Real-World Performance of the MiniMed™ 780G System: First Report of Outcomes from 4120 Users
Open accessAbstract
Background: The MiniMed™ 780G system includes an advanced hybrid closed loop (AHCL) algorithm that provides both automated basal and correction bolus insulin delivery. The preliminary performance of the system in real-world settings was evaluated. Methods: Data uploaded from August 2020 to March 2021 by individuals living in Belgium, Finland, Italy, the Netherlands, Qatar, South Africa, Sweden, Switzerland, and the United Kingdom were aggregated and retrospectively analyzed to determine the mean glucose management indicator (GMI), percentage of time spent within (TIR), below (TBR), and above (TAR) glycemic ranges, system use, and insulin consumption in users having ≥10 days of sensor glucose (SG) data after initiating AHCL. The impact of initiating AHCL was evaluated in a subgroup of users also having ≥10 days of SG data, before AHCL initiation. Results: Users ( N = 4120) were observed for a mean of 54 ± 32 days. During this time, they spent a mean of 94.1% ± 11.4% of the time in AHCL and achieved a mean GMI of 6.8% ± 0.3%, TIR of 76.2% ± 9.1%, TBR <70 of 2.5% ± 2.1%, and TAR >180 of 21.3% ± 9.4%, after initiating AHCL. There were 77.3% and 79.0% of users who achieved a TIR >70% and a GMI of <7.0%, respectively. Users for whom comparison with pre-AHCL was possible ( N = 812) reduced their GMI by 0.4% ± 0.4% ( P = 0.005) and increased their TIR by 12.1% ± 10.5% ( P < 0.0001), post-AHCL initiation. More users achieved the glycemic treatment goals of GMI <7.0% (37.6% vs. 75.2%, P < 0.0001) and TIR >70% (34.6% vs. 74.9%, P < 0.0001) when compared with pre-AHCL initiation. Conclusion: Most MiniMed 780G system users achieved TIR >70% and GMI <7%, while minimizing hypoglycemia, in a real-world condition.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Both generation 1 low-glucose-suspend33 and generation 2 hybrid closed-loop systems improve glycated haemoglobin concentrations, the time in target range and the time below target range in children, adolescents34 and adults.35-37”
From Blood glucose monitoring devices: current considerations (2023) · cites this paper as [37]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“AHCL systems continuously adjust basal insulin delivery every five minutes based on real‐time continuous glucose monitoring (CGM) data and automatically administer correction boluses in response to hyperglycemia, thereby facilitating the maintenance of glucose levels near predefined targets 8 .”
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,237 · cited 1×
“The new parameters are born from the growing use of CGM measurements of time spent in different ranges with consensus on definitions of ranges and current goals.(1) For people with T1DM, except during pregnancy, the goals are to spend >70% of the time between the range of 70–180 mg/dL, <4% below 70 mg/dL, <1% below 54…”
BackgroundSelf-citationState of Type 1 Diabetes Management and Outcomes from the T1D Exchange in 2016–2018
Foster, N. C., Beck, R. W., Miller, K. M., Clements, M. A., Rickels, M. R., DiMeglio, L. A., Maahs, D. M., Tamborlane, W. V., Bergenstal, R., Smith, E., et al. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0384 · Cited by 1,907
Six-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1 Diabetes
Brown, S. A., Kovatchev, B. P., Raghinaru, D., Lum, J. W., Buckingham, B. A., Kudva, Y. C., Laffel, L. M., Levy, C. J., Pinsker, J. E., Wadwa, R. P., et al. (2019) · New England Journal of Medicine · doi:10.1056/nejmoa1907863 · Cited by 1,030
Comparison of Evidence of Treatment Effects in Randomized and Nonrandomized Studies
Ioannidis, J. P. A. (2001) · JAMA · doi:10.1001/jama.286.7.821 · Cited by 919
Holman, N., Knighton, P., Kar, P., O'Keefe, J., Curley, M., Weaver, A., Barron, E., Bakhai, C., Khunti, K., Wareham, N. J., et al. (2020) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(20)30271-0 · Cited by 908
6. Glycemic Targets: Standards of Medical Care in Diabetes—2021
American Diabetes Association (2021) · Diabetes Care · doi:10.2337/dc21-s006 · Cited by 847
Interpretation and Impact of Real-World Clinical Data for the Practicing Clinician
Blonde, L., Khunti, K., Harris, S. B., Meizinger, C., & SKOLNIK, N. (2018) · Advances in Therapy · doi:10.1007/s12325-018-0805-y · Cited by 757
Glucose management indicator (GMI): a new term for estimating A1C from continuous glucose monitoring
Bergenstal, R. M., Beck, R. W., Close, K. L., Grunberger, G., Sacks, D. B., Kowalski, A., Brown, A. S., Heinemann, L., Aleppo, G., Ryan, D. B., et al. (2018) · Diabetes Care · doi:10.2337/dc18-1581 · Cited by 684 · cited 1×
“Users with ‡10 days of sensor glucose (SG) data after AHCL was enabled for the first time (post-AHCL) were included in the analysis, based on previous publications using a similar duration of time to estimate or determine CGM-derived metrics.(13,14) All data available were included, whether the system was in AHCL cont…”
MethodGarg, S. K., Weinzimer, S. A., Tamborlane, W. V., Buckingham, B. A., Bode, B. W., Bailey, T. S., Brazg, R. L., Ilany, J., Slover, R. H., Anderson, S. M., et al. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0421 · Cited by 585
Self-citationWeisman, A., Bai, J.-W., Cardinez, M., Kramer, C. K., & Perkins, B. A. (2017) · The Lancet Diabetes & Endocrinology · doi:10.1016/s2213-8587(17)30167-5 · Cited by 463
Flagged first, then by citations
Research works printed but not linked: 3
- Da Silva J, , Arrieta A, , Castaneda J, et al.: Real-world Performance of the MiniMed™ 780G System: impact of initiating automated basal and correction boluses. Diabetes Technol Ther, 2021; 23:A79–A80.
- Habif S, , Singh H, , Manning M, et al.: Quality of life outcomes and glycemic control from the T:slim X2 Pump with Control-IQ Technology—Real World Observations from the CLIO Study. Diabetes Technol Ther, 2021; 23:A-66.
- Da Silva J, , Velado K, , Agrawal P, et al.: Real-life use and performance of the MiniMed™ 670G System in Europe. Diabetes Technol Ther, 2020; 22:A96–A97.
Cited by
Sorted by citations
Phillip, M., Nimri, R., Bergenstal, R. M., Barnard-Kelly, K., Danne, T., Hovorka, R., Kovatchev, B. P., Messer, L. H., Parkin, C. G., Ambler-Osborn, L., et al. (2023) · Endocrine Reviews · doi:10.1210/endrev/bnac022 · Cited by 443
Self-citationISPAD Clinical Practice Consensus Guidelines 2022: Diabetes technologies: Insulin delivery
Sherr, J. L., Schoelwer, M., Dos Santos, T. J., Reddy, L., Biester, T., Galderisi, A., van Dyk, J. C., Hilliard, M. E., Berget, C., & DiMeglio, L. A. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13421 · Cited by 152
Yogev, D., Goldberg, T., Arami, A., Tejman‐Yarden, S., Winkler, T. E., & Maoz, B. M. (2023) · APL Bioengineering · doi:10.1063/5.0152290 · Cited by 147
A Systematic Review of Commercial Hybrid Closed-Loop Automated Insulin Delivery Systems
Peacock, S., Frizelle, I., & Hussain, S. (2023) · Diabetes Therapy · doi:10.1007/s13300-023-01394-5 · Cited by 122
Predictors of time in target glucose range in real‐world users of the MiniMed 780G system
Castañeda, J., Mathieu, C., Aanstoot, H., Arrieta, A., Silva, J. D., Shin, J., & Cohen, O. (2022) · Diabetes Obesity and Metabolism · doi:10.1111/dom.14807 · Cited by 108
Self-citationPetrovski, G., Campbell, J., Pasha, M., Day, E., Hussain, K., Khalifa, A., & van den Heuvel, T. (2023) · Diabetes Care · doi:10.2337/dc22-1692 · Cited by 108
Forlenza, G. P., DeSalvo, D. J., Aleppo, G., Wilmot, E. G., Berget, C., Huyett, L. M., Hadjiyianni, I., Méndez, J. J., Conroy, L. R., Ly, T. T., et al. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0578 · Cited by 99
Choudhary, P., Arrieta, A., van den Heuvel, T. R. A., Castañeda, J., Smaniotto, V., & Cohen, O. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0433 · Cited by 83
Self-citationMatejko, B., Juza, A., Kieć-Wilk, B., Cyranka, K., Krzyżowska, S., Chen, X., Cohen, O., Da Silva, J., Malecki, M. T., & Klupa, T. (2022) · Diabetes Care · doi:10.2337/dc22-0470 · Cited by 72
Self-citationForlenza, G. P., Carlson, A. L., Galindo, R. J., Kruger, D. F., Levy, C. J., McGill, J. B., Umpierrez, G., & Aleppo, G. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0206 · Cited by 70
Current Status and Emerging Options for Automated Insulin Delivery Systems
Forlenza, G. P., & Lal, R. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0514 · Cited by 70
Unannounced Meal Challenges Using an Advanced Hybrid Closed-Loop System
Shalit, R., Minsky, N. C., Laron-Hirsh, M., Cohen, O., Kurtz, N., Roy, A., Grosman, B., Benedetti, A., & Tirosh, A. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0139 · Cited by 64
Self-citationGrassi, B. A., Gómez, A. M., Calliari, L. E., Franco, D. R., Raggio, M., Riera, F., Castro, M., MCVEAN, J. J. F., van den Heuvel, T. R. A., Arrieta, A., et al. (2023) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15023 · Cited by 58
Self-citationBeato-Víbora, P. I., Chico, A. I., Moreno-Fernández, J., Bellido, V., Nattero‐Chávez, L., César, M. J. P., Martínez‐Brocca, M. A., Giménez, M., Aguilera, E., Climent, E., et al. (2023) · Diabetes Care · doi:10.2337/dc23-1355 · Cited by 57
Kovatchev, B. P., Singh, H., Mueller, L., & Gonder-Frederick, L. A. (2022) · Diabetes Care · doi:10.2337/dc22-1217 · Cited by 49
Kovatchev, B. P., Colmegna, P. H., Pavan, J., C., J. L. D., Villa-Tamayo, M. F., Koravi, C. L. K., Santini, G., Alix, C., Stumpf, M. M., & Brown, S. A. (2025) · npj Digital Medicine · doi:10.1038/s41746-025-01679-y · Cited by 50
Closed-loop insulin delivery: update on the state of the field and emerging technologies
Ware, J., & Hovorka, R. (2022) · Expert Review of Medical Devices · doi:10.1080/17434440.2022.2142556 · Cited by 47
Griffin, T. P., Gallen, G., Hartnell, S., Crabtree, T. S. J., Holloway, M., Gibb, F. W., Lumb, A. N., Wilmot, E. G., Choudhary, P., & Hussain, S. (2023) · Diabetic Medicine · doi:10.1111/dme.15078 · Cited by 46
Emerging Diabetes Technologies: Continuous Glucose Monitors/Artificial Pancreases
ALMURASHI, A. M., Rodriguez, E., & Garg, S. K. (2023) · Journal of the Indian Institute of Science · doi:10.1007/s41745-022-00348-3 · Cited by 55
Barriers and Facilitators to Diabetes Device Adoption for People with Type 1 Diabetes
Tanenbaum, M. L., & COMMISSARIAT, P. V. (2022) · Current Diabetes Reports · doi:10.1007/s11892-022-01469-w · Cited by 47
Lombardo, F., Passanisi, S., Alibrandi, A., Bombaci, B., Bonfanti, R., Delvecchio, M., Candia, F. D., Mozzillo, E., Piccinno, E., Piona, C. A., et al. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0491 · Cited by 41
Real‐world glycaemic outcomes of automated insulin delivery in type 1 diabetes: A meta‐analysis
Yang, Q., Zeng, B., Hao, J., Yang, Q., & Sun, F. (2024) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15718 · Cited by 41
Lepore, G., Rossini, A., Bellante, R., Corsi, A. M., Scaranna, C., Dodesini, A. R., & Trevisan, R. (2022) · Acta Diabetologica · doi:10.1007/s00592-022-01937-5 · Cited by 37
Self-citationA Peek Under the Hood: Explaining the MiniMed™ 780G Algorithm with Meal Detection Technology
Grosman, B., Roy, A., LINTEREUR, L. J., Turksoy, K., Benedetti, A., Cordero, T. L., Vigersky, R. A., MCVEAN, J. J. F., Rhinehart, A. S., & Cohen, O. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0446 · Cited by 36
Self-citationThe changing landscape of automated insulin delivery in the management of type 1 diabetes
Lakshman, R., Boughton, C. K., & Hovorka, R. (2023) · Endocrine Connections · doi:10.1530/ec-23-0132 · Cited by 43
A Comparison of Two Hybrid Closed-Loop Systems in Italian Children and Adults With Type 1 Diabetes
Bassi, M., Teliti, M., Lezzi, M., Iosca, A., Strati, M. F., Carmisciano, L., d’Annunzio, G., Minuto, N., & Maggi, D. C. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2021.802419 · Cited by 35
Arunachalum, S., Velado, K., Vigersky, R. A., & Cordero, T. L. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221088608 · Cited by 35
Benhamou, P., Adenis, A., Lebbad, H., Tourki, Y., Heredia, M., Gehr, B., Franc, S., & Charpentier, G. (2023) · Diabetes, Obesity and Metabolism · doi:10.1111/dom.15008 · Cited by 35
Automated Insulin Delivery for Young People with Type 1 Diabetes and Elevated A1c
Boucsein, A., Zhou, Y., Michaels, V. R., Haszard, J. J., Jefferies, C., Wiltshire, E. J., Paul, R. G., Parry‐Strong, A., Pasha, M., Petrovski, G., et al. (2024) · NEJM Evidence · doi:10.1056/evidoa2400185 · Cited by 34
Schiaffini, R., Deodati, A., Nicoletti, M. C., Carducci, C., Ciampalini, P., Lorubbio, A., Matteoli, M. C., Pampanini, V., Patera, I. P., Rapini, N., et al. (2022) · Acta Diabetologica · doi:10.1007/s00592-022-01886-z · Cited by 33
Tanenbaum, M. L., & COMMISSARIAT, P. V. (2023) · Endocrine Connections · doi:10.1530/ec-23-0193 · Cited by 32
Minimed Medtronic 780G optimizes glucose control in patients with type 1 diabetes mellitus
Pintaudi, B., Gironi, I., Nicosia, R., Meneghini, E., Disoteo, O., Mion, E., & Bertuzzi, F. (2022) · Nutrition Metabolism and Cardiovascular Diseases · doi:10.1016/j.numecd.2022.03.031 · Cited by 32
Bassi, M., Patti, L., Silvestrini, I., Strati, M. F., Ponzano, M., Minuto, N., & Maggi, D. C. (2023) · Frontiers in Endocrinology · doi:10.3389/fendo.2023.1099024 · Cited by 32
Šantová, A., Plachý, L., Neuman, V., Pavlı́ková, M., Petruželková, L., Konečná, P., Venháčová, P., Škvor, J., Pomahačová, R., Neumann, D., et al. (2023) · Frontiers in Endocrinology · doi:10.3389/fendo.2023.1283181 · Cited by 32
The Psychological Implications of Automated Insulin Delivery Systems in Type 1 Diabetes Care
Nefs, G. (2022) · Frontiers in Clinical Diabetes and Healthcare · doi:10.3389/fcdhc.2022.846162 · Cited by 35
Cordero, T. L., Dai, Z., Arrieta, A., Niu, F., Vella, M., Shin, J., Rhinehart, A. S., MCVEAN, J. J. F., Lee, S. W., Slover, R. H., et al. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0123 · Cited by 33
Self-citationPerioperative Care of Patients Using Wearable Diabetes Devices
Cruz, P., McKee, A. M., Chiang, H.-H., McGill, J. B., Hirsch, I. B., Ringenberg, K. J., & Wildes, T. S. (2024) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000007115 · Cited by 32
Asgharzadeh, A., Patel, M., Connock, M. J., Damery, S. L., Ghosh, I., Jordan, M., Freeman, K., Brown, A., Court, R. A., Baldwin, S., et al. (2024) · Health Technology Assessment · doi:10.3310/jypl3536 · Cited by 32
Bhargava, A., Bergenstal, R. M., Warren, M., Thrasher, J. R., Dempsey, M. A., Bode, B. W., LaRocque, J., Carlson, A. L., Keiter, A., Ma, H., et al. (2025) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0586 · Cited by 31
Self-citationBassi, M., Franzone, D., Dufour, F., Strati, M. F., Scalas, M., Tantari, G., Aloi, C., Salina, A., d’Annunzio, G., Maghnie, M., et al. (2023) · Life · doi:10.3390/life13030783 · Cited by 34
Sawyer, A., Sobczak, M., Forlenza, G. P., & Alonso, G. T. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0471 · Cited by 28
Long-term outcomes of an advanced hybrid closed-loop system: A focus on different subpopulations
Beato-Víbora, P. I., Ambrojo-López, A., Fernández-Bueso, M., Gil-Poch, E., & Arroyo-Díez, F. J. (2022) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2022.110052 · Cited by 27
Automated Insulin Delivery Systems as a Treatment for Type 2 Diabetes Mellitus: A Review
Karol, A. B., O’Malley, G., Fallurin, R., & Levy, C. J. (2022) · Endocrine Practice · doi:10.1016/j.eprac.2022.10.001 · Cited by 27
Pihoker, C., Shulman, D. I., Forlenza, G. P., Kaiserman, K. B., Sherr, J. L., Thrasher, J. R., Buckingham, B. A., Kipnes, M. S., Bode, B. W., Carlson, A. L., et al. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0255 · Cited by 27
Self-citationHybrid Closed Loop Using a Do-It-Yourself Artificial Pancreas System in Adults With Type 1 Diabetes
Nanayakkara, N., Sharifi, A., Burren, D., Elghattis, Y., Jayarathna, D. K., & Cohen, N. D. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231153882 · Cited by 28
Aktürk, H. K., Battelino, T., Castañeda, J., Arrieta, A., van den Heuvel, T. R. A., & Cohen, O. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0432 · Cited by 26
Self-citationUse of diabetes technology in children
Schoelwer, M. J., DeBoer, M. D., & Breton, M. D. (2024) · Diabetologia · doi:10.1007/s00125-024-06218-0 · Cited by 27
Recent advances in closed-loop insulin delivery
Ware, J., & Hovorka, R. (2021) · Metabolism · doi:10.1016/j.metabol.2021.154953 · Cited by 28
Exercise and nutrition in type 1 diabetes: Insights from the FinnDiane cohort
Barlovič, D. P., Harjutsalo, V., & Groop, P. (2022) · Frontiers in Endocrinology · doi:10.3389/fendo.2022.1064185 · Cited by 26
Seget, S., Rusak, E., Polańska, J., & Jarosz-Chobot, P. A. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0148 · Cited by 22
McCarthy, O. M., Christensen, M. B., Kristensen, K. B., Schmidt, S., Ranjan, A. G., Bain, S. C., Bracken, R. M., & Nørgaard, K. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0009 · Cited by 23
Eviz, E., Kıllı, N. E., Karakus, K. E., Can, E., Gökçe, T., Mutlu, G. Y., & Hatun, Ş. (2024) · Diabetic Medicine · doi:10.1111/dme.15333 · Cited by 24
Karakus, K. E., Mutlu, G. Y., Gökçe, T., Eviz, E., Can, E., Muradoğlu, S., & Hatun, Ş. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/19322968221106194 · Cited by 23
Piccini, B., Pessina, B., Casalini, E., Lenzi, L., & Toni, S. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13440 · Cited by 22
Lepore, G., Borella, N. D., Castagna, G., Ippolito, S., Bonfadini, S., Corsi, A. M., Scaranna, C., Dodesini, A. R., Bellante, R., & Trevisan, R. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0261 · Cited by 22
Self-citationCYRANKA, K. M., Matejko, B., Juza, A., Kieć‐Wilk, B., Krzyżowska, S., Cohen, O., Silva, J. D., Lushchyk, M. L., Malecki, M. T., & Klupa, T. (2022) · JMIR Formative Research · doi:10.2196/43535 · Cited by 24
Self-citationAl‐Sofiani, M. E., Alharthi, S. K., Albunyan, S., Alzaman, N. S., Klonoff, D. C., & Alguwaihes, A. M. (2024) · Diabetes Care · doi:10.2337/dc23-1968 · Cited by 22
Development and Future of Automated Insulin Delivery (AID) Systems
Garg, S. K., & MCVEAN, J. J. F. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0467 · Cited by 25
Elbarbary, N. S., & Ismail, E. A. R. (2023) · Diabetology & Metabolic Syndrome · doi:10.1186/s13098-023-01184-w · Cited by 20
MiniMed TM 780G Advanced Hybrid Closed-Loop System Study in Pregnant Women with Type 1 Diabetes
Guibert, C., Amoura, L., Rakotoarisoa, L., Plat, F., Sonnet, E., Lablanche, S., Tréglia, C., Sarde, E., Leca, V., Rimareix, F., et al. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0267 · Cited by 21
Bahillo-Curieses, P., Fernández-Velasco, P., Pérez-López, P., Martínez, A. M. V., de la Marca, M. D. L. O. N., & Díaz-Soto, G. (2024) · Endocrine · doi:10.1007/s12020-024-03881-6 · Cited by 24
Pöhlmann, J., Smith‐Palmer, J., Serné, E. H., Tuomaala, A., JENDLE, J. H., Choudhary, P., Bosi, E., Mohan, V., van den Heuvel, T. R. A., Cohen, O., et al. (2025) · Diabetes Obesity and Metabolism · doi:10.1111/dom.70161 · Cited by 21
Self-citationClosed-Loop and Artificial Intelligence–Based Decision Support Systems
Nimri, R., Phillip, M., & Kovatchev, B. P. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.2505 · Cited by 21
Eviz, E., Mutlu, G. Y., Karakus, K. E., Can, E., Gökçe, T., Muradoğlu, S., & Hatun, Ş. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0112 · Cited by 19
Reducing Diabetes Burden in Medtronic's Automated Insulin Delivery Systems
Forlenza, G. P., Dai, Z., Niu, F., & Shin, J. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0459 · Cited by 20
Self-citationMatejko, B., Juza, A., Kieć‐Wilk, B., CYRANKA, K. M., Krzyżowska, S., Cohen, O., Malecki, M. T., & Klupa, T. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0059 · Cited by 18
Self-citationCastañeda, J., Arrieta, A., van den Heuvel, T. R. A., & Cohen, O. (2023) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15139 · Cited by 18
Self-citationMunda, A., Kovacic, C., & Barlovič, D. P. (2022) · Metabolites · doi:10.3390/metabo12111137 · Cited by 17
McCarthy, O. M., Christensen, M. B., Kristensen, K. B., Schmidt, S., Ranjan, A. G., Bain, S. C., Bracken, R. M., & Nørgaard, K. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0542 · Cited by 17
Shannon, N., Castañeda, J., Choudhary, P., Kolassa, R., Keuthage, W., Kroeger, J., Thivolet, C. H., Evans, M. L., Ré, R., Cellot, J., et al. (2023) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15217 · Cited by 18
Self-citationRachmiel, M., Lebenthal, Y., Mazor‐Aronovitch, K., Brener, A., Levek, N., Jacobi‐Polishook, T., Ari, T. B., Abiri, S., Landau, Z., & Pinhas‐Hamiel, O. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0081 · Cited by 17
Automated Insulin Delivery: A Milestone on the Road to Insulin Independence in Type 1 Diabetes
Limbert, C., Kowalski, A. J., & Danne, T. (2024) · Diabetes Care · doi:10.2337/dci24-0007 · Cited by 17
Amigó, J., Ortiz-Zúñiga, Á., de Urbina, A. M. O., Sánchez, M., Dos-Santos, M., Abad, M., Cuadra, F., Simó, R., Hernández, C., & Simó-Servat, O. (2023) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2023.110730 · Cited by 15
Gardner, D. S.-L., Lakkad, M., Qiu, Z., Inoue, Y., CHANDRAN, S. R., & Wherry, K. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0455 · Cited by 16
Mameli, C., Smylie, G. M., Galati, A., Rapone, B., Cardona‐Hernandez, R., Zuccotti, G., & Delvecchio, M. (2023) · European Journal of Pediatrics · doi:10.1007/s00431-023-04833-4 · Cited by 15
Kesavadev, J., Basanth, A., Krishnan, G., Shankar, A., Sanal, G., & Jothydev, S. (2023) · Diabetes Therapy · doi:10.1007/s13300-023-01427-z · Cited by 16
Real-World Diabetes Technology: Overcoming Barriers and Disparities
Messer, L. H., Addala, A., & WEINZIMER, S. A. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.2511 · Cited by 14
Commercial hybrid closed-loop systems available for a patient with type 1 diabetes in 2022
Seget, S., Tekielak, A., Rusak, E., & Jarosz-Chobot, P. A. (2023) · Pediatric Endocrinology Diabetes and Metabolism · doi:10.5114/pedm.2023.126359 · Cited by 14
Malone, S. K., MATUS, A. M., Flatt, A. J., Peleckis, A. J., Grunin, L., Yu, G., Jang, S., Weimer, J. E., Lee, I., Rickels, M. R., et al. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231182406 · Cited by 14
An Automated Insulin Delivery System with Automatic Meal Bolus Based on a Hand-Gesturing Algorithm
Roy, A., Grosman, B., Benedetti, A., Engheta, B., Miller, D. M., Laron-Hirsh, M., Cohen, Y. C., Ré, R., Edd, S. N., Vigersky, R. A., et al. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0529 · Cited by 14
Self-citationCherubini, V., Faragalli, A., Arnaldi, C., Bassi, M., Bonfanti, R., Bracciolini, G., Cardella, F., Bo, S. D., Delvecchio, M., Candia, F. D., et al. (2024) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2024.111621 · Cited by 15
Folk, S. Y. L., Zappe, J., Wyne, K. L., & Dungan, K. M. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241234948 · Cited by 14
Munda, A., Kovacic, C., & Barlovič, D. P. (2024) · Journal of Diabetes and its Complications · doi:10.1016/j.jdiacomp.2024.108795 · Cited by 15
Herguido, N. G., Amuedo, S., Bellido, V., Gallardo, G. L., Losada, F., Pérez-Morales, A., Trillo, C. A. R., & Moreno, A. S. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2022.0287 · Cited by 13
(Hybrid) Closed-Loop Systems: From Announced to Unannounced Exercise
Zimmer, R. T., Auth, A., Schierbauer, J., Haupt, S., Wachsmuth, N. B., Zimmermann, P., Voit, T., Battelino, T., Sourij, H., & Moser, O. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0293 · Cited by 15
Self-citationSingh, R., Imberg, H., Ahmadi, S. S., Hallström, S., JENDLE, J. H., Tengmark, B.-O., Folino, A., Marie, E., & Lind, M. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241242386 · Cited by 13
Dekker, P., van den Heuvel, T. R. A., Arrieta, A., Castañeda, J., Mul, D., Veeze, H. J., Cohen, O., & Aanstoot, H. (2024) · Journal of Diabetes Science and Technology · doi:10.1177/19322968241290259 · Cited by 16
Self-citationAdvances in Automated Insulin Delivery with the Medtronic 780G: The Australian Experience
Halim, B., Abraham, M. B., Manos, G., Arrieta, A., Dai, Z., Vogrin, S., Lu, J., MacIsaac, R. J., Ekinci, E. I., Davis, E. A., et al. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0348 · Cited by 12
Self-citationElbarbary, N. S., Alguwaihes, A. M., ZARIF, H. A., Hassanein, M., Deeb, A., Petrovski, G., Aldahash, R., Amoudi, R. A., Hussain, S., Ibrahim, M. A., et al. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0200 · Cited by 13
van den Heuvel, T. R. A., Castañeda, J., Arrieta, A., Voelker, B., Cohen, O., Liu, M., Garelli, F. D., & Shin, J. (2024) · Diabetes Obesity and Metabolism · doi:10.1111/dom.15868 · Cited by 13
Self-citationGera, S., Iyer, J., Meighan, S. S., March, C. A., & Marks, B. E. (2025) · Endocrine Practice · doi:10.1016/j.eprac.2025.06.004 · Cited by 13
Blood glucose monitoring devices: current considerations
Sly, B., & Taylor, J. (2023) · Australian Prescriber · doi:10.18773/austprescr.2023.013 · Cited by 11
Trandafir, L. M., Moisă, Ș. M., Vlaiculescu, M. V., Butnariu, L. I., Boca, L. O., Constantin, M. M., Lupu, P. M., Brinza, C., Temneanu, O. R., & Burlacu, A. (2022) · Medicina · doi:10.3390/medicina58111671 · Cited by 10
Landau, Z., Lebenthal, Y., Mazor‐Aronovitch, K., Brener, A., Levek, N., Jacobi‐Polishook, T., Ari, T. B., Abiri, S., Haim, A., Nir, J., et al. (2024) · Endocrine · doi:10.1007/s12020-024-03683-w · Cited by 10
Boucsein, A., Zhou, Y., Haszard, J. J., Jefferies, C., Wiltshire, E. J., Styles, S. E., Crocket, H. R., Galland, B. C., Pasha, M., Petrovski, G., et al. (2024) · Journal of Diabetes & Metabolic Disorders · doi:10.1007/s40200-024-01397-4 · Cited by 10
Jalilova, A., Pilan, B. Ş., Demir, G., Özbaran, B., Balkı, H. G., Arslan, E., Köse, S., Özen, S., Darcan, Ş., & Gökşen, D. (2024) · European Journal of Pediatrics · doi:10.1007/s00431-024-05551-1 · Cited by 10
Simplified Meal Management in Adults Using an Advanced Hybrid Closed-Loop System
Minsky, N. C., Shalit, R., Benedetti, A., Laron-Hirsh, M., Cohen, O., Kurtz, N., Roy, A., Grosman, B., & Tirosh, A. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0224 · Cited by 10
Self-citationBombaci, B., Passanisi, S., Calderone, M., Macrì, F., Lombardo, F., & Salzano, G. (2025) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16226 · Cited by 12
Jurado, A. L., Gómez‐Perea, A., Jiménez, A. B. A., Tapia‐Ceballos, L., Becerra‐Paz, I., Lirio, M. F. M., Jabato, F. M., & Gea, I. L. (2023) · Journal of Diabetes · doi:10.1111/1753-0407.13426 · Cited by 10
Newman, C., Hartnell, S., Wilinska, M. E., Alwan, H., & Hovorka, R. (2023) · Journal of Diabetes Science and Technology · doi:10.1177/19322968231187915 · Cited by 10
Metadata
- Authors
- Silva, J. D., Lepore, G., Battelino, T., Arrieta, A., Castañeda, J., Grossman, B., Shin, J., & Cohen, O.
- Affiliations
- Medtronic (Switzerland) (CH)
- Ospedale Papa Giovanni XXIII (IT)
- University of Ljubljana (SI)
- Medtronic (Netherlands) (NL)
- Medtronic, Northridge, California, USA.
- Medtronic (United States) (US)
- Corresponding author
- Cohen, Ohad
- Venue
- Diabetes Technology & Therapeutics
- Publisher
- SAGE Publications
- Published
- 2022-01-31
- Volume
- 24
- Issue
- 2
- Pages
- 113-119
- ISSN
- 1520-9156 (ISSN-L)
- 1557-8593
- PMC
- PMC8817690
- OpenAlex
- W3201039184
- Cited by
- 229
- References
- 26
- Citations by year
- 2026: 29 · 2025: 35 · 2024: 61 · 2023: 67 · 2022: 36
- FWCI
- 30.80
- Type
- journal article
- Language
- en
- Open access
- hybrid
- doi.org
- Licence
- Locations
- Diabetes Technology & Therapeutics (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Education
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Treatment and Management
- Keywords
- MiniMed 780G, hybrid closed-loop insulin delivery, type 1 diabetes, real-world performance, glucose management indicator, time in range, glycemic control, continuous glucose monitoring
- MeSH terms
- Blood Glucose, Diabetes Mellitus, Type 1, Humans, Hypoglycemic Agents, Insulin, Insulin Infusion Systems, Retrospective Studies, Blood Glucose Self-Monitoring
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review