Microdialysis-Based 48-Hour Continuous Glucose Monitoring with GlucoDay ™ : Clinical Performance and Patients' Acceptance
The DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
BACKGROUND: This study was designed to assess clinical performance and patients' acceptance of the minimally invasive microdialysis-based continuous glucose monitoring system Gluco- Day() (Menarini Diagnostics, Florence, Italy) with a targeted monitoring time of 48 h. METHODS: An inpatient sample of 28 patients with diabetes was studied. The analysis of clinical performance was performed using mean absolute differences (MAD) (in percent), Pearson correlations, the Bland-Altman analysis, and Clarke Error Grid Analysis (EGA). GlucoDay glucose values were compared with laboratory standard blood glucose measurements (glucohexokinase assay). The patients' acceptance of the monitoring device was assessed via two self-report scales (pain during application and discomfort while wearing device). RESULTS: A mean (+/- SD) monitoring time of 45.7 +/- 3.3 h with a total of 484 paired readings could be achieved. A correlation of r (average) = 0.91 and a MAD of 19.9% indicated satisfactory to good clinical performance. Of the paired readings, 95.5% fell into the acceptable A and B zones of the EGA. Rather wide 95% limits of agreement were revealed in the Bland-Altman analysis. Whereas virtually no pain was experienced during sensor application, discomfort associated with wearing the device was rather high. All of the participants, however, stated that they would wear the device again. CONCLUSIONS: Satisfactory to good performance of the GlucoDay monitor was observed, indicating the device to be suitable for routine clinical use. In particular, however, the discomfort experienced during wearing requires further improvements in its usability.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312
Cited by
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Glucose sensors: a review of current and emerging technology
Oliver, N. S., Toumazou, C., Cass, A. E. G., & Johnston, D. G. (2009) · Diabetic Medicine · doi:10.1111/j.1464-5491.2008.02642.x · Cited by 584
Corrie, S. R., Coffey, J. W., Islam, J., Markey, K. A., & Kendall, M. A. F. (2015) · The Analyst · doi:10.1039/c5an00464k · Cited by 211
Overview on self-monitoring of blood glucose
Montagnana, M., Caputo, M., Giavarina, D., & Lippi, G. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2009.01.002 · Cited by 116
“Since the 2001, is also available the Glucoday (Menarini Diagnostics, Firenze, Italy); measurements are made every second and the average is provided every 3 min [71].”
MethodFreckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Metadata
- Authors
- Kubiak, T., Wörle, B., Kuhr, B., Nied, I., Gläsner, G., Hermanns, N., Kulzer, B., & Haak, T.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2006-10
- Volume
- 8
- Issue
- 5
- Pages
- 570-575
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2008406255
- Cited by
- 37
- References
- 2
- FWCI
- 2.62
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Microdialysis-Based 48-Hour Continuous Glucose Monitoring with GlucoDay ™ : Clinical Performance and Patients' Acceptance. https://app.cochat.ai/papers/10.1089/dia.2006.8.570
- The paper
- Kubiak, T., Wörle, B., Kuhr, B., Nied, I., Gläsner, G., Hermanns, N., Kulzer, B., & Haak, T. (2006). Microdialysis-Based 48-Hour Continuous Glucose Monitoring with GlucoDay ™ : Clinical Performance and Patients' Acceptance. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2006.8.570
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.