Evaluation of Factors Affecting CGMS Calibration
Open accessThe DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
BACKGROUND: The optimal number/timing of calibrations entered into the CGMS (Medtronic MiniMed, Northridge, CA) continuous glucose monitoring system have not been previously described. METHODS: Fifty subjects with Type 1 diabetes mellitus (10-18 years old) were hospitalized in a clinical research center for approximately 24 h on two separate days. CGMS and OneTouch Ultra meter (LifeScan, Milpitas, CA) data were obtained. The CGMS was retrospectively recalibrated using the Ultra data varying the number and timing of calibrations. Resulting CGMS values were compared against laboratory reference values. RESULTS: There was a modest improvement in accuracy with increasing number of calibrations. The median relative absolute deviation (RAD) was 14%, 15%, 13%, and 13% when using three, four, five, and seven calibration values, respectively (P < 0.001). Corresponding percentages of CGMS-reference pairs meeting the International Organisation for Standardisation criteria were 66%, 67%, 71%, and 72% (P < 0.001). Nighttime accuracy improved when daytime calibrations (pre-lunch and pre-dinner) were removed leaving only two calibrations at 9 p.m. and 6 a.m. (median difference, -2 vs. -9 mg/dL, P < 0.001; median RAD, 12% vs. 15%, P = 0.001). Accuracy was better on visits where the average absolute rate of glucose change at the times of calibration was lower. On visits with average absolute rates <0.5, 0.5 to <1.0, 1.0 to <1.5, and >or=1.5 mg/dL/min, median RAD values were 13% versus 14% versus 17% versus 19%, respectively (P = 0.05). CONCLUSIONS: Although accuracy is slightly improved with more calibrations, the timing of the calibrations appears more important. Modifying the algorithm to put less weight on daytime calibrations for nighttime values and calibrating during times of relative glucose stability may have greater impact on accuracy.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Calibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patients
Choleau, C., Klein, J. C., Reach, G., Aussedat, B., Demaria-Pesce, V., Wilson, G. S., Gifford, R., & Ward, W. K. (2002) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(01)00304-9 · Cited by 76
Cited by
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Weinzimer, S. A., Steil, G. M., Swan, K. L., Dziura, J., Kurtz, N., & Tamborlane, W. V. (2008) · Diabetes Care · doi:10.2337/dc07-1967 · Cited by 567
Diabetes: Models, Signals, and Control
Cobelli, C., Dalla Man, C., Sparacino, G., Magni, L., De Nicolao, G., & Kovatchev, B. P. (2009) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2009.2036073 · Cited by 511
“[200], analyzed the improvement in CGMS sensor accuracy”
Gingras, V., Taleb, N., Roy‐Fleming, A., Legault, L., & Rabasa‐Lhoret, R. (2018) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.13052 · Cited by 113
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Validation of the continuous glucose monitoring sensor in preterm infants
Beardsall, K., Vanhaesebrouck, S., Ogilvy-Stuart, A. L., Vanhole, C., vanWeissenbruch, M., Midgley, P., Thio, M., Cornette, L., Ossuetta, I., Palmer, C. R., et al. (2013) · Archives of Disease in Childhood - Fetal and Neonatal Edition · doi:10.1136/archdischild-2012-301661 · Cited by 82
Deep learning framework for detection of hypoglycemic episodes in children with type 1 diabetes
San, P. P., Ling, S. H., & Nguyen, H. T. (2016) · 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2016.7591483 · Cited by 42
Exogenous Carbohydrate Spares Muscle Glycogen in Men and Women during 10 h of Exercise
HARGER-DOMITROVICH, S. G., McCLAUGHRY, A. E., GASKILL, S. E., & RUBY, B. C. (2007) · Medicine & Science in Sports & Exercise · doi:10.1249/mss.0b013e318157a650 · Cited by 29
Metadata
- Authors
- Diabetes Research In Children Network (Direcnet) Study Group
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2006-06
- Volume
- 8
- Issue
- 3
- Pages
- 318-325
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W1514567538
- Cited by
- 116
- References
- 1
- FWCI
- 6.25
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.