Inaccuracy of seven popular sphygmomanometers for home measurement of blood pressure
The DOI matches its Crossref record.
Abstract
Seven models, available commercially for the self-measurement of blood pressure, were subjected to a validation procedure in which three devices of each model were tested by observers who were trained to a high standard of accuracy. The models were the Omron HEM-400C, the Philips HP5308, the Healthcheck 'Cuffless' CX-5 060020, the Nissei Analogue Monitor, the Philips HP5306/B, the Systema Dr MI-150 and the Fortec Dr MI-100. The validation programme had a number of unique features which included assessment of interdevice variability before and after 1 month of home use, and a new form of analysis, which we term 'clinical', based on the likely influence of three grades of device inaccuracy on patient management. In the main validation phase, one device of each model was compared with simultaneous measurements made by two 'blinded' observers using a standard mercury sphygmomanometer (PyMaH Corporation, New Jersey, USA) in the same arm in 85 subjects with a wide range of blood pressures. Three models (the Healthcheck 'Cuffless' CX-5 060020, the Systema Dr MI-150 and the Fortec Dr MI-100) failed the interdevice variability tests and did not reach the main validation test. Two models (the Omron HEM-400C and the Philips HP5306/B) failed on the criteria set down by the American National Standard for Electronic or Automated Sphygmomanometers, as well as the 'clinical' criteria. The remaining two models (the Nissei Analogue Monitor and the Philips HP5308) were acceptable for the measurement of systolic blood pressure by both methods of analysis but failed in the 'clinical' analysis for diastolic blood pressure. The mercury sphygmomanometer was comfortably within the criteria for both methods of analysis.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Blood pressure measuring devices: recommendations of the European Society of Hypertension
O'Brien, E., Waeber, B., Parati, G., Staessen, J., & Myers, M. G. (2001) · BMJ · doi:10.1136/bmj.322.7285.531 · Cited by 903
Self-citationO??Brien, E., Petrie, J., Littler, W., de Swiet, M., Padfield, P. L., O??Malley, K., Jamieson, M., Altman, D., Bland, M., & Atkins, N. (1990) · Journal of Hypertension · doi:10.1097/00004872-199007000-00004 · Cited by 772
Self-citationOʼBrien, E., Pickering, T., Asmar, R., Myers, M., Parati, G., Staessen, J., Mengden, T., Imai, Y., Waeber, B., Palatini, P., et al. (2002) · Blood Pressure Monitoring · doi:10.1097/00126097-200202000-00002 · Cited by 744
Self-citationVictor, R. G., Haley, R. W., Willett, D. L., Peshock, R. M., Vaeth, P. C., Leonard, D., Basit, M., Cooper, R. S., Iannacchione, V. G., Visscher, W. A., et al. (2004) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2004.02.058 · Cited by 527
Drazner, M. H., Dries, D. L., Peshock, R. M., Cooper, R. S., Klassen, C., Kazi, F., Willett, D., & Victor, R. G. (2005) · Hypertension · doi:10.1161/01.hyp.0000169972.96201.8e · Cited by 320
Dries, D. L., Victor, R. G., Rame, J. E., Cooper, R. S., Wu, X., Zhu, X., Leonard, D., Ho, S.-I., Wu, Q., Post, W., et al. (2005) · Circulation · doi:10.1161/circulationaha.105.568881 · Cited by 195
Reference Values for Self-recorded Blood Pressure: A Meta-analysis of Summary Data
Thijs, L., Staessen, J. A., Celis, H., de Gaudemaris, R., Imai, Y., Julius, S., & Fagard, R. (1998) · Archives of Internal Medicine · doi:10.1001/archinte.158.5.481 · Cited by 143
“Imai et al,34 1993 Population 871 (707) 46 (7-98) 43 HEM 401C 3-28 M 1 (S) James et al,12 1988 NT volunteers 14 24 (?-?) 40 ? 6 M+E 3 (S) Johnson,35 1989 NT 24 20 (19-25) 100 Marshall 28 M+E 1 (S) Joossens et al,36 1971 White-collar workers 39 (?) ? (17-65) 49 ? 63 M+E 2 (L+E) Julius et al,37 1974 NT 49 25 (?-?) 100 ?…”
Home Self Blood Pressure Measurement in General Practice The SMART Study
Chatellier, G. (1996) · American Journal of Hypertension · doi:10.1016/0895-7061(96)00018-0 · Cited by 116
The pursuit of accurate blood pressure measurement: A 35‐year travail
O'Brien, E., & Stergiou, G. S. (2017) · The Journal of Clinical Hypertension · doi:10.1111/jch.13005 · Cited by 39
Metadata
- Authors
- O??Brien, E., Mee, F., Atkins, N., & O??Malley, K.
- Venue
- Journal of Hypertension
- Published
- 1990-07
- Volume
- 8
- Issue
- 7
- Pages
- 621-634
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W2079222449
- Cited by
- 71
- References
- 0
- FWCI
- 3.30
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.