Oxygen Effects on Glucose Measurements with a Reference Analyzer and Three Handheld Meters

Tang, Zuping, Louie, Richard F., Payes, Marta, Chang, Kai-Chiao J., Kost, Gerald J. · 2000 · Diabetes Technology & Therapeutics
Cited by 63

The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.

Abstract

Oxygen may affect glucose meter and reference analyzer measurements. We evaluated the effects of changes in blood oxygen tension (Po2) on Accu-Chek Comfort Curve (Roche Diagnostics, Indianapolis, IN),™ Precision G,™ (Abbott Laboratories, Bedford, MA) and One Touch II® (Lifescan, Milpitas, CA) glucose meter measurements, and on Yellow Springs Instruments (YSI) (Yellow Springs, OH) reference analyzer measurements. Venous blood drawn from healthy volunteers was adjusted to three glucose levels of 80, 200, and 400 mg/dL, each tonometered with six different Po2 levels (40, 80, 160, 240, 320, and 400 torr). To quantitate oxygen effects on reference analyzer measurements, glucose differences between test sample (Po changed) and control (Po 80 torr) were calculated (YSItest-YSIcontrol). The threshold for determination of oxygen effects was ± 2 SD, where 2 SD was from YSI between-day precision evaluated with aqueous controls. To quantitate oxygen effects on glucose meter measurements, paired-differences between glucose meter and reference analyzer measurements were calculated (Meter-YSI). Error tolerances for determination of oxygen effects were (a) within ± 15 mg/dL of reference measurement at glucose levels ≤100 mg/dL, and (b) within ± 15% of reference measurement at glucose levels >100 mg/dL. All differences for the reference analyzer measurements were within ± 2 SD ranges. Increases in Po2 lowered Precision G measurements. Differences exceeding the error tolerances ( ± 15 mg/dL or ± 15%) were mainly at Po2 levels >100 torr. Glucose differences for the Accu-Chek Comfort Curve and One Touch II were within the error tolerances. Changes in Po2 do not affect glucose measurements with the YSI reference analyzer, Accu-Chek Comfort Curve, and One Touch meter systems. High blood Po2 levels (>100 torr) significantly decrease Precision G measurements. We recommend that health care professionals understand oxygen effects and use oxygen-insensitive meter systems for point-of-care glucose testing when patients are on oxygen ventilation or critically ill.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1089/15209150050194215

Cites

1 of 1 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 63

Metadata

Authors
Tang, Z., Louie, R. F., Payes, M., Chang, K.-C. J., & Kost, G. J.
Venue
Diabetes Technology & Therapeutics
Published
2000-10
Volume
2
Issue
3
Pages
349-362
ISSN
  • 1520-9156 (ISSN-L)
  • 1557-8593
OpenAlex
W2076721326
Cited by
63
References
1
FWCI
0.76
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.