Developing a quality control protocol for diffusion imaging on a clinical MRI system
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
This work describes the development of a quality control protocol, which can be implemented to assess the accuracy, precision and reproducibility of the apparent diffusion coefficient (ADC) measurement on a clinical magnetic resonance imaging (MRI) system. The precision and accuracy of the ADC measurement are analysed with regard to MRI system noise, signal reproducibility and differences between nominal and effective b values. Two aqueous test-solutions of CuSO4 and sucrose are prepared for the quality control protocol. ADC measurement with the CuSO4 solution is more sensitive to differences between nominal and effective b values, on account of the solution's high ADC. ADC measurement with the sucrose solution is more sensitive to signal reproducibility due to the solution's low baseline signal intensity. The ADC of the test-solutions is measured on an MRI system at our centre with a sequence used for clinical studies using diffusion imaging. Two parameters, Q and R, are defined for the analysis of the quality control ADC values. The Q parameter is the ratio of the standard deviation of the quality control mean ADC values over time to the optimal standard deviation, as derived from the effect of thermal noise on the ADC measurement uncertainty. Analysis with the Q parameter indicates that signal reproducibility errors contribute to ADC variations on our MRI system when imaging with high b values (b > 500 mm s(-2)), whereas differences between nominal and effective b values have a greater impact on the ADC measurement when imaging with low b values (b < 500 mm s(-2)). The R parameter is defined as the ratio of the directional variation of the ADC quality control values to the uncertainty of the ADC measurement. Analysis with the R parameter shows that the effect of directional variation of the ADC measurement on our MRI system is more pronounced when imaging with low b values. The quality control protocol identified a systematic error, which introduced a small system-induced anisotropy in the ADC measurement. This error is currently taken into account in the analysis of clinical studies employing the diffusion imaging sequence used in this quality control protocol.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Diffusion tensor MR imaging of the human brain
Pierpaoli, C., Jezzard, P., Basser, P. J., Barnett, A., & Di Chiro, G. (1996) · Radiology · doi:10.1148/radiology.201.3.8939209 · Cited by 2,664
Cited by
Radiomics: the process and the challenges
Kumar, V., Gu, Y., Basu, S., Berglund, A., Eschrich, S. A., Schabath, M. B., Forster, K., Aerts, H. J. W. L., Dekker, A., Fenstermacher, D., et al. (2012) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2012.06.010 · Cited by 2,258
“DWI has been proposed as a cancer imaging biomarker, and there are efforts to develop quality control protocols [18,19].”
BackgroundDiffusion-Weighted Magnetic Resonance Imaging as a Cancer Biomarker: Consensus and Recommendations
Padhani, A. R., Liu, G., Mu-Koh, D., Chenevert, T. L., Thoeny, H. C., Takahara, T., Dzik-Jurasz, A., Ross, B. D., Van Cauteren, M., Collins, D., et al. (2009) · Neoplasia · doi:10.1593/neo.81328 · Cited by 1,989
On behalf of the EUSOBI international Breast Diffusion-Weighted Imaging working group, Baltzer, P., Mann, R. M., Iima, M., Sigmund, E. E., Clauser, P., Gilbert, F. J., Martincich, L., Partridge, S. C., Patterson, A., et al. (2020) · European Radiology · doi:10.1007/s00330-019-06510-3 · Cited by 437
Practical Aspects of Assessing Tumors Using Clinical Diffusion-weighted Imaging in the Body
KOH, D.-M., TAKAHARA, T., IMAI, Y., & COLLINS, D. J. (2007) · Magnetic Resonance in Medical Sciences · doi:10.2463/mrms.6.211 · Cited by 211
“This allows the precision and accuracy of the ADC measurement to be determined in relation to the noise of the MR imaging system, signal reproducibility, and diŠerences between nominal and eŠective bvalues.(23) In addition, use of these phantoms will allow early recognition of geometric distortion and other eddy curre…”
MethodMalyarenko, D., Galbán, C. J., Londy, F. J., Meyer, C. R., Johnson, T. D., Rehemtulla, A., Ross, B. D., & Chenevert, T. L. (2013) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.23825 · Cited by 197
Hagiwara, A., Fujita, S., Ohno, Y., & Aoki, S. (2020) · Investigative Radiology · doi:10.1097/rli.0000000000000666 · Cited by 183
Quantitative magnetic resonance imaging phantoms: A review and the need for a system phantom
Keenan, K. E., Ainslie, M., Barker, A. J., Boss, M. A., Cecil, K. M., Charles, C., Chenevert, T. L., Clarke, L., Evelhoch, J. L., Finn, P., et al. (2018) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26982 · Cited by 170
“, using two aqueous test solutions of copper sulfate (CuSO4) and sucrose, developed a quality control protocol to assess the accuracy, precision, and reproducibility of ADC measurement on a clinical MRI system (65).”
MethodKoh, D.-M., Blackledge, M., Collins, D. J., Padhani, A. R., Wallace, T., Wilton, B., Taylor, N. J., Stirling, J. J., Sinha, R., Walicke, P., et al. (2009) · European Radiology · doi:10.1007/s00330-009-1469-4 · Cited by 163
“The mean ADC value over the entire sucrose phantom measured at the two study sites at 22°C [18] were 0.”
MethodSelf-citationDudeck, O., Zeile, M., Pink, D., Pech, M., Tunn, P., Reichardt, P., Ludwig, W., & Hamm, B. (2008) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.21358 · Cited by 159
“These results are in good agreement with previously reported values (11).”
MethodResultDiffusion-MRI in neurodegenerative disorders
Goveas, J., O’Dwyer, L., Mascalchi, M., Cosottini, M., Diciotti, S., De Santis, S., Passamonti, L., Tessa, C., Toschi, N., & Giannelli, M. (2015) · Magnetic Resonance Imaging · doi:10.1016/j.mri.2015.04.006 · Cited by 104
Fieremans, E., De Deene, Y., Delputte, S., Özdemir, M. S., Achten, E., & Lemahieu, I. (2008) · Physics in Medicine and Biology · doi:10.1088/0031-9155/53/19/009 · Cited by 64
Metadata
- Authors
- Delakis, I., Moore, E. M., Leach, M. O., & Wilde, J. P. D.
- Venue
- Physics in Medicine and Biology
- Publisher
- IOP Publishing
- Published
- 2004-03-23
- Volume
- 49
- Issue
- 8
- Pages
- 1409-1422
- ISSN
- 0031-9155 (ISSN-L)
- 1361-6560
- OpenAlex
- W2145103622
- Cited by
- 65
- References
- 1
- FWCI
- 1.34
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.