Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing

Tang, Zuping, Louie, Richard F., Lee, Judith H., Lee, Debra M., Miller, Earl E., Kost, Gerald J. · 2001 · Critical Care Medicine
Cited by 188

The DOI matches its Crossref record; 14 of the 14 references it cites were found in a registry or an index.

Abstract

OBJECTIVES: To determine the effects of different oxygen tensions (Po2) on glucose measurements with glucose dehydrogenase (GD)-based and glucose oxidase (GO)-based test strips, to quantitate changes in glucose measurements observed with different Po2 levels, and to discuss the potential risks of oxygen-derived glucose errors in critical care. DESIGN: Venous blood from healthy volunteers was tonometered to create different oxygen tensions simulating patient arterial Po2 levels. Venous blood from diabetic patients was exposed to air to alter oxygen tensions simulating changes in Po2 during sample handling. Whole-blood glucose measurements obtained from these samples with six glucose meters were compared with reference analyzer plasma glucose measurements. Glucose differences were plotted vs. different Po2 levels to identify error trends. Error tolerances were as follows: a) within +/-15 mg/dL of the reference measurement for glucose levels 100 mg/dL. SETTING AND SUBJECTS: Five healthy volunteers in the bench study and 11 diabetic patients in the clinical study. RESULTS: In the bench study, increases in Po2 levels decreased glucose measured with GO-based amperometric test strips, mainly at Po2 levels >100 torr. At nearly constant glucose concentrations, glucose meter systems showed large variations at low (39 torr) vs. high (396 torr) Po2 levels. Glucose measured with GD-based amperometric and GO-based photometric test strips generally were within error tolerances. In the clinical study, 31.6% (Precision PCx), 20.2% (Precision QID), and 23.0% (Glucometer Elite) of glucose measurements with GO-based amperometric test strips, 14.3% (SureStep) of glucose measurements with GO-based photometric test strips, and 4.6% (Accu-Chek Advantage H) and 5.9% (Accu-Chek Comfort Curve) of glucose measurements with GD-based amperometric test strips were out of the error tolerances. CONCLUSIONS: Different oxygen tensions do not significantly affect glucose measured with the GD-based amperometric test strips, and have minimal effect on GO-based photometric test strips. Increases in oxygen tension lowered glucose measured with GO-based amperometric test strips. We recommend that the effects of different oxygen tensions in blood samples on glucose measurements be minimized by using oxygen-independent test strips for point-of-care glucose testing in critically ill and other patients with high or unpredictable blood Po2 levels.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1097/00003246-200105000-00038

Cites

14 of 14 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Cited by

See all 188

Metadata

Authors
Tang, Z., Louie, R. F., Lee, J. H., Lee, D. M., Miller, E. E., & Kost, G. J.
Venue
Critical Care Medicine
Published
2001-05
Volume
29
Issue
5
Pages
1062-1070
ISSN
0090-3493 (ISSN-L)
OpenAlex
W2041970007
Cited by
188
References
14
FWCI
2.76
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.