A Comparison of the INVOS 4100 and the NIRO 300 Near-Infrared Spectrophotometers
The DOI matches its Crossref record; 18 of the 18 references it cites were found in a registry or an index.
Abstract
UNLABELLED: We determined whether two different devices for measuring near-infrared spectroscopy (NIRS)---the INVOS 4100 and the NIRO 300---produce similar cerebral oxygenation data during the CO(2) challenge test. Nineteen patients anesthetized with sevoflurane, 67% nitrous oxide in oxygen, and fentanyl were studied. A series of measurements of regional cerebral oxygen saturation (rSO(2)), measured by the INVOS 4100, and tissue oxygen index (TOI), measured by the NIRO 300, were performed in the following conditions: 1) normocapnia (PaCO(2), 35--45 mm Hg); 2) hypocapnia (PaCO(2), 25--35 mm Hg); 3) normocapnia; and 4) hypercapnia (PaCO(2), 45--55 mm Hg). Hemodynamic variables, including arterial blood gases and cerebral blood flow velocity, were measured at the same time with transcranial Doppler. The values and percentage changes of rSO(2) and TOI were compared by using regression analysis and Bland and Altman analysis. The rSO(2) showed a significant positive correlation with TOI (r = 0.58, P < 0.01). The percentage change of rSO(2) also showed a significant positive correlation with the percentage change of TOI during the CO(2) challenge (r = 0.85, P < 0.01). Bland and Altman analysis revealed a bias of -0.5% with 2 SD of 15.6% when comparing the rSO(2) value with the TOI value, and it showed a bias of -3.4% with 2 SD of 15.2% when comparing the percentage change of rSO(2) with the percentage change of TOI, indicating unacceptable disagreement of these data. These results indicate that cerebral oxygen saturation and its relative change during the CO(2) challenge may vary depending on the type of NIRS used. Because the measurement technique and algorithm were different in each device, we should carefully consider the clinical application of the values produced by NIRS. IMPLICATIONS: Near-infrared spectroscopy (NIRS) has been proposed as a noninvasive clinical method for assessing cerebral oxygenation. The acceptable reliability and validity of NIRS values have not been established despite their widespread use. The INVOS 4100 and the NIRO 300 can display cerebral oxygen saturation as regional cerebral oxygen saturation and tissue oxygenation index, but they produce differing results.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,856
Jöbsis, F. F. (1977) · Science · doi:10.1126/science.929199 · Cited by 3,899
Cope, M., & Delpy, D. T. (1988) · Medical & Biological Engineering & Computing · doi:10.1007/bf02447083 · Cited by 1,065
Sevick, E. M., Chance, B., Leigh, J., Nioka, S., & Maris, M. (1991) · Analytical Biochemistry · doi:10.1016/0003-2697(91)90339-u · Cited by 360 · cited 1×
“Servic et al. ( 16 ) described a method for determining the ratio of the absorption coefficients by using two wavelengths first.”
MethodNoninvasive Monitoring of Cerebral Oxygenation in Preterm Infants: Preliminary Observations
Brazy, J. E., Lewis, D. V., Mitnick, M. H., & vander Vliet, F. F. J. (1985) · Pediatrics · doi:10.1542/peds.75.2.217 · Cited by 356
Near-Infrared spectroscopy: Theory and applications
Wahr, J. A., Tremper, K. K., Samra, S., & Delpy, D. T. (1996) · Journal of Cardiothoracic and Vascular Anesthesia · doi:10.1016/s1053-0770(96)80107-8 · Cited by 227
Measurement of Optical Path Length for Cerebral Near-Infrared Spectroscopy in Newborn Infants
Wyatt, J. S., Cope, M., Delpy, D. T., van der Zee, P., Arridge, S., Edwards, A. D., & Reynolds, E. O. R. (1990) · Developmental Neuroscience · doi:10.1159/000111843 · Cited by 182
GERMON, T. J., KANE, N. M., MANARA, A. R., & NELSON, R. J. (1994) · British Journal of Anaesthesia · doi:10.1093/bja/73.4.503 · Cited by 170 · cited 1×
“Germon et al. ( 5 ) demonstrated that the INVOS 3100 cerebral oximeter was affected significantly by changes in extracranial blood flow oxygenation, which may affect its reliability in clinical practice.”
BackgroundMeasurement of hemoglobin flow and blood flow by near-infrared spectroscopy
Edwards, A. D., Richardson, C., van der Zee, P., Elwell, C., Wyatt, J. S., Cope, M., Delpy, D. T., & Reynolds, E. O. (1993) · Journal of Applied Physiology · doi:10.1152/jappl.1993.75.4.1884 · Cited by 153 · cited 1×
“Although the effect of extracranial blood flow on the data is still controversial (17), this might lead to the difference between the two devices.”
BackgroundCan Cerebrovascular Reactivity Be Measured With Near-Infrared Spectroscopy?
Smielewski, P. S., Kirkpatrick, P. J., Minhas, P., Pickard, J. D., & Czosnyka, M. (1995) · Stroke · doi:10.1161/01.str.26.12.2285 · Cited by 147 · cited 1×
“CO2-induced cerebrovascular reactivity individually (14,15), it seems likely that their values were not”
BackgroundTerborg, C., Gora, F., Weiller, C., & Röther, J. (2000) · Stroke · doi:10.1161/01.str.31.4.924 · Cited by 141 · cited 1×
“CO2-induced cerebrovascular reactivity individually (14,15), it seems likely that their values were not”
BackgroundNear infrared spectroscopy in adults: Does the Invos 3100 really measure intracerebral oxygenation?
Harris, D. N. F., & Bailey, S. M. (1993) · Anaesthesia · doi:10.1111/j.1365-2044.1993.tb07183.x · Cited by 100
Owen-Reece, H., Elwell, C. E., Harkness, W. F. J., Goldstone, J. C., Delpy, D. T., Wyatt, J. S., & Smith, M. (1996) · British Journal of Anaesthesia · doi:10.1093/bja/76.1.43 · Cited by 62
A comparison of the monitors INVOS 3100 and NIRO 500 in detecting changes in cerebral oxygenation
Grubhofer, G., Tonninger, W., Keznickl, P., Skyllouriotis, P., Ehrlich, M. P., Hiesmayr, M., & Lassnigg, A. M. (1999) · Acta Anaesthesiologica Scandinavica · doi:10.1034/j.1399-6576.1999.430417.x · Cited by 55 · cited 1×
“Several studies (8,9) processed the NIRO 500 data to derive rSo2 to compare the INVOS 3100 and the NIRO 500.”
MethodCho, H., Nemoto, E. M., SANDERS, M. H., Fernandez, K. C., & Yonas, H. (2000) · Journal of neurosurgery · doi:10.3171/jns.2000.93.2.0351 · Cited by 31
A Comparison of the Hamamatsu NIRO 500 and the INVOS 3100 Near-Infrared Spectrophotometers
Gomersall, C. D., Leung, P. L., Gin, T., Joynt, G. M., Young, R. J., Poon, W. S., & Oh, T. E. (1998) · Anaesthesia and Intensive Care · doi:10.1177/0310057x9802600512 · Cited by 23 · cited 1×
“Several studies (8,9) processed the NIRO 500 data to derive rSo2 to compare the INVOS 3100 and the NIRO 500.”
MethodAssessment of spatially resolved spectroscopy during cardiopulmonary bypass
Al-Rawi, P. G., Smielewski, P., Hobbiger, H., Ghosh, S., & Kirkpatrick, P. J. (1999) · Journal of Biomedical Optics · doi:10.1117/1.429911 · Cited by 15 · cited 1×
“(18) indicated that the NIRO 300 detector connection to the patient, being of a low voltage, was more prone to electrical interference from other monitoring equipment.”
Background
Cited by
Event-related functional near-infrared spectroscopy (fNIRS): Are the measurements reliable?
Plichta, M. M., Herrmann, M. J., Baehne, C. G., Ehlis, A.-C., Richter, M. M., Pauli, P., & Fallgatter, A. J. (2006) · NeuroImage · doi:10.1016/j.neuroimage.2005.12.008 · Cited by 358
Rooks, C. R., Thom, N. J., McCully, K. K., & Dishman, R. K. (2010) · Progress in Neurobiology · doi:10.1016/j.pneurobio.2010.06.002 · Cited by 324
Cerebral Near-Infrared Spectroscopy in Adults: A Work in Progress
Ghosh, A., Elwell, C., & Smith, M. (2012) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31826dd6a6 · Cited by 277
Monitoring Neonatal Regional Cerebral Oxygen Saturation in Clinical Practice: Value and Pitfalls
van Bel, F., Lemmers, P., & Naulaers, G. (2008) · Neonatology · doi:10.1159/000151642 · Cited by 258
Naulaers, G., Meyns, B., Miserez, M., Leunens, V., Van Huffel, S., Casaer, P., Weindling, M., & Devlieger, H. (2007) · Neonatology · doi:10.1159/000101063 · Cited by 218
Metadata
- Authors
- Yoshitani, K., Kawaguchi, M., Tatsumi, K., Kitaguchi, K., & Furuya, H.
- Venue
- Anesthesia & Analgesia
- Published
- 2002-03
- Volume
- 94
- Issue
- 3
- Pages
- 586-590
- ISSN
- 0003-2999 (ISSN-L)
- OpenAlex
- W1967137016
- Cited by
- 162
- References
- 18
- FWCI
- 8.03
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.