Use of oscillometric devices in atrial fibrillation: a comparison of three devices and invasive blood pressure measurement
Open accessThe DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
BACKGROUND: The use of automated (oscillometric) blood pressure (BP) devices is not validated in atrial fibrillation (AF) patients. OBJECTIVES: To assess the reliability of three oscillometric BP devices, and the agreement with invasive arterial blood pressure(IBP) in AF patients. METHODS: 48 AF patients with randomized sequences of 10 consecutive BP measurements with two pairs of devices: (1) OmronR7™(wrist) and OmronHEM907™(arm); (2) OmronR7™ and Microlife WatchBPhome(arm). Reliability and agreement of each device were assessed by the intra-class correlation coefficient (ICC) for the continuous BP measurements and Bland & Altman methodology, respectively. In 10 additional AF patients, 10 consecutive measurements with IBP and OmronHEM907™, and IBP and Microlife WatchBPhome were performed. RESULTS: The OmronR7™ was not able to obtain any BP Readings. Arm devices presented better ICC for systolicBP(SBP) than for diastolicBP(DBP) (Omron HEM907™:0.94 [0.90; 0.97] vs. 0.77 [0.67; 0.89]; Microlife WatchBPhome:0.92 [0.88; 0.96] vs.0.79 [0.69; 0.89]).The correlation coefficient between Microlife WatchBPhome and IBP computed using the average of repeated measurements from two to ten measurements improved up to the third and remained stable afterwards. The agreement between IBP and SBP, and IBP and DBP, was moderate as illustrated by a wide limit of agreement [-24; 26](SBP) and [-15;17](DBP) for Microlife WatchBPHome, respectively and [-30; 13](SBP) and [-7; 15](DBP) for OmronHEM907. CONCLUSIONS: BP measurement using the two arm oscillometric devices achieved a high reliability for SBP. The agreement between IBP and arm devices was low but using the average of three consecutive measurements improved the results substantially.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Parati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e328308da66 · Cited by 837 · cited 1×
“Hence, both devices seem adequate in that respect [2].”
BackgroundSelf-citationAccuracy of Cuff-Measured Blood Pressure
Picone, D. S., Schultz, M. G., Otahal, P., Aakhus, S., Al-Jumaily, A. M., Black, J. A., Bos, W. J., Chambers, J. B., Chen, C.-H., Cheng, H.-M., et al. (2017) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2017.05.064 · Cited by 277
Evaluation of a clinical test. I: Assessment of reliability
Kha, K. S., & Chien, P. F. W. (2001) · BJOG: An International Journal of Obstetrics & Gynaecology · doi:10.1111/j.1471-0528.2001.00150.x · Cited by 135
Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis
Stergiou, G. S., Kollias, A., Destounis, A., & Tzamouranis, D. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32835850d7 · Cited by 120
Wiesel, J., Abraham, S., & Messineo, F. C. (2013) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2013.01.331 · Cited by 58 · cited 2×
“Three commonly used BP measuring devices (two arms, and one wrist) were used: the Omron HEM907 (OHEM) (a professional office BP monitor device) is the arm device used in our hospital for BP monitoring in ambulatory and hospitalized patients and two devices developed for self-home BP monitoring by patients: the Microli…”
BackgroundMethod
Cited by
Mancia, G., Kreutz, R., Brunström, M., Burnier, M., Grassi, G., Januszewicz, A., Muiesan, M. L., Tsioufis, K., Agabiti-Rosei, E., Algharably, E. A. E., et al. (2023) · Journal of Hypertension · doi:10.1097/hjh.0000000000003480 · Cited by 3,473
(2022) · PubMed Central · doi:10.6515/acs.202205_38(3).20220321a · Cited by 117
Stergiou, G. S., Kyriakoulis, K. G., Stambolliu, E., Destounis, A., Karpettas, N., Kalogeropoulos, P., & Kollias, A. (2019) · Journal of Hypertension · doi:10.1097/hjh.0000000000002201 · Cited by 67
“Fourteen articles including 12 noninvasive validation studies (781 patients; 5 home, 2 ambulatory, 4 office devices) [25–35] and four invasive ones (179 patients; 2 office, 2 home devices) [16,36,37] were eligible and included in the meta-analysis.”
BackgroundClark, C. E., McDonagh, S. T. J., & McManus, R. J. (2019) · Journal of Human Hypertension · doi:10.1038/s41371-018-0153-z · Cited by 49
Metadata
- Authors
- Halfon, M., Wuerzner, G., Marques-Vidal, P., Taffe, P., Vaucher, J., Waeber, B., Liaudet, L., Ltaief, Z., Popov, M., & Waeber, G.
- Venue
- Blood Pressure
- Publisher
- Informa UK Limited
- Published
- 2017-09-28
- Volume
- 27
- Issue
- 1
- Pages
- 48-55
- ISSN
- 0803-7051 (ISSN-L)
- 1651-1999
- OpenAlex
- W2760793119
- Cited by
- 25
- References
- 5
- FWCI
- 1.09
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Use of oscillometric devices in atrial fibrillation: a comparison of three devices and invasive blood pressure measurement. https://app.cochat.ai/papers/10.1080/08037051.2017.1383852
- The paper
- Halfon, M., Wuerzner, G., Marques-Vidal, P., Taffe, P., Vaucher, J., Waeber, B., Liaudet, L., Ltaief, Z., Popov, M., & Waeber, G. (2018). Use of oscillometric devices in atrial fibrillation: a comparison of three devices and invasive blood pressure measurement. Blood Pressure. https://doi.org/10.1080/08037051.2017.1383852
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.