Reduction in Duration of Hypoglycemia by Automatic Suspension of Insulin Delivery: The In-Clinic ASPIRE Study
Abstract
Background The efficacy of automatic suspension of insulin delivery in induced hypoglycemia among subjects with type 1 diabetes was evaluated. Subjects and methods In this randomized crossover study, subjects used a sensor-augmented insulin pump system with a low glucose suspend (LGS) feature that automatically stops insulin delivery for 2 h following a sensor glucose (SG) value ≤70 mg/dL. Subjects fasted overnight and exercised until their plasma glucose (measured with the YSI 2300 STAT Plus™ glucose and lactate analyzer [YSI Life Sciences, Yellow Springs, OH]) value reached ≤85 mg/dL on different occasions separated by washout periods lasting 3-10 days. Exercise sessions were done with the LGS feature turned on (LGS-On) or with continued insulin delivery regardless of SG value (LGS-Off). The order of LGS-On and LGS-Off sessions was randomly assigned. YSI glucose data were used to compare the duration and severity of hypoglycemia from successful LGS-On and LGS-Off sessions and to estimate the risk of rebound hyperglycemia after pump suspension. Results Fifty subjects attempted 134 sessions, 98 of which were successful. The mean±SD hypoglycemia duration was less during LGS-On than during LGS-Off sessions (138.5±76.68 vs. 170.7±75.91 min, P=0.006). During LGS-On compared with LGS-Off sessions, mean nadir YSI glucose was higher (59.5±5.72 vs. 57.6±5.69 mg/dL, P=0.015), as was mean end-observation YSI glucose (91.4±41.84 vs. 66.2±13.48 mg/dL, P 250 mg/dL. Conclusions Automatic s…
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“when comparing the safety and efficacy of overnight closed-loop delivery of insulin with conventional insulin-pump therapy in adults with type 1 diabetes, closed-loop delivery improved overnight control of glycaemia and reduced the risk of nocturnal hypoglycaemia.98,99”
From Type 1 diabetes (2014) · cites this paper as [98]
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
Mechanisms of Hypoglycemia-Associated Autonomic Failure and Its Component Syndromes in Diabetes
Cryer, P. E. (2005) · Diabetes · doi:10.2337/diabetes.54.12.3592 · Cited by 406
Insulin Pump Therapy With Automated Insulin Suspension in Response to Hypoglycemia
Choudhary, P., Shin, J., Wang, Y., Evans, M. L., Hammond, P. J., Kerr, D., Shaw, J. A. M., Pickup, J. C., & Amiel, S. A. (2011) · Diabetes Care · doi:10.2337/dc10-2411 · Cited by 175
Self-citationPrevention of Hypoglycemia by Using Low Glucose Suspend Function in Sensor-Augmented Pump Therapy
Danne, T., Kordonouri, O., Holder, M., Haberland, H., Golembowski, S., Remus, K., Bläsig, S., Wadien, T., Zierow, S., Hartmann, R., et al. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0084 · Cited by 150
Duration of Nocturnal Hypoglycemia Before Seizures
Buckingham, B. A., Wilson, D. M., Lecher, T., Hanås, R., Kaiserman, K. B., & Cameron, F. J. (2008) · Diabetes Care · doi:10.2337/dc08-0863 · Cited by 111
Self-citationAgrawal, P. K., Welsh, J. B., Kannard, B., Askari, S., Yang, Q., & KAUFMAN, F. R. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500514 · Cited by 86
Self-citationLy, T. T., Hewitt, J., Davey, R. J., Lim, E. M., Davis, E. A., & Jones, T. W. (2011) · Diabetes Care · doi:10.2337/dc10-1042 · Cited by 81
Hypoglycemia Begets Hypoglycemia in IDDM
Cryer, P. E. (1993) · Diabetes · doi:10.2337/diab.42.12.1691 · Cited by 68
Brazg, R., Bailey, T. S., Garg, S. K., Buckingham, B. A., Slover, R. H., Klonoff, D. C., Nguyen, X., Shin, J., Welsh, J. B., & Lee, S. W. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500621 · Cited by 28
Self-citation
Cited by
Sorted by citations
Atkinson, M. A., Eisenbarth, G. S., & Michels, A. W. (2014) · The Lancet · doi:10.1016/s0140-6736(13)60591-7 · Cited by 2,931
Threshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652
Self-citationLy, T. T., Nicholas, J. A., Retterath, A., Lim, E. M., Davis, E. A., & Jones, T. W. (2013) · JAMA · doi:10.1001/jama.2013.277818 · Cited by 411
Insulin delivery methods: Past, present and future
Shah, R., Shah, V. N., Patel, M., & Maahs, D. M. (2016) · International Journal of Pharmaceutical Investigation · doi:10.4103/2230-973x.176456 · Cited by 303
Grunberger, G., Sherr, J., Allende, M., Blevins, T., Bode, B., Handelsman, Y., Hellman, R., Lajara, R., Roberts, V. L., Rodbard, D., et al. (2021) · Endocrine Practice · doi:10.1016/j.eprac.2021.04.008 · Cited by 288
ISPAD Clinical Practice Consensus Guidelines 2018: diabetes technologies
Sherr, J. L., Tauschmann, M., Battelino, T., de Bock, M., Forlenza, G., Roman, R., Hood, K. K., & Maahs, D. M. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12731 · Cited by 197
Adolfsson, P., Riddell, M. C., Taplin, C. E., Davis, E. A., Fournier, P. A., Annan, F., Scaramuzza, A. E., Hasnani, D., & Hofer, S. E. (2018) · Pediatric Diabetes · doi:10.1111/pedi.12755 · Cited by 191
Challenges and recent progress in the development of a closed-loop artificial pancreas
Bequette, B. W. (2012) · Annual Reviews in Control · doi:10.1016/j.arcontrol.2012.09.007 · Cited by 184
Pozzilli, P., Battelino, T., Danne, T., Hovorka, R., Jarosz-Chobot, P. A., & Renard, É. (2015) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.2653 · Cited by 167
Buckingham, B. A., Raghinaru, D., Cameron, F., Bequette, B. W., Chase, H. P., Maahs, D. M., Slover, R. H., Wadwa, R. P., Wilson, D. M., Ly, T. T., et al. (2015) · Diabetes Care · doi:10.2337/dc14-3053 · Cited by 140
Self-citationBosi, E., Choudhary, P., de Valk, H. W., Lablanche, S., Castañeda, J., de Portu, S., Silva, J. D., Ré, R., Vorrink, L., Shin, J., et al. (2019) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(19)30150-0 · Cited by 125
Self-citationRoutine Sensor-Augmented Pump Therapy in Type 1 Diabetes: The INTERPRET Study
Nørgaard, K., Scaramuzza, A. E., Bratina, N., Lalic, N. M., Jarosz-Chobot, P. A., Kocsis, G., Jašinskienė, E., Block, C. E. M. D., Carrette, O., Castañeda, J., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0288 · Cited by 107
Real-Time Improvement of Continuous Glucose Monitoring Accuracy
Facchinetti, A., Sparacino, G., Guerra, S., Luijf, Y. M., DeVries, J. H., Mader, J. K., Ellmerer, M., Benesch, C., Heinemann, L., Bruttomesso, D., et al. (2012) · Diabetes Care · doi:10.2337/dc12-0736 · Cited by 106
Buckingham, B. A., Cameron, F., Calhoun, P. M., Maahs, D. M., Wilson, D. M., Chase, H. P., Bequette, B. W., Lum, J. W., Sibayan, J., Beck, R. W., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0040 · Cited by 104
Self-citationExercise in children and adolescents with diabetes
Robertson, K. J., Riddell, M. C., Guinhouya, B. C., Adolfsson, P., & Hanås, R. (2014) · Pediatric Diabetes · doi:10.1111/pedi.12176 · Cited by 104
Susceptibility of Interstitial Continuous Glucose Monitor Performance to Sleeping Position
Mensh, B. D., Wisniewski, N. A., Neil, B. M., & Burnett, D. R. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700408 · Cited by 97
Current perspectives on physical activity and exercise for youth with diabetes
Pivovarov, J. A., Taplin, C. E., & Riddell, M. C. (2015) · Pediatric Diabetes · doi:10.1111/pedi.12272 · Cited by 101
Danne, T., Tsioli, C., Kordonouri, O., Bläsig, S., Remus, K., Roy, A., Keenan, D. B., Lee, S. W., & KAUFMAN, F. R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0327 · Cited by 91
Self-citationZhong, A., Choudhary, P., McMahon, C., Agrawal, P. K., Welsh, J. B., Cordero, T. L., & KAUFMAN, F. R. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0216 · Cited by 80
Self-citationContinuous Glucose Monitoring: Evidence and Consensus Statement for Clinical Use
Liebl, A., Henrichs, H. R., Heinemann, L., Freckmann, G., Biermann, E., & Thomas, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700227 · Cited by 77
Prediction and prevention of hypoglycaemic events in type-1 diabetic patients using machine learning
Vehí, J., Contreras, I., Oviedo, S., Biagi, L., & Bertachi, A. (2020) · Health Informatics Journal · doi:10.1177/1460458219850682 · Cited by 76
Zaharieva, D. P., Messer, L. H., Paldus, B., O'Neal, D. N., Maahs, D. M., & Riddell, M. C. (2020) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2020.06.003 · Cited by 78
Closed-Loop System in the Management of Diabetes: Past, Present, and Future
Shah, V. N., Shoskes, A., Tawfik, B., & Garg, S. K. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0193 · Cited by 77
Self-citationA review of artificial pancreas technologies with an emphasis on bi‐hormonal therapy
Bakhtiani, P. A., Zhao, L., Youssef, J. E., Castle, J. R., & Ward, W. K. (2013) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12107 · Cited by 65
Improved Real-World Glycemic Control With Continuous Glucose Monitoring System Predictive Alerts
Abraham, S. B., Arunachalam, S., Zhong, A., Agrawal, P., Cohen, O., & McMahon, C. M. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819859334 · Cited by 63
Hypoglycemia Reduction and Changes in Hemoglobin A1c in the ASPIRE In-Home Study
Weiss, R., Garg, S. K., Bode, B. W., Bailey, T. S., Ahmann, A., Schultz, K. A., Welsh, J. B., Shin, J., & for the ASPIRE In-Home Study Group (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0306 · Cited by 54
Self-citationSensor-Augmented Insulin Pumps and Hypoglycemia Prevention in Type 1 Diabetes
Steineck, I. I. K., Ranjan, A. G., Nørgaard, K., & Schmidt, S. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672689 · Cited by 59
Defective Counterregulation and Hypoglycemia Unawareness in Diabetes
Reno, C. M., Litvin, M., Clark, A. L., & Fisher, S. J. (2012) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2012.11.005 · Cited by 56
Khalil, A., Beshyah, S., Awad, S. A., Benbarka, M., Haddad, M., Al-Hassan, D., Kahwatih, M., & Nagelkerke, N. J. D. (2012) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0061 · Cited by 53
Evaluation of a Predictive Low-Glucose Management System In-Clinic
Buckingham, B. A., Bailey, T. S., Christiansen, M. P., Garg, S. K., WEINZIMER, S. A., Bode, B. W., Anderson, S. M., Brazg, R., Ly, T. T., & KAUFMAN, F. R. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0319 · Cited by 53
Self-citationVettoretti, M., & Facchinetti, A. (2019) · BioMedical Engineering OnLine · doi:10.1186/s12938-019-0658-x · Cited by 52
Lane, J. E., Shivers, J. P., & Zisser, H. (2013) · Current Opinion in Endocrinology Diabetes and Obesity · doi:10.1097/med.0b013e32835edb9d · Cited by 50
The Impact of Accelerometer and Heart Rate Data on Hypoglycemia Mitigation in Type 1 Diabetes
Stenerson, M. B., Cameron, F., Wilson, D. M., Harris, B., Payne, S. R., Bequette, B. W., & Buckingham, B. A. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813516208 · Cited by 51
Self-citationEmerging Diabetes Technologies: Continuous Glucose Monitors/Artificial Pancreases
ALMURASHI, A. M., Rodriguez, E., & Garg, S. K. (2023) · Journal of the Indian Institute of Science · doi:10.1007/s41745-022-00348-3 · Cited by 55
Self-citationvan Beers, C. A. J., & DeVries, J. H. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816653411 · Cited by 41
Agrawal, P. K., Zhong, A., Welsh, J. B., Shah, R. V., & KAUFMAN, F. R. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0257 · Cited by 41
Self-citationAbraham, M. B., Davey, R. J., O'Grady, M. J., Ly, T. T., Paramalingam, N., Fournier, P. A., Roy, A., Grosman, B., Kurtz, N., Fairchild, J. M., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0141 · Cited by 40
Social Determinants of Health, Goals and Outcomes in High-Risk Children With Type 1 Diabetes
Hershey, J. A., Morone, J. F., Lipman, T. H., & Hawkes, C. P. (2021) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2021.02.005 · Cited by 40
Elbarbary, N. S. (2016) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.2781 · Cited by 37
A Peek Under the Hood: Explaining the MiniMed™ 780G Algorithm with Meal Detection Technology
Grosman, B., Roy, A., LINTEREUR, L. J., Turksoy, K., Benedetti, A., Cordero, T. L., Vigersky, R. A., MCVEAN, J. J. F., Rhinehart, A. S., & Cohen, O. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0446 · Cited by 36
Technology to Optimize Pediatric Diabetes Management and Outcomes
Markowitz, J. T., Harrington, K. R., & Laffel, L. M. (2013) · Current Diabetes Reports · doi:10.1007/s11892-013-0419-3 · Cited by 38
Sherr, J. L., Palau Collazo, M., Cengiz, E., Michaud, C., Carria, L., Steffen, A. T., Weyman, K., Zgorski, M., Tichy, E., Tamborlane, W. V., et al. (2014) · Diabetes Care · doi:10.2337/dc13-1608 · Cited by 35
Abraham, M. B., de Bock, M. I., Paramalingam, N., O'Grady, M. J., Ly, T. T., George, C., Roy, A., Spital, G., Karula, S., Heels, K., et al. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0364 · Cited by 36
PAULEY, M. E., Tommerdahl, K. L., Snell‐Bergeon, J. K., & Forlenza, G. P. (2022) · Current Cardiology Reports · doi:10.1007/s11886-022-01799-x · Cited by 39
Hanazaki, K., Munekage, M., Kitagawa, H., Yatabe, T., Munekage, E., Shiga, M., Maeda, H., & Namikawa, T. (2016) · Journal of Artificial Organs · doi:10.1007/s10047-016-0904-y · Cited by 33
The Impact of Accelerometer Use in Exercise-Associated Hypoglycemia Prevention in Type 1 Diabetes
Stenerson, M. B., Cameron, F., Payne, S. R., Payne, S. L., Ly, T. T., Wilson, D. M., & Buckingham, B. A. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814551045 · Cited by 31
Self-citationUtility of Continuous Glucose Monitoring in Type 1 and Type 2 Diabetes
Toschi, E., & Wolpert, H. A. (2016) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2016.06.003 · Cited by 35
Katz, M. L., Giani, E., & Laffel, L. M. (2015) · Current Diabetes Reports · doi:10.1007/s11892-015-0692-4 · Cited by 33
Wood, M. A., Shulman, D. I., Forlenza, G. P., Bode, B. W., Pinhas‐Hamiel, O., Buckingham, B. A., Kaiserman, K. B., Liljenquist, D. R., Bailey, T. S., Shin, J., et al. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0209 · Cited by 31
Self-citationHypoglycemia Begets Hypoglycemia: The Order Effect in the ASPIRE In-Clinic Study
Garg, S. K., Brazg, R., Bailey, T. S., Buckingham, B. A., Slover, R. H., Klonoff, D. C., Shin, J., Welsh, J. B., & KAUFMAN, F. R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0219 · Cited by 29
Self-citationChen, E., King, F., Kohn, M. A., Spanakis, E. K., Breton, M. D., & Klonoff, D. C. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0119 · Cited by 28
Self-citationWake, A. D. (2022) · World Journal of Diabetes · doi:10.4239/wjd.v13.i3.161 · Cited by 36
Bakhtiani, P. A., Youssef, J. E., Duell, A. K., Branigan, D. L., Jacobs, P. G., Lasarev, M. R., Castle, J. R., & Ward, W. K. (2014) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2014.09.001 · Cited by 26
Automatic Artificial Pancreas Systems Using an Intelligent Multiple-Model PID Strategy
Batmani, Y., Khodakaramzadeh, S., & Moradi, P. (2021) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2021.3116376 · Cited by 26
Shah, V. N., Moser, E. G., Blau, A., Dhingra, M., & Garg, S. K. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0228 · Cited by 25
Self-citationAlotaibi, A., Khalifah, R. A. A., & McAssey, K. (2020) · Pediatric Diabetes · doi:10.1111/pedi.13088 · Cited by 27
Continuous Glucose Monitoring Adherence
de Bock, M. I., Cooper, M. N., Retterath, A. J., Nicholas, J. A., Ly, T. T., Jones, T. W., & Davis, E. A. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816633484 · Cited by 23
Insulin pump therapy in youth with type 1 diabetes: toward closed-loop systems
Tauschmann, M., & Hovorka, R. (2014) · Expert Opinion on Drug Delivery · doi:10.1517/17425247.2014.910192 · Cited by 22
Do youth with type 1 diabetes exercise safely? A focus on patient practices and glycemic outcomes
Roberts, A. J. C., Yi‐Frazier, J. P., Aitken, K. E., Mitrovich, C., Pascual, M., & Taplin, C. E. (2016) · Pediatric Diabetes · doi:10.1111/pedi.12402 · Cited by 21
Development and Future of Automated Insulin Delivery (AID) Systems
Garg, S. K., & MCVEAN, J. J. F. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0467 · Cited by 25
Self-citationThe role of continuous glucose monitoring in the care of children with type 1 diabetes
Larson, N. S., & Pinsker, J. E. (2013) · International Journal of Pediatric Endocrinology · doi:10.1186/1687-9856-2013-8 · Cited by 21
The Artificial Pancreas: Is it Important to Understand How the β Cell Controls Blood Glucose?
Steil, G. M., & Grodsky, G. M. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700528 · Cited by 20
Abraham, M. B., Nicholas, J. A., Ly, T. T., Roby, H. C., Paramalingam, N., Fairchild, J. M., King, B. R., Ambler, G. R., Cameron, F. J., Davis, E. A., et al. (2016) · BMJ Open · doi:10.1136/bmjopen-2016-011589 · Cited by 19
ARX model for interstitial glucose prediction during and after physical activities
Ugalde, H. M. R., Garnotel, M., Doron, M., Jallon, P., Charpentier, G., Franc, S., Huneker, E., Simon, C., & Bonnet, S. (2019) · Control Engineering Practice · doi:10.1016/j.conengprac.2019.07.013 · Cited by 17
Davis, T., Salahi, A., Welsh, J. B., & Bailey, T. S. (2015) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12542 · Cited by 16
Self-citationWolpert, H. A., Kavanagh, M., Atakov-Castillo, A., & Steil, G. M. (2015) · Diabetic Medicine · doi:10.1111/dme.12823 · Cited by 14
Weiss, R., Garg, S. K., Bergenstal, R. M., Klonoff, D. C., Bode, B. W., Bailey, T. S., Thrasher, J. R., Schwartz, F., Welsh, J. B., KAUFMAN, F. R., et al. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815586014 · Cited by 12
Self-citationVillafuerte, B., Frías, M. M., Martín, M. B. R., Valverde, R. Y., Gómez, M. Á. Á., & Castellanos, R. B. (2017) · Endocrinología Diabetes y Nutrición · doi:10.1016/j.endinu.2017.02.008 · Cited by 15
Automated Insulin Delivery—The Light at the End of the Tunnel
Turksoy, K., Frantz, N., Quinn, L., Dumin, M., Kilkus, J. M., Hibner, B. A., Çınar, A., & Littlejohn, E. (2017) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2017.02.055 · Cited by 12
Klonoff, D. C., Bergenstal, R. M., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A., Welsh, J. B., & Lee, S. W. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700424 · Cited by 12
Self-citationPetruželková, L., Picková, K., Šumnı́k, Z., Šoupal, J., & Obermannová, B. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0459 · Cited by 12
Roze, S., Smith‐Palmer, J., de Portu, S., Saltik, A. Z. O., Akgül, T., & Deyneli, O. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2019.0198 · Cited by 12
Gargallo-Fernández, M., Escalada, J., Peralta, F. G., Moreno, P. R., Martínez, A. M., Botella-Serrano, M., Pérez, C. T., & Fernández, J. L. (2015) · Endocrinología y Nutrición · doi:10.1016/j.endonu.2015.02.004 · Cited by 15
Marçal, D. F. D. S., Alexandrino, E. G., Cortez, L. E. R., & Bennemann, R. M. (2018) · journal of physical education · doi:10.4025/jphyseduc.v29i1.2917 · Cited by 11
Roze, S., Smith‐Palmer, J., Portu, S. D., Delbaere, A., Brouwer, B. D., & de Valk, H. W. (2019) · ClinicoEconomics and Outcomes Research · doi:10.2147/ceor.s186298 · Cited by 11
Predictive Low-Glucose Suspend to Prevent Hypoglycemia
Pinsker, J. E., & Dassau, E. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0064 · Cited by 10
Saiman, Y., & Mahmud, N. (2021) · European Journal of Gastroenterology & Hepatology · doi:10.1097/meg.0000000000002218 · Cited by 9
Modern clinical management helps reducing the impact of type 1 diabetes in children
Scaramuzza, A. E., & Zuccotti, G. (2015) · Pharmacological Research · doi:10.1016/j.phrs.2015.03.001 · Cited by 9
Hypoglycemia Prevention Using Low Glucose Suspend Systems
Bequette, B. W. (2014) · Lecture notes in bioengineering · doi:10.1007/978-3-642-54464-4_3 · Cited by 9
JENDLE, J. H., Adolfsson, P., & Pollock, N. W. (2020) · Diving and Hyperbaric Medicine Journal · doi:10.28920/dhm50.2.135-143 · Cited by 9
Continuous glucose monitoring in pregnant women with type 1 diabetes
Garg, S. K., & Polsky, S. (2017) · The Lancet · doi:10.1016/s0140-6736(17)32449-2 · Cited by 8
Self-citationSperling, M. A., Wolfsdorf, J. I., Menon, R. K., Tamborlane, W. V., Maahs, D. M., Battelino, T., & Phillip, M. (2021) · Elsevier eBooks · doi:10.1016/b978-0-323-62520-3.00021-x · Cited by 11
Kurnaz, E., Aycan, Z., Yıldırım, N., & Çetinkaya, S. Ç. (2017) · The Turkish Journal of Pediatrics · doi:10.24953/turkjped.2017.04.013 · Cited by 8
Swarm hybrid optimization for a piecewise model fitting applied to a glucose model
Acedo, L., Botella, M., Cortés, J. ‐., Hidalgo, J. I., Maqueda, E., & Villanueva, R. J. (2018) · Journal of Systems and Information Technology · doi:10.1108/jsit-10-2017-0103 · Cited by 8
Use of continuous glucose monitoring for sport in type 1 diabetes
Abdulrahman, A., Manhas, J., Linane, H., Gurney, M. E., Fitzgerald, C., & O'Sullivan, E. (2018) · BMJ Open Sport & Exercise Medicine · doi:10.1136/bmjsem-2018-000432 · Cited by 8
The Future of Glucose Monitoring
Garg, S. K. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0421 · Cited by 8
Self-citationClosed-loop insulin delivery in adults with type 1 diabetes
Kumareswaran, K. (2012) · Apollo (University of Cambridge) · doi:10.17863/cam.16644 · Cited by 7
Nutritional Management of Athletes with Type 1 Diabetes: A Narrative Review
Cavallo, M., Fano, M. D., Barana, L., Dozzani, I., Bianchini, E., Pellegrino, M., Cisternino, L., Migliarelli, S., Giulietti, C., Pippi, R., et al. (2024) · Nutrients · doi:10.3390/nu16060907 · Cited by 9
Registration and Real-Life Studies on Automated Insulin Delivery Systems
Aktürk, H. K., Brown, S. A., Beck, R. W., Sherr, J. L., Bergenstal, R. M., Kovatchev, B. P., Vigersky, R. A., Breton, M. D., Hovorka, R., Klonoff, D. C., et al. (2025) · Journal of Diabetes Science and Technology · doi:10.1177/19322968251334993 · Cited by 6
Self-citationRegistration and Real-Life Studies on Automated Insulin Delivery Systems
Aktürk, H. K., Brown, S. A., Beck, R. W., Sherr, J. L., Bergenstal, R. M., Kovatchev, B. P., Vigersky, R. A., Breton, M. D., Hovorka, R., Klonoff, D. C., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2025.0130 · Cited by 6
Self-citationVillafuerte, B., Frías, M. M., Martín, M. B. R., Valverde, R. Y., Gómez, M. Á. Á., & Castellanos, R. B. (2017) · Endocrinología Diabetes y Nutrición (English ed ) · doi:10.1016/j.endien.2017.02.009 · Cited by 7
Experience With the Enlite Sensor in a Multicenter Pediatric Study
Evans, K., Richardson, C., Landry, A., Muileboom, J., Cormack, L., & Lawson, M. L. (2014) · The Science of Diabetes Self-Management and Care · doi:10.1177/0145721714560589 · Cited by 6
Exercise for People with Type 1 Diabetes
GALLEN, I. W. (2014) · Medicine and sport science/Medicine and sport · doi:10.1159/000357344 · Cited by 5
Continuous Glucose Monitoring in 2012
Bode, B. W., & Battelino, T. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.1502 · Cited by 5
Islet cell transplantation: current status in the UK
Choudhary, P., PARROTT, N. R., Birtles, L., & Rutter, M. K. (2012) · Practical Diabetes · doi:10.1002/pdi.1707 · Cited by 5
Yafi, M., Shah, A., & Vélez, K. (2023) · Translational Pediatrics · doi:10.21037/tp-23-145 · Cited by 5
De la bomba de insulina y el monitoreo continuo de glucosa al páncreas artificial
Apablaza, P., Néstor Soto, I., & Ethel Codner, D. (2017) · Revista médica de Chile · doi:10.4067/s0034-98872017000500011 · Cited by 9
Refining the Closed Loop in the Data Age: Research-to-Practice Transitions in Diabetes Technology
Forlenza, G. P., Sankaranarayanan, S., & Maahs, D. M. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2015.0055 · Cited by 5
Gargallo-Fernández, M., Martín, J. E. S., Peralta, F. G., Moreno, P. R., Martínez, A. M., Botella-Serrano, M., Pérez, C. T., & Fernández, J. L. (2016) · PubMed · doi:10.1016/j.endonu.2016.03.002 · Cited by 5
Shah, V. N., Rewers, A. H., & Garg, S. K. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00013-2 · Cited by 4
Self-citation
Metadata
- Authors
- Garg, S., Brazg, R. L., Bailey, T. S., Buckingham, B. A., Slover, R. H., Klonoff, D. C., Shin, J., Welsh, J. B., & Kaufman, F. R.
- Affiliations
- Barbara Davis Center for Childhood Diabetes, Aurora, Colorado.
- Rainier Clinical Research Center (US)
- AMCR Institute (US)
- Stanford Medicine (US)
- Mills Peninsula Health Services (US)
- Medtronic Inc., Northridge, California.
- Corresponding author
- Garg, Satish
- ORCID
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2012-03
- Volume
- 14
- Issue
- 3
- Pages
- 205-209
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W1978849955
- Cited by
- 138
- References
- 9
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 4 · 2023: 4 · 2022: 5 · 2021: 7 · 2020: 6 · 2019: 10 · 2018: 8 · 2017: 11 · 2016: 16 · 2015: 17 · 2014: 16 · 2013: 21 · 2012: 10
- FWCI
- 16.43
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Diabetes Technology & Therapeutics (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- type 1 diabetes, hypoglycemia, low glucose suspend, sensor-augmented pump, exercise-induced hypoglycemia, rebound hyperglycemia, hypoglycemia duration, hypoglycemia severity, continuous glucose monitoring
- MeSH terms
- Adolescent, Adult, Blood Glucose, Diabetes Mellitus, Type 1, Female, Humans, Hypoglycemia, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Male, Middle Aged, Treatment Outcome, Cross-Over Studies, Monitoring, Ambulatory, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review