-Gómez, D. M., Bermúdez, V., et al. (2025) · Journal of Functional Morphology and Kinesiology · doi:10.3390/jfmk10020195 · Cited by 6 · see citing works
Virtual Reality to Reduce Preprocedural Anxiety During Invasive Coronary Angiography
Groenveld, T. D., Breunissen, E. H. W., Bonnes, J., Garms, L. M., Scholten, N., Cetinyurek‐Yavuz, A., van Nunen, L. X., van Wely, M. X., Camaro, C., ten Cate, T. J. F., et al. (2025) · JACC Advances · doi:10.1016/j.jacadv.2025.101976 · Cited by 6 · see citing works
Su, S., Stark, J., Fidler, M., Hössinger, R., & Susilo, Y. O. (2025) · Travel Behaviour and Society · doi:10.1016/j.tbs.2025.101107 · Cited by 6 · see citing works
Ēvelis, K., Koļesovs, A., Rostoka, E., & Plakane, L. (2025) · Scientific Reports · doi:10.1038/s41598-025-22209-6 · Cited by 6 · see citing works
Physical, physiological demands and movement profiles of professional men’s field hockey games
Lin, L., Ji, X., Zhang, L., Weng, H., & Wang, X. (2023) · Kinesiology · doi:10.26582/k.55.1.8 · Cited by 6 · see citing works
Psychophysiological Data Harmonization for the Sustainability of Outdoor Activities
Papale, O., Festino, E., Condello, G., Rocco, F. D., Maio, M. D., Cortis, C., & Fusco, A. (2023) · Sustainability · doi:10.3390/su152215838 · Cited by 6 · see citing works
Reliability and Validity of Kubios HRV Smart Phone Application as Measures of Heart Rate Variability
Gabani, V., & Murugan, S. (2024) · International Journal of Current Research and Review · doi:10.31782/ijcrr.2024.16602 · Cited by 6 · see citing works
The Real-Time Image Sequences-Based Stress Assessment Vision System for Mental Health
Khomidov, M., Lee, D., Kim, C., & Lee, J.-H. (2024) · Electronics · doi:10.3390/electronics13112180 · Cited by 6 · see citing works
Evaluation of a passive back-support exoskeleton during in-bed patient handling tasks
Zheng, L., Alluri, C. S. V., Hawke, A. L., & Hwang, J. (2024) · International Journal of Occupational Safety and Ergonomics · doi:10.1080/10803548.2024.2383077 · Cited by 6 · see citing works
Davis-Wilson, H. C., Hegarty‐Craver, M., Gaur, P., Boyce, M., Holt, J. R., Preble, E. A., Eckhoff, R., Li, L., Walls, H. J., Dausch, D. E., et al. (2024) · JMIR Formative Research · doi:10.2196/53645 · Cited by 6 · see citing works
Recent advances in the tools and techniques for AI-aided diagnosis of atrial fibrillation
Islam, M. S., Islam, M. R., Sanjid-E-Elahi, Abedin, M. A., Dökeroğlu, T., & Rahman, M. (2025) · Biophysics Reviews · doi:10.1063/5.0217416 · Cited by 6 · see citing works
Electrodermal activity as an indicator of challenging road sections
Sudrychova, I., Tomaszek, L., Šaloun, P., Miklošíková, M., & Jánošík, L. (2025) · Journal of Loss Prevention in the Process Industries · doi:10.1016/j.jlp.2025.105557 · Cited by 6 · see citing works
Rogers, B., Fleitas‐Paniagua, P. R., Trpcic, M., Zagatto, A. M., & Murias, J. M. (2025) · European Journal of Applied Physiology · doi:10.1007/s00421-025-05716-2 · Cited by 6 · see citing works
Self-citationEffects of Physiological Load on Kinematic Variables Related to Tennis Serve Performance
Bilić, Z., Očić, M., Dukarić, V., Petrinović, L., & Barbaros, P. (2025) · Sports · doi:10.3390/sports13070197 · Cited by 6 · see citing works
Krause‐Parello, C. A., Friedmann, E., Taber, D., Zhu, H., Quintero, A., & Yount, R. A. (2025) · Behavioral Sciences · doi:10.3390/bs15091180 · Cited by 6 · see citing works
Development of a Real-Time Stress Detection System for Older Adults with Heart Rate Data
Vito, P. D. C. S., Shakeri, G., Ross, J., Yang, X., & Brewster, S. A. (2023) · International Conference on PErvasive Technologies Related to Assistive Environments (PETRA) · doi:10.1145/3594806.3594817 · Cited by 6 · see citing works
Guidelines for conducting biofeedback-enhanced QoE studies in mulsemedia-enhanced virtual reality
da Silveira, A. C., Raisamo, R., Spyridonis, F., Covaci, A., Ghinea, G., & Santos, C. A. S. (2023) · Brazilian Symposium on Multimedia and the Web · doi:10.1145/3617023.3617029 · Cited by 6 · see citing works
Thrower, A., Gibbs, B. B., Alansare, A. B., Sherman, S. A., & Davis, K. K. (2023) · PLoS ONE · doi:10.1371/journal.pone.0294945 · Cited by 6 · see citing works
Fuentes-Barría, H., Aguilera-Eguía, R. A., Polevoy, G. G., Maureira-Sánchez, J., & Angarita, L. (2024) · Journal of Sport and Health Research · doi:10.58727/jshr.102273 · Cited by 6 · see citing works
Vermunicht, P., Buyck, C., Naessens, S., Hens, W., Verberckt, C., Craenenbroeck, E. M. V., Laukens, K., Desteghe, L., & Heidbüchel, H. (2025) · European Heart Journal - Digital Health · doi:10.1093/ehjdh/ztaf084 · Cited by 5 · see citing works
Rogers, B., Murias, J. M., & Fleitas‐Paniagua, P. R. (2025) · European Journal of Applied Physiology · doi:10.1007/s00421-025-06037-0 · Cited by 5 · see citing works
Self-citationCampo, C., Cellucci, F., Treggiari, G., Spedicato, G. A., Lasaponara, S., & Conversi, D. (2026) · Frontiers in Psychology · doi:10.3389/fpsyg.2026.1755043 · Cited by 5 · see citing works
Zhang, Y., Hui, Z., Qi, W., Zhang, J., Wang, M., & Zhu, D. (2024) · PLoS ONE · doi:10.1371/journal.pone.0303517 · Cited by 5 · see citing works
Cassirame, J., Eustache, E., Garbellotto, L., Chevrolat, S., Gimenez, P., & Leprêtre, P. (2024) · European Journal of Applied Physiology · doi:10.1007/s00421-024-05614-z · Cited by 5 · see citing works
Gielen, J., Smets, C., Vidts, N., Schots, S., Stessens, L., Jaspers, A., Romain, M., & Jean‐Marie, A. (2024) · European Journal of Sport Science · doi:10.1002/ejsc.12185 · Cited by 5 · see citing works
Sensory Interventions to Improve Exercise Tolerance
Bird, J. M., Guérin, S. M. R., & Cheval, B. (2025) · Springer eBooks · doi:10.1007/978-3-031-90034-1_8 · Cited by 5 · see citing works
Vermunicht, P., Makayed, K., Buyck, C., Knaepen, L., Giraldo, J. S. P., Naessens, S., Hens, W., Craenenbroeck, E. M. V., Laukens, K., Desteghe, L., et al. (2025) · Digital Health · doi:10.1177/20552076251337598 · Cited by 5 · see citing works
Experience-based risk taking is primarily shaped by prior learning rather than by decision-making
Erdman, A., Gouwy, A., Sananes, G., Salman, M., Eizenstien, L., Eliav, S., & Eldar, E. (2025) · Nature Communications · doi:10.1038/s41467-025-61609-0 · Cited by 5 · see citing works
Diurnal variation of heart rate variability in individuals with spinal cord injury
Saengsuwan, J., Ruangsuphaphichat, A., Brockmann, L., Sirasaporn, P., Manimmanakorn, N., & Hunt, K. J. (2024) · BioMedical Engineering OnLine · doi:10.1186/s12938-024-01256-6 · Cited by 5 · see citing works
Rho, G., Rienzo, F. D., Marinai, C., Giannetti, F., Arcarisi, L., Bufano, P., Zanoletti, M., Righetti, F., Vallati, C., Laurino, M., et al. (2024) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-49062-0_3 · Cited by 5 · see citing works
Fuentes-Barría, H., Aguilera-Eguía, R. A., & Polevoy, G. G. (2024) · Retos · doi:10.47197/retos.v55.105419 · Cited by 5 · see citing works
Moore, A. R., & Butler, B. A. S. (2024) · Research Quarterly for Exercise and Sport · doi:10.1080/02701367.2024.2356893 · Cited by 4 · see citing works
Dergaa, I., Ghouili, H., Clark, C. C. T., Taheri, M., Fessi, M. S., Souissi, N., Guelmami, N., Aissa, M. B., Saad, H. B., Weiss, K., et al. (2024) · Sports Medicine and Health Science · doi:10.1016/j.smhs.2024.11.004 · Cited by 4 · see citing works
Developing a Fatigue Detection Model for Hospital Nurses Using HRV Measures and Machine Learning
Hafiz, W. S., Puspasari, M. A., Fitriani, D. Y., Hanowski, R. J., Syaifullah, D. H., & Arista, S. A. (2025) · Safety · doi:10.3390/safety11020048 · Cited by 4 · see citing works
Kamiya, Y., Saita, K., Kaneko, F., Li, J., & Okamura, H. (2025) · Physiology & Behavior · doi:10.1016/j.physbeh.2025.114969 · Cited by 4 · see citing works
- Authors
- Schaffarczyk, M., Rogers, B., Reer, R., & Gronwald, T.
- Affiliations
- Universität Hamburg (DE)
- University of Central Florida (US)
- MSH Medical School Hamburg – University of Applied Sciences and Medical University (DE)
- Corresponding author
- Schaffarczyk, Marcelle
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2022-08-30
- Volume
- 22
- Issue
- 17
- Pages
- 6536
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC9459793
- OpenAlex
- W4293715556
- Semantic Scholar
- d503adc11de24a90cadc9e05705b05b448ea1071
- Cited by
- 349
- References
- 38
- Citations by year
- 2026: 132 · 2025: 108 · 2024: 75 · 2023: 32 · 2022: 2
- FWCI
- 37.75
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Complementary and alternative medicine › Cardiovascular and exercise physiology
- Keywords
- heart rate variability, electrocardiogram comparison, RR interval, incremental exercise, measurement bias, limits of agreement, recreational athletes, sex differences
- MeSH terms
- Bicycling, Electrocardiography, Exercise Test, Female, Heart Rate, Humans, Male, Exercise
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review