Comparison of Two Continuous Glucose Monitoring Systems, Dexcom G4 Platinum and Medtronic Paradigm Veo Enlite System, at Rest and During Exercise
The DOI matches its Crossref record; 3 of the 3 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Despite technological advances, the accuracy of continuous glucose monitoring (CGM) systems may not always be satisfactory with rapidly changing glucose levels, as is notable during exercise. We compare the performance of two current and widely used CGM systems, Dexcom G4 Platinum (Dexcom) and Medtronic Paradigm Veo Enlite system (Enlite), during both rest and exercise in adults with type 1 diabetes (T1D). RESEARCH DESIGN AND METHODS: Paired sensor and plasma glucose (PG) values (total of 431 data pairs for Dexcom and 425 for Enlite) were collected from 17 adults (37.3 ± 13.6 years) with T1D. To evaluate and compare the accuracy of sensor readings, criteria involving sensor bias (sensor minus PG levels), absolute relative difference (ARD), and percentage of readings meeting International Organization for Standardization (ISO) criteria were considered. RESULTS: Both Dexcom and Enlite performed equally well during the rest period, with respective mean/median biases of -0.12/-0.02 mmol/L versus -0.18/-0.40 (P = 0.78, P = 0.66) mmol/L and ARDs of 13.77/13.34% versus 12.38/11.95% (P = 0.53, P = 0.70). During exercise, sensor bias means/medians were -0.40/-0.21 mmol versus -0.26/-0.24 mmol/L (P = 0.67, P = 0.62) and ARDs were 22.53/15.13% versus 20.44/14.11% (P = 0.58, P = 0.68) for Dexcom and Enlite, respectively. Both sensors demonstrated significantly lower performance during exercise; median ARD comparison at rest versus exercise for both Dexcom and Enlite showed a P = 0.02. More data pairs met the ISO criteria for Dexcom and Enlite at rest, 73.6% and 76.9% compared with exercise 48.2% and 53.9%. CONCLUSION: Dexcom and Enlite demonstrated comparable overall performances during rest and physical activity. However, a lower accuracy was observed during exercise for both sensors, necessitating a fine-tuning of their performance with physical activity.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Effectiveness of Continuous Glucose Monitoring in a Clinical Care Environment
Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2010) · Diabetes Care · doi:10.2337/dc09-1502 · Cited by 313
Longitudinal study of new and prevalent use of self-monitoring of blood glucose
Karter, A. J., Parker, M. M., Moffet, H. H., Spence, M. M., Chan, J., Ettner, S. L., & Selby, J. V. (2006) · Diabetes Care · doi:10.2337/dc06-2073 · Cited by 139
Facchinetti, A., Del Favero, S., Sparacino, G., & Cobelli, C. (2015) · Medical & Biological Engineering & Computing · doi:10.1007/s11517-014-1226-y · Cited by 80 · cited 1×
“8 New generations of CGM are continuously in development to improve accuracy and mitigate the effects of said limitations.”
Cited by
Exercise management in type 1 diabetes: a consensus statement
Riddell, M. C., Gallen, I. W., Smart, C. E., Taplin, C. E., Adolfsson, P., Lumb, A. N., Kowalski, A., Rabasa-Lhoret, R., McCrimmon, R. J., Hume, C., et al. (2017) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(17)30014-1 · Cited by 903
“…accurate for exercise (104,105); however, the lag time in glucose equilibrium with the interstitial space and the rapid turnover in glucose during exercise may impact accuracy (i.e. overestimate glucose value when levels are dropping and underestimate it when levels are rising) (106,107).”
Self-citationContinuous glucose monitoring: a review of recent studies demonstrating improved glycemic outcomes
Rodbard, D. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2017.0035 · Cited by 458
Reddy, M., Jugnee, N., El Laboudi, A., Spanudakis, E., Anantharaja, S., & Oliver, N. (2018) · Diabetic Medicine · doi:10.1111/dme.13561 · Cited by 245
Improving the Clinical Value and Utility of CGM Systems: Issues and Recommendations
Petrie, J. R., Peters, A. L., Bergenstal, R. M., Holl, R. W., Fleming, G. A., & Heinemann, L. (2017) · Diabetes Care · doi:10.2337/dci17-0043 · Cited by 135
Gingras, V., Taleb, N., Roy‐Fleming, A., Legault, L., & Rabasa‐Lhoret, R. (2018) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.13052 · Cited by 113
“Inaccuracies related to the device itself such as errors in signalling, noise filtering or positional stability are additional variables.(69,70) The accuracy of the CGMS has considerably improved,(71) however, errors can be challenging in study protocols; for example, sensor failure due to inaccuracy or loss of signal…”
Self-citation
Metadata
- Authors
- Taleb, N., Emami, A., Suppere, C., Messier, V., Legault, L., Chiasson, J.-L., Rabasa-Lhoret, R., & Haidar, A.
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2016-09
- Volume
- 18
- Issue
- 9
- Pages
- 561-567
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- OpenAlex
- W2461542737
- Cited by
- 62
- References
- 3
- FWCI
- 11.55
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.