Accuracy Evaluation of Five Blood Glucose Monitoring Systems Obtained from the Pharmacy: A European Multicenter Study with 453 Subjects
Open accessThe DOI matches its Crossref record; 4 of the 4 references it cites were found in a registry or an index.
Abstract
BACKGROUND: This multicenter study was conducted to evaluate the performance of five recently introduced blood glucose (BG) monitoring (BGM) devices under daily routine conditions in comparison with the YSI (Yellow Springs, OH) 2300 Stat Plus glucose analyzer. METHODS: Five hundred one diabetes patients with experience in self-monitoring of BG were randomized to use three of five different BGM devices (FreeStyle Lite® [Abbott Diabetes Care Inc., Alameda, CA], FreeStyle Freedom Lite [Abbott Diabetes Care], OneTouch® UltraEasy® [LifeScan Inc., Milpitas, CA], Accu-Chek® Aviva [Roche Diagnostics, Mannheim, Germany], and Contour® [Bayer Vital GmbH, Leverkusen, Germany]) in a daily routine setting. All devices and strips were purchased from local regular distribution sources (pharmacies, four strip lots per device). The patients performed the finger prick and the glucose measurement on their own. In parallel, a healthcare professional performed the glucose assessment with the reference method (YSI 2300 Stat Plus). The primary objective was the comparison of the mean absolute relative differences (MARD). Secondary objectives were compliance with the International Organization for Standardization (ISO) accuracy criteria under these routine conditions and Clarke and Parkes Error Grid analyses. RESULTS: MARD ranged from 4.9% (FreeStyle Lite) to 9.7% (OneTouch UltraEasy). The ISO 15197:2003 requirements were fulfilled by the FreeStyle Lite (98.8%), FreeStyle Freedom Lite (97.5%), and Accu-Chek Aviva (97.0%), but not by the Contour (92.4%) and OneTouch UltraEasy (91.1%). The number of values in Zone A of the Clarke Error Grid analysis was highest for the FreeStyle Lite (98.8%) and lowest for the OneTouch Ultra Easy (90.4%). CONCLUSIONS: FreeStyle Lite, FreeStyle Freedom Lite, and Accu-Chek Aviva performed very well in this study with devices and strips purchased through regular distribution channels, with the FreeStyle Lite achieving the lowest MARD in this investigation.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Passing, H., & Bablok, W. (1983) · cclm · doi:10.1515/cclm.1983.21.11.709 · Cited by 1,513
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
System Accuracy Evaluation of 27 Blood Glucose Monitoring Systems According to DIN EN ISO 15197
Freckmann, G., Baumstark, A., Jendrike, N., Zschornack, E., Kocher, S., Tshiananga, J., Heister, F., & Haug, C. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0128 · Cited by 195 · cited 8×
“Conclusions: FreeStyle Lite, FreeStyle Freedom Lite, and Accu-Chek Aviva performed very well in this study with devices and strips purchased through regular distribution channels, with the FreeStyle Lite achieving the lowest MARD in this investigation.”
BackgroundMethodInfluential(1994) · doi:10.2337/diacare.17.6.595 · Cited by 32
Cited by
Electrochemical Glucose Sensing: Is There Still Room for Improvement?
Witkowska Nery, E., Kundys, M., Jeleń, P. S., & Jönsson-Niedziółka, M. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03151 · Cited by 292
Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237
Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Bailey, T. S., Chang, A., & Christiansen, M. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814559746 · Cited by 168
Comparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127
Freckmann, G., Schmid, C., Baumstark, A., Rutschmann, M., Haug, C., & Heinemann, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815580160 · Cited by 70
Performance of Cleared Blood Glucose Monitors
Klonoff, D. C., & Prahalad, P. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815584797 · Cited by 59
Klaff, L. J., Brazg, R., Hughes, K., Tideman, A. M., Schachner, H. C., Stenger, P., Pardo, S., Dunne, N., & Parkes, J. L. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0069 · Cited by 46
A review of blood glucose monitor accuracy
King, F., Ahn, D., Hsiao, V., Porco, T., & Klonoff, D. C. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0232 · Cited by 37
Boettcher, C., Dost, A., Wudy, S. A., Flechtner-Mors, M., Borkenstein, M., Schiel, R., Weitzel, D., Bechtold-Dalla Pozza, S., Wolf, J., Holl, R. W., et al. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0262 · Cited by 35
Klonoff, D. C., & Reyes, J. S. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700432 · Cited by 29
Yu-Fei, W., Wei-Ping, J., Ming-Hsun, W., Miao-O, C., Ming-Chang, H., Chi-Pin, W., & Ming-Shih, L. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817705143 · Cited by 15
Link, M., Pleus, S., Schmid, C., Freckmann, G., Baumstark, A., Stolberg, E., & Haug, C. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813518859 · Cited by 6
Metadata
- Authors
- Tack, C., Pohlmeier, H., Behnke, T., Schmid, V., Grenningloh, M., Forst, T., & Pfützner, A.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2012-04
- Volume
- 14
- Issue
- 4
- Pages
- 330-337
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2043675644
- Cited by
- 92
- References
- 4
- FWCI
- 6.35
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.