Four Types of Pulse Oximeters Accurately Detect Hypoxia during Low Perfusion and Motion
Abstract
Background Pulse oximeter performance is degraded by motion artifacts and low perfusion. Manufacturers developed algorithms to improve instrument performance during these challenges. There have been no independent comparisons of these devices. Methods We evaluated the performance of four pulse oximeters (Masimo Radical-7, USA; Nihon Kohden OxyPal Neo, Japan; Nellcor N-600, USA; and Philips Intellivue MP5, USA) in 10 healthy adult volunteers. Three motions were evaluated: tapping, pseudorandom, and volunteer-generated rubbing, adjusted to produce photoplethsmogram disturbance similar to arterial pulsation amplitude. During motion, inspired gases were adjusted to achieve stable target plateaus of arterial oxygen saturation (SaO2) at 75%, 88%, and 100%. Pulse oximeter readings were compared with simultaneous arterial blood samples to calculate bias (oxygen saturation measured by pulse oximetry [SpO2] - SaO2), mean, SD, 95% limits of agreement, and root mean square error. Receiver operating characteristic curves were determined to detect mild (SaO2 Results Pulse oximeter readings corresponding to 190 blood samples were analyzed. All oximeters detected hypoxia but motion and low perfusion degraded performance. Three of four oximeters (Masimo, Nellcor, and Philips) had root mean square error greater than 3% for SaO2 70 to 100% during any motion, compared to a root mean square error of 1.8% for the stationary control. A low perfusion index increased error. Conclusions All oximeters…
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“it is known that pulse oximeters underperform under circumstances of low perfusion [33]”
From Investigating the accuracy of blood oxygen saturation measurements in common consumer smartwatches (2023) · cites this paper as [33]
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Beckett, A. H., Rowland, M., & TURNER, P. (1965) · The Lancet · doi:10.1016/s0140-6736(65)91037-8 · Cited by 16,917
Statistical methods for assessing agreement between two methods of clinical measurement
Bland, J. M., & Altman, D. G. (2010) · International Journal of Nursing Studies · doi:10.1016/j.ijnurstu.2009.10.001 · Cited by 5,220
Agreement Between Methods of Measurement with Multiple Observations Per Individual
Bland, J. M., & Altman, D. G. (2007) · Journal of Biopharmaceutical Statistics · doi:10.1080/10543400701329422 · Cited by 1,864
Simple, accurate equations for human blood O2 dissociation computations
Severinghaus, J. W. (1979) · Journal of Applied Physiology · doi:10.1152/jappl.1979.46.3.599 · Cited by 751
Jubran, A. (2015) · Critical Care · doi:10.1186/s13054-015-0984-8 · Cited by 511
Pulse oximetry: fundamentals and technology update
Nitzan, M., Romem, A., & Koppel, R. (2014) · Medical Devices: Evidence and Research · doi:10.2147/mder.s47319 · Cited by 438
Effects of skin pigmentation on pulse oximeter accuracy at low saturation
Bickler, P. E., Feiner, J. R., & Severinghaus, J. W. (2005) · Anesthesiology · doi:10.1097/00000542-200504000-00004 · Cited by 427
Self-citationJubran, A. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2399-x · Cited by 400
The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)
Barker, S. J., & Shah, N. K. (1997) · Anesthesiology · doi:10.1097/00000542-199701000-00014 · Cited by 197
The Effect of Motion on Pulse Oximetry and Its Clinical Significance
Petterson, M. T., Begnoche, V. L., & Graybeal, J. M. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000278134.47777.a5 · Cited by 177
“Motion-Resistant” Pulse Oximetry: A Comparison of New and Old Models
Barker, S. J. (2002) · Anesthesia & Analgesia · doi:10.1097/00000539-200210000-00033 · Cited by 176
Baysal, A., Doğukan, M., & Toman, H. (2015) · Critical Care · doi:10.1186/cc14352 · Cited by 95
Gehring, H., Hornberger, C., Matz, H., Konecny, E., & Schmucker, P. (2002) · Respiratory Care · doi:10.1177/002013240204700105 · Cited by 92
Bickler, P. E., Feiner, J., Lipnick, M. S., Batchelder, P. B., MacLeod, D. B., & Severinghaus, J. W. (2016) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000001421 · Cited by 91
Self-citationNew-Generation Pulse Oximetry in the Care of Critically Ill Patients
Giuliano, K. K., & Higgins, T. L. (2005) · American Journal of Critical Care · doi:10.4037/ajcc2005.14.1.26 · Cited by 64
Batchelder, P. B., & Raley, D. M. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000268495.35207.ab · Cited by 62
Performance of three new-generation pulse oximeters during motion and low perfusion in volunteers
Shah, N., Ragaswamy, H. B., Govindugari, K., & Estanol, L. (2012) · Journal of Clinical Anesthesia · doi:10.1016/j.jclinane.2011.10.012 · Cited by 62
Hummler, H., Engelmann, A., Pohlandt, F., Högel, J., & Franz, A. R. (2006) · Intensive Care Medicine · doi:10.1007/s00134-006-0254-y · Cited by 58
Recent advances in pulse oximetry
Cannesson, M. P., & Talke, P. O. (2009) · F1000 Medicine Reports · doi:10.3410/m1-66 · Cited by 44
A characterization of motion affecting pulse oximetry in 350 patients.
Tobin, R. M., Pologe, J. A., & Batchelder, P. B. (2002) · PubMed · Cited by 38
Influence of pulse oximeter lower alarm limit on the incidence of hypoxaemia in the recovery room
Rheineck-Leyssius, A. T., & Kalkman, C. J. (1997) · British Journal of Anaesthesia · doi:10.1093/bja/79.4.460 · Cited by 34
Issues in the laboratory evaluation of pulse oximeter performance.
Jopling, M. W., Mannheimer, P. D., & Bebout, D. E. (2002) · PubMed · Cited by 21
Comparison of new generation motion‐resistant pulse oximeters
Townshend, J., Taylor, B. J., Galland, B. C., & Williams, S. (2006) · Journal of Paediatrics and Child Health · doi:10.1111/j.1440-1754.2006.00873.x · Cited by 16
Cited by
Sorted by citations
Racial bias in pulse oximetry measurement
Sjoding, M. W., Dickson, R. P., Iwashyna, T. J., Gay, S. E., & Valley, T. S. (2020) · New England Journal of Medicine · doi:10.1056/nejmc2029240 · Cited by 1,114
Current progress of photoplethysmography and SPO2 for health monitoring
TAMURA, T. (2019) · Biomedical Engineering Letters · doi:10.1007/s13534-019-00097-w · Cited by 274
Valbuena, V. S. M., Seelye, S., Sjoding, M. W., Valley, T. S., Dickson, R. P., Gay, S. E., Claar, D., Prescott, H. C., & Iwashyna, T. J. (2022) · BMJ · doi:10.1136/bmj-2021-069775 · Cited by 127
Valbuena, V. S. M., Barbaro, R. P., Claar, D. D., Valley, T. S., Dickson, R. P., Gay, S. E., Sjoding, M. W., & Iwashyna, T. J. (2021) · CHEST Journal · doi:10.1016/j.chest.2021.09.025 · Cited by 118
A review of the effect of skin pigmentation on pulse oximeter accuracy
Al-Halawani, R., Charlton, P. H., Qassem, M., & Kyriacou, P. A. (2023) · Physiological Measurement · doi:10.1088/1361-6579/acd51a · Cited by 119
Metadata
- Authors
- Louie, A., Feiner, J. R., Bickler, P. E., Rhodes, L., Bernstein, M., & Lucero, J.
- Affiliations
- University of California, San Francisco (US)
- Corresponding author
- Lucero, Jennifer
- Venue
- Anesthesiology
- Published
- 2018-03
- Volume
- 128
- Issue
- 3
- Pages
- 520-530
- ISSN
- 0003-3022 (ISSN-L)
- OpenAlex
- W2773802879
- Cited by
- 106
- References
- 24
- Citations by year
- 2026: 6 · 2025: 13 · 2024: 11 · 2023: 13 · 2022: 17 · 2021: 26 · 2020: 9 · 2019: 7 · 2018: 4
- FWCI
- 2.46
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Anesthesiology (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Health Sciences › Medicine › Physiology (medical) › Cardiovascular and exercise physiology
- Keywords
- pulse oximeter, hypoxemia detection, motion artifacts, low perfusion, arterial oxygen saturation, photoplethysmography, pulse oximetry accuracy, device comparison, hypoxia sensitivity
- MeSH terms
- Adult, Algorithms, Hypoxia, Female, Humans, Male, Motion, Oximetry, Oxygen, Reference Values, Reproducibility of Results, Artifacts, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review