Evaluation of the Ohmeda 3700 Pulse Oximeter: Steady-state and Transient Response Characteristics
Open accessThe DOI matches its Crossref record.
Abstract
The authors determined the accuracy of the Ohmeda 3700 (version J) pulse oximeter in healthy volunteers rendered hypoxic (SaO2 from 60-98%) by breathing mixtures of O2 in N2. When equipped with an ear probe, the pulse oximeter reading (y) reliably predicted arterial saturation (x) under steady-state conditions (y = 1.05x - 4.66, r = 0.98) as well as when oxygen saturation was rapidly decreasing (y = 1.05x - 6.38, r = 0.96). Conversely, when equipped with a finger probe, the oximeter tended to significantly underestimate steady-state arterial saturation (y = 1.21x - 19.1, r = 0.98, P less than 0.001). In response to this information, the manufacturer modified the oximeter's software (version XJ1), resulting in improved agreement between oximeter readings and arterial values (y = 0.96x + 4.59, r = 0.99). Despite the close correlation between steady-state oximeter readings and arterial saturation, the 99% prediction limits for both the ear and finger probes (version XJ1) were +/- 8%. Finger probe readings did not reliably reflect radial arterial oxygenation during rapid desaturation (y = 0.55x + 45.2, r = 0.78). This may be related to the time required to "arterialize" the blood in the finger; during acute resaturation, we found that the ear- to finger-probe delay was 24.0 +/- 2.3 s (means +/- SE, P less than 0.001).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Pulse oximetry: Analysis of theory, technology, and practice
Wukitsch, M. W., Petterson, M. T., Tobler, D. R., & Pologe, J. A. (1988) · Journal of Clinical Monitoring · doi:10.1007/bf01617328 · Cited by 306
Oximetry of retinal vessels by dual-wavelength imaging: calibration and influence of pigmentation
Beach, J. M., Schwenzer, K. J., Srinivas, S., Kim, D., & Tiedeman, J. S. (1999) · Journal of Applied Physiology · doi:10.1152/jappl.1999.86.2.748 · Cited by 263
Meta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults
Jensen, L. A., Onyskiw, J. E., & Prasad, N. G. N. (1998) · Heart & Lung · doi:10.1016/s0147-9563(98)90086-3 · Cited by 243
Ries, A. L., Prewitt, L. M., & Johnson, J. J. (1989) · Chest · doi:10.1378/chest.96.2.287 · Cited by 145
The desaturation response time of finger pulse oximeters during mild hypothermia*
MacLeod, D. B., Cortinez, L. I., Keifer, J. C., Cameron, D., Wright, D. R., White, W. D., Moretti, E. W., Radulescu, L. R., & Somma, J. (2005) · Anaesthesia · doi:10.1111/j.1365-2044.2004.04033.x · Cited by 98
Metadata
- Authors
- Kagle, D. M., Alexander, C. M., Berko, R. S., Gluffre, M., & Gross, J. B.
- Venue
- Anesthesiology
- Published
- 1987-03-01
- Volume
- 66
- Issue
- 3
- Pages
- 376-380
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2069535946
- Cited by
- 109
- References
- 0
- FWCI
- 15.38
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Evaluation of the Ohmeda 3700 Pulse Oximeter: Steady-state and Transient Response Characteristics. https://app.cochat.ai/papers/10.1097/00000542-198703000-00018
- The paper
- Kagle, D. M., Alexander, C. M., Berko, R. S., Gluffre, M., & Gross, J. B. (1987). Evaluation of the Ohmeda 3700 Pulse Oximeter: Steady-state and Transient Response Characteristics. Anesthesiology. https://doi.org/10.1097/00000542-198703000-00018
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.