The accuracy of automated blood pressure measuring devices in patients with controlled atrial fibrillation
The DOI matches its Crossref record.
Abstract
OBJECTIVE: To determine whether automated blood pressure measuring devices can measure blood pressure accurately in patients with atrial fibrillation. DESIGN: Comparison of the accuracy of two electronic sphygmomanometers [Takeda UA-751 (Takeda) and Copal UA-251 (Copal)] and two ambulatory blood pressure monitors [Accutracker 1 (Accutracker) and SpaceLabs 90207 (SpaceLabs)] with that of a trained observer using a Hawksley random-zero sphygmomanometer (Hawksley), using the sequential same-arm technique. SETTING: University teaching hospital: medical wards and outpatient department. SUBJECTS: Twenty-eight patients, mean +/- SD age 72 +/- 9 years, blood pressure range 90-158/40-96 mmHg, in atrial fibrillation with a controlled ventricular rate. MAIN OUTCOME MEASURES: The proportion of machine readings > 5 mmHg different from the Hawksley reading was compared with that obtained by three sequential Hawksley measurements. The variability of each measuring method was assessed by determining the SDD for the paired readings from each device. RESULTS: Five per cent of Takeda, 5% of Copal, 14% of Accutracker and 21% of SpaceLabs readings could not be obtained. Sequential testing with the Hawksley resulted in an accuracy at the 5-mmHg level of (systolic/diastolic) 79/79%, compared with 64/54% (P < 0.05 for diastolic) for the Takeda, 68/75% (NS) for the Copal, 50/36% (P < 0.01 for both) for the Accutracker and 50/29% (P < 0.01 for systolic, P < 0.001 for diastolic) for the SpaceLabs. Intrapatient variability, as assessed by SDD, was 8.3/8.6 mmHg for the Hawksley, similar to that for the Copal (7.7/7.3 mmHg) but higher for the Takeda (11.2/19.7 mmHg), the Accutracker (22.4/26.3 mmHg) and the SpaceLabs (7.5/14.8 mmHg). CONCLUSIONS: Accurate measurement of blood pressure with an electronic device is possible in patients who have atrial fibrillation; the Copal UA-251 provides a satisfactory level of accuracy. However, the marked difference between devices and the limited accuracy of the other machines tested here demonstrates the need to ensure that such devices are of proven accuracy in this patient group.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051
O'Brien, E., Asmar, R., Beilin, L., Imai, Y., Mallion, J.-M., Mancia, G., Mengden, T., Myers, M., Padfield, P., Palatini, P., et al. (2003) · Journal of Hypertension · doi:10.1097/00004872-200305000-00001 · Cited by 1,655
Self-citationParati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e328308da66 · Cited by 837
Self-citationBasu, S., & Millett, C. (2013) · Hypertension · doi:10.1161/hypertensionaha.113.01374 · Cited by 267
Oscillometric Blood Pressure Estimation: Past, Present, and Future
Forouzanfar, M., Dajani, H. R., Groza, V. Z., Bolic, M., Rajan, S., & Batkin, I. (2015) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2015.2434215 · Cited by 196
Blood pressure measurement: Part I—Sphygmomanometry: factors common to all techniques
Beevers, G., Lip, G. Y. H., & O'Brien, E. (2001) · BMJ · doi:10.1136/bmj.322.7292.981 · Cited by 183
Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis
Stergiou, G. S., Kollias, A., Destounis, A., & Tzamouranis, D. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32835850d7 · Cited by 120
“Two of the studies have tested the same oscillometric device (SpaceLabs 9020) [14,15] and another two tested oscillometric devices of the same manufacturer (Omron 705CP and Omron 711AC) [16,19], which may have the same or similar algorithm.”
MethodClark, C. E., McDonagh, S. T. J., & McManus, R. J. (2019) · Journal of Human Hypertension · doi:10.1038/s41371-018-0153-z · Cited by 49
“Six studies reported achievement of some, or all, of the standards for the 2010 International Protocol, although none adopted the precise protocol itself [27, 29, 30, 32, 34, 39].”
BackgroundMeasurement of blood pressure in people with atrial fibrillation
on behalf of the Blood Pressure Measurement Working Party of the British and Irish Hypertension Society, Clark, C. E., McDonagh, S. T. J., & McManus, R. J. (2019) · Journal of Human Hypertension · doi:10.1038/s41371-019-0261-4 · Cited by 16
Metadata
- Authors
- Stewart, M. J., Gough, K., & Padfield, P. L.
- Venue
- Journal of Hypertension
- Published
- 1995-03
- Volume
- 13
- Issue
- 3
- Pages
- 297???300
- ISSN
- 0263-6352 (ISSN-L)
- OpenAlex
- W2081797908
- Cited by
- 63
- References
- 0
- FWCI
- 1.18
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.