Accuracy of Continuous Glucose Monitoring During Three Closed-Loop Home Studies Under Free-Living Conditions
Open accessThe DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.
Abstract
OBJECTIVES: Closed-loop (CL) systems modulate insulin delivery based on glucose levels measured by a continuous glucose monitor (CGM). Accuracy of the CGM affects CL performance and safety. We evaluated the accuracy of the Freestyle Navigator(®) II CGM (Abbott Diabetes Care, Alameda, CA) during three unsupervised, randomized, open-label, crossover home CL studies. MATERIALS AND METHODS: Paired CGM and capillary glucose values (10,597 pairs) were collected from 57 participants with type 1 diabetes (41 adults [mean±SD age, 39±12 years; mean±SD hemoglobin A1c, 7.9±0.8%] recruited at five centers and 16 adolescents [mean±SD age, 15.6±3.6 years; mean±SD hemoglobin A1c, 8.1±0.8%] recruited at two centers). Numerical accuracy was assessed by absolute relative difference (ARD) and International Organization for Standardization (ISO) 15197:2013 15/15% limits, and clinical accuracy was assessed by Clarke error grid analysis. RESULTS: Total duration of sensor use was 2,002 days (48,052 h). Overall sensor accuracy for the capillary glucose range (1.1-27.8 mmol/L) showed mean±SD and median (interquartile range) ARD of 14.2±15.5% and 10.0% (4.5%, 18.4%), respectively. Lowest mean ARD was observed in the hyperglycemic range (9.8±8.8%). Over 95% of pairs were in combined Clarke error grid Zones A and B (A, 80.1%, B, 16.2%). Overall, 70.0% of the sensor readings satisfied ISO criteria. Mean ARD was consistent (12.3%; 95% of the values fall within ±3.7%) and not different between participants (P=0.06) within the euglycemic and hyperglycemic range, when CL is actively modulating insulin delivery. CONCLUSIONS: Consistent accuracy of the CGM within the euglycemic-hyperglycemic range using the Freestyle Navigator II was observed and supports its use in home CL studies. Our results may contribute toward establishing normative CGM performance criteria for unsupervised home use of CL.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Epidemiology of severe hypoglycemia in the diabetes control and complications trial
(1991) · The American Journal of Medicine · doi:10.1016/0002-9343(91)80085-z · Cited by 807
Epidemiology of severe hypoglycemia in the Diabetes Control and Complications Trial
(1991) · The American Journal of Medicine · doi:10.1016/0002-9343(91)90605-w · Cited by 807
Closed-loop insulin delivery from bench to clinical practice
Hovorka, R. (2011) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2011.32 · Cited by 342
Self-citationSimulation Environment to Evaluate Closed-Loop Insulin Delivery Systems in Type 1 Diabetes
Wilinska, M. E., Chassin, L. J., Acerini, C. L., Allen, J. M., Dunger, D. B., & Hovorka, R. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400117 · Cited by 244
Self-citationClinical accuracy of a continuous glucose monitoring system with an advanced algorithm
Bailey, T. S., Chang, A., & Christiansen, M. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814559746 · Cited by 168 · cited 1×
“21 The MARD during home use was 11.3%, with 92.4% in CEG Zone A.”
MethodInfluential
Cited by
Continuous glucose monitoring: a review of recent studies demonstrating improved glycemic outcomes
Rodbard, D. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0035 · Cited by 458
Coming of age: the artificial pancreas for type 1 diabetes
Thabit, H., & Hovorka, R. (2016) · Diabetologia · doi:10.1007/s00125-016-4022-4 · Cited by 223
Self-citationOutcome measures for artificial pancreas clinical trials: a consensus report
Maahs, D. M., Buckingham, B. A., Castle, J. R., Cinar, A., Damiano, E. R., Dassau, E., DeVries, J. H., Doyle, F. J., Griffen, S. C., Haidar, A., et al. (2016) · Diabetes Care · doi:10.2337/dc15-2716 · Cited by 211
Self-citationFokkert, M. J., van Dijk, P. R., Edens, M. A., Abbes, S., de Jong, D., Slingerland, R. J., & Bilo, H. J. G. (2017) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2016-000320 · Cited by 209
“9 Whenever a new CGM device becomes available, it is essential to critically evaluate its accuracy and usability.”
BackgroundHeinemann, L., Schoemaker, M., Schmelzeisen-Redecker, G., Hinzmann, R., Kassab, A., Freckmann, G., Reiterer, F., & Del Re, L. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819855670 · Cited by 170
Metadata
- Authors
- Thabit, H., Leelarathna, L., Wilinska, M. E., Elleri, D., Allen, J. M., Lubina-Solomon, A., Walkinshaw, E., Stadler, M., Choudhary, P., Mader, J. K., Dellweg, S., Benesch, C., Pieber, T. R., Arnolds, S., Heller, S. R., Amiel, S. A., Dunger, D., Evans, M. L., & Hovorka, R.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2015-11
- Volume
- 17
- Issue
- 11
- Pages
- 801-807
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2406835988
- Cited by
- 37
- References
- 5
- FWCI
- 5.82
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Accuracy of Continuous Glucose Monitoring During Three Closed-Loop Home Studies Under Free-Living Conditions. https://app.cochat.ai/papers/10.1089/dia.2015.0062
- The paper
- Thabit, H., Leelarathna, L., Wilinska, M. E., Elleri, D., Allen, J. M., Lubina-Solomon, A., Walkinshaw, E., Stadler, M., Choudhary, P., Mader, J. K., Dellweg, S., Benesch, C., Pieber, T. R., Arnolds, S., Heller, S. R., Amiel, S. A., Dunger, D., Evans, M. L., & Hovorka, R. (2015). Accuracy of Continuous Glucose Monitoring During Three Closed-Loop Home Studies Under Free-Living Conditions. Diabetes Technology & Therapeutics. https://doi.org/10.1089/dia.2015.0062
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.