Evaluating the Accuracy and Large Inaccuracy of Two Continuous Glucose Monitoring Systems
Open accessThe DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: This study evaluated the accuracy and large inaccuracy of the Freestyle Navigator (FSN) (Abbott Diabetes Care, Alameda, CA) and Dexcom SEVEN PLUS (DSP) (Dexcom, Inc., San Diego, CA) continuous glucose monitoring (CGM) systems during closed-loop studies. RESEARCH DESIGN AND METHODS: Paired CGM and plasma glucose values (7,182 data pairs) were collected, every 15-60 min, from 32 adults (36.2±9.3 years) and 20 adolescents (15.3±1.5 years) with type 1 diabetes who participated in closed-loop studies. Levels 1, 2, and 3 of large sensor error with increasing severity were defined according to absolute relative deviation greater than or equal to ±40%, ±50%, and ±60% at a reference glucose level of ≥6 mmol/L or absolute deviation greater than or equal to ±2.4 mmol/L,±3.0 mmol/L, and ±3.6 mmol/L at a reference glucose level of <6 mmol/L. RESULTS: Median absolute relative deviation was 9.9% for FSN and 12.6% for DSP. Proportions of data points in Zones A and B of Clarke error grid analysis were similar (96.4% for FSN vs. 97.8% for DSP). Large sensor over-reading, which increases risk of insulin over-delivery and hypoglycemia, occurred two- to threefold more frequently with DSP than FSN (once every 2.5, 4.6, and 10.7 days of FSN use vs. 1.2, 2.0, and 3.7 days of DSP use for Level 1-3 errors, respectively). At levels 2 and 3, large sensor errors lasting 1 h or longer were absent with FSN but persisted with DSP. CONCLUSIONS: FSN and DSP differ substantially in the frequency and duration of large inaccuracy despite only modest differences in conventional measures of numerical and clinical accuracy. Further evaluations are required to confirm that FSN is more suitable for integration into closed-loop delivery systems.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Closed-loop insulin delivery from bench to clinical practice
Hovorka, R. (2011) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2011.32 · Cited by 342
Self-citationFully Integrated Artificial Pancreas in Type 1 Diabetes
Breton, M., Farret, A., Bruttomesso, D., Anderson, S., Magni, L., Patek, S., Dalla Man, C., Place, J., Demartini, S., Del Favero, S., et al. (2012) · Diabetes · doi:10.2337/db11-1445 · Cited by 315
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors
Kovatchev, B. P., Gonder-Frederick, L. A., Cox, D. J., & Clarke, W. L. (2004) · Diabetes Care · doi:10.2337/diacare.27.8.1922 · Cited by 312 · cited 1×
“27 However, none of the above measures provides information about the duration, severity, and incidence of individual large sensor errors.”
Hovorka, R., Kumareswaran, K., Harris, J., Allen, J. M., Elleri, D., Xing, D., Kollman, C., Nodale, M., Murphy, H. R., Dunger, D. B., et al. (2011) · BMJ · doi:10.1136/bmj.d1855 · Cited by 244 · cited 1×
“Subjects attended the research facility on two or three occasions separated by 1–5 weeks to contrast alternative treatments (closed-loop insulindeliveryandconventional treatment 5,7,14,15 or two closed-loop visits.”
Self-citationNovel Use of Glucagon in a Closed-Loop System for Prevention of Hypoglycemia in Type 1 Diabetes
Castle, J. R., Engle, J. M., Youssef, J. E., Massoud, R. G., Yuen, K. C. J., Kagan, R., & Ward, W. K. (2010) · Diabetes Care · doi:10.2337/dc09-2254 · Cited by 232
Kovatchev, B., Cobelli, C., Renard, E., Anderson, S., Breton, M., Patek, S., Clarke, W., Bruttomesso, D., Maran, A., Costa, S., et al. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400611 · Cited by 200
Cited by
Physical activity/exercise and diabetes: A position statement of the American Diabetes Association
Colberg, S. R., Sigal, R. J., Yardley, J. E., Riddell, M. C., Dunstan, D. W., Dempsey, P. C., Horton, E. S., Castorino, K., & Tate, D. F. (2016) · Diabetes Care · doi:10.2337/dc16-1728 · Cited by 2,991
Coming of age: the artificial pancreas for type 1 diabetes
Thabit, H., & Hovorka, R. (2016) · Diabetologia · doi:10.1007/s00125-016-4022-4 · Cited by 223
“lead to incorrect treatment decisions [50] and adversely affect device usability and experience.”
BackgroundSelf-citationHeinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
Limits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
Schrangl, P., Reiterer, F., Heinemann, L., Freckmann, G., & Del Re, L. (2018) · Biosensors · doi:10.3390/bios8020050 · Cited by 46
Metadata
- Authors
- Leelarathna, L., Nodale, M., Allen, J. M., Elleri, D., Kumareswaran, K., Haidar, A., Caldwell, K., Wilinska, M. E., Acerini, C. L., Evans, M. L., Murphy, H. R., Dunger, D. B., & Hovorka, R.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2013-02
- Volume
- 15
- Issue
- 2
- Pages
- 143-149
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2043660159
- Cited by
- 43
- References
- 7
- FWCI
- 4.37
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluating the Accuracy and Large Inaccuracy of Two Continuous Glucose Monitoring Systems. https://app.cochat.ai/papers/10.1089/dia.2012.0245
- The paper
- Leelarathna, L., Nodale, M., Allen, J. M., Elleri, D., Kumareswaran, K., Haidar, A., Caldwell, K., Wilinska, M. E., Acerini, C. L., Evans, M. L., Murphy, H. R., Dunger, D. B., & Hovorka, R. (2013). Evaluating the Accuracy and Large Inaccuracy of Two Continuous Glucose Monitoring Systems. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2012.0245
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.