Accuracy and Acceptability of the 6-Day Enlite Continuous Subcutaneous Glucose Sensor
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
OBJECTIVE: This study evaluated the performance and acceptability of the Enlite(®) glucose sensor (Medtronic MiniMed, Inc., Northridge, CA). SUBJECTS AND METHODS: Ninety adults with type 1 or type 2 diabetes wore two Enlite sensors on the abdomen and/or buttock for 6 days and calibrated them at different frequencies. On Days 1, 3, and 6, accuracy was evaluated by comparison of sensor glucose values with frequently sampled plasma glucose values collected over a 12-h period. Accuracy was assessed at different reference glucose concentrations and during times when absolute glucose concentration rates of change were <1, 1-2, and >2 mg/dL/min. The sensor's ability to detect hypoglycemia or hyperglycemia was evaluated with simulated alerts. Subject satisfaction was evaluated with a 7-point Likert-type questionnaire, with a score of 7 indicating strong agreement. RESULTS: With abdomen sensors under actual-use calibration (mean, 2.8 ± 0.9 times/day), the overall mean (median) absolute relative difference (ARD) values between sensor and reference values were 13.6% (10.1%); the corresponding buttock sensor ARD values were 15.5% (10.5%). With abdomen sensors under minimal calibration (mean, 1.2 ± 0.9 times/day), the mean (median) ARD values were 14.7% (10.8%). Mean ARD values of abdomen sensors at rates of change of <1, 1-2, and >2 mg/dL/min were 13.6%, 12.9%, and 16.3%, respectively. With abdomen sensors, 79.5% and 94.1% of hypoglycemic and hyperglycemic events, respectively, were correctly detected; 81.9% and 94.9% of hypoglycemic and hyperglycemic alerts, respectively, were confirmed. The failure rates for abdomen and buttock sensors were 19.7% and 13.9%, respectively. Mean responses to survey questions for all subjects related to comfort and ease of use were favorable. CONCLUSIONS: The Enlite sensor provided accurate data at different glucose concentrations and rates of change. Subjects found the sensor comfortable and easy to use.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Threshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652
Self-citation
Cited by
Personalized nutrition by prediction of glycemic responses
Zeevi, D., Korem, T., Zmora, N., Israeli, D., Rothschild, D., Weinberger, A., Ben-Yacov, O., Lador, D., Avnit-Sagi, T., Lotan-Pompan, M., et al. (2015) · Cell · doi:10.1016/j.cell.2015.11.001 · Cited by 2,929
Electrochemical glucose sensors in diabetes management: an updated review (2010–2020)
Teymourian, H., Barfidokht, A., & Wang, J. (2020) · Chemical Society Reviews · doi:10.1039/d0cs00304b · Cited by 881
Threshold-based insulin-pump interruption for reduction of hypoglycemia
Bergenstal, R. M., Klonoff, D. C., Garg, S. K., Bode, B. W., Meredith, M., Slover, R. H., Ahmann, A. J., Welsh, J. B., Lee, S. W., & Kaufman, F. R. (2013) · New England Journal of Medicine · doi:10.1056/nejmoa1303576 · Cited by 652
Self-citationGarg, S. K., Weinzimer, S. A., Tamborlane, W. V., Buckingham, B. A., Bode, B. W., Bailey, T. S., Brazg, R. L., Ilany, J., Slover, R. H., Anderson, S. M., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0421 · Cited by 585
Self-citationCappon, G., Vettoretti, M., Sparacino, G., & Facchinetti, A. (2019) · Diabetes & Metabolism Journal · doi:10.4093/dmj.2019.0121 · Cited by 477
Continuous glucose monitoring: a review of recent studies demonstrating improved glycemic outcomes
Rodbard, D. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0035 · Cited by 458
Personalized microbiome-driven effects of non-nutritive sweeteners on human glucose tolerance
Suez, J., Cohen, Y., Valdés-Mas, R., Mor, U., Dori-Bachash, M., Federici, S., Zmora, N., Leshem, A., Heinemann, M., Linevsky, R., et al. (2022) · Cell · doi:10.1016/j.cell.2022.07.016 · Cited by 423
Wearable Continuous Glucose Monitoring Sensors: A Revolution in Diabetes Treatment
Cappon, G., Acciaroli, G., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2017) · Electronics · doi:10.3390/electronics6030065 · Cited by 237
“All these changes All these changes improved usability and accuracy, reaching 13.6% MARD [35]. improved usability and accuracy, reaching 13.6% MARD [35].”
BackgroundInfluentialAssessing Sensor Accuracy for Non-Adjunct Use of Continuous Glucose Monitoring
Kovatchev, B. P., Patek, S. D., Ortiz, E. A., & Breton, M. D. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0272 · Cited by 234
Heinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170
Clinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Bailey, T. S., Chang, A., & Christiansen, M. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814559746 · Cited by 168
Self-citationContinuous glucose monitoring: Current use in diabetes management and possible future applications
Vettoretti, M., Cappon, G., Acciaroli, G., Facchinetti, A., & Sparacino, G. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818774078 · Cited by 96
Freckmann, G., Eichenlaub, M., Waldenmaier, D., Pleus, S., Wehrstedt, S., Haug, C., Witthauer, L., Jendle, J., Hinzmann, R., Thomas, A., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231190941 · Cited by 86
“A few studies, eg, by Zueger et al 64 and Bailey et al, 80 examined the effect of additional and/or timing of calibrations on CGM accuracy.”
BackgroundCappon, G., Vettoretti, M., Marturano, F., Facchinetti, A., & Sparacino, G. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818759558 · Cited by 62
Limits to the Evaluation of the Accuracy of Continuous Glucose Monitoring Systems by Clinical Trials
Schrangl, P., Reiterer, F., Heinemann, L., Freckmann, G., & Del Re, L. (2018) · Biosensors · doi:10.3390/bios8020050 · Cited by 46
Pham-Short, A., Donaghue, K. C., Ambler, G., Garnett, S., & Craig, M. E. (2017) · Scientific Reports · doi:10.1038/srep45286 · Cited by 31
Newbold, R., Benedetti, A., Kimoff, R. J., Meltzer, S., Garfield, N., Dasgupta, K., Gagnon, R., Lavigne, L., Olha, A., Rey, E., et al. (2021) · Chest · doi:10.1016/j.chest.2020.07.014 · Cited by 28
Metadata
- Authors
- Bailey, T. S., Ahmann, A., Brazg, R., Christiansen, M., Garg, S., Watkins, E., Welsh, J. B., & Lee, S. W.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2014-05
- Volume
- 16
- Issue
- 5
- Pages
- 277-283
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2158434788
- Cited by
- 101
- References
- 1
- FWCI
- 9.34
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.