Comparison of Accuracy and Safety of the SEVEN and the Navigator Continuous Glucose Monitoring Systems
The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: This study evaluated the accuracy and safety of two continuous glucose monitoring (CGM) systems, the SEVEN (DexCom, San Diego, CA) and the Navigator (Abbott Diabetes Care, Alameda, CA), with the YSI laboratory measurements of blood glucose (blood glucose meter manufactured by YSI, Yellow Springs, OH), when worn concurrently in adults with type 1 diabetes. DESIGN AND METHODS: Fourteen subjects with type 1 diabetes, 33 +/- 6 (mean +/- SD) years old, were enrolled in this study. All subjects wore both sensors concurrently over three consecutive 5-day CGM sessions (15-day wear). On Days 5, 10, and 15, subjects participated in an 8-h in-clinic session where measurements from the CGM systems were collected and compared with YSI measurements every 15 min. At the end of Day 5 and 10 in-clinic sessions, the sensors were removed, and new sensors were inserted for the following CGM session despite the SEVEN system's recommended use for up to 7 days. RESULTS: The mean absolute relative difference (ARD) for the two CGM devices versus YSI was not different: 16.8% and 16.1% for SEVEN and Navigator, respectively (P = 0.38). In the hypoglycemic region (YSI value <80 mg/dL), the mean ARD for SEVEN was lower than for Navigator (21.5% vs. 29.8%, respectively; P = 0.001). The data analyses were similar when compared with self-monitoring of blood glucose (SMBG) values. Thirteen additional Navigator replacement devices were issued compared to two for the SEVEN. A total of three versus 14 skin reactions were reported with the SEVEN and Navigator insertion area, respectively. CONCLUSIONS: Glucose measurements with the SEVEN and Navigator were found to be similar compared with YSI and SMBG measurements, with the exception of the hypoglycemic range where the SEVEN performed better. However, the Navigator caused more skin area reactions.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Continuous glucose monitoring and intensive treatment of type 1 diabetes
(2008) · New England Journal of Medicine · doi:10.1056/nejmoa0805017 · Cited by 1,559
Epidemiology of severe hypoglycemia in the diabetes control and complications trial
(1991) · The American Journal of Medicine · doi:10.1016/0002-9343(91)80085-z · Cited by 807
Epidemiology of severe hypoglycemia in the Diabetes Control and Complications Trial
(1991) · The American Journal of Medicine · doi:10.1016/0002-9343(91)90605-w · Cited by 807
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312
Critical role of tissue mast cells in controlling long-term glucose sensor function in vivo
Klueh, U., Kaur, M., Qiao, Y., & Kreutzer, D. L. (2010) · Biomaterials · doi:10.1016/j.biomaterials.2010.02.023 · Cited by 48
Cited by
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Diabetes: Models, Signals, and Control
Cobelli, C., Dalla Man, C., Sparacino, G., Magni, L., De Nicolao, G., & Kovatchev, B. P. (2009) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2009.2036073 · Cited by 511
Prospects and limitations of non-invasive blood glucose monitoring using near-infrared spectroscopy
Yadav, J., Rani, A., Singh, V., & Murari, B. M. (2015) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2015.01.005 · Cited by 374
Closed-loop insulin delivery from bench to clinical practice
Hovorka, R. (2011) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2011.32 · Cited by 342
Continuous Glucose Monitoring Systems: A Review
Vashist, S. (2013) · Diagnostics · doi:10.3390/diagnostics3040385 · Cited by 261
Metadata
- Authors
- Garg, S. K., Smith, J., Beatson, C., Lopez-Baca, B., Voelmle, M., & Gottlieb, P. A.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2009-02
- Volume
- 11
- Issue
- 2
- Pages
- 65-72
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2090932625
- Cited by
- 93
- References
- 5
- FWCI
- 8.70
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.