Upper-Extremity Volume Measurements in Women With Lymphedema: A Comparison of Measurements Obtained Via Water Displacement With Geometrically Determined Volume
Abstract
Background and Purpose. Upper-extremity (UE) swelling following breast cancer treatment is a frequent manifestation of lymphedema. In order to document outcomes from lymphedema treatments, reliable, valid, and practical measurements of UE swelling are necessary. The purpose of this study was to compare geometric methods of determining UE volumes with water displacement methods. Subjects. The edematous hand, forearm, and upper arm of 50 women with UE swelling secondary to lymphedema were measured. Methods. Upper-extremity volumes were determined by water displacement using arm and hand volumeters. Displaced water was weighed to determine volume. Circumferential girth measurements were taken. Width and depth measurements of the hand were taken with a tension-controlled caliper. Geometric volume formulas for a cylinder, frustum, rectangular solid, and trapezoidal solid were used to calculate volumes of the arm and hand at different measurement intervals. Results. Intraclass correlation coefficients [2,1] for interrater and intrarater reliability of all water and geometric measurements of the arm and hand were .91 to .99 and .92 to .99, respectively. Water displacement correlated with geometric measurements in the arm (r=.97–.98) and in the hand (r=.81–.91). The limits of agreement (LOA) indicated that water and geometric measurements of arm volume differed by 479 to 655 mL. Scatterplots of the LOA data indicated in that geometric volumes were either larger or smaller than water volumes. The smallest standard error of measurement for the arm measurements was for the 6-cm frustum method at 115 mL; for the hand measurements, the smallest standard error of measurement was for the frustum method at 16 mL. Discussion and Conclusion. Volume of an edematous UE calculated by geometric formulas correlated strongly with volume determined by water displacement. Although strongly correlated, the measurements obtained by the 2 methods did not agree.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
Measurement in Medicine: The Analysis of Method Comparison Studies
Altman, D. G., & Bland, J. M. (1983) · The Statistician · doi:10.2307/2987937 · Cited by 4,300
Introductory Statistics for the Behavioral Sciences.
Greenough, H. P., Lancaster, J. S., Welkowitz, J., Ewen, R. B., & Cohen, J. (1975) · The American Mathematical Monthly · doi:10.2307/2318276 · Cited by 459
Early diagnosis of lymphedema using multiple frequency bioimpedance.
Cornish, B., Chapman, M., Hirst, C., Mirolo, B. R., Bunce, I., Ward, L. C., & Thomas, B. J. (2001) · PubMed · Cited by 312
Lymphedema: current issues in research and management
Petrek, J. A., Pressman, P. I., & Smith, R. A. (2000) · CA: A Cancer Journal for Clinicians · doi:10.3322/canjclin.50.5.292 · Cited by 283
Kaulesar Sukul, D. M. K. S., den Hoed, P. T., Johannes, E. J., van Dolder, R., & Benda, E. (1993) · Journal of Biomedical Engineering · doi:10.1016/0141-5425(93)90062-4 · Cited by 225
Volume measurement in lymphoedema treatment: examination of formulae
SITZIA, J. (1995) · European Journal of Cancer Care · doi:10.1111/j.1365-2354.1995.tb00047.x · Cited by 204
Infrared optoelectronic volumetry, the ideal way to measure limb volume
Tierney, S., Aslam, M., Rennie, K., & Grace, P. (1996) · European Journal of Vascular and Endovascular Surgery · doi:10.1016/s1078-5884(96)80005-0 · Cited by 188
Measuring and representing peripheral oedema and its alterations.
Casley‐Smith, J. R. (1994) · PubMed · Cited by 155
Cited by
Reliability and Validity of Arm Volume Measurements for Assessment of Lymphedema
Taylor, R., Jayasinghe, U. W., Koelmeyer, L., Ung, O., & Boyages, J. (2006) · Physical Therapy · doi:10.1093/ptj/86.2.205 · Cited by 334
Resistance exercise and secondary lymphedema in breast cancer survivors—a systematic review
Keilani, M., Hasenoehrl, T., Neubauer, M., & Crevenna, R. (2016) · Supportive Care in Cancer · doi:10.1007/s00520-015-3068-z · Cited by 93
Metadata
- Authors
- Sander, A. P., Hajer, N. M., Hemenway, K., & Miller, A. C.
- Venue
- Physical Therapy
- Publisher
- Oxford University Press (OUP)
- Published
- 2002-12-01
- Volume
- 82
- Issue
- 12
- Pages
- 1201-1212
- ISSN
- 0031-9023 (ISSN-L)
- 1538-6724
- OpenAlex
- W2110960601
- Cited by
- 317
- References
- 21
- FWCI
- 1.78
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Upper-Extremity Volume Measurements in Women With Lymphedema: A Comparison of Measurements Obtained Via Water Displacement With Geometrically Determined Volume. https://app.cochat.ai/papers/10.1093/ptj/82.12.1201
- The paper
- Sander, A. P., Hajer, N. M., Hemenway, K., & Miller, A. C. (2002). Upper-Extremity Volume Measurements in Women With Lymphedema: A Comparison of Measurements Obtained Via Water Displacement With Geometrically Determined Volume. Physical Therapy. https://doi.org/10.1093/ptj/82.12.1201
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.