A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor
The DOI matches its Crossref record.
Abstract
OBJECTIVE: To assess the clinical accuracy of the Mobil O Graph (version 12) ambulatory blood pressure monitor in an adult population. METHODS: The accuracy of the device was assessed by predefined criteria (British Hypertension Society, BHS) in 85 subjects recruited from the patients and staff in a teaching hospital. A series of same-arm sequential blood pressure measurements were taken: first two observers taking simultaneous mercury readings, followed by a reading with the Mobil O Graph ambulatory monitor. A total of seven readings were taken from each subject in the sitting position. The data were then analysed according to the BHS protocol and the criteria of the Association for the Advancement of Medical Instrumentation (AAMI). RESULTS: The Mobil O Graph ambulatory monitor fulfilled the criteria of the BHS protocol, achieving a grade B for systolic blood pressure (SBP) and a grade A for diastolic blood pressure (DBP). The mean differences were -2+/-8 mmHg for SBP and -2+/-7 mmHg for DBP. The device therefore also passed the AAMI standard (the mean to be within 5+/-8 mmHg). CONCLUSION: The Mobil O Graph ambulatory monitor performed in a satisfactory manner according to the BHS and the AAMI criteria and can therefore be recommended for clinical use in the general population.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Diretrizes Brasileiras de Hipertensão Arterial – 2020
Barroso, W. K. S., Rodrigues, C. I. S., Bortolotto, L. A., Mota-Gomes, M. A., Brandão, A. A., Feitosa, A. D. D. M., Machado, C. A., Poli-de-Figueiredo, C. E., Amodeo, C., Mion, D., et al. (2021) · Arquivos Brasileiros de Cardiologia · doi:10.36660/abc.20201238 · Cited by 1,068
Validation of a Brachial Cuff-Based Method for Estimating Central Systolic Blood Pressure
Weber, T., Wassertheurer, S., Rammer, M., Maurer, E., Hametner, B., Mayer, C. C., Kropf, J., & Eber, B. (2011) · Hypertension · doi:10.1161/hypertensionaha.111.176313 · Cited by 461
Prognostic Influence of Office and Ambulatory Blood Pressures in Resistant Hypertension
Salles, G. F., Cardoso, C. R. L., & Muxfeldt, E. S. (2008) · Archives of Internal Medicine · doi:10.1001/archinte.168.21.2340 · Cited by 293
Coleman, A., Steel, S., Freeman, P., de Greeff, A., & Shennan, A. (2008) · Blood Pressure Monitoring · doi:10.1097/mbp.0b013e3282cb57b6 · Cited by 259
Self-citationEfficacy of Spironolactone Therapy in Patients With True Resistant Hypertension
de Souza, F., Muxfeldt, E., Fiszman, R., & Salles, G. (2010) · Hypertension · doi:10.1161/hypertensionaha.109.140988 · Cited by 234
Acute effects of night-time noise exposure on blood pressure in populations living near airports
Haralabidis, A. S., Dimakopoulou, K., Vigna-Taglianti, F., Giampaolo, M., Borgini, A., Dudley, M.-L., Pershagen, G., Bluhm, G., Houthuijs, D., Babisch, W., et al. (2008) · European Heart Journal · doi:10.1093/eurheartj/ehn013 · Cited by 187
Coleman, A., Freeman, P., Steel, S., & Shennan, A. (2006) · Blood Pressure Monitoring · doi:10.1097/01.mbp.0000189788.05736.5f · Cited by 179
Self-citationPrevalence and Associated Factors of Obstructive Sleep Apnea in Patients with Resistant Hypertension
Muxfeldt, E. S., Margallo, V. S., Guimaraes, G. M., & Salles, G. F. (2014) · American Journal of Hypertension · doi:10.1093/ajh/hpu023 · Cited by 169
Prognostic Value of Nocturnal Blood Pressure Reduction in Resistant Hypertension
Muxfeldt, E. S., Cardoso, C. R. L., & Salles, G. F. (2009) · Archives of Internal Medicine · doi:10.1001/archinternmed.2009.68 · Cited by 155
Salles, G. F., Ribeiro, F. M., Guimarães, G. M., Muxfeldt, E. S., & Cardoso, C. R. L. (2014) · Journal of Hypertension · doi:10.1097/hjh.0000000000000068 · Cited by 34
“Unauth Journal of Hypertension patients in the sitting position, using a digital BP monitor Omron HEM-907 XL (Omron Healthcare, Tokyo, Japan) with a suitably sized cuff.”
BackgroundInfluentialYu, S., Chi, C., Protogerou, A. D., Safar, M. E., Blacher, J., Argyris, A. A., Nasothimiou, E. G., Sfikakis, P. P., Papaioannou, T. G., Xu, H., et al. (2018) · The Journal of Clinical Hypertension · doi:10.1111/jch.13226 · Cited by 9
Metadata
- Authors
- Jones, C. R., Taylor, K., Chowienczyk, P., Poston, L., & Shennan, A. H.
- Venue
- Blood Pressure Monitoring
- Published
- 2000-08
- Volume
- 5
- Issue
- 4
- Pages
- 233-238
- ISSN
- 1359-5237 (ISSN-L)
- OpenAlex
- W2328818748
- Cited by
- 129
- References
- 0
- FWCI
- 0.47
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.