A Multicenter Study of the Accuracy of the One Touch® Ultra® Home Glucose Meter in Children with Type 1 Diabetes
The DOI matches its Crossref record; 14 of the 14 references it cites were found in a registry or an index.
Abstract
Data are not readily available on the accuracy of one of the most commonly used home blood glucose meters, the One Touch Ultra (LifeScan, Milpitas, California). The purpose of this report is to provide information on the accuracy of this home glucose meter in children with type 1 diabetes. During a 24-h clinical research center stay, the accuracy of the Ultra meter was assessed in 91 children, 3-17 years old, with type 1 diabetes by comparing the Ultra glucose values with concurrent reference serum glucose values measured in a central laboratory. The Pearson correlation between the 2,068 paired Ultra and reference values was 0.97, with the median relative absolute difference being 6%. Ninety-four percent of all Ultra values (96% of venous and 84% of capillary samples) met the proposed International Organisation for Standardisation (ISO) standard for instruments used for self-monitoring of glucose when compared with venous reference values. Ninety-nine percent of values were in zones A + B of the Modified Error Grid. A high degree of accuracy was seen across the full range of glucose values. For 353 data points during an insulin-induced hypoglycemia test, the Ultra meter was found to have accuracy that was comparable to concurrently used benchmark instruments (Beckman, YSI, or i-STAT); 95% and 96% of readings from the Ultra meter and the benchmark instruments met the proposed ISO criteria, respectively. These results confirm that the One Touch Ultra meter provides accurate glucose measurements for both hypoglycemia and hyperglycemia in children with type 1 diabetes.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“and provides some margin for the limited accuracy of glucose monitoring devices at low plasma glucose concentrations [5].”
From Preventing hypoglycaemia: what is the appropriate glucose alert value? (2009) · cites this paper as [5]
In the citing paper
“and within 15% of the reference value only 70% of the time [5].”
From Preventing hypoglycaemia: what is the appropriate glucose alert value? (2009) · cites this paper as [5]
In the citing paper
“at glucose levels ≤3.9 mmol/l (70 mg/dl), monitor values were within 10% of the reference blood glucose concentration only 55% of the time”
From Preventing hypoglycaemia: what is the appropriate glucose alert value? (2009) · cites this paper as [5]
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Defining the Relationship Between Plasma Glucose and HbA1c
Rohlfing, C. L., Wiedmeyer, H.-M., Little, R. R., England, J. D., Tennill, A., & Goldstein, D. E. (2002) · Diabetes Care · doi:10.2337/diacare.25.2.275 · Cited by 987
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
(2001) · The Journal of Pediatrics · doi:10.1067/mpd.2001.118887 · Cited by 460
Predictors of glycemic control and short-term adverse outcomes in youth with type 1 diabetes
Levine, B.-S., Anderson, B. J., Butler, D. A., Antisdel, J. E., Brackett, J., & Laffel, L. M. B. (2001) · The Journal of Pediatrics · doi:10.1067/mpd.2001.116283 · Cited by 279
Rosilio, M., Cotton, J.-B., Wieliczko, M.-C., Gendrault, B., Carel, J.-C., Couvaras, O., Ser, N., Gillet, P., Soskin, S., Garandeau, P., et al. (1998) · Diabetes Care · doi:10.2337/diacare.21.7.1146 · Cited by 224
Kuwa, K., Nakayama, T., Hoshino, T., & Tominaga, M. (2001) · Clinica Chimica Acta · doi:10.1016/s0009-8981(01)00426-0 · Cited by 158 · cited 1×
“16 Our comparisons of venous and capillary samples were done at random times throughout the day, so that some were fasting and others postprandial.”
Ellison, J. M., Stegmann, J. M., Colner, S. L., Michael, R. H., Sharma, M. K., Ervin, K. R., & Horwitz, D. L. (2002) · Diabetes Care · doi:10.2337/diacare.25.6.961 · Cited by 133 · cited 1×
“13,14 Previous studies have shown capillary glucose levels to be higher than venous glucose levels after a liquid glucose load, 15–17 although values tend to be similar in the fasting state.”
Dorchy, H., Roggemans, M.-P., & Willems, D. (1997) · Diabetes Care · doi:10.2337/diacare.20.1.2 · Cited by 125
Passey, R. B., Gillum, R. L., Fuller, J. B., Urry, F. M., & Giles, M. L. (1977) · Clinical Chemistry · doi:10.1093/clinchem/23.1.131 · Cited by 114 · cited 1×
“Reference glucose levels were determined using a hexokinase enzymatic method 7,8 (as used in the Diabetes Control and Complications Trial).”
Jungheim, K., & Koschinsky, T. (2002) · Diabetes Care · doi:10.2337/diacare.25.6.956 · Cited by 114
Differences between capillary and venous blood glucose during oral glucose tolerance tests
Larsson-Cohn, U. (1976) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.3109/00365517609081941 · Cited by 65 · cited 1×
“13,14 Previous studies have shown capillary glucose levels to be higher than venous glucose levels after a liquid glucose load, 15–17 although values tend to be similar in the fasting state.”
Capillary-venous differences in blood glucose values during the oral glucose tolerance test.
Eriksson, K. F., Fex, G. A., & Trell, E. (1983) · Clinical Chemistry · doi:10.1093/clinchem/29.5.993a · Cited by 23
Cited by
Cryer, P. E., Axelrod, L., Grossman, A. B., Heller, S. R., Montori, V. M., Seaquist, E. R., & Service, F. J. (2009) · The Journal of Clinical Endocrinology & Metabolism · doi:10.1210/jc.2008-1410 · Cited by 1,277
A history of blood glucose meters and their role in self-monitoring of diabetes mellitus
Clarke, S. F., & Foster, J. R. (2012) · British Journal of Biomedical Science · doi:10.1080/09674845.2012.12002443 · Cited by 456
Impact of Exercise on Overnight Glycemic Control in Children with Type 1 Diabetes Mellitus
Children, T. D. R. I. (2005) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2005.04.065 · Cited by 264
Self-citationThe Diabetes Research in Children Network (DirecNet) Study Group (2006) · Diabetes Care · doi:10.2337/diacare.29.01.06.dc05-1192 · Cited by 183
The Accuracy of Point-of-Care Glucose Measurements
Rebel, A., Rice, M. A., & Fahy, B. G. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600228 · Cited by 144
Metadata
- Authors
- The Diabetes Research In Children, N.
- Corresponding author
- The Diabetes Research In Children N
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2003-12
- Volume
- 5
- Issue
- 6
- Pages
- 933-941
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W4244218647
- Cited by
- 44
- References
- 14
- Citations by year
- 2023: 1 · 2022: 1 · 2020: 1 · 2019: 1 · 2018: 1 · 2017: 1 · 2014: 2 · 2013: 1 · 2012: 3
- FWCI
- 3.26
- Type
- journal article
- Language
- en
- Open access
- closed
- Licence
- www.liebertpub.com/nv/resources-tools/text-and-data-mining-policy/121/ (tdm)
- Locations
- Diabetes Technology & Therapeutics (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Deutsches Krebsforschungszentrum
- National Institutes of Health
- National Center for Research Resources (RR 00059, RR-00070-41, M01RR000059, M01 RR00069, M01RR000070, RR06022, M01RR000069)
- NIH Clinical Center
- Eunice Kennedy Shriver National Institute of Child Health and Human Development (U10HD041919, U10HD041908, HD-041906-01, HD041918-01, HD041919-0, U10HD041918, U10HD041915, HD 041908-01, HD041915, U10HD041906)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Keywords
- type 1 diabetes, children, meter accuracy, self-monitoring of blood glucose, capillary blood glucose, hypoglycemia, hyperglycemia, insulin-induced hypoglycemia test
- MeSH terms
- Adolescent, Blood Glucose, Calibration, Child, Child, Preschool, Diabetes Mellitus, Type 1, Female, Humans, Male, Reference Values, Self Care, Blood Glucose Self-Monitoring, Reproducibility of Results, Age of Onset
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.